Fluke Victoreen 190 User manual

Victoreen 190
Survey and Count Rate Meter
Operators Manual
March 2005
Manual No. 190001 Rev. 8
©2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A.
All product names are trademarks of their respective companies

Fluke Biomedical
Radiation Management Services
6045 Cochran Road
Cleveland, Ohio 44139
440.498.2564
120 Andrews Road
Hicksville, New York 11801
516.870.0100
www.flukebiomedical.com/rms

i
Table of Contents
Section 1: General Information................................................................................... 1-1
1.1 Product Description ..................................................................................... 1-1
1.2 Specifications............................................................................................... 1-2
Section 2: Theory of Operation................................................................................... 2-1
2.1 General Circuit Description .......................................................................... 2-1
2.2 Operational Considerations and Applications .............................................. 2-1
Section 3: Operation.................................................................................................... 3-1
3.1 Unpacking and Inspection............................................................................ 3-1
3.2 Storage ........................................................................................................ 3-1
3.3 Battery Installation/Replacement ................................................................. 3-1
3.4 Probe Holder Installation.............................................................................. 3-2
3.5 Adapter Module Installation/Replacement ................................................... 3-3
3.6 Power........................................................................................................... 3-4
3.7 Readout ....................................................................................................... 3-4
3.8 Internal Controls .......................................................................................... 3-4
3.9 High Voltage Adjustment ............................................................................. 3-6
3.10 Detector Sensitivity Adjustment ................................................................... 3-8
3.11 Jam Current Adjustment .............................................................................. 3-8
3.12 Changing the Response Time (RESP TIME)............................................... 3-8
3.13 Resetting the Integrate Value (RESP TIME)................................................ 3-9
3.14 Changing the Operation Mode (MODE)....................................................... 3-9
3.15 External Controls ....................................................................................... 3-11
3.16 Operational Checkout ................................................................................ 3-11
Section 4: Maintenance ............................................................................................... 4-1
4.1 Maintenance ................................................................................................ 4-1
Section 5: Calibration .......................................................................................5-1
5.1 Calibration.................................................................................................... 5-1
Section A: Appendix A.................................................................................................A-1
A-1 Applicable Drawings and Bills of Materials .................................................. A-1
Section B: Appendix B.................................................................................................B-1
B-1 Supplemental Data for the Model 190AC & 190F ........................................ B-1
Section C: Appendix C.................................................................................................C-1
C-1 Adapter Bracket Assembly Kit Installation Instructions ................................C-1

(Blank page)

General Information
Product Description 1
1-1
Section 1
General Information
1.1 Product Description
Model 190 is a portable, battery-operated general-purpose survey meter for
use with most scintillation and Geiger-Mueller (GM) radiation detector,
proportional counting probes, and neutron probes. The design combines
CMOS microprocessor technology, automatic range changing, and a Liquid
Crystal Display (LCD) into an easy to use instrument. Refer to Figure 1-1 for
a general view.
The Model 190 is housed in a lightweight, plastic case designed for portability
and durability. Measurement results are indicated on the front panel LCD.
The display includes:
•A 51 element analog bar graph near the top of the display to provide
real time response.
•A scale multiplier, located below the bargraph relates the analog bargraph full scale to the units
displayed.
•Detector range bargraph consisting of five horizontal bars that indicate the quantity of the dynamic
range of the detector in use.
•A 16 character alphanumeric display that (1) provides a digitized average of the bargraph value or
integrated reading with elapsed time, and (2) displays operational information (e.g., such as
response time changes or changes in the units of the display).
•A backlight, activated by either of two front panel switches that provides for operation in dark
environments.
The analog bargraph emulates an analog meter movement and is especially useful in survey applications.
The digitized value is intended for quantitative analysis and data logging applications where an
interpretation of the average reading of the analog bargraph is required.
The following controls are available from the front panel:
•Light (2) activates the LCD backlight.
•Mode changes the display units.
•Log, if activated, logs currently displayed data or changes current location identifier, depending on
the currently active logging mode (i.e., timed or manual logging).
The optional Infrared Communicator, Model 190-
1A, is required to activate the Log function.
NOTE
Figure 1-1. Model 190 Top

