Guildline 5031 User manual

TECHNICAL MANUAL
FOR
MODEL 5031
AIR BATH
NOTICE
The contents and information contained in this manual are proprietary to
Guildline Instruments. They are to be used only as a guide to the operation
and maintenance of the equipment with which this manual was issued, and
may not be duplicated or transmitted by any means, either in whole or in
part, without the written permission of Guildline Instruments.
TM5031-D1-00
15 May, 2019


TM5031-D1-00TM5031-D1-00
15 May, 2019
i
TABLE OF CONTENTS
1. INTRODUCTION...................................................................................................1-1
1.1. SCOPE................................................................................................................................................... 1-1
1.2. GENERAL DESCRIPTION................................................................................................................ 1-2
1.3. FRONT PANEL INDICATORS AND CONTROLS......................................................................... 1-3
1.3.1. Power..................................................................................................................................................1-3
1.3.2. Heater ON and OFF............................................................................................................................1-3
1.3.3. High / Low..........................................................................................................................................1-3
1.3.4. Cooling ...............................................................................................................................................1-3
1.3.5. Temperature Set..................................................................................................................................1-3
1.4. REAR PANEL CONNECTORS AND CONTROLS......................................................................... 1-4
1.4.1. Line Input Connector..........................................................................................................................1-4
1.4.2. Voltage Selector..................................................................................................................................1-4
2. SPECIFICATIONS................................................................................................2-1
2.1. GENERAL SPECIFICATIONS.......................................................................................................... 2-1
3. OPERATING INSTRUCTIONS.............................................................................3-1
3.1. INSTALLATION.................................................................................................................................. 3-1
3.2. OPERATING NOTES.......................................................................................................................... 3-2
3.2.1. Heater Control Setting ........................................................................................................................3-2
3.2.2. Bath Temperature Selection................................................................................................................3-3
4. MAINTENANCE....................................................................................................4-1
4.1. PREVENTIVE MAINTENANCE....................................................................................................... 4-1
4.1.1. Cleaning..............................................................................................................................................4-1
4.1.2. Temperature Check.............................................................................................................................4-1
4.2. CORRECTIVE MAINTENANCE...................................................................................................... 4-1
4.2.1. Preliminary Checks.............................................................................................................................4-1
4.2.2. Temperature Control...........................................................................................................................4-1
4.2.3. Over Temperature...............................................................................................................................4-3
5. PRINCIPLES OF OPERATION ............................................................................5-1
5.1. INTRODUCTION ................................................................................................................................ 5-1
5.2. TEMPERATURE SENSITIVE OSCILLATOR................................................................................ 5-2
5.3. AMPLIFIER AND DETECTOR......................................................................................................... 5-2

TM5031-D1-00TM5031-D1-00
15 May, 2019
ii
5.4. HEATER DRIVER............................................................................................................................... 5-3
5.5. OVER TEMPERATURE CIRCUIT................................................................................................... 5-3
5.6. POWER SUPPLY................................................................................................................................. 5-4

TM5031-D1-00TM5031-D1-00
15 May, 2019
iii
TABLE OF FIGURES
FIGURE 1-1: FRONT VIEW....................................................................................................................................1-1
FIGURE 1-2: CONTROL PANEL............................................................................................................................1-3
FIGURE 5-1: HEATER CONTROL UNIT ..............................................................................................................5-1
FIGURE 5-2: TEMPERATURE SENSITIVE OSCILLATOR.................................................................................5-2
FIGURE 5-3: AMPLIFIER AND DETECTOR........................................................................................................5-2
FIGURE 5-4: HEATER DRIVER.............................................................................................................................5-3
FIGURE 5-5: OVER TEMPERATURE CIRCUIT...................................................................................................5-3
FIGURE 5-6: POWER SUPPLY...............................................................................................................................5-4

TM5031-D1-00TM5031-D1-00
15 May, 2019
iv
LIST OF TABLES
TABLE 2-1: GENERAL SPECIFICATIONS...........................................................................................................2-1
TABLE 3-1: SUGGESTED CONTROL SETTINGS ................................................................................................3-3
TABLE 4-1: CONTROL CIRCUIT BOARD TESTING..........................................................................................4-2
TABLE 4-2: TRANSISTOR VOLTAGES................................................................................................................4-3

Section 1
15 MAY, 2019
29 October, 2018
1-1
1. INTRODUCTION
Figure 1-1: Front View
1.1. SCOPE
This manual contains technical specifications, a detailed description, maintenance information,
and diagrams for the Guildline Instruments model 5031 Air Bath.

