INSTALLATION
CALIBRATION
Toensuresafeandefficientoperation,itisessentialtoread
each of these warnings and precautions, and to follow all
instructions listed in this manual.
1. Improper use or installation of this product can
cause serious bodily injury or death.
2. DONOTsmoke near meteror use meternear an open
flamewhendispensingflammablefluids. Firecould
result.
3. Do not exceed 50 PSI / 3.5 BARS line pressure.
4. CAUTION:DONOTINSTALLADDITIONALFOOT
VALVEORCHECKVALVEDURINGINSTALLATION
WITHOUTPRESSURERELIEFVALVE.CRACKING
MAYRESULT.
5. Thisproductshould notbe used forfluid transferinto
aircraft.
6. This product is not suited for use with fluids for human
consumption,includingpotablewater.
SAFETY INSTRUCTIONS
Owner's Operation & Safety Manual
SERIES 900
METER
Models 901, 901N, 901T, 9011.5,
901N1.5, 901T1.5, 901MK300,
901MK4200
Metersare furnishedforhorizontalpiping;lefttorightflow.
Flow ports can be changed to any of four positions for
horizontalorverticalpipingandforeitherdirectionofflow.
1. Determinedirectionforfluidtoflow.
2. Installmeterobservingdirectionalarrowoncasting.
3. Removefourscrews(item28).
4. Rotate meter cover assembly (item 37) to desired
orientation.
5. Replacefourscrews(item28).
TheFill-RiteSeries900meterscanbecalibratedforeither
U.S. gallons or liters. Calibration is required upon
installation, after disassembly, after significant wear or
whenmeteringadifferentviscosityfluid. Dependingonthe
model, Series 900 meters are factory calibrated in either
U.S.gallons or litersusing mineral spirit. Calibrationmust
be done between 6 and 40 GPM (23 and 151 LPM).
Meter calibration can be easily changed by following the
calibrationprocedure. AcontainerofKNOWNvolumewill
beneededforthecalibrationprocedure. Forthe900Series
meter,afivegalloncontainer or larger shouldbeused.
For accurate measurement and to prevent meter damage,
meterandpipingmustalwaysbefilledwithliquidandfreeof
air. Metershouldbecalibratedperinstructionsinthismanual
prior to its use.
1. Stopflow of liquid.
2. Resetregisterto “0”.
3. Meterisreadyforuse. Startflowofliquid. Donotexceed
50PSI of linepressure.
OPERATING INSTRUCTIONS
MAINTENANCE
Metershouldoperatemaintenancefree. However,certain
liquids can dry out while in the meter housing, causing the
metertostop. If thishappens,meter should bethoroughly
cleaned (seeinstructionsbelow).
CleaningInstructions:
Runa flushing fluidthroughmeter. For a morethorough
cleaning, disassemble meter per “ASSEMBLY /
DISASSEMBLY”section, “Meter ChamberAssembly”
subsection. Rinseallmetercomponents.Recalibratemeter
followingcalibrationinstructionsabove.
Storage:
Ifmeteristobestoredforaperiodoftime,cleanthoroughly.
Thiswill help protectmeterfrom damage.
TROUBLESHOOTING GUIDE
Counter is reading high or low:
Checkcalibrationandrecalibrateif necessary. Checkforair
inproductandrepairairleaks. Measuringchamberorgears
could be sticking. Correct by cleaning or replacing internal
meteringcomponents.
Shaft seal leakage:
Possible causes are dirty or damaged seals. Correct by
cleaningo-ringandseatareaorreplacing seal.
Gasket leakage:
Correctloosegasketbytighteningjoints. Cleandirtygaskets
andseatarea. Replacedamagedgaskets. Ifcausedbyhigh
pressure,installpressurereliefvalvetoallowhighpressureto
bleed back to tank.
Low flow capacity:
Cleancloggedmeterchamber;cleanorreplacescreensand
filtersin piping.
Meterbody cracks:
