Attach with correct orientation, and right side up.
Incorrect attaching will cause wrong temperature data.
Put the sensor on the connector pin firmly. To avoid
damaging the sensor, do not bend or deform it.
When using the Sensor-life Counter function (described later),
reset the sensor-damage value in measuring mode after
replacing with a new sensor.
Be sure to attach the sensor properly
before starting measurement. When the
alarm (shown left) appears, it means
that the sensor is not attached properly or measured tip's
temperature is out of range. Check for a break in the sensor, or
check the settings of the device.
The sensor life "200 times when counted in units of 3-seconds'
use at temperature 350˚C(662˚F)." is just a tentative indication.
The actual sensor-life can be changed by your operating condi-
tions. When replacing the sensor, reset the damaged value.
Figure out the sensor life and change its set value for effective use.
Flux residue on the sensor chip, or a stained tip will reduce
the accuracy of the measuring sensor. Clean them regu-
larly with CP series (solder wick). Too little solder on the chip
may result in wrong measuring data. Use the proper solder
amount for your tip size.
The sensor reception will become worse because of the deg-
radation of the chip after repeated use. Replace the degraded
sensor with a new one to ensure accurate measuring. Please
use the below Sensor-Alarm Function for replacement timing.
To turn off the device, return to measuring
mode. In setting mode, the power off (& Auto
Power-Off) function does not work.
TAIYO ELECTRIC IND.CO.,LTD.
T M -10 0
KEEP THIS MANUAL FOR FUTURE REFERENCE
Thank you for buying Tip Thermometer TM-100.
Read this Owner's Operation Manual before using
your tip thermometer.
NOTE
Careless handling may result in injury to
yourself or to others, physical damage,
or a system failure of this device.
A note or word of advice.
In this manual, the following safety marks are used.
SAFETY MARK DEFINITIONS
Too much pressure to the sensor risks
damage. Apply the tip on the chip gently.
1. The measuring tip is very hot. Handle with care.
Careless handling may result in fire or personal injury.
2.
Be sure not to heat the plastic case such as by
contact with the hot tip, except to the sensor, or by
blowing warm air onto the body. Excess heat will
melt or damage the case.
3.
Do not open the device. Sensitive parts might be
damaged.
4.
The LED can flash even when power is off. This is not
a system error. This is an effect of static electricity.
5.
Avoid impacts to prevent damage to, or deterioration of,
the product during transportation. Be especially careful
to avoid deformation or breakage from dropping.
Sensor (3pcs)
TM-100S
Sensor Probe (for solder pot)
TM-100SP
1. SPECIFICATIONS
2. PACKAGE CONTENTS / NAMES OF PARTS
3. HOW TO ATTACH / DETACH THE SENSOR
4. HOW TO MEASURE THE TEMPERATURE
5. FUNCTION AND SETTING
6. REPLACEMENT PARTS / OPTION
7. OPERATIONS
NOTE
NOTE
NOTE
CAUTION
CAUTION
CAUTION
CAUTION
NOTE
NOTE
NOTE
Printed in Japan, JANUARY 2014
MODEL TM-10 0
Power Supply AA battery LR6 x 4pcs: 6V
Dimension
83(W) ✕41(H) ✕140(D)mm
Weight 150g (w/o battery)
Temperature Resolutions
1˚C
Ranges
w/sensor (TM-100S): 0-550˚C(32-932˚F)
w/optional probe (TM-100SP): 0-700˚C(32-1292˚F)
Accuracy ±0.5% of the full scale (w/o sensor error)
Display LED indication
Operating Environment 0-50˚C, 20-85%RH (no condensation)
Model TM-100S
Sensor Type Type K ( CA type )
Size 9(W)
✕
45(H)
✕
3(D)mm
Weight
0.8g
SENSOR
UNIT
HOW TO ATTACH
1. Attach the red sheath side of
the sensor to the connecting
pin on the red silicon plate of
the device.
A. Measuring Mode
After power-on, always start with the measuring mode. Measuring
mode has the following functions. (Digit change/setting can be config-
ured in setting mode.)
Peak-Hold Function (factory default setting: OFF)
In measuring mode, press the PEAK HOLD/MODE SET➌ button, then
the HOLD ON lamp ➏will light up. Retains display of peak temperature.
Press the PEAK HOLD/MODE SET➌button again to unset the peak
hold function.
Automatic Power-Off function
(factory default setting:
60 seconds
)
The device will power off automatically when 100˚C or lower is continu-
ously measured for more than the preset time. The setting time for the
Auto Power-Off can be changed in units of 30 seconds. (MAX 300sec-
onds) When the displayed temperature is under 100˚C in measuring
mode, the count time will be reset as the result of pressing SENSOR
RESET➋button within the preset time. And extend the time until Auto
Power-Off.