Section 1
15 MAY, 2019
29 October, 2018
1-2
1.2. GENERAL DESCRIPTION
The model 5031 Air Bath is a bench top high precision, variable temperature bath which
provides 0.08 m3(2.9 cu. ft.) of usable temperature controlled chamber. The internal temperature
is maintained by a solid state control system. The regulated temperature can be selected from a
digital switch control on the front control panel. The temperature range is from 15.00 C to
50.00 C with a resolution of 0.01 C and 0.03 C stability. The design assures minimal
thermal gradients within the chamber.
The air bath chamber is of stainless steel construction. This shielded enclosure is tied to earth
ground through the power cord to reduce electromagnetic interference. The circulation fan,
cooling unit and heater are separated from the air bath chamber to provide full and unimpeded
access to the bath chamber.
The design of the bath employs a fixed rate of cooling from a thermoelectric module and
balances this to the heating/cooling level required by turning a heater off and on rapidly. This
provides the required heating/cooling to balance the heat losses to ambient or the heat generated
by apparatus within the chamber.
To protect the contents of the chamber from overheating due to a malfunction of the temperature
control, there is an overriding over-temperature control that will disable the heating system if the
temperature exceeds 55 C 4 C.
Uses for the Programmable Air Bath include:
•Maintaining a constant temperature environment for reference resistors such as the Guildline
Instruments 9330 series resistors.
•Calibration of thermometers.
•Thermal stressing of precision materials.

Section 1
15 MAY, 2019
29 October, 2018
1-3
Figure 1-2: Control Panel
1.3. FRONT PANEL INDICATORS AND CONTROLS
The front panel indicators and controls are illustrated in Figure 1-2. They are provided to indicate
power status, heater operation and temperature status.
1.3.1. Power
The POWER switch controls AC line power to the air bath.
1.3.2. Heater ON and OFF
The HEATER indicator lamps indicate whether or not the heater is energised; alternating
illumination indicates that the bath has reached the pre-set operating temperature.
1.3.3. High / Low
This switch, in the HIGH position, energises the heater with full-wave current. In the
LOW position a diode in series with the heater provides half-wave operation and so
reduces the heater power to half the value obtained on the HIGH position. When
operating in the LOW position only one electrode of the neon lamps will glow, while in
the HIGH position both will be illuminated.
1.3.4. Cooling
Operation of the cooling switch will energise the circuit supplying current to the cooling
modules and also operate the fan, which cools their associated heat sink.
1.3.5. Temperature Set
The temperature setting is controlled by means of the four-digit thumbwheel switch. The
table enclosed with this manual indicates the control settings for temperatures over the
entire range of 15 C to 50 C. The effect of a one step change of these dials on the
temperature set point will vary depending upon their settings. The rate of change of
resistance setting vs. operating temperature is approximately -4%/C. The total resistance
of the 4-digit TEMPERATURE SET control is 200 more than the value indicated. For
example, if the operating temperature is 23 C when the control is set to 4000 (i.e.
resistance is 4200 ), then changing the control setting to 3832 (4000 - 4% of 4200) will
produce an operating temperature of approximately 24 C.

Section 1
15 MAY, 2019
29 October, 2018
1-4
1.4. REAR PANEL CONNECTORS AND CONTROLS
The rear panel connectors and controls include power entry, and line voltage selection.
1.4.1. Line Input Connector
The Line input connector is a standard male 3-prong AC connector, with an integral fuse. A
2 amp time delay fuse is provided for 115 VAC operation and a 1 amp time delay fuse is
provided for 230 VAC operation.
1.4.2. Voltage Selector
The line input voltage selector is a two-position switch that allows setting for a nominal line
voltage of either 115 VAC, or 230 VAC. The actual limits of the line voltage are
90-132 VAC or 200-264 VAC, 47-63 Hz.