Victoreen 190
Operators Manual
1-2
•Rate/Integ toggles the alphanumeric display between the digitized rate value and the integrate
value.
•Resp Time changes the response time on the first range or resets the integrate value and elapsed
time (if pressed and held for three seconds).
•Audio toggles the click output associated with counting events in the detector on/off and
acknowledges alarms.
•On/Off turns instrument power on/off when permitted by the microprocessor (i.e., when Logging is
inactive or all logged data has been uploaded to a computer).
The batteries (four) are housed in a separate compartment accessed from the instrument back panel. All
four 9-volt alkaline batteries are required to operate the unit for approximately 200 hours. Visual and
audible low battery indications are provided whenever battery power is low (four minute intervals for the
audible and continuous for the visual after a battery test).
Communication with the Model 190 is accomplished through an optional infrared two-way adapter, Model
190-1A, which mounts at the top of the instrument. The adapter provides positive keying with the
instrument for stable connection and communication.
1.2 Specifications
Radiation Detected: Alpha, Beta, Gamma, or X-ray depending on detector used.
Operating Rang: Rate Mode: CPM: 0-100; 0-1000; 0-10,000; 0-100,000; 0-1,000,000, and 0-
10,000,000. CPS: 0-1, 0-10, 0-100, 0-1000, 0-10,000, 0-100,000, and 0-166,666. μR/hr, mR/hr, R/hr,
μSv/hr, mSv/hr, Gy/hr, dpm, Bq/cm2, & μCi/cm2: Integrate Mode: Counts: 0-100,000,000; displayed
as 4 digits up to 9999 counts, 3 digits plus a K (for thousand) or an M (for Million) at higher counts.
Seconds: 0-65535; displayed as 4 digits up to 9999 seconds, 3 digits in minutes and tenths of minutes
above 9999 seconds. μR, mR, R, cts, d, μSv, mSv, Bq, & μCi: depends on conversion factor of the
probe used.
Accuracy: Within 10% of reading between 10% and 100% of full-scale indication on any range, exclusive
of energy response.
Ranges: Six (6) decades are available for calibration; the number of usable decades is detector
dependent.
Deadtime Correction: Provided above 1000 CPM.
Jam Detection: Adjustable level for determining jam condition; provided in probe adapter modules.
Alarm: Can be set on any decade at 10% to 100% (in increments of 10%) of full scale.
Detector Probes: Part No.: 489-4, 489-35, 489-50, 489-55, 489-60, 489-110C, 491-30, 491-40, 493-50,
425-110A, 425-200, 489-120, 489-200, RP-1 & PRN
The Model 190 may be ordered with a BNC probe
adapter (P/N 190070) to allow for compatibility with
other manufacturer’s probes.
Warm-up Time: 15 seconds (diagnostic testing).
NOTE

General Information
Specifications 1
1-3
Response Time(Bargraph): Selectable; 24, 12, 6, and 3 seconds.
24 second response time selected:
Scale Range (CPM) Response (seconds)
1 0-100 24
2 0-1000 12
3 0-10K 3
4 0-100K 1.5
5 0-1M 1.5
6 0-10M 0.7
12 second response time selected:
Scale Range (CPM) Response (seconds)
1 0-100 12
2 0-1000 6
3 0-10K 3
4 0-100K 1.5
5 0-1M 1.5
6 0-10M 0.7
6 second response time selected:
Scale Range (CPM) Response (seconds)
1 0-100 6
2 0-1000 3
3 0-10K 1.5
4 0-100K 1.5
5 0-1M 0.7
6 0-10M 0.7
3 second response time selected:
Scale Range (CPM) Response (seconds)
1 0-100 3
2 0-1000 1.5
3 0-10K 0.7
4 0-100K 0.7
5 0-1M 0.7
6 0-10M 0.7

