page 8
B 600 M
RUN MODUS : N FIXED
B 600 - EXPLANATION OF TEST PARAMETERS
Press ( Next )
to move down in this menu.
Press
SELECT ACTUAL RUN PARAMETERS
MST( STEP ) : 4 sec
LIFTANGLE : 50 °
N ( STEP ) : / h
MS WINDOW : 100 °
21600
to move up in this menu.
Press ( Menu + ) or ( Menu - )
to change a value in the marked line.
Press and hold ( MENU + ) to encrease
Press and hold ( MENU - ) to decrease
B 600 M
RUN MODUS : AUTMATCL
MST( STEP ) : sec
MST( FINE ) 4
Measuring time : ( MST )
The result of time deviation will be
desplayed after the set value of ( MST )
and will be the meanvalue of the ( MST )
SELECT ACTUAL RUN PARAMETERS
LIFTANGLE : 50 °
MS WINDOW : 100 ° va
are:
4, 8, 12, 30, 60, sec.
values in MST ( FINE ) are:
2, up to 960 sec. in steps of 2
Lift angle: (50 - 52°)
50
B 600 M
50
RUN MODUS : AUTMATCL
MST( FINE ) : 4 sec
LIFTANGLE : °
°
The value of lift angle depends on
the construction of the watch. It is used
in the B 600 to calculate the amplitude.
Most watches have a lift angle of 50°-52°.
It can be set from 30° up to 70°.
If the value of lift angle is set lower than
the value of the measured watch, the
result of amplitude will be lower than
SELECT ACTUAL RUN PARAMETERS
it should be. If it is set higher the result of
amplitude will be higher.
Beat frequency N (STEP) or N (FINE)
21600
50
B 600 M
RUN MODUS : AUTMATCL
MST( STEP ) : 4 sec
LIFTANGLE : 50 °
MS WINDOW: 100°
the beat frequency is automatically
detected and set from the B 600 to a
standard beat. Standard beats are :
3600, 4800, 7200, 7920, 8100,
8640. 9360, 9440, 9720, 9760,
14400, 16200, 18000, 19800, 21600,
Operating - Manual B 600 M / Q
SELECT ACTUAL RUN PARAMETERS
,
In RUN MODUS ( N FIXED)
The beat can be set manually from
3240 up to 39600 in N ( FINE )