H53.0.X1.6C-07 Operating Manual GFTH 95 Page 4 of 6
2. If there is a risk whatsoever involved in running it, the device has to be switched off im-
mediately and to be marked accordingly to avoid re-starting.
Operator safety may be a risk if:
- there is visible damage to the device
- the device is not working as specified
- the device has been stored under unsuitable conditions for a longer time.
In case of doubt, please return device to manufacturer for repair or maintenance.
3. Do not use these products as safety or emergency stop devices or in any other applica-
tion where failure of the product could result in personal injury or material damage.
Failure to comply with these instructions could result in death or serious injury and mate-
rial damage.
3 Product Specification
3.1 Scope of supply
The scope of supply includes:
Device GFTH 95, incl. 9V battery block
Operation manual
3.2 Operation and maintenance advice
•Battery operation:
If „BAT„ appears in the left hand corner of the display, the battery is used up and needs to be replaced.
However, the instrument can still be operated for a short period of time. If „bAt“ is displayed in large let-
ters the battery is finally used up and has to be replaced. Measuring is no more possible.
The battery has to be taken out, when storing device above 50 °C.
We recommend taking out battery if device is not used for a longer period of time.
The humidity and temperature sensor are located in the protruding sensor head. Make sure that the
openings are not soiled. In case of soiling do not try to remove the dirt. Improper use may damage the
sensors. Prevent any mechanical impact on the device as this may also lead to a destruction of the
sensors (supporting material glass or ceramics)!
In the sensor area there is ESD-sensitive. Never touch or hold sensor head!
In order to guarantee accurate measurements the device has to be adjusted to ambient temperature.
If necessary wait for the temperature of the device to adjust to the ambient temperature.
If this is not possible, please proceed as follows:
Move device around while holding it at an arm's length (fanning), to speed up the exchange of air and
the temperature adjustment process. As soon as the display value is steady, readings can be made.
This procedure shall be used for temperature and for humidity measurements.
If you hold the device during measurements your body heat as well as your breath will change the
temperature and the humidity. In order to minimise these factors, hold the device at as great a distance
from the sensor as possible. Also try to prevent your breath coming into contact with the sensor. For
optimum measuring results put down device and read display from a certain distance as soon as a
constant measuring value is displayed.
Always take into account that any humidity measurements in an open space cannot be made to an ac-
curacy of more than 0.1% due to factors such as movement in the air, temperature changes etc.
A user-calibration at the users site is not possible. To ensure highest accuracies it should be returned
to the manufacturer every 12 months (when used in clean air) for checking or, if necessary: recalibra-
tion.
If desired a calibration certificate can be made (acc. ISO 9000 ff). Price upon request.
Probe tube and case are not 100% proof. If there are pressure differences between probe tube and
case, there may be made wrong measuring.
For measuring in channels with over/under pressure or strong air flow we recommend our GMH3330 or
GMH3350 with humidity probe TFS0100 or TFS0100E.