
012019-GPSIMOD-IOM REV C
OPERATION
1. Once the voltage selector switch has been set, HV wire(s) connected and bar(s) mounted, turn the power switch to the “ON”
position. When the switch is turned “On” the “Power On” light will illuminate letting the end user know power is supplied and
the GPS-iMOD system is energized. Note: If a door switch, fan interlock switch or air flow switch are in series with the power,
the system may not turn on until all safeties are closed. When power is supplied and the internal or optional remote mounted
GPS-iDetect-P is sensing output, the “Plasma On” light will illuminate.
2. The internal BAS Alarm Contacts will close proving system operation to the BMS.
2. Using a standard non-contact voltage meter, one can place it near the ion needles and prove there is ion output. An
optional ion meter can be purchased from GPS and actual values may be measured. A permanent mount ion detector with BAS
interface may be provided as an option for 24/7 output monitoring.
STARTUP/TESTING
1. Once the entire system is mounted and wired, energize the system by turning the on/off switch to the “on” position. The
green power “ON” LED and “PLASMA ON” LED should illuminate. If the PLAMA ON does not illuminate, confirm the red jumper
is installed between C and NO on the GPS-iDetect-P terminal block. If an external sensor is used, confirm it is on and operating.
2. Using a high voltage probe similar to a BK Precision Plus 28A, connect the probe to a multi-meter, connect the ground clamp,
and measure the AC high voltage (AC not DC) at the 6 inch modular stingers. Insert the tip of the probe into one of the brush
clusters and confirm the voltage is greater than 4,000 VAC. Please note, most probes provide a 1000:1 step down. As an
example, a display of 4.0 would be 4.0 x 1000 = 4,000 VAC. The typical range of voltage is 4,000VAC to 7,000VAC and the actual
voltage will depend on the length and quantity of bars attached.
MAINTENANCE
The GPS-iMOD system has been designed for minimum maintenance. Below are the steps to ensure a long trouble-free life:
1. On an annual basis, turn off power and use isopropyl alcohol and a nylon (wire free) brush to gently clean the needles.
2. Use a soft cloth with isopropyl alcohol and wipe any debris off the GPS-iMOD outer bar and spaces between needle
housings.
3. Note, in smoking applications, the GPS-iMOD will require more frequent cleaning based on the filter efficiency prior to the
GPS-iMOD system.
TROUBLESHOOTING
1. Power supply “Power On” light not illuminated when the power switch is in the “On” position.
1A. Check that all safeties are closed and there is primary power applied to the power supply. If light will still not illuminate,
remove power and energize after five minutes. The GPS-iMOD system uses an internal auto-reset circuit breaker. Either a
voltage surge or high temperature/load condition can trip the circuit breaker. If the “Power On” light is off and the “Plasma
On” light is “On”, the “Power On” light may have burned out. Contact your local Representative or the GPS factory to have
your power supply repaired or replaced.
2. No Ionization Output.
2A. Confirm the power supply is operating properly as shown in step 1A above. Confirm the HV cables are inserted and
secured properly. Confirm the needles are clean and free of debris.