AEMC 400D-6 User manual

DIGITAL FLEXPROBE®
ENGLISH User Manual
400D-6 400D-10 400D-244000D-14 4000D-24
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Statement of Compliance
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
certifies that this instrument has been calibrated
using standards and instruments traceable to
international standards.
We guarantee that at the time of shipping your
instrument has met its published specifications.
An NIST traceable certificate may be
requested at the time of purchase, or obtained
by returning the instrument to our repair and
calibration facility, for a nominal charge.
The recommended calibration interval for this
instrument is 12 months and begins on the date of
receipt by the customer. For recalibration, please
use our calibration services. Refer to our repair
and calibration section at .
Serial #: ________________________________
Catalog #: ______________________________
Model #: ______________________________
Please fill in the appropriate date as indicated:
Date Received: _________________________________
Date Calibration Due: _______________________
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

2
Digital Flexprobe
®
Models 400D & 4000D
Table of Contents
1. INTRODUCTION............................................................................... 4
1.1 International Electrical Symbols................................................5
1.2Denitionof MeasurementCategories .....................................5
1.3ReceivingYourShipment..........................................................6
1.4OrderingInformation.................................................................6
1.4.1AccessoriesandReplacementParts............................6
2. PRODUCT FEATURES ...................................................................... 7
2.1Description................................................................................7
2.2ControlFeatures .......................................................................8
3. OPERATION .................................................................................. 10
3.1 MeasurementPrinciple ...........................................................10
3.2 Use .........................................................................................10
3.2.1Connection..................................................................10
3.2.2Measurement.............................................................11
3.2.3FreezingtheMeasurement......................................... 11
3.2.4SearchforMaximum..................................................12
3.2.5AutoPowerOFF .........................................................12
3.2.6LowBattery.................................................................12
3.2.7Disconnecting .............................................................12
4. SPECIFICATIONS........................................................................... 13
4.1ReferenceConditions .............................................................13
4.2 Electrical .................................................................................13
4.3VariationsinRangeofUse .....................................................14
4.4TypicalFrequencyResponseCurves .....................................15
4.5PowerSupply..........................................................................15
4.6 Environmental.........................................................................16
4.7Mechanical..............................................................................16
4.8Safety......................................................................................16
4.9ElectromagneticCompatibility ................................................16
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Digital Flexprobe
®
Models 400D & 4000D
3
5. MAINTENANCE ............................................................................. 17
5.1Cleaning..................................................................................17
5.2ReplacingtheBatteries...........................................................17
5.3RepairandCalibration ............................................................18
5.4TechnicalandSalesAssistance..............................................18
5.5 Limited Warranty.....................................................................19
5.6WarrantyRepairs....................................................................19
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

4
Digital Flexprobe
®
Models 400D & 4000D
CHAPTER 1
INTRODUCTION
Thank you for purchasing an AEMC®Digital Flexbprobe.
For best results and for your safety, read the enclosed operating
instructions carefully and comply with the precautions for use.
PRECAUTIONS BEFORE USE
This instrument is protected against voltages of not more than
1000V with respect to ground in measurement CAT III or 600V in
CAT IV between the sensor and the conductor that measures the
current.
The protection provided by the instrument may be impaired if the
instrument is used other than as specified by the manufacturer.
• Do not exceed the rated maximum voltage and current or the
measurement category.
• Observe the conditions of use; temperature, relative humidity,
altitude, level of pollution, and location.
• Before each use, check the integrity of the insulation on the
sensor, cable and housing. Do not use the instrument if it is
open, damaged, poorly assembled, or if its accessories appear
damaged.
• The sensor must not be applied to or removed from uninsulated
conductors at dangerous voltages.
• Use personal protection equipment systematically.
• All troubleshooting and metrological checks must be
performed by competent and accredited personnel.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Digital Flexprobe
®
Models 400D & 4000D
5
1.1 International Electrical Symbols
Signifies that the instrument is protected by double or reinforced insulation.
CAUTION - Risk of Danger! Indicates a WARNING and that the operator must
refer to the user manual for instructions before operating the instrument in all
cases where this symbol is marked.
Must not be applied to or removed from bare conductors at dangerous voltages.
Type B current sensor as per IEC 61010‑2‑032.
Important instructions to read and understand completely.
Important information to acknowledge.
Battery.
Compliance with the Low Voltage & Electromagnetic Compatibility European
directives (73/23/CEE & 89/336/CEE).
In the European Union, this product is subject to a separate collection system
for recycling electrical and electronic components In accordance with directive
WEEE 2002/96/EC.
1.2 Definition of Measurement Categories
CAT IV: For measurements performed at the primary electrical supply
(<1000V) such as on primary overcurrent protection devices,
ripple control units, or meters.
CAT III: For measurements performed in the building installation at the
distribution level such as on hardwired equipment in fixed instal-
lation and circuit breakers.
CAT II: For measurements performed on circuits directly connected to
the electrical distribution system. Examples are measurements
on household appliances or portable tools.
1.3 Receiving Your Shipment
Upon receiving your shipment, make sure that the contents are consis-
tent with the packing list. Notify your distributor of any missing items. If
the equipment appears to be damaged, file a claim immediately with the
carrier and notify your distributor at once, giving a detailed description of
any damage. Save the damaged packing container to substantiate your
claim.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

