Fluke 187 User manual

Page 1 of 3, Document #2153
187/189 Extended Specifications
©2000 Fluke Corporation
Rev. A-8/00
Accuracy Accuracy Dual Display AC or AC+DC5
20 Hz- 45 Hz- 1kHz-
Function Range Resolution DC 45 Hz 1 kHz 20 kHz
DC mV 50.000 mV 0.001mV 0.1% +20 2% + 80 0.5% + 40 6% + 40
500.00 mV 0.01 mV 0.03% + 2 2% + 80 0.5% + 40
3000.0 mV 0.1 mV 0.025% + 5 2% + 80 0.5% + 40
DC V 5.0000V 0.0001V 0.025% + 1032% + 80 0.5% + 40 2% + 40
50.000V 0.001V 0.03% + 332% + 80 0.5% + 40
500.00V 0.01V 0.1% + 232% + 80 0.5% + 40 Not specified
1000.0V 0.1V 0.1% + 232% + 80 0.5% + 40 Not specified
DC µA 500.00 µA 0.01 µA 0.25% + 20 1% + 20 1% + 20 2% + 40
5.000 µA0.1µA 0.25% + 2 1% + 10 0.75% + 10 2% + 40
DC mA 50.000 mA 0.001 mA 0.15% + 10 1% + 20 0.75% + 20 2% + 40
400.00 mA 0.01 mA 0.15% + 2 1% + 10 1% + 10 3% + 40
DC A 5.0000A 0.0001A 0.5% + 10 2% + 20 2% + 20 6% + 40
10.000A40.001A 0.5% + 2 1.5% + 10 1.5% + 10 5% + 40
Detailed Specifications
Accuracy is specified for a period of one year after calibration, at 18°C to 28°C (64°F to 82°F) with
relative humidity to 90%. Accuracy specifications are given as ±([% of reading] + [number of least
significant digits])
Accuracy
45 Hz- 20 Hz- 1 kHz- 10 kHz- 20 kHz-
Function Range
Resolution
1 kHz 45 Hz 10 kHz 20 kHz 100 kHz
AC mV 1,2 50,000mV 0.001mV 0.4% + 40 2% + 80 5% + 40 5.5% + 40 15% + 40
500.00 mV 0.01 mV 0.4% + 40 2% + 80 5% + 40 5.5% + 40 8% + 40
3000.0 mV 0.1 mV 0.4% + 40 2% + 80 0.4% + 40 1.5% + 40 8% + 40
AC V 1,2 5.0000V 0.0001V 0.4% + 40 2% + 80 0.4% + 40 1.5% + 40 8% + 40
50.000V 0.001V 0.4% + 40 2% + 80 0.4% + 40 1.5% + 40 8% + 40
500.00V 0.01V 0.4% + 40 2% + 80 0.4% + 40
Not specified Not specified
1000.0V 0.1V 0.4% + 40 2% + 80 0.4% + 40
Not specified Not specified
dBV -56 to -6 0.01 dB 0.1 dB 0.2 dB 0.5 dB 0.5 dB 1.4 dB
-6 to +34 0.01 dB 0.1 dB 0.2 dB 0.1 dB 0.2 dB 0.8 dB
+34 to +60 0.01 dB 0.1 dB 0.2 dB 0.1 dB
Not specified Not specified
Fluke 187/189
True-rms Digital Multimeter
Extended Specifications
Nominal Specifications
Function Ranges/Description
DC Voltage 0 to 1000V
DC Current 0 to 10A (20A for 30 seconds)
AC Voltage, True-rms 2.5 mV to 1000V - 100 kHz bandwidth
AC Current, True-rms 25 µA to 10A (20A for 30 seconds)
Resistance 0 to 500 Mohms
