GUIDO RAYOS X NESTOMAT 6050 User manual

GUIDO RAYOS X, S.A.
NESTOMAT 6050
1
MAN-012
Sept.00
Ed. 1 / Rev. 3
INDEX
Disclaimer
Symbols 4
Specifications 5
I Description 6
I.1 Patient Bed. Positioning of patient 6
I.2 I.V. Stand 6
I.3 Operation controls 10
II Operation 10
II.1 To operate 10
II.2 To preheat 10
II.3 To heat 12
A Air control mode 12
B Servo control mode 13
II.4 Test timer 14
II.5 Oxygen monitoring (Optional) 15
II.6 Alarms 16
A Air control mode 17
1 Sensor alarm 17
2 Low temperature alarm 17
B Servo control mode 18
1 Sensor alarm 18
2 Low temperature alarm 18
3 High temperature alarm 18
C Power failure 19
II.7 Sensor activation and deactivation 19
II.8 Software language 20
II.9 Change of display type 21
II.10 On/Off Switch (SW) activation 21
II.11 Change of temperature alarm range (High/Low) 21
III Maintenance 22
IV Optional Accessories 23
IV.1 Phototherapy assembly 23
IV.2 Resuscitaion / Aspiration Assembly 23

GUIDO RAYOS X, S.A.
NESTOMAT 6050
2
MAN-012
Sept.00
Ed. 1 / Rev. 3
Using the NEOPUFF Infant resuscitator 24
Cleaning 26
Maintenance 27
Functional schematic 28
Assembly diagram 28
Product specifications 29
IV.3 Shelves 30
IV.4 Other optional accessories 30
IV.5 Pulseoximetry module 30
Theory of operation 30
Front Panel Displays, Indicators and Keys 32
Alarms and indicators 33
Starting pulseoximetry 34
Measuring Pulse rate and SaO234
Setting and interpreting alarms 35
Adjusting the audio volume 35
V Installation 36
VI Service 39
VI.1 Instruments required 39
VI.2 Adjustments 39
A Air temperature 40
B Patient temperature 40
C Oxygen concentration 40
Electronic diagram 42
Warranty

GUIDO RAYOS X, S.A.
NESTOMAT 6050
3
MAN-012
Sept.00
Ed. 1 / Rev. 3
DISCLAIMER
The safety devices and other controls provided in this equipment will perform reliably
when operated, maintained, and repaired in accordance with the instructions of this
manual.
Safety devices must be checked periodically and reset, repaired, or replaced as
necessary to ensure that they will operate reliably. Equipment and parts that are broken,
missing, badly worn, distorted or contaminated should be replaced with appropriate
GUIDO RAYOS X parts. The equipment or its components should not be modified
without the approval of the manufacturer.
The Manufacturer disclaims all responsability for any malfunction of this equipment
resulting from faulty operation, maintenance or repair, or if any of its components are
damaged or modified by anyone other than the manufacturer.

GUIDO RAYOS X, S.A.
NESTOMAT 6050
4
MAN-012
Sept.00
Ed. 1 / Rev. 3
SYMBOLS
SYMBOL DESCRIPTION
In compliance with the required european norms CEI 601-
1-2 y CEI 601-1
Altern current
Atention, read the attached documents
Fuse
Off
On

GUIDO RAYOS X, S.A.
NESTOMAT 6050
5
MAN-012
Sept.00
Ed. 1 / Rev. 3
SPECIFICATIONS
Description Radiant warmer
Model NESTOMAT 6050
Alimentación External AC : 220 V ó 125 V 50/60 Hz
Consumo AC 0,25 A
Air control temperature YES
Air temperature range 20 - 40 ºC
Air temperature digital readout YES
Patient temperature control YES
Patient temperature range 20 - 40 ºC
Patient temperature digital YES
readout
Electrosthatic discharges Class : I
protection Type : BF
Fuses Internal : 1 of 2A
External : 3 of 2A and 3 of 5A
Operation modes Preheating
Air control mode
Servo control mode
Special functions Oxygen
Alarms Air sensor, skin sensor, oxygen sensor, high or low temperature, high or
low oxygen concentration, power failure, heating failure, motor error.
Optional Phototherapy assembly, resuscitation/aspiration assembly, shelf monitors,
instruments shelf, holding assembly for two gas cyllinders, oxygen tent
hood with security valve, oxygen servoregulator, head frame, tray for
X-ray cassettes, drawer assembly.

