FORESIGHT TIMS MVP User manual

060000-91, Rev. 1

ii TIMS MVP™Setup and Configuration Guide
TIMS MVP™Setup and Configuration Guide
Rev. 1 September 2020
TIMS Medical, TIMS, Foresight Imaging, and the Foresight Imaging logo are registered trademarks of
Foresight Imaging LLC. TIMS, TIMS Medical, TIMS MVP, the TIMS logo, TIMS Medical Video Platform,
AccuStream, and Auto-SYNC are trademarks of Foresight Imaging LLC. All other products, service
names, and registered trademarks are trademarks of their respective companies.
Copyright © 2020 by Foresight Imaging, LLC.
All rights reserved.
The information in this document has been carefully checked and is believed to be accurate.
However, Foresight Imaging assumes no responsibility for inaccuracies. All specifications are subject
to change without notice.

TIMS MVP™Setup and Configuration Guide iii
Regulatory Information
Conforms with ANSI/AAMI ES60601-1, CAN/CSA-C22.2 No. 60601-1, FCC Part 15 Class A
Refer to ANSI/AAMI ES60601-1 for requirements applicable to the TIMS MVP™System.
Intended Use
There are no known contraindications to use of the TIMS MVP™System.

iv TIMS MVP™Setup and Configuration Guide
Explanation of Symbols
Symbol
Explanation
Manufacturer
European Authorized Representative
Serial Number
Date of Manufacture
Caution, Consult Accompanying Documents
Catalogue Number
Caution
Temperature Operation Limitations
Humidity Operation Limitations
Atmospheric Pressure Operations Limitations
Alternate Current (AC)
Direct Current (DC)
Antenna
Standby Switch

TIMS MVP™Setup and Configuration Guide v
Handling the Medical Equipment
Warning:
Use proper hygiene when using the TIMS MVP™System within the patient
environment.
Please ensure that your medical facility is in strict compliance and adherence to all
applicable laws and regulations regarding the handling of electronic equipment with
respect to cleanliness, disinfection and sterile environments and all other use and
care instructions when operating this equipment within the patient environment.
To avoid electrical shock, this equipment must only be connected to a supply main
with protective earth.
Do not position unit such that it is difficult to disconnect the device.
Environmental
Temperature, Operating:
Storage:
10°to 40°C
-40°to +70°C
Relative Humidity, Operating:
Storage Humidity:
30 to 75% rH
10 to 95% rH
Atmospheric pressure, Operating:
Storage:
700 to 1060 hPa
500 to 1060 hPa
Standby button operation should be pressed for < 1 second.
Power Supply
Warning:
The Power Supply (MeanWell model GSM120A12-R7B)
shipped with the TIMS MVP™System is a medical grade power
supply and is the only power supply to be used with the TIMS
MVP™System computer. Do not use any other power supply
with the TIMS MVP™System as this may result in unsafe
operation.
Input voltage rating of 100/240Vac, 50/60 Hz, 1.4-0.7A

vi TIMS MVP™Setup and Configuration Guide
Servicing Equipment
Warning: No user serviceable parts.
No modification of this equipment is allowed.
Electromagnetic Compatibility (EMC)
This device has been tested for compliance with the EMC requirements. The guidelines in this
section will help you to make sure that your medical equipment will meet the requirements of the
standard.
Warning:
Medical equipment should not be used, stacked, or located on or around
equipment that may create electromagnetic inferences.
Cables and Accessories
Use of all included cables and accessories is required for proper use of this equipment. Any cable or
accessory not released with the device may compromise safety and performance.
Any cable included or provided by TIMS is the proper and required length needed to operate the
equipment.
Replacement parts, such as cables and accessories, must be purchased through TIMS to ensure
proper compliance requirements.
Warning:
Using other manufacturer’s cables and accessories may affect EMC and
product performance. Unauthorized use of these items will void warranty
and may cause injury harm to you, others and/or the equipment.
Emissions
This equipment does not contain any RF communications equipment.
Note: Mobile RF communications equipment can affect medical electrical equipment.
Failure to use this equipment in the specified type of shielded location could result in degradation of
the performance of this equipment, interference with other equipment or interference with radio
services”.