Conductance 0 to 500 Nanosiemens
Capacitance 0.001 nF to 50 mF
Diode Test 3.1V
Temperature -200°C to 1350°C (-328°F to 2462°F)
Frequency 0.5 Hz to 1000 kHz
Accuracy (Basic DC V) 0.025%
(Basic AC V) 0.4%
Features
Feature Description
Dual Displays 50,000 count primary display
5,000 count secondary display
Backlight with 2 Bright white backlight for clear readings
brightness selections in poorly lighted areas
Fast Autorange Meter automatically selects best range -
instantly
AC+DC True RMS, ac Choices for AC only, AC and DC dual
rms specified to 100 kHz display, or AC+DC readings
dBm, dBV User selectable impedance references
for dBm
AutoHOLD
Holds readings on display
Continuity/Open Test Beeper sounds for Ohms readings below
threshold or for momentary
open circuit indication
Fast Bar Graph 51 Segments for peaking and nulling
Duty Cycle/Pulse Width Measure time a signal is on or off in %
or milliseconds
MIN MAX/Fast MIN MAX Record Maximum, Minimum, and Average
with elapsed and values. Real Time for MAX or MIN, elapsed
Real Time Stamp time for AVG.
Fast MIN MAX captures peaks to 250 µsec.
Closed Case Calibration No internal adjustments needed
Battery/Fuse Battery or fuse replacement without
Access Door voiding calibration
Hi-Impact Overmold Case Integrated Protective Holster provides
superior impact protection for your meter
Accuracy
45 Hz- 20 Hz- 1 kHz- 20 kHz-
Function Range
Resolution
1 kHz 45 Hz 20 kHz 100 kHz
AC µA 500.00 µA 0.01 µA 0.75% + 20 1% + 20 0.75% + 20 6% + 40
5000.0 µA0.1µA 0.75% + 5 1% + 5 0.75% + 10 2% + 40
AC mA 50.000 mA 0.001 mA 0.75% + 20 1% + 20 0.75% + 20 9% + 40
400.00 mA 0.01 mA 0.75% + 5 1% + 5 1.5% + 10 4% + 40
AC A 5.0000A 0.0001A 1.5% + 20 1.5% + 20 6% + 40 Not specified
10.000A40.001A 1.5% + 5 1.5% + 5 5% + 10 Not specified
1
For the 5,000 count mode, divide the number of least significant digits (counts) by 10.
2 A residual reading of 8 to 180 digits with leads shorted, will not affect stated
accuracy above 5% of range.
320 counts in dual display DC or AC + DC
410A continuous up to 35°C, less than 10 minutes 35°to 55°. 20A overload for
30 seconds maximum
5See AC conversion notes for AC mV and V.
Above specifications are subject to change without notice.