GUIDO RAYOS X, S.A.
NESTOMAT 6050
6
MAN-012
Sept.00
Ed. 1 / Rev. 3
I. DESCRIPTION
The NESTOMAT Radiant Warmer provide a controlled source of radiant heat and
illumination. The heat is generated by one 500 watt quartz heat tube located above the
procedure table. The diverging energy emitted from the heating element is reflected
back toward the table surface with a parabolic reflector, this one focuses the diverging
waves of energy into parallel waves which are aimed toward the patient.
A lamphouse contains one standard 20 watt halogen lamp for illumination of the work
area.
I.1 PATIENT BED. POSITIONING OF PATIENT.
The infant should be placed on the procedure table directly on the mattress or with a
single cloth covering between the infant and the mattress cover. The NESTOMAT
Radiant Warmer has capability (ten locked positions) to place the head up or the head
down. Grasp the locking puller and pull outwards. Reposition the table and release
knob. The table will lock in position. Tilting puller is located under the front side of the
table.
The table has hinged, transparent panels on the sides and front wich can be positioned
upright to retain the infant or downward to ease accessibility to the infant. The air
temperature sensor is located at the rear panel.
A space for a X-ray cassette tray is provided at the patient platform; the radiant heater
shroud can be swivelled, during operation, to allow X-ray tratments, just only pushing
the knob located in the rear side of the shroud.
I.2 I.V. STAND.
The I.V. stand is mounted in the accessories rear rail, its height is adjustable.
I.3 OPERATION CONTROLS.
All operation controls are located on the front and the rear of the upper column.

GUIDO RAYOS X, S.A.
NESTOMAT 6050
7
MAN-012
Sept.00
Ed. 1 / Rev. 3
(1) HEAT/CALOR Indication of heating device operation
ON/OFF Switch the heating control
DISPLAY Displays information, messages, etc.
(2) SERVO Patient Temperature operation mode
AIR Air Temperature operation mode
START/STOP Start / Stop operation
1 ... 0 Numerical sensitive keypad
ENTER Confirm the set values
PRE Preheating
O2Select Oxygen readout (optional)
(3) HIGH High temperature alarm
LOW Low temperature alarm
SENSOR Sensor failure alarm
POWER Power failure alarm
O2High or Low oxygen alarm
SIL Silence alarms

GUIDO RAYOS X, S.A.
NESTOMAT 6050
8
MAN-012
Sept.00
Ed. 1 / Rev. 3
1
2
3
4
5
6
7
TEST TIMER
1 I/O Start up of the timer
2 Indication of operation
3LCD Display
4 Indicator of timer operation
5 START/STOP
SIL. Start / stop operation and silence alarm
6 Indication of time
7 Audible alarm. End of time
SENSORS AIRE/AIR Air temperature sensor
% O2Oxygen sensor (Optional)
PACIENTE/SKIN Skin temperature sensor

GUIDO RAYOS X, S.A.
NESTOMAT 6050
9
MAN-012
Sept.00
Ed. 1 / Rev. 3
1 I/O Start up phototherapy
2 Treatment clock with zero reset
3 Total operation clock
4 Illumination lamp start up
5 I/O NESTOMAT On / Off switch
6 Power on led

GUIDO RAYOS X, S.A.
NESTOMAT 6050
10
MAN-012
Sept.00
Ed. 1 / Rev. 3
II.OPERATION
IMPORTANT: BEFORE TO USE THE RADIANT WARMER WITH A
PATIENT, IT IS NECCESSARY TO CHECK ITS GOOD PERFORMANCE CONDITIONS,
AS WELL AS THOSE FROM THE ACCESSORIES AND OPTIONS.
II.1 TO OPERATE
To start operation, plug the cord to the mains (check voltage) and press
the I/O switch located at the right side of the trolley.
Press the ON/OFF key at the front panel, the LCD will display "GUIDO
RAYOS X - VER. x.x", immediately afterwards the Radiant warmer will
perform a selftest of all its circuitry.
II.2 TO PREHEAT
CAUTION:BEFORE TO USE THE RADIANT WARMER WITH THE
NEWBORN, PREHEAT THE UNIT FOR A BETTER COMFORT OF THE NEWBORN AND
AN ACCURATE PERFORMANCE. PREHEATING PERIOD OF TIME DEPENDS ON
ENVIRONMENTAL TEMPERATURE. IF PREHEATING MODE IS NOT DESIRED, THE
OPERATION MODE, AIR OR SERVO, CAN BE DIRECTLY SELECTED.
Pressing any key the LCD will display,
AIR=__._ ºC
SELECT MODE
Press the key “PRE” to start the preheating until to reach aproximately
the factory preset temperature (28 ºC). In the meantime the LCD will
display
AT=__._ ºC (__._ ºC)
PREHEATING
After “=“ symbol we can find present temperature.
It is possible to modify preset preheating temperature in order to reach a
nearly temperature of the normal operation temperature. Proceed with
following steps to change preset preheating temperature:
• Turn on the radiant warmer:
GUIDO RAYOS X
VER X.X