TIMS MVP™Setup and Configuration Guide vii
Product Disposal Information
This product should not be disposed of with household waste. It should be deposited at an
appropriate facility to enable recovery and recycling.
Table 1 –Guidance and MANUFACTURER’S declaration –
ELECTROMAGNETIC EMISSIONS –for all ME EQUIPMENT and ME
SYSTEMS
Guidance and manufacturer’s declaration - electromagnetic emissions
The TIMS MVP™System
needsspecialprecautionsregardingEMC(ElectroMagnetic
Compatibility) and needsto beinstalledand put intoservice according tothe
EMC
informationprovided inthismanual.
When
TIMS MVP™System
isexposed to electromagnetic (EM) disturbances, the
system
mayshowabnormal behavior. Forinstance,themeasuredtraces may
be become
obscured, the software may crash or, incase ofvery high-level
transient voltage events
likefor instance ESD, partsofthesystem may even
become defective.
Emissions test
Compliance
Electromagnetic environment —guidance
RF emissions
CISPR 11
Group 1
The TIMS MVP™System
uses RF energy
only
foritsinternalfunction.Therefore,
its
RFemissions arevery lowand
are notlikely
to cause any
interferenceinnearby
electronic
equipment.
RF emissions
CISPR 11
Class A
The EMISSIONS characteristics of the TIMS
MVP™System make it suitable for Class A
use in industrial areas and hospitals.
(CISPR 11 class A). If it is used in a residential
environment (for which CISPR 11 class B is
normally required) this equipment might not
offer adequate protection to radio-
frequency communication services. The user
might need to take mitigation measures,
such as relocating or re-orienting the
equipment.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies

viii TIMS MVP™Setup and Configuration Guide
Table 2 –Guidance and MANUFACTURER’S declaration –
electromagnetic IMMUNITY –for all ME EQUIPMENT and ME
SYSTEMS
(see 5.2.2.1 f)
Guidance and manufacturer’s declaration —electromagnetic immunity
The TIMS MVP™System is intended for use in the electromagnetic environment
specified below. The customer or the user of the TIMS MVP™System should assure that
it is used in such an environment.
When the TIMS MVP™System is exposed to electromagnetic (EM) disturbances, the
system may show abnormal behavior. For instance, the measured traces may be
become obscured, the software may crash or, in case of very high-level transient
voltage events like for instance ESD, parts of the system may even become defective.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±1 kV for
input/output lines
±2 kV for AC and
DC powerlines
±1 kV for
input/output
lines
±2 kV for AC and
DC powerlines
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 differential
mode
±2 kV common
mode
±1 differential
mode
±2 kV common
mode
Mains power quality should
be that of a typical
commercial or hospital
environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
0 % UT
(100 % dip in UT)
for 0,5 cycle at
0°,45°,90°,135°1
80°,225°,270°
and 315°
70% UT
(30% dip in UT)
for 25 cycles
70% UT
(30% dip in UT for
25 cycles)
0 % UT
(100 % dip in UT)
for 0,5 cycle at
0°,45°,90°,135°1
80°,225°,270°
and 315°
70% UT
(30% dip in UT)
for 25 cycles
0% UT
(100% dip in UT
for 5 sec)
Mains power quality should
be that of a typical
commercial or hospital
environment, If the user of
the TIMS MVP™System
requires continued
operation during power
mains interruptions, it is
recommended that the
TIMS MVP™System be
powered from an
uninterruptible power
supply or a battery.