Section 2
15 MAY, 2019
29 October, 2018
2-1
2. SPECIFICATIONS
2.1. GENERAL SPECIFICATIONS
SPECIFICATIONS
5031
Chamber Temperature Range
15 °C to 50 °C, (Minimum to 6 °C below ambient)
Temperature Set Point
Accuracy
± 0.06 °C over 1 year within ± 5°C of ambient
± 0.1 °C over 1 year outside ± 5°C of ambient
± 0.2 °C over 1 year set point over 40°C
Set Point Resolution
0.01 °C
Temperature Stability
± 0.015 °C over 24 hours within ± 2 °C ambient
± 0.03 °C over 24 hours within ± 5 °C ambient
± 0.06 °C over 24 hours outside ± 5 °C ambient
± 0.2 °C over 24 hours set point over 40°C
Temperature Uniformity
± 0.2 °C relative to chamber center, 5 cm minimum from walls for +15
°C to +40 °C
Temperature Attenuation
± 0.04 °C/°C of ambient temperature
Heating Rate
6 °C/hour
Cooling Rate
5 °C/hour, above ambient temperature
2 °C/hour, below ambient temperature
Cold Power On Stabilization
6 hours to within ±0.1 °C of set point
Over Temperature Protection
Automatic shutdown if temperature exceeds 55 °C ± 4 °C
Maximum Power Dissipation of unit under test (set point above
ambient)
5W maximum
Table 2-1: General Specifications
NOTES:
1. These specifications are applicable with the air bath chamber empty.
2. The temperature set point shall be no more than 6 °C below the ambient temperature.
3. 24 hour set point accuracy and stability are defined as the deviation of the mean hourly value from the 24
hour mean for a single ambient temperature point at one point in the air bath chamber (typically the center).
The peak to peak stability can be higher (±0.06 °C typical) depending on the temperature selected.
4. Temperature uniformity is relative to the center of the air bath chamber and at least 5 cm from the walls of
the chamber for +15 °C to +40 °C.

Section 3
15 MAY, 2019
29 October, 2018
3-1
3. OPERATING INSTRUCTIONS
3.1. INSTALLATION
This instrument was thoroughly tested and inspected before shipment and should be free from
damage when received. Inspect it carefully, verify that all items on the packing list are present
and check the instrument operation as soon as possible. Refer to the warranty card at the front of
this manual if any damage or deficiencies are found.
The installation procedure is as follows:
1. Place the instrument on a stable bench capable of supporting the weight of the air bath.
CAUTION
2. Set the controls as follows:
- POWER: OFF
- COOLING: OFF
- TEMPERATURE SET: 9999
3. Ensure that the correct fuse has been installed –2 A 250 V (T) should be used for a nominal
line input of 115 V and 1 A 250 V (T) should be used for a nominal line input of 230 V.
Only fuses with the rated current and specified type should be used for replacement.
4. Plug the supplied moulded line cord into the 3-pin power receptacle on the rear panel of the
instrument. Ensure that the other end of the line cord is plugged into a wall socket or
extension cord that has a protective or safety ground. Where 3-contact power supply outputs
are not available, a suitable protective ground connection must be made before switching the
instrument power on. Any interruption of the protective ground may possibly render the
instrument unsafe.
Ensure that air flow through the rear door is not obstructed.
Allow at least 25 cm clearance.
The 5031 has been shipped with the line input voltage selector set
to 230 V. The line input selector must be set to the correct line
voltage before power is applied to the instrument. The settings
available are 115 V and 230 V. Remove the warning label
positioned across the power line input socket after setting the
correct operating voltage.