6
Digital Flexprobe
®
Models 400D & 4000D
1.4 Ordering Information
Digital FlexProbe®(MiniFlex) Model 400D-6 w/6 ft Lead
(TRMS, 4AAC, 40AAC, 400AAC)............................................Cat. #2153.30
Digital FlexProbe®(MiniFlex) Model 400D-10 w/6 ft Lead
(TRMS, 4AAC, 40AAC, 400AAC)............................................Cat. #2153.31
Digital FlexProbe®(MiniFlex) Model 4000D-14 w/6 ft Lead
(TRMS, 40AAC, 400AAC, 4000AAC)......................................Cat. #2153.32
Digital FlexProbe®(MiniFlex) Model 4000D-24 w/6 ft Lead
(TRMS, 40AAC, 400AAC, 4000AAC)......................................Cat. #2153.35
All models include one Digital FlexProbe®, two AAA batteries and a product user manual.
1.4.1 Accessories
Multifix Multi-Position Magnetic Mounting Accessory........Cat. #5000.44
Soft Carrying Case ..............................................................Cat. #2118.65
Small Classic Tool Bag........................................................Cat. #2133.72
Order Accessories and Replacement Parts Directly Online
Check our Storefront at www.aemc.com for availability
Digital FlexProbe®(MiniFlex) Model 400D-24w/6 ft Lead
(TRMS, 4AAC, 40AAC, 400AAC)............................................Cat. #2153.36
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Digital Flexprobe
®
Models 400D & 4000D
7
CHAPTER 2
PRODUCT FEATURES
2.1 Description
An ideal addition to the electrician’s tool kit, the Digital FlexProbe®series
can be used for TRMS AC current measurements from 20mA to 4000A,
and are rated 600V CAT IV. They provide a welcomed solution when
accessing electrical conductors is difficult and in tight places.
Two models are available. The Model 400D, available with a 6", 10", or 24" sensor, has a measurement range starting at 20mA, and is
designed for work in residential, commercial and light industrial
applications. It can be used to check electrical distribution systems up
to 400A.
The Model 4000D, available with a either a 14" or 24" sensor, has a
measurement range starting at 100mA and can be used on higher-
power industrial installations, as well as electrical utilities for
measurements up to 4000A.
Although they are high-performance instruments, the Digital
FlexProbe®series are very simple to use: two buttons are all it takes
to start the instrument, deactivate the auto power-off, HOLD the value
on the display or store the maximum value (MAX HOLD). The values are
read directly on the built-in 4000-count display.
They are ergonomically designed for comfortable handheld use
even when bulky gloves are required. The optional articulating,
magnetic Multifix mounting system accessory makes it simple to hang
on a wall, door, table edge or clip onto a belt.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

