Gill Instruments WindObserver II User manual

WindObserver II
Ultrasonic Anemometer
(Pipe Mount)
Part 1390-PK-042
User Manual
Doc No. 1390-PS-0009
Issue 04
Gill Instruments Limited
Saltmarsh Park,
67 Gosport Street,
Lymington,
Hampshire.
SO41 9EG
UK
Tel: +44 (0) 1590 613500
Fax: +44 (0) 1590 613501
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备
Tel:18966887800 Web:www.xavery.cn Email:[email protected]

Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount Page 2 Issue 04
Doc. No. 1390-PS-0009 August 2009
Figure 1. WindObserver II Anemometer (Pipe Mount)
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备
Tel:18966887800 Web:www.xavery.cn Email:[email protected]

Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount Page 3 Issue 04
Doc. No. 1390-PS-0009 August 2009
Contents
1. FOREWORD............................................................................................5
2. INTRODUCTION......................................................................................5
3. FAST TRACK SET-UP............................................................................5
4. SPECIFICATION .....................................................................................6
5. PRE-INSTALLATION..............................................................................7
5.1. Equipment supplied..........................................................................................7
5.2. Installation requirements...................................................................................7
6. INSTALLATION..............................................................................................8
6.1. Installation Guidelines.......................................................................................8
6.2. Bench system test ............................................................................................9
6.3. Cabling.............................................................................................................9
6.4 . Connections..................................................................................................10
Factory Default Output Settings...........................................................................10
Connecting to a PC with a RS422 or RS485 input ...............................................11
Networking units..................................................................................................12
Connecting to a PCI Unit with RS422 input..........................................................13
Indoor Power and Communications Interface (Part 1189-PK-021)........................13
PCI Electrical Power Requirements.....................................................................13
PCI Connector Pin and Cable assignments.........................................................14
Anemometer connector – 15 way ........................................................................14
Auxiliary Analogue Outputs – 9 way (not applicable to Pipe Mount unit)...............14
DC Supply – 4 way..............................................................................................14
RS 232 Output – 9 way .......................................................................................15
RS422 Network In – 9 way Skt............................................................................15
RS422 Network Out – 9 way Plg..........................................................................15
6.5. Mechanical installation....................................................................................16
Location..............................................................................................................16
Orientation..........................................................................................................16
Mounting bracket.................................................................................................16
6.6. Power supplies...............................................................................................16
Alignment............................................................................................................17
7USING WITH THE GILL WINDDISPLAY ..............................................18
Connecting to a Gill WindDisplay.........................................................................19
8MESSAGE FORMATS & OPTIONS......................................................20
8.1. Wind Speed format.........................................................................................20
8.2. Output formats................................................................................................20
Low wind speeds.................................................................................................20
8.3. Networking......................................................................................................20
8.4. Units...............................................................................................................21
8.5. Output rate......................................................................................................22
Sample Average..................................................................................................22
8.6. Averaging .......................................................................................................22
8.7. Options...........................................................................................................22
Speed of Sound (SOS) and Temperature............................................................22
Heating...............................................................................................................22
45Offset............................................................................................................22
Vertical Output Padding.......................................................................................22
8.8. ASCII message format (UV and Polar) ...........................................................23
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备
Tel:18966887800 Web:www.xavery.cn Email:[email protected]

Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount Page 4 Issue 04
Doc. No. 1390-PS-0009 August 2009
ASCII UV format.................................................................................................23
ASCII Polar format .............................................................................................24
ASCII Polled (UV and Polar)................................................................................24
8.9. Binary message format (UV, Polar and Tunnel)..............................................25
Binary UV format.................................................................................................25
Binary Polar format..............................................................................................26
Binary Tunnel format...........................................................................................26
8.10. NMEA Format.................................................................................................27
9. CONFIGURING......................................................................................28
9.1 Configuring using HyperTerminal....................................................................28
9.2 Entering Configuration mode .......................................................................... 28
9.3 Returning to Measurement mode....................................................................28
9.4 Checking the configuration .............................................................................29
9.5. Changing settings...........................................................................................29
9.6. Configuration Settings.....................................................................................30
Ax – Speed of Sound (SOS) and Temperature....................................................30
BX - Baud Rate...................................................................................................30
Cx – Not Applicable.............................................................................................30
Dx – Diagnostic and Configuration Command......................................................31
Ex - Duplex mode................................................................................................31
Fx - Data and parity options.................................................................................31
Gx to Gxxx – Averaging (Digital Outputs Only) ....................................................31
Hx - Heating........................................................................................................32
Jx – Not Applicable..............................................................................................32
Kx – NMEA Settings............................................................................................32
Lx - ASCII Message terminator............................................................................32
Mx to MXX – Message Format ............................................................................32
Nx - Node Address..............................................................................................33
Ox – ASCII Output Format (Output String Padding).............................................33
Px - Output rate...................................................................................................33
Q – Measurement Mode......................................................................................33
Tx – Not Applicable.............................................................................................33
Ux – Digital Output Units.....................................................................................33
Vx - Vertical Output Padding ...............................................................................34
Xx - 45Alignment Offset/Inverted Operation.......................................................34
Yx - Not Applicable.............................................................................................34
Zx – Not Applicable.............................................................................................34
10. MAINTENANCE & FAULT-FINDING ....................................................35
10.1 . Cleaning .......................................................................................................35
10.2. Servicing.........................................................................................................35
10.3. Fault-finding....................................................................................................35
10.4. Returning unit.................................................................................................36
10.5 . Status (error) codes......................................................................................36
10.6. Bench test.......................................................................................................36
Integrity Check Chamber (ICC)............................................................................37
Integrity & Zero Wind Check................................................................................37
11. APPENDICES........................................................................................38
11.1 . Glossary & Abbreviations..............................................................................38
11.2 . Guarantee.....................................................................................................39
11.3 . Principle of operation....................................................................................40
11.4 . Electrical Conformity.....................................................................................41
EC DECLARATION OF CONFORMITY ACCORDING.........................................41
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备
Tel:18966887800 Web:www.xavery.cn Email:[email protected]

Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount Page 5 Issue 04
Doc. No. 1390-PS-0009 August 2009
1. FOREWORD
Thank you for purchasing the WindObserver II (Pipe Mount) manufactured by Gill
Instruments Ltd.
The unit has no customer serviceable parts and requires no calibration or maintenance.
To achieve optimum performance we recommend that you read the whole of this manual
before proceeding with use. Do NOT remove black “rubber” transducer caps.
Gill products are in continuous development and therefore specifications may be subject to
change and design improvements without prior notice.
The information contained in this manual remains the property of Gill Instruments and
should not be copied or reproduced for commercial gain.
2. INTRODUCTION
The Gill WindObserver II (Pipe Mount) wind sensor is a very robust, lightweight unit with
no moving parts, outputting wind speed and direction. The units of wind speed, output rate
and formats are all user selectable.
The WindObserver II (Pipe Mount) is supplied with de-icing heating on (recommended if
icing is likely).
The WindObserver II (Pipe Mount) can be used in conjunction with a PC, data-logger or
other device, provided it is compatible with the RS422 output. Multiple units can be
networked if required.
Alternatively, the WindObserver II (Pipe Mount) is designed to connect directly to the
Gill WindDisplay unit to provide a complete wind speed direction system without any
configuration required by the user.
The output message format can be configured in Polar, UV (2-axis), NMEA (0183
Version 3), Tunnel formats, and as either a Continuous output or Polled (requested by host
system), detailed in full in Section 8 Message Formats & Options.
WindObserver II (Pipe Mount) may be configured using WindCom software which is
available, free of charge, from the Gill website www.gill.co.uk, or by using standard
communications software on a PC. This is explained in Section 9 Configuring.
3. FAST TRACK SET-UP
If you are in a hurry to try out the WindObserver II (Pipe Mount), are familiar with Gill
equipment, and coupling to a PC using RS422, go to the following sections:
Section 6 Installation
Section 8 Message Formats & Options
Section 9 Configuring
After you have successfully set up the WindObserver II (Pipe Mount), we strongly advise
that you then go back and read the rest of the manual to ensure that you get the best results
from the WindObserver II (Pipe Mount).
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备
Tel:18966887800 Web:www.xavery.cn Email:[email protected]

Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount Page 6 Issue 04
Doc. No. 1390-PS-0009 August 2009
4. SPECIFICATION
Measurement
Output 1, 2, 4, 5, 8, 10Hz
Parameters UV, Polar, NMEA, Tunnel
Units m/s, Knots, MPH, KPH ft/min
Averaging Flexible 1-3600 seconds
Wind Speed
Range 0 - 65m/s (0-145mph)
Accuracy 2%
Resolution 0.01m/s
Offset ±0.01m/s
Direction
Range 0 - 359°
Accuracy ± 2°
Resolution 1°
Sonic Temperature Output
Range -40°C to + 70°C
Anemometer Status Supplied as part of standard message
Starting Threshold 0.01 m/s
Dead Band Wind Direction None
Power Requirement
Anemometer only 9-30 V DC (40mA @ 12VDC)
Heating - Optional Allow 3A @ 22v to 30v max AC rms or DC
Digital Output
Communication RS422, full duplex, network facility
Baud rates 1200, 2400, 4800, 9600, 19200, 38400
Formats 8 data, odd, even or no parity
Dimensions
Size 405mm x 210mm
Weight 1.5kg
Materials
External Construction Stainless Steel 316
Environmental
Moisture protection IP66 (NEMA4X)
Operating temperature -55°C to +70°C (heated version)
Humidity 0% to 100% RH
Precipitation 300mm/hr
EMC BS EN 50081-1: 1992 (Emissions class B)
BS EN 50082-2: 1992 (Immunity)
FCC class A
Icing MILSTD810E Method 521.1 Procedure 1
Standards Traceable to UK national standards
Site Calibration None required. (See Section 10.6 Zero Wind Check)
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备
Tel:18966887800 Web:www.xavery.cn Email:[email protected]

Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount Page 7 Issue 04
Doc. No. 1390-PS-0009 August 2009
5. PRE-INSTALLATION
5.1. Equipment supplied
WindObserver II (Pipe Mount) RS422 output and de-icing
and Installation kit
and WindObserver II (Pipe Mount) User Manual (this manual)
and Integrity Check Chamber (ICC) (Optional)
5.2. Installation requirements
Host system - One of the following:
PC with an internal or external interface compatible with the RS422 output from
the WindObserver II (Pipe Mount).
Networking - Multiple WindObserver II (Pipe Mount) units can be networked
Software - One of the following:
WindCom – A user-friendly package providing easy configuring of the
WindObserver II (Pipe Mount), for use on PCs running under Windows™ 98, 2000
or XP. Available to customers as a free download from www.gill.co.uk .
HyperTerminal (for Windows™ 9x and later), or Terminal (Windows™ 3.n),
normally already installed on a PC.
Other Terminal Emulation software packages
Cable and Junction box
To connect between the WindObserver II (Pipe Mount) and the host system. The unit is
supplied with a 4 metre flying unterminated cable.
See Section 6.4 for connection details
Mounting Pole
The WindObserver II (Pipe Mount) is designed to mount on a pole of nominal pipe size
diameter 50mm (See Fig 2.Page 17).
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备
Tel:18966887800 Web:www.xavery.cn Email:[email protected]

Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount Page 8 Issue 04
Doc. No. 1390-PS-0009 August 2009
6. INSTALLATION
Do NOT remove the black “rubber” transducer caps. Warranty is void if the
blue security seal is damaged or removed.
All the time the WindObserver II (Pipe Mount) is not in its final location, it
should be protected from damage by keeping it in the original packaging as
long as possible, treating it as a delicate instrument.
If an Intrinsically Safe (IS) device is required, the I.S. WindObserver must be
used - the WindObserver II (Pipe Mount) is NOT an IS device.
6.1. Installation Guidelines
The WindObserver II (Pipe Mount) has been designed to meet and exceed the stringent
standards listed in its specification. Operating in diverse environments all over the world,
WindObserver II (Pipe Mount) requires no calibration or adjustment whatsoever.
As with any sophisticated electronics, good engineering practice should be followed to
ensure correct operation.
Always check the installation to ensure the WindObserver II (Pipe Mount) is not
affected by other equipment operating locally, which may not conform to current
standards, e.g. radio/radar transmitters, boat engines, generators etc.
Guidelines –
Avoid mounting in the plane of any radar scanner – a vertical separation of
at least 2m should be achieved.
Radio transmitting antennas, the following minimum separations (all round)
are suggested
VHF IMM – 1m
MF/HF – 5m
Satcom – 5m (avoid likely lines of sight)
Ensure the product is correctly earthed in accordance with this manual
Use cables recommended by Gill, keeping the length below the maximum allowed
(See Section 6.3) Where the cables are cut and re-connected (junction boxes, plugs
and sockets) the cable screen integrity must be maintained, to prevent the EMC
performance being compromised.
Earth loops should not be created – earth the system in accordance with the
installation guidelines. (See Section 6.4)
Ensure the power supply operates to the WindObserver II (Pipe Mount)
specification at all times.
Avoid turbulence caused by surrounding structures that will affect the accuracy of the
WindObserver II (Pipe Mount) such as trees, masts and buildings. The World
Meteorological Organisation makes the following recommendation:
The standard exposure of wind instruments over level open terrain is 10m above
the ground. Open terrain is defined as an area where the distance between the
sensor and any obstruction is at least 10 times the height of the obstruction.
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备
Tel:18966887800 Web:www.xavery.cn Email:[email protected]

Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount Page 9 Issue 04
Doc. No. 1390-PS-0009 August 2009
6.2. Bench system test
Note: Prior to physically mounting the WindObserver II (Pipe Mount) in its final location,
we strongly recommend that a bench system test be carried out to confirm the system is
configured correctly, is fully functional and electrically compatible with the selected host
system and cabling (preferably utilising the final cable length). The required data format,
units, output rate, and other options should also all be configured at this stage. If an
Integrity Check Chamber (ICC) has been purchased refer to Section 10.6.
6.3. Cabling
Cable type
A RS422 compatible cable should be used, with the number of twisted pairs matching the
application.
Generic description
No. of pairs
Gill ref. Belden
ref. Batt Electronics
6 026-03508 9806 91021
Cable length
The typical maximum length at 9600 baud is 1km using the recommended cable. If any
problems of data corruption are experienced (due to, for example, a high local electrical
‘noise’ level), then a lower baud rate should be used. Alternatively, a thicker or higher
specification cable can be tried. . If a heated unit is used on longer cables than supplied,
consideration should be given to powering the heater using lower gauge wire or
paralleling wires to reduce volt drops.
Cabling and junction box
The WindObserver II (Pipe Mount) unit is fitted with a flying 6 pair cable attached (4
metre long). This must be terminated in a suitable manner to prevent moisture ingress.
If any cable is likely to be exposed to mechanical damage, it must be enclosed in a suitable
conduit or cable tray. The cable must be securely fixed with cable clamps or equivalent at
regular intervals such that the WindObserver cable gland does not support the cable
weight.
The gland area at the base of the WindObserver II (Pipe Mount) should not be directly
exposed to moisture, as whilst the gland is sealed when mated, the anemometer is vented
to air at the base to avoid pressure build up.
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备
Tel:18966887800 Web:www.xavery.cn Email:[email protected]

Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount Page 10 Issue 04
Doc. No. 1390-PS-0009 August 2009
6.4. Connections
Important Do NOT join any of the cores of the cable together. Joining cores
could damage the unit permanently. Any cores not used should be isolated.
Do NOT connect the unit’s 0V, heating –ve or digital 0V to the screen or earth.
The screens of each pair are joined together inside the anemometer - these should be
joined to the cable screen(s) of any further cable run. Avoid long grounding loops. Digital
OV should be used in conjunction with RS422 TX RX lines in order to improve noise
immunity.
Conductor colour Designation
Pair 1 green RS422_TXB (+)
black RS422_TXA (-)
Pair 2 white RS422_RXB (+)
black RS422_RXA (-)
Pair 3 red Supply V+
black Supply V-
Pair 4 blue Digital 0v
black Not used (isolate)
Pair 5 yellow Heater +
black Heater –
Pair 5 brown Heater +
black Heater –
Screens Chassis ground
Earthing or grounding
To ensure correct operation, and for maximum protection against lightning, the
anemometer MUST be correctly earthed (grounded) via its mountings. Inadequate
Earthing will degrade anemometer performance, particularly in the presence of radio
frequency interference.
The unit MUST be connected to an appropriate grounding point with a minimum of 6mm²
copper wire, via the M8 base screw. The cable screens must be joined with any cable
screen continuing from the unit’s cable via a junction box. The primary earth for the
anemometer must be provided via the base screws and not via the cable screens. See Fig.2.
Factory Default Output Settings
The factory default settings are:
A0 B3 C1 E1 F1 G0000 H2 J1 K1 L1 M2 NA O2 P5 T1 U1 V1 X2 Y1 Z1
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备
Tel:18966887800 Web:www.xavery.cn Email:[email protected]

Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount Page 11 Issue 04
Doc. No. 1390-PS-0009 August 2009
Connecting to a PC with a RS422 or RS485 input
The PC requires either an internal RS422 interface card, or an external RS422/485 to
RS232 converter, which will plug into the standard 9 way or 25 way RS232 serial port of
the PC or host equipment.
WindObserver II (Pipe Mount)
6 Pair Cable PC with RS422 or
RS485 input
Signal
names Pair
no. Colour Cable – twisted
pairs Signal names
TXB (+)
TXA (–) 1
1 Green
Black RXB (+)
RXA (–)
RXB (+)
RXA (-)
DIG 0V
2
2
4
White
Black
Blue
TXB (+)
TXA (–)
Sig. Gnd
Connected Internally Screen
wire Ground (Earth)
Supply V+
Supply V- 3
3 Red
Black
+
-
DC Power
supply 9 – 30V
See Section 6.6
Heater +
Heater - 5
5 Yellow & Brown
Black & Black
+
-
Heater power
supply
See Section 6.6
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备
Tel:18966887800 Web:www.xavery.cn Email:[email protected]

Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount Page 12 Issue 04
Doc. No. 1390-PS-0009 August 2009
Networking units
Before coupling units into a network:
Each device must be configured with a unique Unit Identifier (letter B to Z)
It must be configured to a tri-state polled mode M3 or M4.
See Section 9 - Configuring
WindObserver II (Pipe Mount)
6 Pair Cable PC with RS485 card
Signal
names Pair
no. Colour Cable – twisted
pairs Signal names
TXB (+)
RXB (+) 1
2 Green
White T/RXB (+)
TXA (-)
RXA (-)
DIG 0V
1
2
4
Black
Black
Blue
T/RXA (-)
Sig. Gnd
Connected Internally Screen
wire Ground (Earth)
Supply V+
Supply V- 3
3
Red
Black
+
-
DC Power
supply 9 – 30V
See Section 6.6
Heater +
Heater - 5
5
Yellow & Brown
Black & Black
+
-
Heater power
supply
See Section 6.6
Note: Each unit in the network will require its own power supplies.
Please refer to section 8.3.
To Next Unit
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备
Tel:18966887800 Web:www.xavery.cn Email:[email protected]

Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount Page 13 Issue 04
Doc. No. 1390-PS-0009 August 2009
Connecting to a PCI Unit with RS422 input
WindObserver II (Pipe Mount)
6 Pair Cable PCI 15 Way D Type
Connector
Signal
Names Pair
No. Colour Pin Nos.
Signal Names
TXB(+) 1 Green 2 RXB(+)
TXA(–) 1 Black 9 RXA(–)
RXB(+) 2
3 TXB(+)
RXA(-) 2 Black 10 TXA(–)
Digital 0v
4
Blue
4 Signal Ground
Supply V+
3 Red 5 POWER V +
Supply V-
3 Black 11 POWER V-
Heater +
Heater –
5
5
Yellow & Brown
Black & Black
+
–
Heater power
supply
See Section 6.6
Indoor Power and Communications Interface (Part 1189-PK-021)
Environmental
The desktop PCI is intended for use in a protected environment and provides interfacing to
a single anemometer. The PCI is not intended for outdoor use.
PCI Electrical Power Requirements
1) 100Vac – 120Vac, 10VA for the 115V switch position
2) 200 Vac – 250Vac, 10VA for the 230V switch position
3) 9-30Vdc, 120mA max (anemometer)
8-15Vdc, 50mA max (interface)
3 Fuses protect the unit:
1. Externally accessible 315mA(T) for 115/230v operation
2. Internally accessible 500mA (T) AUX
500mA (T) ANEM
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备
Tel:18966887800 Web:www.xavery.cn Email:[email protected]

Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount Page 14 Issue 04
Doc. No. 1390-PS-0009 August 2009
PCI Connector Pin and Cable assignments
Anemometer connector – 15 way
Pin Designation
1 Chassis
2 RS422_RXB (+)
3 RS422_TXB (+)
4 Digital 0V, Signal ground
5 Supply V+
6 Not used
7 Not used
8 Not Used
9 RS422_RXA (-)
10 RS422_TXA (-)
11 Supply V-
12 Not used
13 Not Used
14 Not used
15 Not used
Auxiliary Analogue Outputs – 9 way (not applicable to Pipe Mount unit)
Pin Designation
1 Earth
2 Analogue Ground 1
3 Analogue Ground 2
4 Analogue Ground 3
5 Analogue Ground 4
6 Analogue Output 1
7 Analogue Output 2
8 Analogue Output 3
9 Analogue Output 4
DC Supply – 4 way
Pin Designation
1 Interface V+
2 Interface V-
3 Anemometer V-
4 Anemometer V+
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备
Tel:18966887800 Web:www.xavery.cn Email:[email protected]

Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount Page 15 Issue 04
Doc. No. 1390-PS-0009 August 2009
RS 232 Output – 9 way
Pin Designation
1 DCD
2 RX data
3 TX data
4 DTR
5 Signal Ground
6 DSR
7 RTS
8 CTS
9 RI
RS422 Network In – 9 way Skt
Pin Designation
1 Not connected
2 Not connected
3 Not connected
4 RS422_RXA (-)
5 Signal ground
6 RS422_TXA (-)
7 RS422_TXB (+)
8 RS422_RXB (+)
9 Chassis
RS422 Network Out – 9 way Plg
Pin Designation
1 Not connected
2 Not connected
3 Not connected
4 RS422_RXA (-)
5 Signal ground
6 RS422_TXA (-)
7 RS422_TXB (+)
8 RS422_RXB (+)
9 Chassis
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备
Tel:18966887800 Web:www.xavery.cn Email:[email protected]

Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount Page 16 Issue 04
Doc. No. 1390-PS-0009 August 2009
6.5. Mechanical installation
Before installing, a bench system test is recommended (see Section 6.2).
All the time the WindObserver II (Pipe Mount) is not in its final location, it
should be protected from damage by keeping it in the original packaging as
long as possible, treating it as a delicate instrument.
Location
It is important to ensure that the WindObserver II (Pipe Mount) is mounted in a position
clear of any structure, which may obstruct the airflow or induce turbulence.
Do NOT mount the WindObserver II (Pipe Mount) in close proximity of high-powered
radar or radio transmitters. A site survey may be required if there is any doubt about the
strength of external electrical noise. (See Section 6.1)
Orientation
The WindObserver II (Pipe Mount) must be mounted vertically with the cable exit at the
bottom.
Mounting bracket
It is recommended that the WindObserver II (Pipe Mount) mounting pipe be manufactured
from stainless steel 316, to prevent possible galvanic corrosion.
Stainless steel 316 fixings (supplied) must be used.
6.6. Power supplies
All WindObserver II (Pipe Mount) units
Voltage 9 to 30v DC
Current 60mA max. 50mA average
Heating Voltage 22V to 30V (max) AC RMS or DC.
Current allow for 3A.
The AC Supply must be isolated from Mains Supply.
The heating module requires a separate power supply.
Heater Cable length should be minimised to avoid cable volt drops and ensure
maximum voltage received at the Anemometer.
Each transducer is heated independently and will be active when ambient
temperature drops below approximately +15ºC each transducer will de-activate
when +25ºC threshold is reached.
The WindObserver II has reverse polarity protection.
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备
Tel:18966887800 Web:www.xavery.cn Email:[email protected]

Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount Page 17 Issue 04
Doc. No. 1390-PS-0009 August 2009
Alignment
The anemometer should be set to point North (or to another known reference direction),
using the North alignment indicator on the base of the instrument. See Figure 2.
Figure 2 North marker and Mounting
(See fig. 4, on page 31 for u, v definition)
4M Cable
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备
Tel:18966887800 Web:www.xavery.cn Email:[email protected]

Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount Page 18 Issue 04
Doc. No. 1390-PS-0009 August 2009
7 USING WITH THE GILL WINDDISPLAY
The WindObserver II is designed to link directly to the Gill WindDisplay unit to provide a
complete wind speed and direction system.
When coupled to a WindDisplay, the WindObserver II (Pipe Mount) can be used as
supplied,
however if a fault occurs, the WindDisplay may lock into the last valid reading. Re-
configuring the Wind Observer (Pipe Mount) to Fixed Field Output (O2) will ensure
that any fault is flagged on the WindDisplay.
After coupling to a WindDisplay, the Wind Speed units and the Averaging period can be
selected using the WindDisplay controls. See the WindDisplay User Manual.
Note that although the WindDisplay can display wind speed in various units, these are
calculated within the WindDisplay. The data coming to the WindDisplay must be in
metres/sec (the factory default output setting).
Figure 3 The WindDisplay
The WindDisplay has been designed to provide anemometer power (approximately 15V
@40mA).
Note the WindObserver II (Pipe Mount) de-icing heater will need a separate power supply.
(See Section 6.6).
NOTE: - If the WindDisplay is configured for NMEA mode then the WindObserver II
(Pipe Mount) must also be configured for NMEA mode and 4800-baud operation
(configuration settings M5 and B2).
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备
Tel:18966887800 Web:www.xavery.cn Email:[email protected]

Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount Page 19 Issue 04
Doc. No. 1390-PS-0009 August 2009
Connecting to a Gill WindDisplay
For further details see Section 7 Using With The Gill WindDisplay,
and the WindDisplay User Manual for the method of operation.
Notes
Use the WindObserver II (Pipe Mount) as supplied in the factory default settings – i.e. do
not reconfigure.
The WindDisplay can provide power for the WindObserver II (Pipe Mount).
WindObserver II (Pipe Mount)
6 pair cable WindDisplay
Signal names Pair
no. Colour Cable –twisted
pairs Signal
names Terminal
nos.
TXB (+)
TXA (–) 1 green
black TXD +
TXD – RS422+
RS422-
CONNECTED INTERNALLY Screen
wire Ground (Earth)
Supply V+
Supply V- 3 red
black +
O 2
1
Heater +
Heater - 5 & 5 yellow & brown
black & black
+
-
Heater power
supply
See Section 6.6
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备
Tel:18966887800 Web:www.xavery.cn Email:[email protected]

Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount Page 20 Issue 04
Doc. No. 1390-PS-0009 August 2009
8 MESSAGE FORMATS & OPTIONS
On first applying power to the WindObserver II (Pipe Mount), it will be in ‘Measurement
Mode’, and it will output wind measurement information in one of nine configurations as
described below.
Setting the output format, units, other parameters, options and the communication settings
are all carried out in the alternative ‘Configuration Mode’.
See Section 9 CONFIGURING for details of how this is done.
The factory default settings are shown here in bold, and for convenience some
‘Configuration codes’ (as used to set the configuration) are shown in blue boxes.
For example M3.
8.1. Wind Speed format
The wind speed measurements can be output in one of the following formats:
UV, Polar, Customer formats (NMEA and Tunnel).
8.2. Output formats
The UV and Polar wind speed parameters are output in either ASCII or binary.
These parameters can be transmitted continuously or polled from the user.
Polar is also available in continuous NMEA format.
Output format Output Tri-state o/p Configuration code
Continuous No M1
ASCII UV Polled Yes M3
Continuous No M2
ASCII Polar Polled Yes M4
Binary UV Continuous No M7
Binary Polar Continuous No M8
Binary Tunnel Continuous No M6
NMEA Continuous No M5
Low wind speeds
Whilst the wind speed is below 0.05 metres/sec, the wind direction will not be calculated,
in CSV mode no wind direction figure is shown, in Fixed Field mode the last known good
direction figure is shown. All other output parameters will continue to update at the output
rate.
8.3. Networking
Gill Proprietary Network
Each anemometer connected to the network MUST be given a unique Unit Identifier
(Letters A to Z), and set to a polled tri-state mode (UV or Polar) (M3 or M4).
In these modes the communications transmit line is tri-state unless the anemometer is
responding to a command.
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备
Tel:18966887800 Web:www.xavery.cn Email:[email protected]
Other manuals for WindObserver II
2
This manual suits for next models
1
Table of contents
Other Gill Instruments Measuring Instrument manuals

Gill Instruments
Gill Instruments HS-50 User manual

Gill Instruments
Gill Instruments 1360-PK-022 User manual

Gill Instruments
Gill Instruments WindObserver II User manual

Gill Instruments
Gill Instruments Windobserver 65 User manual

Gill Instruments
Gill Instruments WindObserver II User manual

Gill Instruments
Gill Instruments MetPak II User manual

Gill Instruments
Gill Instruments R3-50 User manual

Gill Instruments
Gill Instruments 1561-PK-020 User manual