INSTRUCTION MANUAL FOR DIGITAL INDICATOR (302-012)
Tolerance setting “TOL”
a) Under “ SET-TOL” mode , press button MODE to set lower tol.
While “ ” icon is blinking. And the last digit is blinking now, then press button MODE to modify
the number from 0-9. Then press button ZERO/ON/OFF to blink the left digit , the set the
intended value as above process till finishing all digits.
b) Press button SET again to move to upper tol. Setting, while
“”icon is blinking, set the intended value as same procedure as
lower tol. Setting .
Presetting value “CAL”
a) Under “SET-CAL” mode, press button to set the present MODE value. The last digit is blinking and
can be modified from 0 to 9 by pressing button MODE .
b) Press button ZERO/ON/OFF to blink the left digit, finish the setting for all other digits per same
procedure.
c) Press button to SET quit. The setting can be applied to both metric and inch mode.
d) If L –press button ZERO/ON/OFF, the preset value goes back to zero and preset mode is quitted.
Changing between metric / inch system “mm-in”
a) Under “SET-mm-in” mode , press button MODE to change metric/ inch system. “mm” is blinking
on metric system , and “in” is blinking on inch system.
b) Press button SET to quit.
Measuring Direction Setting “dir”
a) Under “SET-dir” mode , press button MODE to change the measuring directly. Sign “+” stands for
normal direction . sign “-”stands for reverse direction. “ “displayed on reverse direction.
b) Press button SET to quit