TIMS MVP™Setup and Configuration Guide ix
0% UT
(100% dip in UT
for 5 sec)
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE UTis the a.c. mains voltage prior to application of the test level.

xTIMS MVP™Setup and Configuration Guide
Table 4 –Guidance and MANUFACTURER’S declaration –
electromagnetic IMMUNITY –for ME EQUIPMENT and ME SYSTEMS
that are not LIFE-SUPPORTING
(see 5.2.2.2)
Guidance and manufacturer’s declaration —electromagnetic immunity
The TIMS MVP™System is intended for use in the electromagnetic environment specified below. The
customer or the user of the TIMS MVP™System should assure that it is used in such an environment.
When the TIMS MVP™System is exposed to electromagnetic (EM) disturbances, the system may show
abnormal behavior. For instance, the measured traces may become obscured, the software may crash
or, in case of very high-level transient voltage events like for instance ESD, parts of the system may
even become defective.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment —guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the TIMS MVP™System, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
Conducted RF
IEC61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
Radiated RF
IEC 61000-4-3
6 Vrms for ISM
bands
3 V/m
80 MHz to 2.7 GHz
27 V/m
385 MHz
28 V/m
450 MHz
9 V/m
710/745/780 MHz
28 V/m
810/870/930 MHz
28 V/m
1720/1845/1970
MHz
28 V/m
2450 MHz
9 V/m
5240/5500/5785
MHz
3 V/m
80 MHz to 800 MHz
800 MHz to 2,7 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in metres
(m).
The TIMS MVP™System is fairly sensitive to
conducted and radiated RF. Disturbance of the
TIMS MVP™System trace is possible at and
below the specified test level. It may be
necessary to relocate the TIMS MVP™System
or to apply shielding.
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,a
should be less than the compliance level in
each frequency range.b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

TIMS MVP™Setup and Configuration Guide xi
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be
considered. Ifthe measured field strength in the location in which the TIMS MVP™System is used exceeds the applicable RF
compliance level above, the TIMS MVP™System should be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the TIMS MVP™System.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

xii TIMS MVP™Setup and Configuration Guide
Table 6 –Recommended separation distances between portable
and mobile RF communications equipment and the ME
EQUIPMENT or ME SYSTEM –for ME EQUIPMENT and ME SYSTEMS
that are not LIFE-SUPPORTING
(see 5.2.2.2)
Recommended separation distances between portable and mobile RF communications
equipment and the TIMS MVP™System
The TIMS MVP™System is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the TIMS MVP™System
can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the TIMS MVP™System
as recommended below, according to the maximum output power of the communications
equipment.
WARNING:
Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any
part of the TIMS MVP™System, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
Rated maximum
output power of
transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
0,01
0.12
0.12
0.23
0,1
0,37
0,37
0,74
1
1,17
1,17
2,33
10
3,70
3,70
7,37
100
11,70
11,70
23,30
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.

TIMS MVP™Setup and Configuration Guide xiii
FCC Compliance Information
FCC NOTICE
WARNING
Changes or modifications to this unit not expressly approved by the party responsible for
compliance could void the user’s authority to operate this equipment.
NOTE: This equipment has been tested and found to comply with the limits for a Class A digital
device, pursuant to Part 15 of the FCC rules. The limits are designed to provide reasonable
protection against harmful interference when the equipment is operated in a commercial
environment. This equipment generates, uses, and can radiate radio frequency energy and, if not
installed and used in accordance with the instruction manual, may cause harmful interference to
radio communications. Operation of this equipment in a residential area is likely to cause harmful
interference in which case the user will be required to correct the interference at his/her own
expense.
Users should periodically inspect the TIMS MVP™System for signs of worn or frayed cables. If worn
or frayed cables are found, contact technical support immediately. Also, periodically inspect the
TIMS MVP™System cable connections to ensure that all cables are securely attached.
There are no user replaceable or serviceable parts in the TIMS MVP™System. Please contact
technical support with any questions.