Section 3
15 MAY, 2019
29 October, 2018
3-2
Where the moulded plug on the line cord supplied with the instrument does not match the
local power outlet socket, the plug can be removed and replaced with one that does fit the
local service. The plug should be re-wired as follows:
Brown wire- Line input
Blue wire- Neutral input
Green/Yellow Ground (safety)
5. Set the POWER switch to the ON position and observe that the HEATER OFF lamp lights
and the circulating fans operate.
6. Set the COOLING switch to the ON position and observe that the fan on the rear door
operates and that both cooling fins behind the rear door become warm.
7. Set the TEMPERATURE SET switches appropriately, according to the table enclosed with
this manual, for the desired air bath temperature. Set the HIGH/LOW and COOLING
switches as recommended in Table 3.1 for that set point temperature.
8. When the inner chamber reaches its operating temperature, the HEATER lamps will indicate
periodic cycling. Due to the thermal lagging, it will take several hours to raise the inner
chamber to the operating temperature. The bath should be allowed to stabilise at the operating
temperature for at least 6 hours before use.
3.2. OPERATING NOTES
Operation of the Model 5031 Air Bath comprises of selecting control settings and monitoring the
air temperature. The following paragraphs contain suggested procedures for the use of the
controls.
3.2.1. Heater Control Setting
The HIGH/LOW and COOLING switches control the amount of power available to heat
the Air Bath chamber. Refer to Table 3.1 for the settings to be used under various
conditions. If a large temperature increase is required quickly, select HIGH and switch
COOLING off until the HEATER ON and HEATER OFF lamps light alternately then
change to the setting recommended in Table 3.1. Optimum stability is achieved when the
HEATER ON and OFF lamps light approximately an equal amount of time. Table 3.1
also shows when the COOLING control should be ON. Normally it will be used only if
the operating temperature is less than 10 C above ambient or if a large temperature
decrease is required quickly.

Section 3
15 MAY, 2019
29 October, 2018
3-3
RANGE
COOLING
HIGH/LOW
15 C to 20 C
ON
LOW
20 C to ambient+10 C
ON
HIGH
ambient +10 C to 40 C
OFF
LOW
40 C to 50 C
OFF
HIGH
Table 3-1: Suggested Control Settings
3.2.2. Bath Temperature Selection
The operating temperature is determined by the TEMPERATURE SET control. A table
attached to this manual provides the settings required for the full range of the bath.
Temperatures not on the table must be arrived at by interpolation between the “dial
settings” presented in the table.
A more precise setting of the bath temperature can be achieved by inserting a
thermometer in through the side access hole and making a direct measurement of the air
temperature within the chamber. The air bath temperature can then be adjusted by
changing the TEMPERATURE SET control by –4%/C of the TEMPERATURE SET
resistance. For example, if the desired temperature is 25.00 C but the actual temperature
is 25.16 C when the TEMPERATURE SET control is set to 3670 (i.e. the
TEMPERATURE SET resistance is 3670+200=3870 ), then adjust the
TEMPERATURE SET control by +25 (i.e. (–0.16)(-0.04)3870 ) to 3695.

Section 4
15 MAY, 2019
29 October, 2018
4-1
4. MAINTENANCE
4.1. PREVENTIVE MAINTENANCE
4.1.1. Cleaning
The instrument requires very little maintenance other than periodic dusting of the outside
and inside of the enclosure. Remove accumulated dust from the air filter on the rear door.
4.1.2. Temperature Check
Insert a thermometer in through the side access hole and make a direct measurement of
the air temperature within the chamber. Compare this temperature to the expected
temperature for the TEMPERATURE SET control setting as indicated by the enclosed
temperature setting table.
4.2. CORRECTIVE MAINTENANCE
Symptoms appear at the beginning of each of the following paragraphs as a guide to detailed
fault location. Refer to Section 5 of this manual for the PRINCIPLES OF OPERATION.
4.2.1. Preliminary Checks
Check POWER –Either the HEATER ON or HEATER OFF lamp should be lit when the
instrument is connected to the AC line and the POWER switch is set to “ON”. If not,
check that the circulation fans inside the chamber are running. If they are not, then a
general loss of power is indicated. Check the line input fuse.
Check the HEATER ON/OFF indicators - They should indicate the cycling on and off of
the heater.
4.2.2. Temperature Control
If the temperature is low and the HEATER indicator is lit, check the continuity and
resistance of the heater circuit.
If the temperature is low and the HEATER indicator is not lit, check the operation of the
control circuit as follows:
1) Short out between WT103 and WT104 and the HEATER indicator should light. If
not, then check for a fault in the control circuit. If the indicator lights, then check the
resistance of R2 (the TEMPERATURE SET resistance) and RT1 (the thermistor).
RT1 is equal to Rl+R103 (200 ) at the control temperature, and is larger than
R2+R103 if the temperature is lower. It has a temperature coefficient of
approximately -4%/C.

