Genlantis FirstResponder E4110FRS User manual

FirstResponder® Sterilizer
INTRODUCTION
The FirstResponder Sterilizer* is a unique and innovative sterilization system. This portable sterilizer from Genlantis generates high-levels of
Triatomic Oxygen (TO) to clean and sterilize first responder vehicles. The unit will eliminate surface and airborne bacteria, mold, fungi, spores and
viruses, and eliminate odors as well. The FirstResponder Sterilizer is designed to allow maximum placement flexibility and usage.
The TO produced by the Genlantis FirstResponder Sterilizer readily gives up its single oxygen atom, which combines with other molecules and
neutralizes them via a powerful and natural oxidative process. Best of all, the FirstResponder Sterilizer contains a scrubbing feature, which converts
TO back to oxygen at the end of a sterilization cycle, hence removing TO gas from the environment rapidly and efficiently.
PRECAUTIONS AND WARNINGS
(a) DEVICE IS NOT INTENDED FOR USE IN ANY OCCUPIED SPACES. TO may cause respiratory irritation; make sure the FirstResponder
device finishes its sterilization and scrub cycles in full.
(b) Read all instructions before using the device, which must be used ONLY as intended and described in this manual.
(c) Do NOT operate the FirstResponder® Sterilizer without the scrub filters.
(d) DO NOT Immerse the FirstResponder® Sterilizer in water or operate at high humidity levels above 70%.
(e) Do not operate this appliance if it has a damaged cord or plug, if it is not working properly, or if it has been damaged or dropped.
(f) Do not attempt to repair the unit unless you are a Genlantis qualified repair technician.
SPECIFICATIONS
Cat#:
E4110
FRS
Cat#: E422
0
FRS
Electrical 120 VA, 60 Hz. AC 230 VA, 50 Hz. AC
Fuse
Type 239,
250V, 2 Amp
Type 239, 250V, 2 Amp
Dimensions (Width x Depth x Height)
20.62" x 16.87" x 8.12" (52.4 x 42.8 x 20.6 cm)
20.62" x 16.87" x 8.12" (52.4 x 42.8 x 20.6 cm)
Weight 35.0 lbs. (15.8 kg.) 35.0 lbs. (15.8 kg.)
Ozone Fan
220 CFM
220 CFM
Ozone
Output
16,000 mg/hour (mgph)
16,000 mg/hour (mgph)
De-ozonation (scrub) Fans 4 @ 110 cfm (440 CFM total) 4 @ 110 cfm (440 CFM total)
Warranty
1 year parts and labor
1 year parts and labor
* Disclaimer: The Genlantis FirstResponder Sterilizer is not a replacement or substitute for autoclaves, and should not be used to sterilize equipment for use directly
in medical or surgical applications of any kind. The FirstResponder sterilizing cycles can eliminate most surface contaminants, especially bacteria and fungus that
exist under normal use, however heavier concentrations of contaminants MUST be removed or wiped clean first before using the FirstResponder Sterilizer.
CAREFULLY READ AND FOLLOW ALL INSTRUCTIONS FOR OPERATION OF THE FIRSTRESPONDER STERILIZER. Avoid walking into any first
responder vehicle or space while the FirstResponder Sterilizer is still running AND before it completes its full sterilization and scrub cycles.
†Genlantis makes no claims and offers no guarantees of any kind that the FirstResponder Sterilizer will eliminate 100% of all contaminants under all circumstances.
Whenever and wherever possible, users should test an FirstResponder sterilized area for presence of trace contaminants if absolute cleanliness is desired or
required.
* Patents Pending. Registered under EPA Establishment Number 91845-CA-001
OPERATION
Catalog #
Description/Content
Content
s
E4110FRS
FirstResponder
®
Sterilizer,
110V
1 Sterilizer Device;
2 Scrub Filters;
1 Power Cord;
1 Spare Fuse;
1 Remote Control;
Spare Battery for remote control
E4220FRS FirstResponder® Sterilizer,
220V
Storage: Keep unit stored in shipping container while not in use to protect filters and electronics from contamination and dust.
Related
Products
Catalog #
FirstResponder® Mobile Gear Sterilizer
E600PCM
SteriZAP™ UV
-
C Room Sterilizer, 110V
E500UVCT
SteriZAP™ UV
-
C Desktop Sterilizer, 110V
E500UVCD
FirstResponder® Handheld UV
-
C Sterilizer, 110V
E500UVCH
FirstResponder® Duffle Bag
Sterilizer, medium
E600DBM
FirstResponder® Duffle Bag Sterilizer, large with
wheels
E600DBLW
6044 Cornerstone Court West, Suite E.
San Diego, CA 92121
Toll Free (888) 428-0558
Fax: (858) 623-9494
Web: http://www.genlantis.com