Victoreen 190
Operators Manual
1-4
Front Panel Controls
Eight push-button switches: LIGHT (2): either activates the LCD backlight;
MODE: Changes the display units;
LOG: Stores currently displayed data or changes current location identifier, depending on the
currently active logging mode;
RATE/INTEG: Toggles the alphanumeric display between the digitized rate value and the integrate
value;
RESP TIME: Changes the response time on the first range or resets the integrate value and elapsed
time;
AUDIO: Toggles the click output associated with counting events in the detector, turns Audio on/off,
and acknowledges alarm conditions;
ON/OFF: Turns instrument power on/off when permitted by the microprocessor (i.e., when data
logging is inactive or all logged data has been read by a PC).
Internal Controls
Main unit: LCD contrast adjustment and +5 V Power Supply, Probe adapter modules: probe high
voltage, threshold, and detector jam current controls
Environmental Effects
Operating temperature range: 14 F to 140 F (-10 C to 60 C) with alkaline batteries.
Relative humidity range: 0 to 95% non-condensing.
Temperature dependence: Less than 5% of full scale change in the reading when the operating
temperature is in the 20 C to 50 C range.
Batteries
Four 9-volt batteries installed provide up to 200 hours of continuous operation.
Readout
Backlit liquid crystal display includes:

Theory of Operation
General Circuit Description 2
2-1
Section 2
Theory of Operation
2.1 General Circuit Description
A block diagram of the circuit of the Model 190 is shown in Figure 2-1. The CMOS microprocessor is a TI
7000 series microprocessor, capable of addressing 64k bytes of memory. A 32k byte PROM contains the
operating program and an 8k byte RAM provides storage for operational and data logging information.
The microprocessor addresses the display via two display drivers. The microprocessor senses the
keypad directly, with the exception of the ON/OFF switch. The communications is direct from the
microprocessor through its ports, light emitting diodes, and photosensitive transistors. The
microprocessor reads and writes to both EEPROMs and reads count data directly from the probe adapter
module
A probe adapter module connection is provided on the back of the instrument. The probe adapter module
retains the calibration information for a specific probe and contains high voltage and threshold
adjustments for the probe. A portion of the circuit in the probe adapter module provides an adjustable
threshold for current drain through the detector. This circuit is used to detect a jam condition in the
detector and to signal the instrument. The pulse shaping circuit in the module accepts a wide range of
detector input amplitudes and pulse decay times.
Communication with the Model 190 is accomplished through an infrared two-way adapter that mounts at
the top of the instrument. The infrared receiver and transmitter are viewed through two small holes in the
upper edge of the decal covering the LCD.
2.2 Operational Considerations and Applications
Control and Configuration
The Model 190 makes use of two data bases for operation; one is stored in an EEPROM on the main
circuit board and one is stored in an EEPROM in the probe adapter module. The instrument EEPROM
determines which set of configuration and control data codes will be used to define the operating modes.
These codes may be user-modified with the optional Communicator (P/N 190-1A) to tailor the instrument
for a specific application.
For example, any or all of the front panel buttons may be disabled or the unit may be configured to
operate in only one Display Mode. Refer to the Instruction Manual provided with the Communicator for
further details.
Display Calculations
The Model 190 retains all data internally as both counts and counts per minute. All other displays are
calculated from these data sets and stored calibration information in the EEPROMS, allowing the operator
the capability to switch back and forth between the various display modes without affecting the database
used for calculations.

Victoreen 190
Operators Manual
2-2
The counts per minute mode does not require a
calibration adjustment because its time base is
based on the highly accurate crystal clock of the
microprocessor. All other displayed values (e.g.,
mR/h and Sv/h) can be derived from the counts per
unit time value if a suitable calibration constant is
used.
Above 384,000 CPM, the counts in a time slot are counted and the value multiplied by the appropriate
factor to determine the actual count rate, which may cause a reduced accuracy specification on the fifth
and sixth scales.
Figure 2-1. Model 190 Block Diagram
NOTE
Micro-
processor
PROM
RAM
EEPROM
Analog
Display
Driver
Digital
Display
Driver
Front
Panel
Keypad
Counter
HV
Supply
EEPROM
Probe
GM Tube
Power
Supply
Display
Communicator
Microprocessor I/O
and BUS