8
Digital Flexprobe
®
Models 400D & 4000D
2.2 Control Features
MAX
(>2s)
HOLD
4000 A
6
0
0
V
C
A
T
I
V
1
0
0
0
V
C
A
T
I
I
I
P
MAX
HOLD
A
Digital
FlexProbe®
MODEL 400
D
TRMS CURRENT METER
1
2
3
5
6
7
4
1. Flexiblesensor
- Model400D(6", 10", 24")
- Model4000D(14", 24")
2. Sensoropening/closinglever
3. Shieldedlead
4. Protectivehousing
5. LCDdisplay
6. ON/OFFbutton
7. HOLDbutton
Miniflex example shown
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Digital Flexprobe
®
Models 400D & 4000D
9
On the back of the protective housing is a notch for the attachment of a
belt clip (optional).
With the Multifix magnetic mounting accessory, you can position your
Digital FlexProbe®anywhere, leaving both hands free. The Multifix can
be used to:
• Carry the Digital Flexprobe®on a belt
• Attach it to a metal surface using the built-in magnet
• Attach it to a door top or the edge of a table
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

10
Digital Flexprobe
®
Models 400D & 4000D
CHAPTER 3
OPERATION
3.1 Measurement Principle
The flexible sensor is based on the Rogowski coil.
This principle combines:
• Excellent linearity with no saturation effect (therefore no heating)
• Light-weight (no magnetic circuit)
3.2 Use
3.2.1 Connection
1. Pressthelockingclip(s)toopenthesensor.
2. Placethesensoraroundtheconductorthroughwhichthecurrentto
bemeasuredows(onlyoneconductorinthesensor),thenclosethe
sensor.
3. Inordertooptimizemeasurementquality,itisbesttocenterthecon-
ductorinthecoilandmaketheshapeofthecoilasnearlycircularas
possible.
4. Pressthe buttontoturntheinstrumenton.
3.2.2 Measurement
Read the measurement on the display. The current is given in Arms.
A
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Digital Flexprobe
®
Models 400D & 4000D
11
If the measurement exceeds the display capacity (4000A), the device dis-
plays 3999, blinking.
A
If the measurement is too low (see § 3.2), the device displays dashes.
A
If the edges of the signal are too steep, or its peak factor is too large, the
device displays OL.
A
3.2.3 Freezing the Measurement
If you want to freeze the display of a measurement, press the HOLD
button. The HOLD symbol is displayed.
HOLD
A
The device continues to make measurements, but the display is frozen.
To release it, press the HOLD button again.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

12
Digital Flexprobe
®
Models 400D & 4000D
3.2.4 Search for Maximum
To search for a maximum, for example a spike lasting at least 100ms,
press the HOLD (MAX > 2s) button for more than two seconds.
The Max symbol will display and the instrument will begin measuring.
M
AX
A
The Digital Flexprobe®compares each new measurement to the one
displayed. If the new measurement is greater than the old, it replaces it
in the display.
To return to the real-time display mode, press the HOLD (MAX > 2s)
button again.
3.2.5 Auto Power OFF
If there is no activity after 10 minutes, the Digital Flexprobe®turns off
automatically (unless the MAX function is active) to preserve battery life.
To deactivate the Auto Power OFF feature, press the and HOLD
buttons simultaneously when turning the instrument on.
To reactivate automatic switching off, switch the device off, then back on.
3.2.6 Low Battery
When the battery voltage drops and the remaining battery life of the
instrument is approximately one hour, the symbol blinks on the
display.
When the battery voltage is too low to guarantee the accuracy of a
measurement, the symbol lights steadily. The batteries must then be
replaced (see § 5.2).
3.2.7 Disconnecting
Turn off the device by pressing the button. Press the yellow opening
lever to open the flexible sensor. Remove the sensor from the conductor.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Digital Flexprobe
®
Models 400D & 4000D
13
CHAPTER 4
SPECIFICATIONS
4.1 Reference Conditions
Quantity of influence Reference values
Temperature 73.4 ± 5.4°F (23 ± 3°C)
Relative humidity 45 to 75% RH
Frequency of the signal measured 40 to 65Hz
Peak factor of the signal measured √2
Conductor diameter ≤5mm
Battery voltage 2.8 to 3.2V
External electric field none
External DC magnetic field (earth field) <40 A/m
External AC magnetic field none
Position of the conductor centered in the measurement coil
Shape of the measurement coil nearly circular
4.2 Electrical
Model 400D-6 / 400D-10 / 400D-24 Display Range 4A 40A 400A
Measurement Range 0.020 to 3.999A 4.00 to 39.99A 40.0 to 399.9A
Measurement Range (max) 0.100 to 3.999A 4.00 to 39.99A 40.0 to 399.9A
Resolution 1mA 10mA 100mA
Accuracy ±(2% + 10ct) ±(1.5% + 2ct) ±(1.5% + 2ct)
Model 4000D-14 / 4000D-24
Display Range 40A 400A 4000A
Measurement Range 0.20 to 39.99A 40.0 to 399.9A 400 to 3999A
Measurement Range (max) 1.00 to 39.99A 40.0 to 399.9A 400 to 3999A
Resolution 10mA 100mA 1A
Accuracy ±(2% + 10ct) ±(1.5% + 2ct) ±(1.5% + 2ct)
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

