3.Regular checking /adjusting
3-1 Checking and adjusting of the circular level
(1)Adjust the leveling screws to center the bubble in the circular level.
(2)Turn the instrument 360°. If the bubble is still in the center, it means
that the circular level is workable.
(3)If the bubble shifts away from the center, please adjust according to
following methods:
Move the bubble one half of the shift to the center of the circular level
by adjusting the leveling screws. Then get the bubble in the center of
the circular level by adjusting the circular level adjust-screw with the
hexagonal wrench. Repeat the above checking and adjusting, until the
bubble does not shift when the instrument turns to any direction. (See
fig.7)
3-2 Checking of the compensator
(1)Center the bubble in the circular level, and aim the telescope at a
target about 70m away.
(2)Turning the leveling screws to make the bubble slide out about 1/4 of
the circular to any direction, if there is no deviation between the image
of target and the horizontal cross-lines, it means that the working range