Fisher Scientific Fisherbrand accumet AB330 User manual

Operations Manual Electrochemistry Meters
68X002128 | Revision 01 | September 2021
accumet AB330 Conductivity
Meter User Manual


Contents
Chapter 1................................................................. 6
Introduction.....................................................6
1.1 Intended Use ........................................6
1.1.1 Intended Use .............................................6
1.1.2 Non-Intended Use .....................................6
1.1.3 Safety ........................................................6
1.1.4 Power Connection .....................................6
1.2 Cleaning Instructions (Meter Only)......... 6
Chapter 2................................................................. 7
Getting Started................................................7
2.1 Electrode Holder Installation..................7
2.2 Universal Power Supply Assembly ........7
2.3 Keypad Functions................................. 8
2.3.1 Setup Menu Scrolling.................................8
2.3.2 Setup Menu Value Changes.......................8
2.3.3 Mode Shortcut Key....................................8
2.3.4 Date and Time View Shortcut Key..............8
2.4 Display Overview ..................................9
2.5 Meter Connections .............................10
2.6 Electrode Information..........................10
2.7 Meter Startup Sequence..................... 10
Chapter 3............................................................... 11
Setup Menu ..................................................11
3.1 Setup Menu Overview.........................11
3.2 Setup Menu Items ..............................12
3.2.1 View Logs................................................12
3.2.1a Viewing the Data Log...........................12
3.2.1b Viewing the Calibration Log..................13
3.2.2 Log Export Type ......................................14
3.2.3 Export All Logs ........................................14
3.2.4 Temperature Calibration or Manual
Temperature Input ............................................14

3.2.4a Temperature Calibration.......................14
3.2.4b Manual Temperature Input...................15
3.2.5 Measure Mode ........................................15
3.2.6 Nominal Cell Constant .............................15
3.2.7 Temperature Compensation Coefficient ...16
3.2.8 Reference Temperature ...........................16
3.2.9 TDS Factor ..............................................17
3.2.10 Read Type .............................................17
3.2.11 Timed Interval ........................................17
3.2.12 Calibration Due Alarm ............................18
3.2.13 Temperature Units .................................18
3.2.14 Set Date Format ....................................18
3.2.15 Set Date Value .......................................19
3.2.16 Set Time Value.......................................20
3.2.17 Audio Mode ...........................................20
3.2.18 Sleep Mode ...........................................20
3.2.19 Clear Data .............................................21
3.2.20 Factory Reset ........................................21
Chapter 4............................................................... 22
Calibration.....................................................22
4.1 Conductivity Calibration ......................22
4.1.1 Conductivity Calibration Overview ............22
4.1.2 Conductivity Calibration Procedure ..........23
4.1.3 Conductivity Calibration Example .............23
4.2 TDS Calibration................................... 24
4.3 Salinity Calibration .............................. 24
4.4 Temperature Calibration...................... 25
Chapter 5............................................................... 26
Measurements .............................................. 26
5.1 Read Types ........................................26
5.1.1 Continuous ..............................................26
5.1.2 Auto-Read ...............................................26
5.1.3 Timed ......................................................26
5.2 Conductivity, Salinity and TDS
Measurements..........................................26
5.3 Calculating TDS Factor .......................26

5.4 Calculating Temperature Coefficients.. 27
5.5 Data Viewing.......................................27
5.6 Data Exporting.................................... 27
5.6.1 Remote Commands.................................27
Chapter 6............................................................... 29
Troubleshooting ............................................ 29
6.1 Meter Troubleshooting........................29
6.2 Measurement Troubleshooting............29
6.3 Technical Support...............................29
Chapter 7............................................................... 30
Meter Info......................................................30
7.1 Meter Specifications.......................... 30
7.2 Meter Dimensions...............................31
7.3 Warranty............................................. 31
7.3.1 Return of Items ........................................31
Chapter 8............................................................... 32
Regulatory Compliance ................................. 32
8.1 European Union.................................. 32
8.2 Product Safety....................................32
8.3 Electromagnetic Compatibility............. 32
8.3.1 FCC Statement (USA) ..............................32
8.4 Canadian ISED IC Notice ....................32
8.5 Environmental Compliance .................32
8.5.1 REACH - Europe ...................................32
8.5.2 RoHS - Europe ......................................32
8.6 WEEE Compliance..............................33
Chapter 9............................................................... 34
User Notes.................................................... 34
Chapter 10............................................................. 36
Ordering Info .................................................36