Sensor-Life Counter Alarm
(PAT.P) (factory default setting: 200 times)
Weigh and integrate the sensor damage by our own calculational
procedure by counting in units of 3-seconds' use at temperature
350˚C(662˚F). The alarm lamp➎will flash when the counter exceeds
the set number of counts, to indicate the approx. replacement-timing of
the sensor. To turn off this function, set the preset value at 0.
To reset (return to 0) the integrated value of the damage, press and hold
the SENSOR RESET➋for over 3 seconds in measuring mode.
Battery Alarm
The battery lamp will flash when the battery voltage drops lower than speci-
fied value to indicate the approx. replacement-timing of the battery.
2. Press the attaching/detaching button on the left. The connecting pin & the
red silicon plate will slightly move toward the opposite side. Then attach
the white sheath side of the sensor to the other connecting pin of the
device.
HOW TO DETACH THE SENSOR
Hold the device firmly and turn it
upside-down. Press the left button
several times, then the sensor will
detach.
1. Press POWER to display the set-unit Celsius or
Fahrenheit. When the room temperature appears the
device is ready for measurement. (measuring mode)
2. Put the tip to the sensor chip, applying fresh solder to both parts.
Please read when the displayed temperature is stable.
Power Sensor Reset and Setting
Battery Alarm Lamp (Flash)
Peak Hold / Mode
Sensor-life Alarm Lamp (Flash)
Hold Lamp (light up)
Red silicon plate
Sensor attaching
/detaching button
Press
Red
Key
Operation Turn Display Function & Range
Set the time (seconds)
until Auto-Power OFF.
Ranges: 0-300 (in units of
30seconds) Default setting:
60s. Turns off at 0.
Set the temperature
display in Celsius (˚C) or
Fahrenheit (˚F).
The setting ˚C or ˚F Lights up.
Default setting: ˚C.
Set the sensor-life value.
Ranges: 0-500 (in units of
50 times) Default setting:
200 times. Turns off at 0.
Display the sensor-
damage value.
Resetting displays 0. Default
setting 0.
PEAK HOLD ON/OFF
Extension of Automatic
Power-Off time
Sensor Damage
Reset (Return to 0)
➔➔➔
➔
Return
to start
Shift
Setting value
change
B. Setting Mode
To enter setting mode, press and hold the PEAK HOLD/MODE SET➌
button for over 3 seconds in measuring mode. In setting mode, every
pressing of the PEAK HOLD/MODE SET➌ button will shift the set-
B.Setting Mode
A. Measuring Mode
PEAK HOLD/
MODE SET
Press the
After power-on, always start with the measuring mode.
To switch between setting/measuring mode, press and
hold
PEAK HOLD/MODE SET
for over 3 seconds.
SENSOR RESET
SENSOR RESET
Press the
CALIBRATION
Accuracy assured for 1 year from the day of purchase. After 1 year
we also offer calibration for a fee. Please contact your nearest
distributor.
■ FRONT PANEL
TM-10 0SP
ting function. To change the setting value, press theSENSOR
RESET➋button.To return to measuring mode, press and hold the
PEAK HOLD/MODE SET➌button for over 3 seconds in setting
mode.
The enclosed batteries are for testing only.
*Over 1000˚F, temperature is indicated by F-1 and the last three digits of degrees ˚F.
Sensor attaching
/detaching button
Red silicon plate Sensor
connecting pin
Battery case
Name plate label (English)
Red mark
Serial number label
TOP
OPEN
BOTTOM
■SENSOR
■UNIT
■ AA battery LR6 1.5V 4pcs
■ Instruction manual
(Japanese/English, English/Chinese)
■ User registration card / Warranty card
(Japan only)
■ Name plate label (Chinese)
Chip
White sheath
FRONT BACK
Red sheath
PAT.P
3pcs
TIP THERMOMETER
4. HOW TO MEASURE THE TEMPERATURE
3. HOW TO ATTACH / DETACH THE SENSOR
5. FUNCTION AND SETTING
1. SPECIFICATIONS
TABLE OF CONTENTS 2. PACKAGE CONTENTS / NAMES OF PARTS
6. REPLACEMENT PARTS / OPTION
7. OPERATIONS
Please refer to [7. OPERATIONS]
for each function.
Turn over
Press
Press the Press the Press and hold the
over3seconds.
Protective clear seals on the key panel & silver plate.
The sensor and the batteries are consumable items.
OPERATION MANUAL