Install pressure relief valve to allow high pressure to bleed
back to tank.
Nutatingdiscbreaks:
Avoidflowsurgebyputtingshut-offvalveonoutletofmeter;
placemeterascloseaspossibletopump;keeppipingfullof
liquid.
Procedure for Calibration:
1. Forthemostaccuratecalibration,installthemeterinthe
application.Fill acontainer toa knownvolume withthe
liquidtobemeasured.
2. Ifmeteramountisincorrect,turncalibrationscrew(item
31) counterclockwise for less liquid, or clockwise for
moreliquid.
3. Repeat steps 1 and 2 until calibration is acceptable.
WHENORDERINGREPAIRPARTS,BESURETO
GIVEREPLACEMENTPARTNUMBER,DATEOF
MANUFACTUREANDMETERSERIESNUMBER.
THISWILLENSURETHATTHECORRECT
REPLACEMENTPARTISSUPPLIED.
TOLLFREECUSTOMERCARENUMBER
8006342695
2
FLUIDCOMPATIBILITY
The900 is compatiblewith the followingfluids:
•Diesel Fuel, Gasoline, Kerosene, Mineral Spirits,
Heptane,andHexane
The900isNOTcompatiblewiththefollowingfluids:
•Bleach, Hydrochloric Acid, Ink, Sulfuric Acid, and Salt
Water
If in doubt about compatibility of a specific fluid, contact
supplier of fluid to check for any adverse reactions to the
followingwettedmaterials.
ASSEMBLY&DISASSEMBLY
Meterconsistsofachamberhousing,measuringchamber,
gear train, counter assembly and cover. Meter can be
completelydisassembledwithoutdisturbingpiping.
CounterAssembly
For access to counter assembly, remove reset knob (item
15) by grasping edges and pulling firmly. Remove two
screws (item 14) and lift bezel (item 11) off. Remove two
screws(item12)todetachcounterface(item13). Remove
two screws (item 9) to extract counter (item 10).
Reassemblebyreversingprocedure.
Meter Chamber Assembly
To expose meter chamber assembly, gear train and seal,
remove four screws (item 28). Meter chamber assembly
consistsofupper and lowerchambers,anutatingdiscand
four screws. Meter chamber assembly (item 3) can be
dislodgedbyremovingfourscrews(item5). Reassembleby
reversingprocedure.
If replacement of any components of the meter chamber
assembly is required, the complete assembly must be
replacedduetotheprecisemethodofitsconstruction.This
assuresa properfit anda correctlyoperating chamber.
Gear Train and Seal
To disassemble gear train and seal, remove two screws
(item8)andgearframe(item6). Removeclustergear(item
18), washer (item 19), and shaft (item 17). Remove drive
gear(item24)andwashers(item23)byrotatingandpulling
drivegear. RemoveO-ringseal(item25).
When reassembling seal, lubricate O-ring with oil or
petroleumjellyandreplaceincover.Placewasherondrive
gearshaft. Carefully rotateand push shaftthrough O-ring
andcovertopreventdamagetoO-ring. Shaftmustthenbe
guided into pinion bevel (item 27) if counter has not been
removed. Replaceremaining partsto complete assembly
byreversingdisassemblyprocedure.
REPAIR
Metersneedingrepairshouldbetakentoanauthorizedrepair
shop.Metersmustbethoroughlytriple-rinsedbeforebeing
takenin forrepair.
8.5"
21,6 cm
6.5"
16,5 cm
5.5"
14,0 cm
900KTF8119 Meter Repair Kit, U.S. Gallon (Standard Seals)
(Includes items 2-4, 6, 17-19, 21-25, 27, 30, 32)
900KTF8120 Meter Repair Kit, U.S. Gallon
(Fluorocarbon Seals)
(Includes items 2-4, 6, 17-19, 21-25, 27, 30, 32)
900KTF8121 Meter Repair Kit, Liter (Standard Seals)
(Includes items 2-4, 6, 17-19, 21-25, 27, 30, 32)
900KTF8123 Meter Repair Kit, Liter (Fluorocarbon Seals)
(Includes items 2-4, 6, 17-19, 21-25, 27, 30, 32)
900 SERIES REPAIR PARTS KITS
PRIOR TO SERVICE, ADHERE TO FOLLOWING