14
Digital Flexprobe
®
Models 400D & 4000D
4.3 Variations in Range of Use
Quantity of influence Range of influence
Battery voltage 1.8 to 2V
Temperature 32 to 122°F (0 to 50°C)
Relative humidity 10 to 90% RH
Frequency response
10 to 20Hz
20 to 30Hz
30 to 400Hz
400 to 1000Hz
1000 to 3000Hz
Position of the conductor in the
sensor (f<400 Hz) Any position on the interior perimeter of the sensor
Adjacent conductor
carrying alternating current
Conductor touching the exterior
perimeter of the sensor
Peak factor 1.4 to 3.5 limited to 6000Apeak
Serial mode rejection ratio in AC 0 to 400ADC
Common mode rejection, 50/60 Hz 0 to 600Vrms
Influence of a 50/60Hz external
magnetic field 0 to 400A/m
Accuracy
Typical Maximum
< 1ct ±(2% + 1ct)
± 0.25% / 10°C ±( 0.5% / 10°C + 2cts)
0.2% ± (0.3% + 2cts)
See § 3.4
± (5% + 1ct)
± (1% + 1ct)
± (0.5% + 1ct)
± (6% + 1ct)
‑3 dB typical
± 0.5 % ± (1.5% ± 1ct)
Away from opening: 55 dB
At opening: 55 dB
Away from opening: ≥45 dB
At opening: ≥45 dB
at 16.66Hz: ± (2 % + 1ct)
at 50Hz: ± (0.5 % + 1ct)
at 440Hz: ± (30 % + 1ct)
± (6% + 1ct)
± (3% + 1ct)
—
< 1ct ≥50 dB
< 1ct ≥60 dB
Housing: 43 dB
Sensor: 50 dB
Housing: ≥30 dB
Sensor: ≥40 dB
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Digital Flexprobe
®
Models 400D & 4000D
15
4.4 Typical Frequency Response Curves
(at 39AAC)
-50 %
-40 %
-30 %
-20 %
-10 %
0 %
10 %
zH00001zH0001zH001zH01
4.5 Power Supply
The device can be powered by either:
• Two 1.5V (AAA) alkaline batteries
• Two NiMH storage batteries of the same size
The nominal operating voltage is between 1.8V and 3.2V.
The battery life in continuous operation is:
• 70 hours with super-alkaline batteries
• 50 hours with NiMH storage batteries having a capacity
of 1200 mAh
The low battery condition is acknowledged by a blinking symbol on
the display. When lit steadily, the batteries must be replaced (see § 5.2).
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