GUIDO RAYOS X, S.A.
NESTOMAT 6050
11
MAN-012
Sept.00
Ed. 1 / Rev. 3
• Press any key :
AIR=__._ ºC
SELECT MODE
• Press AIR ,
AIR=__._ ºC
PSET AIR=28.0 ºC
• Change old preset 28.0ºC temperature to new preheating temperature with
numerical keypad.
• By pressing ENTER, you will confirm selection ,
AT=__._ ºC
PRESS START
• Despite it is displayed “PRESS START”, you have to press “ENTER”. This is
because this is an special feature of NESTORET 5050 .
AIR=__._ ºC
SELECT MODE
• Finally, press PRE. ,
AT=__._ ºC (__._)
PREHEATING
Once the preheating temperature is reached, the LCD will display "UNIT
READY", to be used within a patient, displaying the temperature inside
the Radiant warmer as well as advising to the medical personnel by
means of an acoustic signal. Then the Operation Mode should be selected
just afterwards the newborn is placed into the Radiant warmer.
IMPORTANT:BY PRESSING AGAIN THE PRE. KEY, THE LCD WILL
DISPLAY THE REGISTERED PARAMETERS, SUCH AS AIR TEMPERATURE (AT),
PATIENT TEMPERATURE (ST).
NOTE: TO SET OXYGEN AND RELATIVE HUMIDITY VALUES, SEE CHAPTER
II.5.

GUIDO RAYOS X, S.A.
NESTOMAT 6050
12
MAN-012
Sept.00
Ed. 1 / Rev. 3
II.3. TO HEAT
If Preheating Mode was not selected, proceed as follows:
Press any key, the LCD will display,
AIR=__._ ºC
SELECT MODE
To select the Operation Mode press AIR or SERVO key. AIR key
corresponds to Air Control Mode; SERVO key corresponds to Servo
Control Mode (patient skin temperature).
A) AIR CONTROL MODE:
NOTE: THE DIGITAL DISPLAY INFORMS THE TEMPERATURE OF THE AIR
WHERE THE NEWBORN IS PLACED.
Pressing AIR key, the LCD will display ,
AIR=__._ ºC
PSET AIR=__._
By means the numerical keypad, set the heating device working range
between 20 and 40 ºC.
NOTE: WHEN SETTING TEMPERATURE ENTER ALWAYS 3 DIGITS, EVEN IN
CASE OF NO DECIMALS. AS EXAMPLE PRESS 3, THEN 7, THEN 0 TO SET 37
DEGRESS.
Then press ENTER to confirm, the LCD will display ,
AT=__._ (__._)
PRESS START
Then press START to start the heating. The heat lamp will indicate the
operation of the heating device. By reaching the selected heating range,
the heating devices will modulate to keep the temperature stable.
The LCD will display the air registered temperature, as well as will
inform, too, the set temperature, displaying ACTIVED UNIT to inform
the heating device is in operation.