FirstResponder® Sterilizer Manual
VKM180403 Genlantis Page 2 of 5
6044 Cornerstone Court W., Ste. E. San Diego, CA 92121 Toll Free (888) 428-0558 U.S. & Canada (858) 457-1919 www.genlantis.com
2018 Gene Therapy Systems, Inc. (DBA: Genlantis).
I. Positioning the Unit
The FirstResponder® Sterilizer is portable and ready to plug in and use directly
out of the box. Its effectiveness depends on how the generated is dispersed
through ambient air. We recommend that users place the unit as centrally as
possible in the vehicle or room, or closest to known or suspected heavy
contamination areas.
II. Preparing the Area to Be Sterilized
1. Clean area or cabinet that needs sterilization from heavy or excess
contaminants using wipes or cleaning solutions like soap, ethanol, bleach,
etc.
2. Seal room or vehicle and verify that there are no gaps or openings. An
airtight space will prevent TO gas from escaping and will allow the buildup
of TO molecules. If any spaces cannot be closed off, it is necessary to
seal openings with plastic sheeting and/or tape.
3.
4. Use a copy of the “CAUTION” page in this manual (Page 7), or make your own note and prepare to affix on the outside of the first
responder vehicle or room door indicating that a sterilization cycle is in progress. This is necessary to avoid disrupting a sterilization cycle
and to eliminate accidental human exposure to high levels of TO.
III. Operating the Unit
General Notes
a. Different materials will react differently to TO. Consult Appendix A on page 4 for
guidance.
b. For safety reasons, the unit will not operate if the lid is open.
c. The unit can be set to operate between 00:01 and 99:99 minutes for either of the TO
or scrub cycles.
d. We recommend users keep a log of hours used; this will help with knowing when to
replace the Scrub Filter (recommended after each 3,000 hours).
To set the run time for the ozone cycle:
For recommended sterilization times based on size of area, see Table 1 below.
a. Press the [OZONE TIME] button; the digital display will light up and read either “0000” or a previously set time.
b. Press [TIME SELECT] button; the left-most digit will blink;
c. Press [SET TIME] up or down arrow button repeatedly until desired time is selected.
d. Press [TIME SELECT] button to move to next digit, repeat Step c, then press [SET TIME] to move to the next digit.
e. Repeat Step d. one last time to set the last digit; the display now should read the ozone time with no blinking digits.
To set the run time for the scrub cycle:
For recommended scrubbing times based on size of area, see Table 1 below.
f. Press the [SCRUB TIME] button; the digital display will read either “0000” or a previously set time.
g. Press [TIME SELECT] button; the left-most digit will blink;
h. Press [SET TIME] up or down arrow button repeatedly until desired time is selected.
i. Press [TIME SELECT] button to move to next digit, repeat Step c, then press [SET TIME] to move to the next digit.
j. Repeat Step d. one last time to set the last digit; the display now should read the ozone time with no blinking digits.
k. If needed or desired, set a lab timer with the total sterilization and scrubbing cycle times as a reminder before you run the unit. Once
the ozone and scrub cycles have been set, press the [RUN] button; the green light will turn on and you will hear the machine operate.
Immediately seal the area that is being sterilized.
l. Upon sterilization and scrub cycles completion, the FirstResponder Sterilizer will automatically turn off and the digital display will be
blank; users can enter the vehicle or room that is being sterilized. If needed, the FirstResponder Sterilizer can be run again with current
or readjusted settings for further cleaning cycles.
TRAINING VIDEO
Genlantis provides an online training video for the Ozilla®
EMS Sterilizer, which is identical in operation and usage as
the FirstResponder Sterilizer. The video is available at one of
these two locations:
1. On Youtube.com by clinking on, copying and pasting, or
typing this link: https://youtu.be/DJ8VE8sndoc.
2. By visiting this Genlantis web page:
http://www.genlantis.com/ozilla-ems-sterilizer.html and
clicking on the “VIDEO AND BROCHURE” tab to the right
of the main product “DESCRIPTION” tab.
NOTE: Consult Table 1 on Page 5 for suggestions on
FirstResponder sterilization times.
Figure 1: Front panel of FirstResponder Sterilizer

