
3
Cleanallpipelinestoremovedirtandscalebeforeinstallation.Apply
a minimum amount of pipe compound to the male threads of the
fittingonly. DoNotuseteflontapeasasealant. Startwiththethird
thread back and work away from the end of the fitting to avoid
contaminating the transducer. Install the transducer in the air line.
Theinletandoutletportsarelabeledontheendsof thetransducer.
Tightenconnectionssecurely. Avoidundersizedfittingsthatwilllimit
the flow through the transducer and cause pressure drop down
stream.
Electric Connection
Make connections as shown in Figure 8.
Pneumatic Connection
NOTE: Oil free air is required. Use a filter to remove dirt and
liquid in the air line ahead of the transducer. If an air
line lubricator is used, it MUST be located down-
stream to avoid interference with transducer perfor-
mance.
The user is responsible for insuring that the environ-
ment in which the unit will be installed and the
operatinggasarecompatiblewiththematerialsinthe
transducer.
1.
2.
3.
Shut off the valve that is supplying air to transducer.
It is not necessary to remove the Transducer from
the air line.
Remove the Orifice Assembly from the unit. For
detailed information see Figure 1 “T6100 Calibration
Configuration on page 1.
Clean with alcohol and dry with compressed air.
To clean the Orifice, use the following procedure:
MAINTENANCE
Parts must be completely dry before reassembling.
If the standard maintenance procedure does not
correct the trouble, install the appropriate Service
Kit below:
20781-1
Service Kits also include Gasket, O-Rings, and
Membranes to replace "W" Option product.
NOTES:
The information set
forth in the foregoing
Operation and Mainte-
nance Instructions
shall not be modified
or amended in any re-
spect without prior
written consent of
Fairchild Industrial
Products Company. In
addition, the informa-
tion set forth herein
shallbefurnishedwith
each product sold in-
corporatingFairchild's
unit as a component
thereof.
LEGAL NOTICE:
Figure 8.
1
TROUBLE-SHOOTING
Table 1. Trouble-Shooting.
Problem Solution (check)
Supply Pressure
Clogged Orifice
Input Signal
Pneumatic Connections
Zero and Span Adjust
Supply Pressure Low
OutputLeakage
DC Signal
Loose Wires or Connections
Liquid in Air Supply
Dirt in Magnet Gap
NoOutput
Leakage
Low or Improper
SpanAdjust
Erratic Operation
Failure of Transducer could result in out-
put pressure increasing to supply pres-
sure possibly causing personal injury or
damage to equipment.
WARNING:
SPLIT RANGE OPERATION
All units have the capability to be split ranged or set for any
output in the range as long as the Output Span is equal to
or greater than the minimum Span.