GUIDO RAYOS X, S.A.
NESTOMAT 6050
13
MAN-012
Sept.00
Ed. 1 / Rev. 3
AT=__._ ºC (__._)
UNIT ACTIVE
The selected heating range can be modified at any time by pressing
START/STOP and after ENTER key; you will find displayed following
message :
AIR=__._ ºC
SELECT MODE
Then set the new temperature by the keypad as previously explained and
confirm by pressing ENTER key. The LCD will display the messages
described above. Then press START key.
B) SERVO CONTROL MODE:
ATTENTION: BEFORE SELECTING SERVO CONTROL MODE
PROCEED TO PREHEAT THE RADIANT WARMER AS
EXPLAINED PREVIOUSLY IN THE II.2. TO PREHEAT
CHAPTER.
Before anything display on the screen following message :
AIR=__._ ºC
SELECT MODE
Press SERVO key, the LCD will display ,
SKIN=__._ ºC
PSET SKIN=__._
Set the patient temperature between 20 and 40 ºC by means of the
keypad, confirm by pressing ENTER key.
NOTE: WHEN SETTING TEMPERATURE ENTER ALWAYS 3 DIGITS, EVEN IN
CASE OF NO DECIMALS. AS EXAMPLE PRESS 3, THEN 7, THEN 0 TO SET 37
DEGRESS.
The LCD will display ,
ST=__._ ºC (__._)
PRESS START

GUIDO RAYOS X, S.A.
NESTOMAT 6050
14
MAN-012
Sept.00
Ed. 1 / Rev. 3
NOTE: ST MEANS PRESENT TEMPERATURE. TEMPERATURE IN BRACKETS
MEANS PRESET TEMPERATURE.
Then press START to start the heating. The heat lamp will indicate the
operation of the heating device. By reaching the selected patient
temperature, the heating device will modulate to keep the temperature
stable.
The LCD will inform the patient temperature and the selected one. By
displaying UNIT ACTIVE will inform the heating device is on
operation.
The set patient temperature can be modified at any time by pressing
START/STOP and after ENTER key; you will find displayed following
message :
AIR=__._ ºC
SELECT MODE
Then set the new temperature by the keypad as previously explained and
confirm by pressing ENTER key. The LCD will display the messages
described above. Then press START key.
ATTENTION: IF A MOMENTARY PAUSE WITHOUT
CHANGING THE SET PATIENT TEMPERATURE IS
REQUIRED, PRESS STOP KEY. PRESS START KEY TO
CONTINUE OPERATION. THE RADIANT WARMER WILL
OPERATE ON THE SAME OPERATION MODE AND
PARAMETERS ALREADY SET.
II.4 TEST TIMER.
This timer has a visual and audible alarms after 1, 5 and 10 minutes (1 + 4 + 5) since it
is switched on.
To operate the timer:
• Press I/O key to activate TEST TIMER device.
• Press START/STOP key to iniciate first cycle (1 minute).
• When the display shows 1 minute, the device will stop and alrmas (visual and
audible) will be activated.
• Press START/STOP key to silence alarms and begin next cycle (4 minutes).
• When the display shows 5 minutes, the device will stop and alrmas (visual and
audible) will be activated.
• Press START/STOP key to silence alarms and begin last cycle (5 minutes).
• When the display shows 10 minutes, the device will stop and alrmas (visual and
audible) will be activated. Press START/STOP key to silence alarms; the device
will be disconected automatically.

GUIDO RAYOS X, S.A.
NESTOMAT 6050
15
MAN-012
Sept.00
Ed. 1 / Rev. 3
II.5 OXYGEN MONITORING (OPTIONAL).
NESTOMAT 6050 Radiant Warmer monitors the oxygen percentage (optional) when
used with a tent hood (optional).
To set the alarm level for above mentioned parameter, proceed as follows (do the
calibration with the sensor out of the tent hood, as it has to be done at a concentration of
oxygen of 21%, which is the oxygen concentration of air) :
NOTE: WHEN SELECTING YOU MUST START FROM “SELECT MODE” SCREEN.
AIR=__._ ºC
SELECT MODE
Press the O2/HUM key, the LCD will display ,
AIR=__._ ºC
1=OXYG, 2=HUMID
IMPORTATNT: OPTION 2=HUMID IS NOT OPERATIVE FOR RADIANT WARMERS.
To set the Oxygen range press 1, the LCD will display
AIR=__._ ºC
OX: 1=PRG, 2=CAL
Then, press 1again ,
AT=__._ ºC (__._)
PRESET OX %=21.0
By means of the keypad set the desired value. Press ENTER key to
confirm.
Once again you will find following screen :
AIR=__._ ºC
1=OXYG, 2=HUMID
Finally press ENTER to confirm values.
To calibrate the Oxygen Sensor, press the O2/HUM. key, the LCD will
display ;
AIR=__._ ºC