Theory of Operation
Operational Considerations and Applications 2
2-3
Deadtime Effects
The deadtime correction is limited to a maximum multiplier of 3.7 above 106counts per minute (cpm) 5.4
from 105to 106cpm, and 1.6 from 104to 105 cpm. Larger correction factors are permitted at higher count
rates because of increased counting statistics. The display will indicate “OVER RANGE” if the deadtime
correction is greater than the values listed. This typically happens only on scale 6. In the event of an
overrange indication due to the deadtime, the bar graph portion of the display continues to function and
may be used to observe increased or decreased radiation even though numerical values are not
displayed. The bargraph displays relative values in this mode.
Low Battery Indications
Once every 256 seconds the instrument checks the amount of energy remaining in the batteries. When
the values reach a threshold, the instrument simultaneously displays a LOW BATTERY message and
sounds the sounder for a minimum of one second. The amount of time remaining before the unit stops
operating depends on previous use of the instrument. If the instrument is used for longer continuous
periods of time, there will be a slower decrease in available battery life at the battery end point; if the
instrument is on for shorter periods of time, the battery fails more precipitously at the battery end point.
The display may prematurely indicate a low battery
if the count rate is high and/or the backlight and/or
Alarm are operational simultaneously.
Response Times
The response times stated in Specifications in Section 1 - Introduction apply to the bargraph display.
Since the digital data is updated only once each second, a filter is used to smooth the readings. There is
approximately two seconds additional response time (10-90% of change) for the digital display.
Range Changing
A hysteresis is built into the range changing of the Model 190 to prevent excessive range changing near a
range change threshold. In addition, internal readings above 10% of the current full scale are allowed for
one full second prior to downscaling. Any reading above the current full scale will cause a range change
to the next higher scale.
NOTE

Victoreen 190
Operators Manual
(Blank page)

Operation
Unpacking and Inspection 3
3-1
Section 3
Operation
3.1 Unpacking and Inspection
Upon receipt of the unit:
1. Inspect the carton(s) and contents for damage. If damage is evident, file a claim with the carrier
and notify Fluke Biomedical, Radiation Management Services at 440.248.9300.
2. Remove the contents from the packing material.
3. Verify that all items listed on the packing list have been received and are in good condition.
If any of the listed items are missing or damaged,
notify Fluke Biomedical at 440.248.9300.
3.2 Storage
If the unit is to be stored prior to use, pack it in the original container(s), if possible, and store in an
environment free of corrosive materials, fluctuations in temperature and humidity, and vibration and
shock.
3.3 Battery Installation/Replacement
Four 9-volt alkaline batteries are supplied with the Model 190. The battery compartment is located on the
back of the instrument. The compartment will hold up to four batteries. All four batteries must be in place
for proper operation and to avoid damage to the instrument. Use the following procedure to
install/replace the batteries:
To prevent battery leads from shorting on the
battery compartment’s conductive coating, Ensure
that all four batteries are INSTALLED at all times.
Unit power must be left ON and batteries replaced
one at a time to prevent data loss when the log
mode is activated and logged data is to be
retrieved.
1. Loosen the two-quarter turn fasteners securing the battery compartment cover to the back panel.
NOTE
CAUTION
CAUTION

Victoreen 190
Operators Manual
3-2
2. Remove the battery compartment cover to gain access to the batteries.
3. Replace the batteries one at a time, observing proper polarity (refer to Figure 3-1)
The unit may make beeping sounds while inserting
the batteries.
4. Replace the battery compartment cover, securing it with the two-quarter turn fasteners.
Figure 3-1. Battery Installation (For 190AC and 190F only three 9-volt batteries are
installed.)
3.4 Probe Holder Installation
Extreme caution should be used when connecting
the probe to the meter. Improper connection may
result in injury, damage to the instrument, or
damage to the probe.
1. Attach the included Velcro strap through the strap slot in the mounting bracket.
Two Velcro straps are included. Use the strap that
fits the type of probe being utilized.
2. Attach the blade type-mounting bracket to the Model 190 Survey Meter, by performing the
following:
a. Remove and discard screw (B) from the Model 190.
b. Place the bracket in the correct position on the Model 190.
NOTE
WARNING
NOTE