6 Chapter 1 | Introduction
Chapter 1
Introduction
Thank you for purchasing the Fisherbrand™accumet™
AB330 conductivity bench meter. The accumet AB300
meters are designed to provide the important functions
needed for routine measurements with simplified,
convenient, easy-to-use functionality.
The accumet AB330 conductivity meters are capable of
measuring conductivity, salinity, TDS and temperature and
offer a large, backlit display for clear viewing with small
footprint to conserve laboratory bench space.
Ensure important data is preserved with a 500-point data
log with date/time stamp that can also be exported to a
printer or computer using the meter communication port.
All accumet AB330 conductivity meters include a meter-
attached electrode holder and universal 100-240V power
adapter with wall plugs for US/Japan, Euro, UK/Singapore,
Australia/New Zealand and China. Additional meter
accessories can be viewed at www.fishersci.com.
1.1 Intended Use
1.1.1 Intended Use
This device is a bench laboratory meter intended for use in
a typical, indoor, controlled, laboratory environment. This
device should only be used for analytical testing in
accordance with these instructions.
1.1.2 Non-Intended Use
This device is not a Medical Device. It is not intended to be
used to diagnose, treat, cure, or prevent disease.
1.1.3 Safety
CAUTION: This symbol, in the context of
a CAUTION, indicates a potentially
hazardous situation which if not avoided
could result in minor to moderate injury or
damage to the equipment.
Note: This symbol, in the context of a
Note, indicates to follow the manufacturer
specified instructions, notes and
requirements set out in the instruction
manuals.
CAUTION: Before using or maintaining
this product, please be sure to read the
manual carefully. Failure to follow these
instructions may cause the product to
malfunction.
CAUTION: Use this product only in the
way described in the product literature
and in this manual. Before using it, verify
that this product is suitable for its
intended use. If the product is used in a
manner not specified by the manufacturer,
the protection provided by the product
may be impaired.
CAUTION: Do not modify system
components. Use OEM exact
replacement equipment or parts. Before
use, confirm that the product has not
been altered in any way.
Note: Unauthorized repair of your meter
will invalidate the warranty. Contact
accumet Technical Support at 1-888-358-
4706 or accumet@fishersci.com for
additional information.
1.1.4 Power Connection
See the Universal Power Supply Assembly section for
complete instructions on assembling and installing the
included meter power supply.
Note: Position the meter so the operator
can access the power connection and
unplug the power adapter in the event of
a fault so the hazard of electrocution can
be removed.
1.2 Cleaning Instructions
(Meter Only)
Wipe the exterior surfaces (except the display panel) using
a lint free cloth dampened in clean water.
Wipe the display panel with a clean microfiber cloth
dampened with clean water, taking caution to wipe lightly
to avoid scratching the meter display.
The meter keypad can also be cleaned using a clean
microfiber cloth dampened with IPA (isopropyl alcohol) for
disinfection purposes.

accumet AB330 Conductivity Meter 7
Chapter 2
Getting Started
2.1 Electrode Holder
Installation
The meter-attached electrode holder (Catalog Number 13-
637-671) can be installed on either side of the meter.
Unpack the base plate and stand with electrode
holder from the meter box.
Turn the meter over, with the meter display facing
down, on a clean dry surface.
Identify the side of the meter that the holder will be
installed on and remove the screw between the
circles on that side of the meter.
Align the base plate of the holder with the circles
on the meter.
Replace the screw from step 3 to attach the base
plate to the meter.
Turn the meter over, with meter display facing up.
Insert the stand with electrode holder into the
metal post on the base plate.
Place the electrodes into the electrode holder.
2.2 Universal Power
Supply Assembly
A universal power adapter (Catalog Number 13-637-010)
with US/Japan, Euro, UK/Singapore, Australia/New
Zealand and China wall plug plates is included with the
meter. This universal power adapter is specifically for use
with this meter. Use of other power adapters can damage
the meter and will void the warranty.
Unpack the power supply provided with the meter.
Select the appropriate wall plug plate for the
power outlet that will be used.
Slide the appropriate wall plug plate into the
groove on the back of the power adapter.
Connect the assembled power adapter to a power
outlet and meter input.
a. To connect the power adapter with the meter,
insert the power adapter plug and twist to lock
the connection.
Note: The power adapter plug has two prongs that allow
the power adapter to be locked onto the meter. These two
prongs must be properly aligned when connecting it to
ensure it is fully connected and locked onto the meter.
You are now ready to turn on the meter.
CAUTION: Use of a surge protector or
uninterrupted power supply (UPS) is
recommended, as an unintended power
surge of electricity to the meter may
damage the meter and void the warranty.

