
Customer Service, Sales and Technical Support:
Phone: 1 (800) 394-4562 • 1 (201) 599-1400
Fax: 1(201)599-1406•E-mail:mail@globescientic.com
©2019GlobeScienticInc.Allrightsreserved.TheGlobeScienticlogoisaregisteredtrademarkofGlobeScienticInc.GCM-24Guide19107
GCM-24
Microcentrifuge
Quick Start Guide
For the full User Manual, please visit www.globescientic.com
CONTROL PANEL
Speed Display
Temperature Display
Time Display
1
2
34
5
IMPORTANT: Plug unit into appropriate main supply based
on the rated voltage.
Pulse � � � � � � � � � � � � � Press to accelerate and hold at the set
speed. The centrifuge stops immediately
once this button is released.Radius
Door� � � � � � � � � � � � � � Press to open the lid. This button is not
enabled while the centrifuge is running.
Start / Stop � � � � � � � Press to start the centrifuge. Press to
stop if centrifuge is already running.
Select� � � � � � � � � � � � � Press to choose the parameter you wish
to modify.
Adjustment Knob Rotate clockwise to increase parameter
values. Rotate counterclockwise to
decrease parameter values. Press to
switch between speed and RCF.
OPERATION
1
2
3
4
5
Rotor Tube Adapter Maximum Maximum Centrifugal Maximum
Type Type Required Speed (rpm) Radius (cm) RCF (×g)
2.0/1.5mL Tube — 15000 8.5 21380
0.2mL PCR Tube GCM-24-AD2 15000 6.9 17350
0.5mL PCR tube GCM-24-AD5 15000 7.6 19100
0.5mL PCR tube 15000 8.5 21380
0.2mLPCRStrip GCM-24-AD52 15000 7.6 19100
0.2mLPCRStrip — 6.5/7.2 16350/18100
GCM-24-05ML
(optional)
GCM-24-2ML
GCM-24-PCR
(optional)
The centrifuge rotor can separate samples whose density is lower than 2.0g/ml. If the separated sample’s density is over 2.0g/ml, please
calculate the allowable speed using the following formula:
Allowable Speed (rpm) = Maximum Speed × (2�0(g/ml) / Sample Density (g/ml))1/2