
● The Fitbit Flow has not been tested for MRI scanner compatibility and should not
be used in the vicinity of an MRI scanner.
● Do not pull on the oxygen tube. Doing so could dislodge or otherwise move the
manual resuscitator within the enclosure.
● The Fitbit Flow has been tested for 24 hours of continuous use. If using longer
than 24 hours, check for signs of wear and tear, particularly on the manual
resuscitator.
● The Fitbit Flow has not been characterized under leakage test conditions. If there
is significant patient circuit leakage, the device may under-deliver tidal volume to
the patient.
● Provide sufficient space around the Fitbit Flow for adequate air flow. Do not
position next to objects that may block the flow of air. Failure to do so may cause
the equipment to overheat, thereby interfering with patient ventilation.
● Do not block the gas intake port or emergency intake port. This may interfere with
patient ventilation.
● Do not add any attachments or accessories to the Fitbit Flow that contravene the
instructions for use of the Fitbit Flow, as the Fitbit Flow might not function
correctly, leading to the risk of patient death or serious deterioration of health.
● The Fitbit Flow shall not be used in a hyperbaric chamber. Such use might cause
the Fitbit Flow to not function correctly, causing patient death or serious
deterioration of health.
● The Fitbit Flow does not support the delivery of gas anesthesia agents, and is not
intended for use with such medications. Such use might cause the Fitbit Flow to
not function correctly, causing patient death or serious deterioration of health.
● The Fitbit Flow shall not be used with inlet gases other than medical air or oxygen,
which are not specified for use (e.g. helium or mixtures with helium). Such use
might cause the Fitbit Flow to not function correctly, causing patient death or
serious deterioration of health.
● It is the responsibility of the clinician to ensure that the oxygen source is
appropriately selected with respect to the range of pressure, flow rate and oxygen
concentration.
● Breathing system filters and heat and moisture exchangers may require more
frequent replacement to prevent increased resistance and blockage.
● When moving the Fitbit Flow, disconnect it from the patient and turn off the
device.
● Lock the stand’s caster wheels whenever the Fitbit Flow is in use.
● Do not attempt to move the Fitbit Flow when it is connected to a patient.
● Do not cover or obstruct the alarm button. The operator must be able to see the
alarm button when the device is connected to the patient, particularly in noisy
environments where the operator may not hear the alarm.
● Frequent alarms are possible. When an alarm is triggered, always check the cause
and respond accordingly. Failure to respond to alarms can result in serious harm
or death.
● Failure to follow the cleaning and maintenance instructions could result in
equipment damage and potentially harm the patient.
● Do not attempt to clean or re-use single use accessories.
● Do not reuse single-use components. Doing so increases the risk of
cross-contaminations between patients.
For use only while FDA’s Emergency Use Authorization is in effect.
10