FirstResponder® Sterilizer Manual
VKM180403 Genlantis Page 3 of 5
6044 Cornerstone Court W., Ste. E. San Diego, CA 92121 Toll Free (888) 428-0558 U.S. & Canada (858) 457-1919 www.genlantis.com
2018 Gene Therapy Systems, Inc. (DBA: Genlantis).
Table 1: Recommended Sterilization and Scrubbing Times
Enclosure Type Examples
Subtype
Recommended Minimum
Sterilization Time (hours)
Recommended Minimum
Scrubbing Time**
Ambulance (initial sterilization)
EMS Types I, II, and III
4 hours
30 minutes
Ambulance (subsequent sterilizations)
EMS Types I, II, and III
1 hour
30 minutes
Police
vehicle (initial sterilization)
Sedan (e.g. Ford Taurus)
30 minutes
10 minutes
Police vehicle (initial sterilization)
SUV (e.g. Ford Explorer)
45 minutes
10 minutes
**NOTE: Minimum Scrub Times recommended can be extended by running additional scrub cycles if needed or desired.
Remote Control Operation
The FirstResponder® Remote Control allows users to switch the Sterilizer [RUN] cycle from “Ozone” to
“Scrub”. This allows users to prematurely terminate the sterilization cycle when needed and the unit will start
to remove generated TO gas for the amount of time that has been programmed by the user.
To activate the Scrub cycle, pull remote control antenna out (see figure to the right), point the remote in the
direction of the FirstResponder, preferably while having direct line of sight, then continuously press the
unlock button for 3-5 seconds until the green [RUN] button on the unit starts to flash and the Scrub Cycle
time appears on the display.
NOTE: The remote lock button does not have any function at this time and pressing it will result in no action.
IV. Post Sterilization Steps
During the sterilization process, TO gas will penetrate into hidden areas and crevices. Even after scrub cycles, some TO gas may remain
trapped in certain areas, like in AC vents for example. To avoid exposure to residual TO gas, follow these steps:
a. Open all doors and windows to air out of the first responder vehicle before entering.
b. Set the first responder vehicle ventilation/AC fan(s) to maximum and run for 3-5 minutes to purge the trapped TO gas.
c. If in a room, open windows and let ambient air vent out gradually or use fans to circulate air more effectively.
TROUBLESHOOTING
Problem
Solution
Unit does not start when [OZONE TIME] or
[SCRUB TIME] buttons are pressed.
Check that power cord is attached on the back of the unit;
Check that the FirstResponder Sterilizer case is closed and latched tight.
Sometimes during shipping, the ozone ceramic plates may get dislodged. Check the plates and
seat them properly by removing them and reinserting them back tightly in their respective slots.
Ozone smell is not
as noticeable as usual, or there
is no detectable ozone smell.
Clean the ceramic ozone plates as described in the MAINTENTANCE section. If plates are older
than two years replace with new ones (Cat# E40016P).
Power does not come on.
Check fuse on the ba
ck side of the unit; and if blown, replace with
type 239, 250V, 3 Amp fuse.
MAINTENANCE
WARNING: The FirstResponder® Sterilizer generates high voltage current inside the machine. MAKE SURE TO UNPLUG UNIT
from power source before performing any parts cleaning or replacements.
(a) Scrub Filter Replacement: Replace the scrub filter according to the following guidelines:
Time Used Hours Used
Light Use 12 months ≤ 500 hours
Moderate use 9 months ≤ 1,000 hours
Heavy use 6 months ≤ 3,000 hours
To change the filters, unlatch and open the unit’s lid. Locate and cut the zip tie that is holding the two filters down (towards front of lid
opening). Insert new filters and replace with new zip tie (supplied). Close and latch lid tightly.
(b) Ozone Plates Cleaning: Clean the ozone plates at the beginning of each new season, or when you suspect that ozone production
has decreased. To clean the ozone plates, open the unit lid and pull out the sixteen ozone generator plates. i) wash the plates in
warm soapy water, then air dry or use a hair dryer; ii) clean the contact area of the flat stainless steel holders with a Q-tip moistened
with alcohol and dry with tissue; iii) to avoid arcing or damaging the FirstResponder Sterilizer, make sure plates are completely dry
before re-inserting into the unit.
Antenna
Unlock
button
Lock
button
FirstResponder Sterilizer
Remote Control