8 Chapter 2 | Getting Started
2.3 Keypad Functions
Key
Function
In Continuous measure mode, press to hold a
reading on the display and then press again
to release a reading on the display.
In Auto-Read measure mode, press to start a
new reading.
Press to escape the Setup Menu and
Calibration mode without saving changes.
In Continuous measure mode, press to save
the reading to the data log and export to an
external device if one is connected.
Press to enter the Setup Menu.
Press to exit the Setup Menu.
Press to enter the Calibration mode.
Press while in the Calibration mode to save
and end the calibration when performing a
one to four point conductivity/TDS calibration.
Press to scroll through menu items in the
Setup Menu; the list is cyclical.
In the Measure mode, press and hold (long
press) for three seconds as a shortcut key to
change the measure type from conductivity to
TDS to salinity.
Press to change a setting in the Setup Menu.
Press to scroll through a list or change a
numeric value.
In the Calibration mode, press to manually
change the calibration value.
In the Measure mode, press and hold (long
press) for three seconds as a shortcut key to
view current date and time setting.
Press to save a setting in the Setup Menu.
Press to accept a calibration point in the
Calibration mode.
Press to power on the meter. When the meter
is on, press to turn the backlight off or on.
When the meter is on, press and hold for
three seconds to power off the meter.
2.3.1 Setup Menu Scrolling
To scroll through the Setup Menu list:
•Press the “Menu” key to enter the Setup Menu.
•The first Setup Menu item will always be show.
•Press the key to scroll to the second item or
press the key to scroll to the last item.
•The item list is cyclical and will scroll from the last
item to the first item as the key is pressed.
2.3.2 Setup Menu Value Changes
For all numeric value changes, press the or key to
edit a value:
•Press the key once to increase the value by one
least significant digit/unit.
•Press the key once to decrease the value by one
least significant digit/unit.
•Press and hold the key to quickly increase the
numeric value.
•Press and hold the key to quickly decrease the
numeric value.
2.3.3 Mode Shortcut Key
In the main measure mode, press and hold (long press) the
or key for approximately three seconds to change the
main measure mode to conductivity, salinity or TDS.
2.3.4 Date and Time View
Shortcut Key
In the main measure mode, press and hold (long press) the
or key for approximately three seconds to view the
current date and time settings.

accumet AB330 Conductivity Meter 9
2.4 Display Overview
Icon
Description
Main Measurement Field:
Displays conductivity, salinity
or TDS readings in Measure
mode.
2nd Measurement Field:
Displays nominal cell constant
in Measure mode.
3rd Measurement Field:
Displays temperature value as
°C or °F in Measure mode.
4th Measurement Field:
Displays linear temperature
compensation coefficient
value in Measure mode.
Press to scroll Icon: On-
screen text prompt shown
when keys can be used to
scroll through a list; for
example, scroll through list of
Setup Menu options.
Press to change Icon:
On-screen text prompt shown
when keys can be used to
change a value; for example,
to change the temperature
value in the Setup Menu.
Press CAL to save & end
Icon:
On-screen text prompt
shown when in calibration
mode and calibration can be
saved and ended, when one
to four points are complete.
Press Enter to accept Icon:
On-screen text prompt shown
when changes must be saved
by pressing the “Enter” key on
the keypad.
Icon
Description
Calibration Icon: Indicates meter is in
the calibration mode; also shown with
the Setup Menu Icon and Memory Icon
when viewing the calibration log.
Setup Menu Icon: Indicates meter is in
the Setup Menu mode.
Measure Icon: Indicates meter is in the
Measure mode.
Memory Icon: Indicates meter is in the
Setup Menu and viewing the data log or
calibration log.
Log Saved Icon: Indicates meter is
actively saving data to the data log or
calibration log.
Continuous Icon: Indicates the current
meter read type is Continuous and
measurement values are continuously
updated on the display.
Stopwatch Icon: Indicates meter
measurement stability criteria for current
measurement is stabilizing/not ready
and meter is actively reading.
Checkmark Icon: Indicates meter
measurement stability criterial for current
measurement is stable/ready and meter
has determined the measurement is
complete.
Timed Icon: Indicates the current meter
read type is Timed, measurement values
are continuously updated on the display
and logged/exported according to the
selected time intervals.
Auto-Read Icon: Indicates the current
meter read type is Auto-Read,
measurement value is updated on the
display until stable and then
measurement is logged/exported and
locked on display until the “Read” key is
pressed again.
Hold Icon: Indicates meter is in
Continuous read type, the “Hold” key
has been pressed and the measurement
is locked on the display until the “Hold”
key is pressed again.
MAN Icon: Indicates no temperature
probe is connected to the meter, use
the Setup Menu to enter the sample
temperature.
ATC Icon: Indicates a temperature
probe is connected to the meter and
actively sending temperature readings.

