Installation
The cabinet must only be used indoors and be placed in an area that is dry and well ventilated. It
must not be positioned in direct or indirect sunlight, near warm appliances or locations with
drafts.
The cabinet is designed to perform in an ambient temperature of between +10°C (minimum) and
+32°C (maximum) and a maximum ambient humidity of 60%.
By placing the cabinet in an environment above 60% ambient humidity it may be necessary to
wipe the glass door of condensation.
The cabinet can be installed freestanding, against a wall or built in. If built in refer to instructions
below.
Important!
The cabinet must have a minimum 30 mm uninterrupted ventilation and free air circulation
underneath and on top of the cabinet.
The drawings on Page 4 show how much air space is needed.
Fig 1. Shows the cabinet freestanding or against a wall.
Fig 2. Shows the cabinet build in.
The dimensions must be maintained to ensure adequate air circulation.
If the cabinet is installed inside a construction and the ventilation dimensions cannot be strictly
observed good air circulation must be provided by making a large hot-air escape channel in the
construction - with particular attention given to the top ventilation of the cabinet.
IMPORTANT !!
In order to ensure good ventilation and not to void the Warranty contact ICS PACIFIC if further
information is required.