Operation
Probe Holder Installation 3
3-3
Do not use the long screw (B), removed from the
Model 190, in screw position (A). Doing so will
damage the instrument.
c. Use the .63” screw included with the kit to secure in screw hole (A).
d. Use the 1 1/2” screw included with the kit to secure in screw hole (B).
Figure 3-2 Probe Holder Installation
3.5 Adapter Module Installation/Replacement
The Model 190 is shipped with a MHV compatible adapter module, (unless BNC compatibility was
requested on the purchase order). If a probe was ordered with the instrument, it is connected to the
probe adapter. Use the following procedure to replace a probe and adapter:
1. Loosen the captive Philips screw from the top panel of the adapter.
2. Unplug the probe adapter module (with detector, is applicable) from the unit.
3. Plug the replacement adapter and detector into the unit, being careful not to damage the connector
pins. It may be necessary to tilt the adapter slightly when plugging it into the unit.
CAUTION
MODEL
S/N REV
VICTOREEN
MADE IN U.S.A.
Screw
Hole (A)
Screw (B)
MODEL
S/N REV
VICTOREEN
MADE IN U.S.A.
Strap
Slot
Mounting
Bracket

Victoreen 190
Operators Manual
3-4
We recommend that the probe be secured to the
mating probe adapter receptacle using heat shrink
tubing.
4. Secure the Philips screw located on the top panel of the adapter.
3.6 Power
The Model 190 is powered using four 9-volt alkaline batteries. The batteries (four) are housed in a
separate compartment accessed from the instrument back panel. All four 9-volt alkaline batteries are
required to operate the unit for approximately 200 hours. Visual and audible low battery indications are
provided whenever battery power is low (four minute intervals for the audible and continuous for the visual
after a battery test).
3.7 Readout
Measurement results are indicated on the front panel LCD. The display includes an analog bargraph,
scale multiplier, detector range bargraph, and a 16 character alphanumeric display (refer to Figure 3-3).
Each of the 51 elements in the bargraph represents 2% of the scale span. The analog bar graph is
divided into 5 sections with scale notations of 0, 2, 4, 6, 8, and 10. The scale sections are marked by
longer bars. The bars are arranged in an arc with a radius of 3.75 inches (9.5 cm.). Scale length is 2
inches (5 cm.).
The 51 element analog bargraph, located near the top of the display, provides real time response. It is
designed to emulate an analog meter movement, making it especially useful in survey applications. The
16 character alphanumeric display at the bottom of the LCD displays units, an average digitized value of
the reading, error messages, integrated value (with elapsed time), operational parameters (e.g., response
time), low battery indication, probe type and serial number, and data logging information.
A scale multiplier, located below the bargraph relates the analog bargraph full scale to the units displayed.
The detector range bargraph consists of five horizontal bars that indicate the decade of the dynamic
range in cpm of the detector in use. Most detectors will cover a dynamic counting range of five decades.
A few detectors (e.g., those with short dead times) can reach the sixth decade of operation. The sixth
decade is indicated by all five bars being lit.
The bargraph may be disabled using the Model
190-1A Communicator.
In the Alarm condition, the entire display blinks on and off with a period of two seconds
NOTE
NOTE

Operation
Internal Controls 3
3-5
3.8 Internal Controls
Use extreme care and ensure that you are properly
grounded prior to opening the unit for voltage
adjustments. Failure to comply could destroy the
instrument.
Display Contrast Adjustment (RP1)
The Model 190 display is designed for optimum viewing at approximately 60ofrom the plane of the display
and directed toward the bottom of the instrument. The Display Contrast Adjustment (RP1) is factory set
for optimum viewing and Fluke Biomedical does not recommend that it be further adjusted. However; if
the display contrast requires adjustment, turn RP1 to vary the angle of maximum contrast of the 16-
character display. The contrast for the upper part of the display is fixed. Refer to Figure 3-4 for RP1
location.
Voltage Adjustment (RP2)
A +5 volt adjustment is provided on the main circuit board. Refer to Figure 3-3 for adjustment location.
Use the following procedure:
1. Connect the - lead of a DVM to the groundside of R18.
2. Connect the + lead of the DVM to Z10-12.
3. Adjust RP2 for 5.00 volts +0.02 volts.
Figure 3-3. Model 190 Display
CAUTION
0246
810
X0.0.0.010 MILLION
X0.100 THOUSAND