INSTRUCTIONS:
Meters must be triple-rinsed and accompanied by a
note indicating the chemicals which have been
pumpedthroughtheunit. Metersnotadheringtothese
specifications may be refused service.
901 METER KIT
901MK300 901 Meter Kit for FR300 Gallon
(Includes 901 meter with 800F4390 1 x 2 Nipple,
900F8082 Meter Flange)
901LMK300 901 Meter Kit for FR300 Liter
(Includes 901 meter with 800F4390 1 x 2 Nipple,
900F8082 Meter Flange)
901NMK300 901 Meter Kit for FR300 Gallon Nickel plated
(Includes 901 meter with 800F4390 1 x 2 Nipple,
900G9478 Nickel plated Meter Flange)
901MK4200 900 Meter Kit for FR4200 Gallon
(Includes 901 meter with 304F7924 1 x 5 Nipple,
1200F6732 1 x 4 Nipple, 304F7885 Elbow,
305F0998 Outlet Fitting)
901LMK4200 900 Meter Kit for FR4200 Liter
(Includes 901 meter with 304F7924 1 x 5 Nipple,
1200F6732 1 x 4 Nipple, 304F7885 Elbow,
305F0998 Outlet Fitting)
901NMK4200 900 Meter Kit for FR4200 Gallon Nickel plated
(Includes 901 meter with 304F7924 1 x 5 Nipple,
1200F6732 1 x 4 Nipple, 304F7885 Elbow,
900G9477 Nickel plated Outlet Fitting)
Ryton Aluminum Stainless Steel
Flourocarbon Buna N Polyester
Teflon® Nickel
ITM. PART
NO. NO. DESCRIPTION QTY..
..
.
3
ITM. PART
NO. NO. DESCRIPTION QTY..
..
.
900 SERIES METER PARTS LIST
900F9118 rev7
1 900F8051 Meter Housing 1
900F8102 Meter Housing-Nickel Plated Opt.
900F8101 Meter Housing-Teflon® Coated Opt.
2 300F7744 Inlet/Outlet Gasket Buna-N 2
300F7788 Inlet/Outlet Gasket Fluorocarbon Opt.
3 900F8087 Meter Chamber Assembly 1
4 900F8067 Cover Gasket Buna-N 1
900F8068 Cover Gasket Fluorocarbon Opt.
5 900G8886 #10-24 x 5/8 TORX PH 4
6 900F8066 Gear Frame 1
7 900F8053 Meter Cover (Includes Item 20) 1
900F8104 Meter Cover - Nickel Plated Opt.
900F8103 Meter Cover - Teflon® Coated Opt.
8 VP1400F8822 #10-24 1/2 TORX 2
9 900F4007 #8-32 x 5/16 PHMS, ACR II, TT 2
10 900F8070 Counter Assembl
y
- U.S. Gallon 1
900F8071 Counter Assembl
y
- Liter Opt.
11 900F8069 Bezel 1
12 35F1397 #4 x 3/8 PHTS 2
13 900F8072 Counter Face 1
900F8089 Counter Face - Litre Opt.
14 900G8504 #8 x 5/8 ZN/PL PH OVAL 2
15 800G8870 Knob 1
17 800F3820 Shaft - Cluster Gear 1
18 800F3841 Cluster Gear - U.S. Gallon 1
800F3843 Cluster Gear - Liter Opt.
19 800F3830 Washer 1
20 900F8063 Driver Pinion Shaft (Included w Item 7) 1
21 900F8065 Retaining Ring 1
22 900F8064 Driver Pinion 1
23 800F3980 Washer 3
24 800F3845 Drive Gear - U.S. Gallon 1
800F3846 Drive Gear - Liter Opt.
25 800F4191 O-Ring (5-106) Fluorocarbon 1
27 800F3959 Pinion Bevel 1
28 700F2810 5/16-18 x 7/8 HHCS 4
29 900F8158 Seal Screw 1
30 800F4449 O-Ring (-012) Fluorocarbon 1
31 900F8160
A
d
j
ustment Screw (Includes Item 32) 1
32 900F8159 O-Ring (-010)(inc. in item 31) Fluorocar 1
33 1200F6721 1/4-20 x 3/4 HHCS (1" meters) 8
34 900F8076 1" Meter Flange 2
900F8106 1" Meter Flange - Nickel Plated Opt.
900F8105 1" Meter Flange - Teflon® Coated Opt.
35 900F8092 1-1/2" Meter Flange Opt.
900F8110 1-1/2" Meter Flange Nickel Plated Opt.
900F8109 1-1/2" Meter Flange Teflon® Coated Opt.
36 900F8091 1/4-20 x 1-1/2 HHCS (1-1/2" meters) Opt.
37 900F8045 Meter Cover Assembly U.S. Gallon 1
900F8047 Meter Cover Assembl
y
-U.S. Gallon Opt.
Nickel Plated
900F8046 Meter Cover Assembl
y
-U.S. Gallon Opt.
Teflon® Coated
900F8048 Meter Cover Assembl
y
- Liter Opt.
900F8050 Meter Cover Assembl
y
- Liter Opt.
Nickel Plated
900F8049 Meter Cover Assembly - Liter Opt.
Teflon® Coated