xiv TIMS MVP™Setup and Configuration Guide
Preface
The TIMS MVP™User’s Guide consists of two parts:
•This TIMS MVP™Setup and Configuration Guide provides hardware setup and software
configuration information for IT & biomedical engineering professionals.
•The TIMS MVP™Operator’s Guide provides instructions for technicians and medical
professionals to work with the TIMS MVP System in everyday operational use.
What Is TIMS MVP?
TIMS MVP™is an integrated hardware/software system powered by the TIMS Medical Video
Platform (MVP), a software architecture upon which all TIMS MVP medical products are built for
delivering high performance medical video applications. TIMS MVP™provides for the recording,
editing, labeling, review, analysis, and archiving of medical video throughout the healthcare
enterprise. This includes fluoroscopy, endoscopy, cardiology, angiography, ultrasound, surgical
video, and any other medical video modality. All studies can be archived to PACS/VNA in DICOM
format and exported as DICOM or MP4 video for sharing, distribution, and integration into reports
and presentations. Any DICOM study can be imported into TIMS MVP™for full use within the
system.
What Is IDEA Auto-SYNC?
IDEA Auto-SYNC is an application that enables you to optimize the performance of TIMS MVP™with
your modalities. Although TIMS MVP™provides predefined common hardware profiles (CHP) files
for its Foresight Imaging video capture boards, you are recommended always to create your own
customized CHP files. You do this using Auto-SYNC.
About This User’s Guide
The content of this user’s guide may differ in some respects from the functionality of your TIMS
MVP™. Your system’s appearance, hardware configuration, and the layout of your PC and modality
connections may not be exactly the same as described and depicted in this guide.
In addition, some of the functionality described in this guide depends on which TIMS MVP™product
configuration you are using. See next page and margin notes throughout this guide.
This user’s guide describes hardware setup and configuration for the TIMS 2000 and TIMS 2000 SP.
Some features are not present on TIMS 500 and TIMS Review. See TIMS System Product Line and
margin notes throughout this guide.

TIMS MVP™Setup and Configuration Guide xv
Consult the TIMS MVP™Release Notes
Latest details about what is new or different in the current TIMS MVP™release are contained in the
Release Notes.
Open the Release Notes from the Start menu > TIMS MVP > TIMS MVP Release Notes.

xvi TIMS MVP™Setup and Configuration Guide
TIMS Product Line
This user’s guide describes the full range of functionality available with the following TIMS MVP™
configurations: TIMS 2000, SP EN TIMS 2000 SP, TIMS 500, and TIMS Review. The key differences
are bolded in the following summary. Where functionality is not applicable to all versions, the
applicable product version is noted in the margin (left).
TIMS 2000 EN
✓Static & streaming capture
✓For medical device modalities
✓DICOM send and DICOM receive
✓Extensive review, editing, labeling, and analysis tools
✓Foresight Imaging video capture hardware
✓Small footprint computer
✓Support and maintenance (1 year)
✓LCD monitor and typical onsite installation
TIMS 2000 SP
✓Synched audio
✓Static & streaming capture
✓For medical video modalities
✓DICOM send and DICOM receive
✓Extensive review, editing, labeling, and analysis tools
✓Foresight Imaging video capture hardware
✓Small footprint computer
✓Support and maintenance (1 year)
✓LCD monitor and typical onsite installation

TIMS MVP™Setup and Configuration Guide xvii
TIMS 500
✓Static capture only
✓For medical video modalities
✓DICOM send and DICOM receive
✓Extensive review, editing, labeling, and analysis tools
✓Foresight Imaging video capture hardware
✓Small footprint computer
✓Support and maintenance (1 year)
✓LCD monitor and typical onsite installation
TIMS Review
✓DICOM send and DICOM receive
✓Extensive review, editing, labeling, and analysis tools
✓Computer not included
.