GUIDO RAYOS X, S.A.
NESTOMAT 6050
16
MAN-012
Sept.00
Ed. 1 / Rev. 3
1=OXYG, 2=HUMID
NOTE: DUE TO THE WEAR OF THE OXYGEN CELL SENSOR IT IS
RECOMMENDED TO PERFORM PERIODICALLY ITS CALIBRATION PROCEDURES.
Check the Oxygen Sensor is pluged on the front panel and its housing is
not located inside the Tent Hood, allow to stabilize.
Press 1, the LCD will display,
AIR=__._ ºC
OX: 1=PRG, 2=CAL
Press 2, the LCD will display ,
OX CALIBR (21 %)
PRESS ENTER
By pressing ENTER, the Oxygen Sensor is already calibrated with a new
reference.
New screen displayed ;
AIR=__._ ºC
OX: 1=PRG, 2=CAL
II.6. ALARMS
The NESTOMAT 6050 Alarm System informs to the medical attendant of any
incidence may occur. The alarm is visible and audible.
NOMENCLATURE OF SYMBOLS IN ALARM
MESSAGES
+ Excess temperature alarm
- Failing temperature alarm
= Temperautre O.K.
* Desactive sensor
A) AIR CONTROL MODE

GUIDO RAYOS X, S.A.
NESTOMAT 6050
17
MAN-012
Sept.00
Ed. 1 / Rev. 3
1. SENSOR ALARM:
In case of air sensor failure or it gets disconnected or out of range; the
SENSOR Alarm will activate the corresponding LED and buzzer, the
alarm cannot be silenced. The LCD will inform to the attendant
displaying ,
* ALARM *
AIR SENSOR
The Radiant warmer is now out of service. Call to any authorized Service
engineer.
2. LOW TEMPERATURE ALARM:
Is activated when temperature reaches aproximately 1 ºC below the preset
temperature (See paragrap II.10 to change preset temperature alarm
range). "LOW" alarm LED will lit, to silence acoustic alarm press
SILENCE key. The LCD will display
* ALARM *
AT - RH OX
The alarm can be silenced by pressing the SILENCE key. If SILENCE
was pressed and the alarm conditions remain during a period of ten
minutes, the alarm will be activated again.
3. HIGH TEMPERATURE
Is activated when the temperature reaches aproximately 1 ºC higher than
the preset temperature (See paragrap II.10 to change preset temperature
alarm range). "HIGH" alarm LED will lit and the LCD will display ,
* ALARM *
AT + RH OX
to silence acoustic alarm press SILENCE key. If SILENCE was pressed
and the alarm conditions remain during a period of ten minutes, the alarm
will be activated again.
Check if heat sources are close to the Radiant warmer location, if positive
keep them away. If the alarm is still effective unplug the unit from the
power source, remove the patient from the radiant warmer and call to the
authorised service engineer.
B) SERVO CONTROL MODE
1. SENSOR ALARM:

GUIDO RAYOS X, S.A.
NESTOMAT 6050
18
MAN-012
Sept.00
Ed. 1 / Rev. 3
Informs when a failure of the skin probe or if it gets disconnected or out
of range. The Alarm will activate the corresponding LED and buzzer.
The LCD will display ,
* ALARM *
SKIN SENSOR
being possible to silence. The Radiant warmer is now out of service. The
Radiant warmer can be used on Air Mode if no any circunstances or
alarms avoid it, in case it is absolutely needed switch to Air Operation
Mode and call to the authorized service engineer.
2. LOW TEMPERATURE
Is activated when temperature reaches aproximately 1 ºC below the preset
temperature (See paragrap II.10 to change preset temperature alarm
range). "LOW" alarm LED will lit and the LCD will display,
* ALARM *
ST - RH OX
to silence acoustic alarm press SILENCE key. If SILENCE was pressed
and the alarm conditions remain during a period of ten minutes, the alarm
will be activated again.
IMPORTANT: CHECK THE CONDITIONS OF THE PATIENT. CHECK THIERE IS
NOT AIR FLOW IN THE WARD. CHECK THE HEAT LAMP IS ON (THE
MICROPROCESSOR CONTROLS THE HEATING DEVICES); IF CONTRARY, UNPLUG
THE UNIT FROM THE POWER SOURCE, REMOVE THE PATIENT FROM THE
RADIANT WARMER AND CALL TO THE AUTHORISED SERVICE ENGINEER.
3. HIGH TEMPERATURE
Is activated when the temperature reaches aproximately 1 ºC higher than
the preset temperature (See paragrap II.10 to change preset temperature
alarm range). "HIGH" alarm LED will lit and the LCD will display ,
* ALARM *
ST + RH OX
to silence acoustic alarm press SILENCE key. HEAT lamp will switch
off. If SILENCE was pressed and the alarm conditions remain during a
period of ten minutes, the alarm will be activated again.
Check the conditions of the patient. Check if heat sources are close to the
Radiant warmer location, if positive keep them away.
If the alarm is still effective, unplug the unit from the power source,
remove the patient from the radiant warmer and call to the authorised
service engineer.