Other manuals for 900 Series

1

This manual suits for next models

8

Other FILL-RITE Measuring Instrument manuals

FILL-RITE 806C Use and care manual

FILL-RITE

FILL-RITE 806C Use and care manual

FILL-RITE 900CD Series User manual

FILL-RITE

FILL-RITE 900CD Series User manual

FILL-RITE 901 Use and care manual

FILL-RITE

FILL-RITE 901 Use and care manual

FILL-RITE TT10A User manual

FILL-RITE

FILL-RITE TT10A User manual

FILL-RITE 900 Series User manual

FILL-RITE

FILL-RITE 900 Series User manual

FILL-RITE 900DB Use and care manual

FILL-RITE

FILL-RITE 900DB Use and care manual

FILL-RITE FR1118-A10 User manual

FILL-RITE

FILL-RITE FR1118-A10 User manual

FILL-RITE 806C Use and care manual

FILL-RITE

FILL-RITE 806C Use and care manual

FILL-RITE 806C Use and care manual

FILL-RITE

FILL-RITE 806C Use and care manual

FILL-RITE TS20A Guide

FILL-RITE

FILL-RITE TS20A Guide

FILL-RITE TT10A Use and care manual

FILL-RITE

FILL-RITE TT10A Use and care manual

Popular Measuring Instrument manuals by other brands

Hach TPM870 user manual

Hach

Hach TPM870 user manual

Palmer MicroCALL+ user manual

Palmer

Palmer MicroCALL+ user manual

Burster 9163 manual

Burster

Burster 9163 manual

Siemens SITRANS F M MAG 5000 operating instructions

Siemens

Siemens SITRANS F M MAG 5000 operating instructions

Fuji Electric ZSS-8 Series instruction manual

Fuji Electric

Fuji Electric ZSS-8 Series instruction manual

SUPMEA SUP-LDG user manual

SUPMEA

SUPMEA SUP-LDG user manual

GEISMAR Amber operating manual

GEISMAR

GEISMAR Amber operating manual

PRECISION DIGITAL PD213 operating instructions

PRECISION DIGITAL

PRECISION DIGITAL PD213 operating instructions

Keysight Technologies MXA Series Installation note

Keysight Technologies

Keysight Technologies MXA Series Installation note

Laserworld EL-250 RBV manual

Laserworld

Laserworld EL-250 RBV manual

Sigma 980 Operating and maintenance manual

Sigma

Sigma 980 Operating and maintenance manual

Hioki FT6380 instruction manual

Hioki

Hioki FT6380 instruction manual

Tektronix P6243 instructions

Tektronix

Tektronix P6243 instructions

Hilti PX 10 operating instructions

Hilti

Hilti PX 10 operating instructions

ITRON X61 installation instructions

ITRON

ITRON X61 installation instructions

Agilent Technologies B1500A Training manual

Agilent Technologies

Agilent Technologies B1500A Training manual

Selec MFM374 operating instructions

Selec

Selec MFM374 operating instructions

Gill WindSonic user manual

Gill

Gill WindSonic user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.