FirstResponder® Sterilizer Manual
VKM180403 Genlantis Page 4 of 5
6044 Cornerstone Court W., Ste. E. San Diego, CA 92121 Toll Free (888) 428-0558 U.S. & Canada (858) 457-1919 www.genlantis.com
2018 Gene Therapy Systems, Inc. (DBA: Genlantis).
(c) Ozone Plates Replacement: Replace every two years or so depending on usage, and especially when you notice a decrease in
sterilization efficiency.
(d) General cleaning: wipe the unit exterior with a clean soft cloth tissue that’s lightly moistened with water or mild cleaning solution. DO
NOT SPRAY the unit directly with any liquid cleaner or immerse in water or solution of any kind.
FIRST AID: if users experience eye or breathing irritations, immediately take the following steps:
(i) Turn the sterilization cycle off by pressing [RUN] button;
(ii) Remove affected person from the area and administer first aid as necessary and described in the Material Safety Data Sheet;
(iii) Run a scrub cycle as indicated in this Owner’s Manual, Section III “Operating the Unit” on Page 3; alternatively, turn the FirstResponder
device off and air the affected area until the odor of ozone is no longer detectable or noticeable.
LIMITED LICENSE: The FirstResponder® Sterilizer is Patent Pending. The purchase price paid for the FirstResponder® Sterilizer grants end users a non-transferable, non-
exclusive license to use the unit in a research laboratory setting, and for sterilizing purposes only as described in this manual. This use license excludes and without limitation,
resale, repackaging, or modification of the unit in any way and without prior notification of and approval by Genlantis.
Under no circumstances shall the FirstResponder Sterilizer be used on food or drink products intended for human or animal consumption. At all times, care and attention should
be exercised in handling and using the FirstResponder Sterilizer by following the instructions in the manual, using common sense practices, wearing protective clothing and
eyewear when appropriate. NEVER OPERATE THE FIRSTRESPONDER STERILIZER IN A ROOM OR AREA WHERE PEOPLE OR ANIMALS ARE PRESENT.
Purchasers may refuse this license by returning the FirstResponder Sterilizing unit unused and its original packaging. By keeping or using the FirstResponder device, users agree
to be bound by the terms of this license. The laws of the State of California shall govern the interpretation and enforcement of the terms of this license
APPENDIX A
WARNING: Different materials respond differently to TO molecules, and some common first responder gear and triage materials
may be damaged by exposure. To minimize material damage, carefully read and adhere to the following guidelines:
SE = Strong effect: Materials are most susceptible to triatomic oxygen (TO) degradation; these materials should be removed from areas to be sterilized.
MEH = Moderate Effect High: Materials are noticeably affected by TO and exposure should be limited to short periods of time only. Frequent exposure of these materials to TO
will require they be discarded or replaced.
MEL = Moderate Effect Low: Materials are not affected by infrequent, brief exposure, but TO can deteriorate materials found under MEL. Prolonged exposure to high levels of