Page 2 of 3, Document #2153
187/189 Extended Specifications
©2000 Fluke Corporation
Rev. A-8/00
Detailed Specifications (continued)
Accuracy is specified for a period of one year after calibration, at 18°C to
28°C (64°F to 82°F) with relative humidity to 90%.
Accuracy specifications are given as ±([% of reading] + [number of least
significant digits])
Function Range Resolution Accuracy
Resistance1500.00Ω0.01Ω0.05% + 106
5.0000 kΩ0.0001 kΩ0.05% + 2
50.000 kΩ0.001 kΩ0.05% + 2
500.00 kΩ0.01 kΩ0.05% + 2
5.0000 MΩ0.0001 MΩ0.15% + 42
30.000 MΩ0.001 MΩ1% + 42
100.0 MΩ0.1 MΩ3% + 211
500.0 MΩ0.1 MΩ10% + 211
Conductance 50.00 nS 0.01 nS 1% + 10
Capacitance31.000 nF 0.001 nF 2% + 5
10.00 nF 0.01 nF
100.0 nF 0.1 nF
1.000 µF 0.001 µF
10.00 µF 0.01 µF 1% + 5
100.0 µF0.1µF
1.000 µF1µF
10.0 mF 0.01 mF
50.00 mF 0.01 mF53% + 10
Diode Test13.1000V 0.0001V 2% + 20
Frequency 500.00 Hz 0.01 Hz4
5.0000 kHz 0.0001 kHz ±(0.005% + 1)
50.000 kHz 0.001 kHz
999.99 kHz 0.01 kHz
Duty Cycle
10.00 to 90.00%
0.1% ±(Voltage Range/
Input Voltage) x 300
counts)9,10
Pulse Width5499.99 ms 0.01 ms ±(3% x(Voltage range/
input voltage)+ 1 count)
9,10
999.9 ms 0.1 ms
Temperature -200 to 0.1°C±(1% of reading + 1°C)7,12
+1350°C
-328 to 0.1°F±(1% of reading +
+2462°F1.8°F)7,12
Min-Max-Avg Response: 100 ms to 80% Specifed accuracy ±12
counts for changes
>200 ms in duration.
(±40 digits in AC for
changes >350 ms and
inputs >25% of range)
Fast Min-Max 250 µs8Specified accuracy ±100
counts for changes
>250 µs in duration8
1For the 5,000 count mode, divide the number of least significant digits
(counts) by 10.
2For relative humidity greater than 70%, resistance accuracy is 0.5%
over 1 MΩand 2.5% over 10 MΩ.
3For film capacitor or better, using Relative mode (REL ∆) to zero
residual on 1.1 nF range.
4Reading will be 0.00 for signals below 0.5 Hz.
5Least significant digit not active above 10 mF.
6Using relative mode (REL ∆) to zero residual reading.
7For ambient temperature changes of ±5°C, rated accuracy applies after
1 hour.
8For repetitive peaks, 2.5 ms for single events.
9Frequency greater than 5 Hz, except for VDC, 500 mV dc and
3000 mV dc functions; 0.5 Hz to 1.0 + 0.
10 Range/input ratios also apply to current functions.
11 To ensure stated accuracy, switch to conductance mode and verify that
the open circuit reading is less than 0.10 nS.
12 Accuracy specification is relative to the user - adjustable temperature
offset, and assumes ambient temperature stable to ±1°C
Above specifications are subject to change without notice.
Frequency Counter Sensitivity
Approximate VAC Sensitivity
(RMS Sine Wave)1
15 Hz to 500 Hz
2
VAC Approximate VCD VDC
Input Range
100 kHz
2
Bandwidth
3
Trigger Levels
1
Bandwidth
3
50 mV 5 mV 10 mV 1 MHz -5 mV & 5 mV 1 MHz
500 mV 20 mV 20 mV 1 MHz 5 mV & 65 mV 1 MHz
3000 mV 500 mV 2000 mV 800 kHz 140 mV & 200 mV 90 kHz
5V 0.5V 2.0V 950 kHz 1.4 V & 2.0 V 14 kHz
50V 5V 5.0V 1 MHz 0.5 V & 6.5 V > 400 kHz
500V 20V 20V 1 MHz 5 V & 65 V > 400 kHz
1000V 100V 100V > 400 kHz 5 V & 65 V > 400 kHz
1Maximum imput = 10 x Range (1000 V max). Noise at low frequencies and amplitudes may
affect accuracy.
2Usable at reduced sensitivity to 0.5 Hz and 1000 kHz.
3Typical frequency bandwidth with full scale (or maximum 2 x 107V-Hz product)
RMS sine wave.
Memory and PC Communication
Functions (Fluke 189 Only)
The Fluke 189 adds the following capabilities:
Interval LOGGING At least 288 intervals (specified by user in Setup)
may be recorded to internal memory. These
values may be viewed using the VIEW MEM
function on the meter. Up to 700 unstable event
values (similar to AutoHold) are automatically
added to LOGGING memory for viewing only
through the optional FlukeView®Forms PC
software. Additional intervals will be logged if
the signal is stable.
Reading SAVE Up to 100 readings may be saved by the user in a
memory separate from LOGGING memory. These
readings may be viewed using VIEW MEM.
Burden Voltage (A, mA, µA)
Function Range Burden Voltage (typical)
mA- µA 500.00 µA 102 µV / µA
5,000 µA 102 µV / µA
50.000 mA 1.8 mV / mA
400.00 mA 1.8 mV / mA
A 5.0000 A 0.04 V / A
10.000 A 0.04 V / A

