Rev. 6-501 Model P53 pH/ORP Analyzer (universal-moun 1/2 DIN)
5
CONDENSED OPERATING INSTRUCTIONS
This manual con ains de ails for all opera ing aspec s of he ins rumen . The following con-
densed ins ruc ions are provided o assis you in ge ing he ins rumen s ar ed up and
opera ing as quickly as possible. These condensed instructions only pertain to basic pH
measurement operation using a GLI Differential pH sensor. To measure ORP, use a con-
ven ional combina ion elec rode or use specific fea ures of he ins rumen , refer o he
appropria e sec ions in his manual for ins ruc ions.
A. CONNECTING SENSOR/CONFIGURING SENSOR TYPE AND TEMPERATURE ELEMENT
1. Af er properly moun ing he analyzer (PART TWO, Sec ion 2), connec he GLI Differ-
en ial Technique pH sensor, ma ching wire colors o erminals as indica ed:
Sensor
Wire Colors
Analyzers wi h
“B” Prefix Serial No.
Analyzers wi h
“A” or No Le er Prefix Serial No.
Whi e Terminal #13 on TB1 Terminal #13 on TB1
Black Terminal #14 on TB1 Terminal #14 on TB1
Yellow Terminal #16 on TB1 Terminal #16 on TB1
Shield Terminal #18 on TB1 Terminal #11 on TB1
Green Terminal #20 on TB1 Terminal #20 on TB1
Red Terminal #22 on TB1 Terminal #22 on TB1
2. The analyzer is fac ory-se o use a GLI Differen ial Technique pH sensor. To use a
conven ional combina ion elec rode, you mus change he sensor ype (see PART
THREE, Sec ion 4.2, subheading “SELECT SENSOR Type”).
3. The analyzer is fac ory-se for au oma ic empera ure compensa ion using he NTC 300
ohm empera ure elemen buil in o all GLI Differen ial sensors (excep GLI’s pure wa er
pH sensor sys em 6006P4-2000 which uses a P 1000 RTD). When you wan fixed
MANUAL empera ure compensa ion or if you are using a sensor wi h a differen em-
pera ure elemen , you mus change he empera ure elemen ype (see PART THREE,
Sec ion 4.2, subheading “Selec TEMP ELEMENT Type”).
B. CONNECTING LINE POWER
Important: Follow ins ruc ions in PART TWO, Sec ion 3.7 o connec line power o he analyzer.
C. ADJUSTING DISPLAY CONTRAST
Ambien ligh ing condi ions may make i necessary o adjus display con ras o improve
visibili y. Wi h he MEASURE screen displayed, press and hold he ENTER key and
simul aneously press he
×
×
or
Ø
Ø
key un il a aining he desired con ras .
D. CONFIGURING BUFFER TYPE/CALIBRATING THE ANALYZER
The analyzer mus be calibra ed so ha measured values will correspond o ac ual process
values. Before calibra ing for he firs ime, selec he buffer se you in end o use. Then,
calibra e using he recommended “2 POINT BUFFER” me hod which provides he mos ac-
cura e pH measuremen s.
1. The analyzer is fac ory-se for he common 4.00, 7.00, and 10.00 pH buffer se . For
DIN 19267 s andard value buffers you mus change he buffer se (see PART THREE,
(con inued on nex page)
NOTE: For GLI Differen ial
sensors ha have wo
shield wires, always
connec he ou er
shield o an ear h
ground erminal and
he inner shield o he
“shield” erminal.