9. Specification
Absolute maximum ratings:
Performance data Limit values
Connection
between min. max. min. max. Notes
Supply voltage 3 and 4 207 VAC 244 VAC 0 VAC 253 VAC
Relay output
1 and 2
1 and 3,
2 and 3
5 A,
ohm resistive load
Output 3
(NPN, open collector) 5 and 6 0 VDC 28 VDC,
I < 30mA
Input 0 – 50mV, TC 7 and 6 -1 VDC 10 VDC,
I < 10mA
Absolute maximum ratings must not be exceeded (not even for a short period of time)!
Meas. Input: Thermocouple probe type J, K, N, S, T
Standard signal 0-50mV
Meas. range: (thermocouple probe)
NiCr-Ni (type K): -270 ... +1350 °C or -454 ... +2462 °F
Pt10Rh-Pt (type S): - 50 ... +1750 °C or - 58 ... +3182 °F
NiCrSi-NiSi (type N): -270 ... +1300 °C or -454 ... +2372 °F
Fe-CuNi (type S): -170 ... + 950 °C or -274 ... +1742 °F
Cu-CuNi (type T): -270 ... + 400 °C or -454 ... + 752 °F
Display range: (for voltage measuring)
-1999 ... 9999 Digits, initial value, terminal value and decimal point position arbitrary.
Recommended range: < 2000 Digits
Accuracy: (at nominal temperature)
Standard signals: < 0.3% FS ±1Digit
Thermocouple probe: < 0.3% FS ±1Digit (with type S: < 0.5% FS ±1Digit)
Point of comparison:±1°C ±1Digit (at nominal temperature)
Meas. frequency: approx. 4 measures / sec.
Outputs: 2 Relay outputs, switching to 230V, 1 NPN-output (open collector) switching to ground
Relay-output: closing contact
Breaking capacity:5A, 230VAC,ohm resistive load
NPN-Output: NPN, open collector
Breaking capacity: 30mA, max. 28VDC
Output functions:2-point, 3- point, 2- point with alarm, 3- point with alarm, min-/max-alarm.
Switching points: arbitrary
Display: approx. 10 mm high, 4-digit red LED-display
Operation: 3 push buttons
Power supply:230V, 50/60Hz
Power consumption: approx. 2VA
Nominal temp.: 25°C
Operating ambient: -20 to +50°C
Relative humidity: 0 to 80 % RH (non condensing)
Storing temp.: -30 to +70°C
Enclosure: main housing: fibre-glass-reinforced noryl
front view-panel: polycarbonate
Dimensions: 24 x 48 mm (front-panel cut out).
Installation depth: approx. 65 mm (incl. screw-in/plug-in clamps)
Panel mounting: via VA-spring-clip.
Panel thickness: possible from 1 to approx. 10 mm possible.
Panel cut-out: 21.7+0.5 x 45+0.5 mm (H x W)
Connection: via screw-in/plug-in clamps: 4-pol. for mains-supply and relay connection, 4-pol. for measuring
input and alarm output. Conductor cross-selection: from 0.14 to 1.5 mm² (input/alarm) and 0.14
to 2.5mm² (mains/relay).
Protection class: front IP54
EMC: EN61326 +A1 +A2 (appendix A, class B), additional errors: < 1% FS
When connecting long leads adequate measures against voltage surges have to be taken.