GUIDO RAYOS X, S.A.
NESTOMAT 6050
19
MAN-012
Sept.00
Ed. 1 / Rev. 3
C) POWER FAILURE
Is activated when failure of the power from the mains. The alarm circuit
includes a rechargeable battery and operation conditions of it must be
checked periodically.
To switch off the alarm operation, if desired, refer to chapter II.6.
"SENSORS ACTIVATION AND DEACTIVATION".
II.7. SENSORS ACTIVATION AND DEACTIVATION
The applications program of the NESTOMAT 6050 Radiant warmer allows to the user,
at any moment, to deactivate or activate the sensors of the monitored parameters, if
required for calibration procedures or non essential parameters cancellation (Oxygen
and/or Relative Humidity).
If the Radiant warmer is switched off, proceed as follows:
Press I/O switch, located at the side of the trolley.
Press ON/OFF key at the front panel. The LCD will display ,
GUIDO RAYOS X
VER X.X
then type sequentially:
• SILENCE key
• 7388
• SILENCE key.
• ENTER
then the LCD will display ,
MAINT: 0 SW, 1 LNG
2 SEN, 3 DIS, 4 RNG
Press 2key and the LCD will display all coded available sensors and their status as
follows,
1RH 2AT 3ST 4HT
5FA 6OX 7A2 8AI
(at the right side of each coded sensor "+" or "-" will display, meaning activated or
deactivated respectively)
(RH=Relative Humidity - Non operative, AT=Air Temperature, ST=Skin
Temperature,HT=Heating Device Temperature, FA=Fan - Non operative, OX=Oxygen
- Optional, A2 and A1 are auxiliary options for future implements)
To activate (+) or deactivate (-) press the corresponding figure key of the sensor.
Proceed with any sensor you want to activate or deactivate, then press ENTER to

GUIDO RAYOS X, S.A.
NESTOMAT 6050
20
MAN-012
Sept.00
Ed. 1 / Rev. 3
confirm.
The LCD will display ,
MAINT.: 0 SW, 1 LNG
2 SEN, 3 DIS, 4 RNG
pressing again ENTER the LCD will display the initial message. You can proceed to
set the Operation Mode, etc. as above described.
II.8. SOFTWARE LANGUAGE
The applications program of your NESTOMAT 6050 is delivered from factory in two
languages: english and spanish. At factory your radiant warmer has been preset to
english, if you want to switch to the language, proceed as follows:
Press I/O switch, located at the side of the trolley.
Press ON/OFF key at the front panel. The LCD will display ,
GUIDO RAYOS X
VER X.X
then type sequentially:
• SILENCE key
• 7388
• SILENCE key.
• ENTER
then the LCD will display,
MAINT.: 0 SW, 1 LNG
2 SEN, 3 DIS, 4 RNG
Press 1key
the LCD will display ,
1=ENGLISH VER
2=SPANISH VER
press the corresponding key to the selected language and confirm by pressing ENTER.
II.9. CHANGE OF DISPLAY TYPE
Press 3 key when applications programm is selected, display type will be activated,
Table of contents
Popular Food Warmer manuals by other brands

Vulcan-Hart
Vulcan-Hart VB150 ML-126552 Installation and operation manual

APW Wyott
APW Wyott DWC13-N Installation and operating instructions

Lanaform
Lanaform Hand Warmer instruction manual

Subaru
Subaru J601SCA000 installation instructions

Broil King
Broil King PROFESSIONAL CDR-1TB Use & care manual

Johnson
Johnson Otti Instructions for use