10 Chapter 2 | Getting Started
Icon
Description
Computer Export Icon: Indicates meter
is actively exporting measurement data
or calibration data to a computer.
Printer Export Icon: Indicates meter is
actively exporting measurement data or
calibration data to a printer.
2.5 Meter Connections
Input
Function
Power
Universal power supply
GND Ground
Export Connection to printer or computer
Conductivity Connection to conductivity electrode
2.6 Electrode Information
The accumet AB330 conductivity meter is compatible with
2-cell and 4-cell conductivity electrodes with a DIN
connector. The accumet AB330 conductivity meter can be
purchased in a meter only configuration (purchase
electrode separately) or kitted with a conductivity
electrode.
Refer to the electrode instructions for specific care and
maintenance procedures. The following are general
recommendations for electrode preparation.
Remove the packaging from the electrode and
save for storage.
Rinse the electrode using distilled water.
Connect the electrode to the meter.
2.7 Meter Startup
Sequence
When the meter is powered on, it displays a set of startup
screens, starting with all segments lit screen, followed by
the meter info screen and then the meter self-test screen.
The meter info screen shows the meter model number and
current software revision.
After the meter info screen, the meter performs a series of
internal self-tests to verify that the meter is operating
correctly and then the self-test results are shown.
Finally, the meter proceeds to the main measure mode.

accumet AB330 Conductivity Meter 11
Chapter 3
Setup Menu
3.1 Setup Menu Overview
Use the meter Setup Menu to customize meter settings.
In the Measure Mode, press the “Menu” key to
access the meter Setup Menu.
The first Setup Menu item (View Logs, Data Log)
will be shown. To scroll through the Setup Menu
list, press the or key.
a. Press the key to scroll to the second item.
b. Press the key to scroll to the last item.
c. The list is cyclical, so continue to press the
key to scroll from the last item to the first item.
To change a setting within a Setup Menu item,
press the or key.
a. For numeric value changes:
i. Press the key once to increase the value
by one least significant digit/unit.
ii. Press the key once to decrease the
value by one least significant digit/unit.
iii. Press and hold the key to quickly
increase the value.
iv. Press and hold the key to quickly
decrease the value.
Once a setting is changed, press the “Enter” key
to save the change.
When viewing data logs or calibration logs, press
the “Menu” key to go back to the main Setup
Menu list.
Press the “Read” key at any time to exit the Setup
Menu and return to the main measure mode.
#
Setup Menu Item
Description
1 View Logs
View up to 500 data log points
and active conductivity, salinity,
TDS and temperature
calibrations
2 Log Export Type
Set the export type as computer
or printer
3 Export All Logs
Send all saved data logs and
calibration logs to the selected
export device
4
Temperature
Calibration
or
Manual
Temperature
Input
When an ATC probe is
connected, use the Temp.
Calibration menu to perform a
temperature offset calibration
When no ATC probe is
connected, use the Manual
Temp. Input menu to enter the
sample temperature value
5 Measure Mode
Set the main measure mode to
conductivity, salinity or TDS
6
Nominal Cell
Constant
Set the nominal cell constant for
the conductivity electrode in use
7
Temperature
Compensation
Coefficient
Set temperature compensation
off or enter the value of the linear
coefficient value used for
temperature compensation
8
Reference
Temperature
Set the reference temperature
for temperature compensation
9
TDS Factor
Enter the TDS factor value
10 Read Type
Set the read type to define how
measurements are performed
and when measurements are
saved and exported
11 Timed Interval
When Timed is set as the Read
Type, set the time interval to be
used to automatically save and
export measurements
12 Calibration Due
Alarm
Set the calibration due alarm
interval; an alarm is triggered if a
calibration is not performed
within the specified time interval
13
Temp. Units
Set the temp. units as °C or °F
14 Set Date Format
Set the date format as month-
day-year (MM.DD.YYYY) or day-
month-year (DD.MM.YYYY)
15
Set Date Value
Set the day, month and year
16
Set Time Value
Set the time in AM/PM format
17
Audio Mode
Set the audible beep on or off
18
Sleep Mode
Set the sleep mode on or off
Setup Menu Item
Setting Options for
Setup Menu Item