16
Digital Flexprobe
®
Models 400D & 4000D
4.6 Environmental
Operating Temperature: 32° to 122°F (0° to 50°C)
Storage Temperature (without batteries): -4° to 158°F (-20° to +70°C)
Operating Relative Humidity: 80% RH to 122°F (50°C)
Storage Relative Humidity: 90% RH up to 113°F (45°C) The sensor can withstand a temperature of 194°F (90°C) Indoor use / Level of pollution: 2 / Altitude: < 2000m
4.7 Mechanical
Dimensions: 3.94 x 2.36 x 0.79" (100 x 60 x 20mm)
Cable Length: 6 ft (1.83m)
Sensor Length: 400D-6: 6" (170mm)
400D-10: 10" (250mm)
400D-24: 24" (610mm) 4000D-14: 14" (350mm)
4000D-24: 24" (610mm)
Sensor Diameter: 400D-6: Ø 1.77" (45mm)
400D-10: Ø 2.75" (70mm)
400D-24: Ø 8" (190mm)4000D-14 : Ø 3.94" (100mm) 4000D-24 : Ø 8" (190mm)
Weight: 0.29 lbs (130g) approx (MiniFlex)
Index of Protection: IP 40 per IEC 60529
IK 04 per IEC 50102
V0 (per UL 94)
The flexible coil is resistant to oils and aliphatic hydrocarbons.
4.8 Safety
Electrical safety per IEC 61010-2-032 for type B sensors
Rated 600V CAT IV
4.9 Electromagnetic Compatibility
Emissions and immunity in an industrial setting compliant with IEC
61326-1 for portable devices.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Digital Flexprobe
®
Models 400D & 4000D
17
CHAPTER 5
MAINTENANCE
Use only factory specified replacement parts. AEMC®will not be held
responsible for any accident, incident, or malfunction following a repair
done other than by its service center or by an approved repair center.
5.1 Cleaning
Disconnect the instrument from any source of electricity.
• Use a soft cloth lightly moistened with soapy water.
• Wipe with a moist cloth and then dry with a dry cloth.
• Do not allow water or other foreign substances into the case.
• Never use alcohol, solvents or hydrocarbons.
5.2 Replacing the Batteries
Disconnect the instrument from any source of electricity.
The battery must be replaced when the symbol flashes or remains
steady on the display.
1. Useascrewdrivertounscrewthetwoclosingscrewsonthebackof
thehousing.
2. Replacetheoldbatterywithanewbattery(1.5VorAAAsuper-alkaline
batteries).
3. Closethehousing;makesurethatitiscompletelyandcorrectlyclosed.
4. Screwbothscrewsbackin.
Used batteries must not be treated as ordinary household waste.
Take them to the appropriate recycling collection point.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

18
Digital Flexprobe
®
Models 400D & 4000D
5.3 Repair and Calibration
To ensure that your instrument meets factory specifications, we recommend
that it be scheduled back to our factory Service Center at one-year intervals
for recalibration, or as required by other standards or internal procedures.
For instrument repair and calibration:
You must contact our Service Center for a Customer Service Authorization
Number (CSA#). This will ensure that when your instrument arrives, it will be
tracked and processed promptly. Please write the CSA# on the outside of the
shipping container. If the instrument is returned for calibration, we need to
know if you want a standard calibration, or a calibration traceable to N.I.S.T.
(Includes calibration certificate plus recorded calibration data).
Ship To:
(Or contact your authorized distributor)
Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are
available.
NOTE: You must obtain a CSA# before returning any instrument.
5.4 Technical and Sales Assistance
If you are experiencing any technical problems, or require any assistance with
the proper operation or application of your instrument, please call, fax or e-
mail our technical support team:
Contact:
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Digital Flexprobe
®
Models 400D & 4000D
19
5.5 Limited Warranty
The Digital FlexProbe®Models 400D and 4000D are warranted to the owner
for a period of two years from the date of original purchase against defects
in manufacture. This limited warranty is given by AEMC®Instruments, not by
the distributor from whom it was purchased. This warranty is void if the unit
has been tampered with, abused or if the defect is related to service not per-
formed by AEMC®Instruments.
.
Please print the online Warranty Coverage Information for your records.
What AEMC®Instruments will do:
If a malfunction occurs within the warranty period, you may return the
instrument to us for repair, provided we have your warranty registration
information on file or a proof of purchase. AEMC®Instruments will, at its
option, repair or replace the faulty material.
5.6 Warranty Repairs
What you must do to return an Instrument for Warranty Repair:
First, request a Customer Service Authorization Number (CSA#) by phone
or by fax from our Service Department (see address below), then return the
instrument along with the signed CSA Form. Please write the CSA# on the
outside of the shipping container. Return the instrument, postage or shipment
pre-paid to:
Ship To:
Caution: To protect yourself against in-transit loss, we highly recommend
you insure your returned material.
NOTE: You must obtain a CSA# before returning any instrument.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
Other manuals for 400D-6
1
This manual suits for next models
4
Table of contents
Other AEMC Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Tektronix
Tektronix SG 505 instruction manual

SI Analytics
SI Analytics handylab pH LF 12 operating manual

Geokon
Geokon 6150F instruction manual

MP Biomedicals
MP Biomedicals MPQuant user manual

Ecotech
Ecotech Serinus 51 user manual

PCB Piezotronics
PCB Piezotronics 339A30/NC Installation and operating manual