Page 3 of 3, Document #2153
187/189 Extended Specifications
©2000 Fluke Corporation
Rev. A-8/00
Safety Information
Function Description
Safety 1000 V, AC/DC, maximum voltage between any
terminal and earth ground.
Complies with ANSI/ISA-S82.01-94, CSA C22.2
No 1010.1-92 to 1000 V Overvoltage Category III
and to 600V Overvoltage Category IV.
Certification agencies (approvals/listings pending):
UL per standard UL 3111 (pending)
CSA per standard CSA/CAN C22.2 No. 1010.1-92
TÜV per standard EN 61010 Part 1-1993 (pending)
Surge Protection 8 kV peak per IEC 1010.1-92
Fuse Protection 440 mA, 1000 V FAST Fuse, 11A, 1000V FAST Fuse
Markings UL, CSA, TUV, CE (VDE pending)
Input Impedance
Function Input Impedance (Nominal)
Volts, mV 10 MΩ, < 100 pF
Common Mode Rejection Ratio Normal Mode Rejection
DC Volts, mV >100 dB at dc, 50 Hz or 60 Hz ±0.1% >90 dB at 50 Hz or 60 Hz ±0.1%
AC Volts, mV > 90 dB dc to 60 Hz
Open Circuit Test Voltage Full-Scale Voltage
To 5 MΩ30 to 500 MΩ+ nS
Ohms < 5V 500 mV 3.1V
Diode Test < 5V 3.1000V
Typical Short-Circuit Current
500Ω5 kΩ50 kΩ500 kΩ5 MΩ30 MΩ
Ohms 1 µA 100 µA10µA1µA0.1µA0.1µA
Diode Test 1.0 mA typical
General Specifications
Function Description
Display Digital: 50000/5000 counts primary, 5000 counts secondary,
updates 4/second.
Analog: 51 segments, updates 40/second.
Operating Temperature -20°C to +55°C
Storage Temperature -40°C to +60°C
Temperature Coefficient 0.05 x (specified accuracy)/°C (<18°C or >28°C)
Relative Humidity 0% to 90% (0°C to 55°C)
0% to 70% (35°C to 55°C)
Altitude Operating: 2000 meters
Storage: 10000 meters
Battery Type 4 AA Alkaline, NEDA 15A or LR6
Battery Life 72 Hours typical (with backlight off)
Shock Vibration Per MIL-T-PRF 28800 for Class II instruments
EMC Susceptibility and Emissions: Commercial Limits per EN61326-1
Size 10.0 cm X 20.3 cm X 5.0 cm (3.94" X 8.00" X 1.97")
(Not Including Accessory Mount)
Weight 545 grams (1.2 lbs.)
Case Sealing IP-42 per IEC 529, Section 3
Warranty Lifetime
Calibration Interval 1 year
Fluke. Keeping your world up and running.
Above specifications are subject to change without notice.
Other manuals for 187
3
This manual suits for next models
1
Table of contents
Other Fluke Multimeter manuals

Fluke
Fluke 106 User manual

Fluke
Fluke 789 User manual

Fluke
Fluke 87 III Series Operating and installation instructions

Fluke
Fluke 28 II Ex User manual

Fluke
Fluke 1508 User manual

Fluke
Fluke 8808A Parts list manual

Fluke
Fluke 1550B User manual

Fluke
Fluke 2063513 User manual

Fluke
Fluke 8060A User manual

Fluke
Fluke 17B User manual

Fluke
Fluke 87 Series IV User manual

Fluke
Fluke 289 User manual

Fluke
Fluke 116 User manual

Fluke
Fluke 373 Use and care manual

Fluke
Fluke 27-II Use and care manual

Fluke
Fluke 345 User manual

Fluke
Fluke 106 User manual

Fluke
Fluke 3000 FC User manual

Fluke
Fluke Biomedical MPS450 Operators User manual

Fluke
Fluke 175 User manual