12 Chapter 3 | Setup Menu
#
Setup Menu Item
Description
19 Clear Data
Erase all data logs or erase all
calibration logs
20 Factory Reset
Erase all data logs, calibration
logs and settings and return the
meter to its factory default state.
3.2 Setup Menu Items
3.2.1 View Logs
View up to 500 data log points and active conductivity,
salinity, TDS and temperature calibrations.
When the “Menu” key is pressed, the View Logs setup
menu item is always shown first.
Press the or key to scroll through the data
and calibration log options.
Press key
Press key
Press key
Press key
Press key
Press the “Enter” key to view the display log.
3.2.1a Viewing the Data Log
Press the “Enter” key to view the display log.
For the data log, press the or key to scroll
through the data log points. Scroll through data
log points when viewing the date/time screens or
the measurement data screens.
Press key
For the data log, press the or key to change
the view from the date/time screen to the
measurement data screen.

accumet AB330 Conductivity Meter 13
Press or key
If desired, press the “Log/Export” key to export the
individual data log point being viewed.
Press the “Menu” key to return to the main Setup
Menu list.
3.2.1b Viewing the Calibration Log
Press the “Enter” key to view the display log.
For the calibration log, press the or key to
change the view from the date/time screen to the
calibration data screen.
Press or key
For the calibration log, press the or key to
scroll through the calibration log data.
Press key
Press key
Press key
Press key
If desired, press the “Log/Export” key to export the
individual calibration log being viewed.
Press the “Menu” key to return to the main Setup
Menu list.

14 Chapter 3 | Setup Menu
3.2.2 Log Export Type
Set the log export type as computer or printer. If computer
is selected, the logs are exported in CSV format. If printer
is selected, the logs are exported in list format.
Press the or key to scroll through the Setup
Menu list until the Log Export Type item is shown.
Press the or key to change the setting to
computer or printer.
Press the “Enter” key to save the change.
3.2.3 Export All Logs
Send all saved data logs and calibration logs to the
selected export device.
Press the or key to scroll through the Setup
Menu list until the Export All Logs item is shown.
Press the “Enter” key to export all saved logs to
the selected export device.
3.2.4 Temperature Calibration or
Manual Temperature Input
3.2.4a Temperature Calibration
When an ATC probe is connected, use the Temperature
Calibration menu to perform a one-point temperature offset
calibration, up to ±5.0 °C.
Press the or key to scroll through the Setup
Menu list until the Temp. Calibration item is shown.
The temperature value will blink, indicating the
reading from the ATC probe is being measured.
Once the reading is stable, the value will stop
blinking and the checkmark icon will be shown.
Press the or key to adjust the temperature
value in the 3rd field. The temperature offset value
will be shown in the 2nd field.
a. Press the key to increase the value by 0.1°.
Press and hold the key to quickly increase.
b. Press the key to decrease the value by 0.1°.
Press and hold the key to quickly decrease.
Press the “Enter” key to save the change.