TO will require they be discarded or replaced.
ZE = Zero Effect: materials are not affected by TO; sterilize as often as needed.
Unlisted Materials: if a laboratory material is not listed in the table below, err on the side of caution and remove these unlisted materials from areas to be sterilized with the
FirstResponder Mobile Gear Sterilizer.
Genlantis is not responsible for damage arising from failure to follow these guidelines.
Material SE MEH MEL ZE
Buna-N (nitrile) / Fiber reinforced plastic (FRP) / Magnesium… X
Natural Rubber / Nylon / Steel / Zinc. X
Acetal resin (Delrin™) / Cast Iron / Galvanized Steel… X
Chlorosulfonated polyethylene (CSPE) (Hypalon™)… X
Thermoplastic elastomer (Hytrel™) / Monel™ / Neoprene (Polychlorprene™)… X
Polyamide / Polypropylene / Glass filled polypropylene (GFPP). X
ABS Plastic / Acrylic / Aluminum / Brass / Bronze / Copper… X
Ethylene propylene diene monomer (EPDM) / Flexelene™ tubing… X
Low-density polyethylene (LDPE) / Poly(vinyl chloride) (PVC)… X
Polyacrylate / Polyethylene / Polysulfide / Tygon™. X
Butyl / Chlorinated polyvinyl chloride (CPVC) / Cross linked polyethelene (PEX)… X
ehtylene propylene / Ethylene acrylic elastomer (Vamac™) / X
Ethylene propylene rubber (EPR) / Fluorosilicone / Glass… X
High-density polyethylene (HDPE) / Polycarbonate / PVDF… X
Perfluoroelastomer (Kalrez™, Chemraz™, Viton™) / Silicone… X
Polychlorotrifluoroethene(PCTFE) / Polyether ether ketone (PEEK)… X
Polytetrafluoroethylene (PTFE) / Polyurethane / Stainless Steel… X
Superalloys (Hastelloy™, Inconel) / Thermoplastic elastomer (TPE) / Titanium. X
Hastelloy™
-
C is a trademark of Haynes International
Chemraz™ is a registered trademark of Green, Tweed and Co.
Tygon™ is a registered trademark of Saint-Gobain Corporation
Monel™ is a registered trademark of Special Metals Corporation
Delrin™,Hypalon™, Hytrel™, Hypalon™, Polychloroprene™, Vamac™, Viton™, and Kalrez™ are trademarks of DuPont™

Photocopy this page of the manual and post on the door of an area or cabinet during every ozone sterilization cycle.
Sterilization Cycle in progress
Do NOT open door until this sign is removed
For users with remote control, press the remote button to immediately cancel the Ozone cycle and
shift the device into the Scrub cycle. Allow the FirstResponder device to fully run the Scrub cycle (the
digital display will go blank) before entering space, or opening cabinet that is being sterilized.
Name:________________________________________
Date:_________________________________________
Time Sterilization Cycle Started:____________
Expected Cycle Completion:________________
This manual suits for next models
1
Table of contents
Other Genlantis Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

PASCO
PASCO TD-8565 Product guide

Analytik Jena
Analytik Jena CyBio Well vario OL3381-25-300 operating manual

Analytik Jena
Analytik Jena Biometra TAdvanced Series operating manual

Civco
Civco ASTRA TEE A6172-01 installation manual

Kinematica
Kinematica POLYTRON PT 2500 E operating instructions

Ametek
Ametek FlarePro user manual