Victoreen 190
Operators Manual
3-6
Figure 3-4. Display Contrast (RP1) and +5 (RP2) Adjustment Locations
Probe Adapter Module Adjustments
The probe adapter module has high voltage (RP3), detector sensitivity (RP2), and detector jam current
(RP1) adjustments located within it. Use the following procedure to make the appropriate adjustments:
1. Loosen the Philips screw on the top panel of the adapter.
2. Unplug the adapter from the unit.
3. Remove the flat head screw from the bottom panel of the adapter.
4. Remove the top panel of the adapter to gain access to the inner components.
There is a potential shock hazard when the adapter
cover is removed and the unit is operating. Use an
insulated screwdriver when making adjustments,
being careful not to touch any of the components
other than the potentiometers specified.
5. Plug the adapter back into the Model 190 with the component side of the circuit board exposed.
+5 V Ad
j
ustment (RP2)
Dis
p
la
y
Contrast Ad
j
ustment (RP1)
CAUTION

Operation
Internal Controls 3
3-7
It may be necessary to tilt the adapter slightly when
attempting to plug it into the unit.
6. Make the required voltage, threshold, and jam current adjustments as outlined in the following
paragraphs.
The Model 190 must be turned ON while making
the probe adapter adjustments.
7. When the adjustments are completed, unplug the adapter from the unit.
8. Position the back panel on the adapter.
9. Replace the flat head screw removed in step 3.
10. Plug the adapter back into the unit.
11. Secure adapter to unit by tightening the captive screws located on the top panel of the adapter.
3.9 HIGH VOLTAGE ADJUSTMENT
Determine which module you have (surface mount
or non-surface mount) before continuing, and
ensure that the correct figure is referenced.
Refer to Figure 3-5 for location of the high voltage adjustment (RP3). Use the following procedure to
adjust the detector high voltage:
1. Turn unit power OFF.
2. Locate the 3-pin jumper connector JP1 (refer to Figure 3-5 for jumper location).
3. Position the jumper as required for either a surface mount or non-surface mount module by
referring to Figure 3-6.
4. Connect either a DVM with a high resistance probe (> 100Mohm) or an electrostatic voltmeter to
the MHV or BNC connector.
5. Turn the Model 190 on.
6. Adjust RP3 for the required voltage.
NOTE
NOTE
NOTE

Victoreen 190
Operators Manual
3-8
Figure 3-5. Probe Adapter Adjustments
Table of contents
Other Fluke Measuring Instrument manuals

Fluke
Fluke 845AR User manual

Fluke
Fluke i1010 User manual

Fluke
Fluke VR101S User manual

Fluke
Fluke 312 User manual

Fluke
Fluke VT Plus HF User manual

Fluke
Fluke 972A User manual

Fluke
Fluke 789 Owner's manual

Fluke
Fluke i3000s Flex Series User manual

Fluke
Fluke 1550 User manual

Fluke
Fluke 365 Use and care manual

Fluke
Fluke i3000s Flex Series User manual

Fluke
Fluke 80i-2010s User manual

Fluke
Fluke 55XXA/COIL 50 User manual

Fluke
Fluke TPS FLEX 18-TF-II Configuration guide

Fluke
Fluke 421D Instruction manual

Fluke
Fluke 1742 Configuration guide

Fluke
Fluke FIBER QUICKMAP User manual

Fluke
Fluke FI-500 FiberInspector Micro User manual

Fluke
Fluke 700G01 User manual

Fluke
Fluke IRR2-BT User manual