accumet AB330 Conductivity Meter 15
3.2.4b Manual Temperature Input
When no ATC probe is connected, use the Manual
Temperature Input menu to enter the sample temperature
value, from -5.0 °C to 105.0 °C.
Press the or key to scroll through the Setup
Menu list until the Manual Temperature Input item
is shown.
Press the or key to adjust the temperature
value.
a. Press the key to increase the value by 0.1°.
Press and hold the key to quickly increase.
b. Press the key to decrease the value by 0.1°.
Press and hold the key to quickly decrease.
Press the “Enter” key to save the change.
3.2.5 Measure Mode
Set the measure mode to conductivity, salinity or TDS.
Press the or key to scroll through the Setup
Menu list until the Measure Mode item is shown.
Press the or key to set the measure mode.
Press the “Enter” key to save the change.
3.2.6 Nominal Cell Constant
Set the nominal value for the cell constant (K) of the
conductivity electrode being used.
The entered nominal cell constant (K) is used to determine
conductivity, salinity and TDS measurement values from
the conductance reading. Once a calibration is performed,
the entered nominal cell constant (K) multiplied by the
calculated calibration factor is used to determine
conductivity, salinity and TDS measurement values from
the conductance reading.
The entered nominal cell constant (K) is also used for
automatic conductivity standard recognition during
conductivity calibrations.
Press the or key to scroll through the Setup
Menu list until the Nominal Cell Constant item is
shown.
Press the or key to set the cell constant (K)
value.

16 Chapter 3 | Setup Menu
a. Press the key to increase the value by
0.001. Press and hold the key to quickly
increase.
b. Press the key to decrease the value by
0.001. Press and hold the key to quickly
decrease.
Press the “Enter” key to save the change.
3.2.7 Temperature Compensation
Coefficient
Set temperature compensation off or enter the value of the
linear coefficient value for temperature compensation to
apply a constant percent correction factor to every degree
change in temperature.
To set the temperature compensation off, set the value to
0.0 %/°C.
To set a linear temperature compensation value, enter a
value of 0.0 to 10.0 %/°C.
Press the or key to scroll through the Setup
Menu list until the Temperature Compensation
Coefficient item is shown.
Press the or key to set the coefficient value.
a. Press the key to increase the value by 0.01.
Press and hold the key to quickly increase.
b. Press the key to decrease the value by 0.01.
Press and hold the key to quickly decrease.
Press the “Enter” key to save the change.
3.2.8 Reference Temperature
Set the reference temperature for temperature
compensation to 15°C, 20°C, 25°C or 30°C.
Press the or key to scroll through the Setup
Menu list until the Reference Temperature item is
shown.
Press the or key to set the reference
temperature value.
Press the “Enter” key to save the change.

accumet AB330 Conductivity Meter 17
3.2.9 TDS Factor
Enter the TDS factor value used to calculate the TDS (total
dissolved solids) value shown in the main measure mode.
Press the or key to scroll through the Setup
Menu list until the TDS Factor item is shown.
Press the or key to set the TDF factor value.
a. Press the key to increase the value by 0.01.
Press and hold the key to quickly increase.
b. Press the key to decrease the value by 0.01.
Press and hold the key to quickly decrease.
Press the “Enter” key to save the change.
3.2.10 Read Type
Set the Read Type as Continuous, Auto-Read or Timed to
define how measurements are performed and when
measurements are saved and exported.
Continuous:
In the Continuous measure mode, measurements are
continuously updated on the display and the stopwatch
(stabilizing) or checkmark (stable) icon indicates the
measurement stability status. Press the “Log/Export” key
to save a measurement to the data log and export.
Auto-Read:
In the Auto-Read measure mode, press the “Read” key to
start a measurement. When the measurement is stable, the
checkmark (stable) icon is shown and the measurement is
locked on the display until the “Read” key is pressed again.
The stable measurement is automatically saved to the data
log and exported.
Timed:
In the Timed measure mode, the measurement values are
continuously updated on the display. Measurements are
automatically saved to the data log and exported at the
selected time intervals, from 5 seconds to 60 minutes, the
entire time the meter is in the measurement mode.
Press the or key to scroll through the Setup
Menu list until the Read Type item is shown.
Press the or key to set the read type.
Press the “Enter” key to save the change.
3.2.11 Timed Interval
When Timed is set as the Read Type, set the time interval
from 5 seconds to 60 minutes. This time interval is used to
automatically save and export measurements.
Press the or key to scroll through the Setup
Menu list until the Timed Interval item is shown.