xviii TIMS MVP™Setup and Configuration
Guide
Table of Contents
TIMS MVP™Setup and Configuration Guide ................................................................................ ii
Regulatory Information...............................................................................................................iii
Intended Use .....................................................................................................................................iii
Explanation of Symbols .....................................................................................................................iv
Handling the Medical Equipment.......................................................................................................v
Environmental ....................................................................................................................................v
Power Supply......................................................................................................................................v
Servicing Equipment..........................................................................................................................vi
Electromagnetic Compatibility (EMC) ...............................................................................................vi
Cables and Accessories......................................................................................................................vi
Emissions...........................................................................................................................................vi
Product Disposal Information...........................................................................................................vii
Table 1 –Guidance and MANUFACTURER’S declaration – ELECTROMAGNETIC EMISSIONS
–for all ME EQUIPMENT and ME SYSTEMS......................................................................................vii
Table 2 –Guidance and MANUFACTURER’S declaration – electromagnetic IMMUNITY
–for all ME EQUIPMENT and ME SYSTEMS.....................................................................................viii
Table 4 –Guidance and MANUFACTURER’S declaration – electromagnetic IMMUNITY
–for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING..........................................x
Table 6 –Recommended separation distances between portable and mobile RF communications
equipment and the ME EQUIPMENT or ME SYSTEM –for ME EQUIPMENT and ME SYSTEMS that
are not LIFE-SUPPORTING ................................................................................................................xii
FCC Compliance Information...........................................................................................................xiii
Preface.....................................................................................................................................xiv
What Is TIMS MVP?.........................................................................................................................xiv
What Is IDEA Auto-SYNC?................................................................................................................xiv
About This User’s Guide ..................................................................................................................xiv
Consult the TIMS MVP™Release Notes...........................................................................................xv
TIMS Product Line.....................................................................................................................xvi
Table of Contents.................................................................................................................... xviii
Glossary of Acronyms.................................................................................................................... xxiv
Get Connected............................................................................................................................ 1
Site Survey ....................................................................................................................................... 1
Typical TIMS MVP™System Installation............................................................................................ 2
TIMS MVP™System Checklist....................................................................................................2
Typical System Unit ....................................................................................................................... 4
Front View (Front Panel)............................................................................................................4
Rear View (Backplane)............................................................................................................5
Indicate PC Connections.................................................................................................................... 5
Power Supply Connection ................................................................................................................. 6
Power Supply Connection ................................................................................................................. 7

TIMS MVP™Setup and Configuration Guide xix
Modality Connections ....................................................................................................................... 7
To Make Modality Connections .................................................................................................... 7
Connecting Your Video Source to AccuStream HD+........................................................................ 11
DVI-IN Video Sources .................................................................................................................. 11
Optional Trigger...................................................................................................................... 12
S-Video (Y/C) Color Video Sources or Composite Color Video.................................................... 12
Optional Trigger...................................................................................................................... 13
HDMI Video Sources.................................................................................................................... 13
Optional Trigger...................................................................................................................... 13
Triggers............................................................................................................................................ 13
FI Video Splitter Isolator.................................................................................................................. 14
Network Setup................................................................................................................................. 15
Set Computer Name, Domain/Workgroup ................................................................................. 15
If Static IP Address Is Required.................................................................................................... 16
Verifying Audio................................................................................................................................ 18
Verify the Speaker Setup............................................................................................................. 21
Test Your Microphone and Speakers for Use with TIMS MVP™................................................. 21
Getting Started ..........................................................................................................................27
Configuring Channels....................................................................................................................... 27
Configuring Channel Settings ...................................................................................................... 27
Add a New Channel................................................................................................................. 28
Delete a Channel..................................................................................................................... 35
Edit a Channel......................................................................................................................... 36
Duplicate a Channel to Create a New Channel....................................................................... 36
Rename a Channel.................................................................................................................. 37
Configure the TIMS MVP™Worklist................................................................................................ 39
Configuring Worklist Settings...................................................................................................... 40
Open to Worklist..................................................................................................................... 40
Override the Application Entity Title ...................................................................................... 41
Remove the AE Title Override................................................................................................. 42
Add Worklist Local Settings .................................................................................................... 42
Add a Worklist............................................................................................................................. 44
Delete a Worklist......................................................................................................................... 45
Edit a Worklist ............................................................................................................................. 46
Duplicate a Worklist .................................................................................................................... 47
Configuring DICOM Destinations..................................................................................................... 49
Configuring DICOM Server Settings ............................................................................................ 49
Override the Application Entity Title ...................................................................................... 50
Remove the AE Title Override................................................................................................. 52
Add DICOM Server Local Settings........................................................................................... 52
Add a DICOM Server.................................................................................................................... 53