18 Chapter 3 | Setup Menu
Press the or key to adjust the time value.
a. Press the key to increase the time value by
one second. Press and hold the key to
quickly increase.
b. Press the key to decrease the time value by
one second. Press and hold the key to
quickly decrease.
Press the “Enter” key to save the change.
3.2.12 Calibration Due Alarm
Set the calibration due alarm interval from 0 hours (off) to
60 hours. An alarm is triggered if a calibration is not
performed within the specified time interval.
Press the or key to scroll through the Setup
Menu list until the Calibration Due Alarm item is
shown.
Press the or key to adjust the alarm value.
a. Press the key to increase the alarm value by
one hour. Press and hold the key to quickly
increase.
b. Press the key to decrease the alarm value
by one hour. Press and hold the key to
quickly decrease.
Press the “Enter” key to save the change.
3.2.13 Temperature Units
Set the temperature units as °C (Celsius) or °F (Fahrenheit).
Press the or key to scroll through the Setup
Menu list until the Temperature Units item is
shown.
Press the or key to set the temperature
units.
Press the “Enter” key to save the change.
3.2.14 Set Date Format

accumet AB330 Conductivity Meter 19
Set the date format as month-day-year (MM.DD.YYYY) or
day-month-year (DD.MM.YYYY).
Press the or key to scroll through the Setup
Menu list until the Set Date Format item is shown.
Press the or key to set the date format.
Press the “Enter” key to save the change.
3.2.15 Set Date Value
Set the month, day and year values.
The date format used for this Setup Menu item will match
the setting in the Set Date Format item. For this example,
the month-day-year (MM.DD.YYYY) format is shown.
Press the or key to scroll through the Setup
Menu list until the Set Date Value item is shown.
The month value will flash. Press the or key
to adjust the month.
a. Press the key to increase by one month.
Press and hold the key to quickly increase.
b. Press the key to decrease by one month.
Press and hold the key to quickly decrease.
Press the “Enter” key to save the change.
The day value will flash. Press the or key to
adjust the day.
a. Press the key to increase by one day. Press
and hold the key to quickly increase.
b. Press the key to decrease by one day.
Press and hold the key to quickly decrease.
Press the “Enter” key to save the change.
The year value will flash. Press the or key to
adjust the year.
a. Press the key to increase by one year. Press
and hold the key to quickly increase.
b. Press the key to decrease by one year.
Press and hold the key to quickly decrease.
Press the “Enter” key to save the change.

20 Chapter 3 | Setup Menu
3.2.16 Set Time Value
Set the time in AM/PM format.
Press the or key to scroll through the Setup
Menu list until the Set Time Value item is shown.
The AM/PM setting will flash. Press the or
key to set the AM or PM time.
Press the “Enter” key to save the change.
The hour value will flash. Press the or key to
adjust the hours.
a. Press the key to increase by one hour.
Press and hold the key to quickly increase.
b. Press the key to decrease by one hour.
Press and hold the key to quickly decrease.
Press the “Enter” key to save the change.
The minutes value will flash. Press the or key
to adjust the minutes.
a. Press the key to increase by one minute.
Press and hold the key to quickly increase.
b. Press the key to decrease by one minute.
Press and hold the key to quickly decrease.
Press the “Enter” key to save the change.
3.2.17 Audio Mode
Set the audible beep on or off. The audible beep is used
each time a meter key is pressed, when a measurement
stabilizes and when an alarm is triggered.
Press the or key to scroll through the Setup
Menu list until the Audio Mode item is shown.
Press the or key to set the audio mode.
Press the “Enter” key to save the change.
3.2.18 Sleep Mode
Set the sleep mode on or off. When the meter sleep mode
is on, the meter will enter sleep mode when no keys are
pressed for 20 minutes. Once the meter is in sleep mode,
press the “Power” key to resume using the meter. Make
sure sleep mode is off when taking timed measurements.
Press the or key to scroll through the Setup
Menu list until the Sleep Mode item is shown.
Press the or key to set the sleep mode.
Press the “Enter” key to save the change.
Table of contents
Other Fisher Scientific Measuring Instrument manuals

Fisher Scientific
Fisher Scientific Isotemp 228 User manual

Fisher Scientific
Fisher Scientific accumet AB200 User manual

Fisher Scientific
Fisher Scientific Accumet Basic AB30 User manual

Fisher Scientific
Fisher Scientific accumet AP 110 accumet AP 115accumet AP 125 User manual

Fisher Scientific
Fisher Scientific accumet AE150 User manual

Fisher Scientific
Fisher Scientific XL15 User manual

Fisher Scientific
Fisher Scientific accumet XL150accumet XL200accumet XL250accumet XL500accumet... User manual