xx TIMS MVP™Setup and Configuration Guide
Edit a DICOM Server.................................................................................................................... 57
Delete a DICOM Server................................................................................................................ 59
Duplicate the Settings for a DICOM Server................................................................................. 60
Advanced Settings .....................................................................................................................61
Labels for Series Description Tags................................................................................................... 61
Add a Series Description Name and Label .................................................................................. 62
Edit a Label .................................................................................................................................. 65
Delete a Label.............................................................................................................................. 66
Duplicate a Label ......................................................................................................................... 68
Windows Active Directory............................................................................................................... 69
System Requirements for TIMS Review ...................................................................................... 69
Configuring the Active Directory Server...................................................................................... 69
Access Active Directory TIMS MVP™configuration page ........................................................... 70
Configuring Active Directory ....................................................................................................... 71
Active Directory Tests for Admin and User Groups................................................................ 72
Configuring Security...................................................................................................................74
Change the Administrator Password........................................................................................... 74
Set the Admin Password ............................................................................................................. 74
Configuring the Study Lists.............................................................................................................. 76
Select Fields for the Study Dashboard ........................................................................................ 76
Purging Old Studies ..................................................................................................................... 78
Changing the TIMS MVP™Studies Directory .............................................................................. 79
Import and Export TIMS MVP™Configuration File ......................................................................... 80
Export Configuration................................................................................................................... 80
Import Configuration................................................................................................................... 81
Retain Worklist UID Received from MVP ........................................................................................ 82
Migrate TIMS3 Config...................................................................................................................... 83
Export TIMS MVP Current Settings ............................................................................................. 83
Migrate TIMS3 Settings............................................................................................................... 84
Creating a Custom Hardware Profile...........................................................................................88
Using IDEA Auto-SYNC..................................................................................................................... 88
Image Requirements ....................................................................................................................... 88
Limitations....................................................................................................................................... 88
Using IDEA Auto-SYNC..................................................................................................................... 90
Starting IDEA Auto-SYNC ................................................................................................................. 92
Recommended Calibration Images ............................................................................................. 92
Creating CHP Files............................................................................................................................ 94
Location of CHP Files ....................................................................................................................... 94
Windows 10 (64-bit) Installation................................................................................................. 95
Windows 7 (64-bit) Installation................................................................................................... 95
Using Auto-SYNC Advanced Options............................................................................................... 95
This manual suits for next models
4
Table of contents
Popular Medical Equipment manuals by other brands

Novametrix Medical Systems
Novametrix Medical Systems 515 Service manual

EDAN
EDAN SE-601A Service manual

Graham Field
Graham Field LUMEX TENDERCLOUD AQTC8035 Installation & operation instructions

Stryker
Stryker iSwitch Operation and maintenance manual

ResMed
ResMed VS Ultra Patient manual

Abiomed
Abiomed AB5000 quick start guide

Envirotainer
Envirotainer Releye RAP user manual

Medivators
Medivators 2-8-611HAN Hookup connection guide

Halyard
Halyard On-q QUIKBLOC Instructions for use

Monitex
Monitex CAPSULE MIXER CM-II Instructions for use

Iridex
Iridex PASCAL Synthesis Operator's manual

laerdal
laerdal VitalSim Sounds Trainer Directions for use