Section 4
15 MAY, 2019
29 October, 2018
4-2
Control circuit board testing - Measure the voltage at the terminals as indicated in Table
4-1.
Between
And
Should Read
Components To Check
7
16
-12 V D.C
CR106, CR107
9
7
-3.2 V
see NOTE 1
10
7
-3.2 V
see NOTE 1
12
7
-3.2 V
see NOTE 1
14
7
-0.3 V
when heater off
-0.7 V
when heater on: see NOTE 2
Table 4-1: Control Circuit Board Testing
NOTE 1: If this reading is incorrect, then check the voltage across C101, 500 pF capacitor. –It
should read 8.2V ± 5%. If not, then check R107, CR101, CR102 and C101. If correct,
then check R102, R105, C103 and Q101.
NOTE 2: If this reading is not obtained, check Q7.
Check the operation of the temperature sensitive oscillator. If the resistance of RT1 is
greater than the resistance of Rl, then the oscillator should be running. Check with an
oscilloscope between TP1 (collector of Q102) and Jl-7. The waveform should be as
shown. If not, then check the following components:
Q101, Q102, C104, T101, C102, R103, R125, R104, R110
TP1
If oscillation is observed, then check between TP2 (coil Q103) and Jl-7. The waveform
should be as shown. If not, then check the following components:
Q103, C106, R117, R112, R106

Section 4
15 MAY, 2019
29 October, 2018
4-3
TP2
If this waveform is observed, then measure the D.C. voltage between TP3 (CR104 anode)
and Jl-7. It should read -2.5 volts. If not, then check C105, C107, CR103 and CR104. If
it is correct, then check Q104, Q105, R121, R122 and R114.
Note: Both of the above indicated waveforms are typical, if RT1 is at ambient
temperature of about 25 °C.
As a further guide to trouble shooting, typical D.C. voltage at the transistor terminals are
given in Table 4-2.
DEVICE
VOLTAGE READINGS (Jl-7 Common)
HEATER ON
HEATER OFF
E
B
C
E
B
C
Q101
-3.05
-3.2
-5.2
-3.05
-3.2
-5.2
Q102
-5.4
-5.2
-3.05
-5.4
-5.2
-3.05
Q103
-2.95
-3.05
-5.3
-2.95
-3.05
-5.3
Q104
-1.4
-2.0
-1.45
———--Variable-——-
Q105
-75
-1.42
-1.45
———--Variable-——-
Q7
0
-.75
-.25
0
-.25
-12.8
Table 4-2: Transistor Voltages
4.2.3. Over Temperature
If the temperature within the chamber exceeds 55 °C, then the thermal breaker will trip.
Allow the enclosure to cool to less than 50 °C. Check that the temperature sensitive
switch S2 is open. If this condition is satisfied, then reset the switch. The reset switch is
located inside the back panel of the inner chamber. The switch may be accessed by
opening the rear door and removing the back panel of the inner chamber. WARNING:
Unplug main power cord before removing back panel.

Section 4
15 MAY, 2019
29 October, 2018
4-4
If the HEATER lamps do not indicate heater cycling when the temperature approaches
normal, then check the operation of the control circuit as follows:
- Locate pins A and B of P2.
- Short pin A to pin B. The HEATER indicator should go out. If it does, then
check the resistances of RT1 and R2 as detailed in section 4.2.2. Replace them
as a matched pair. If not, then check at TP1 and TP2 with an oscilloscope
(common to J1-7). With P2-A and P2-B shorted, there should be no oscillation.
Check the voltage at TP3. It should be zero. Refer to Table 4-2 for voltage
measurements to aid in locating the fault in the heater driver.

Section 5
15 MAY, 2019
15 May, 2019
5-1
5. PRINCIPLES OF OPERATION
5.1. INTRODUCTION
The heater control unit in the model 5031 enclosure consists of the following functional blocks:
a) Temperature Control - consisting of a sensor, oscillator, amplifier, detector and
heater
b) Overtemperature Circuit
c) Power Supply
Figure 5-1: HEATER CONTROL UNIT

Section 5
15 MAY, 2019
15 May, 2019
5-2
5.2. TEMPERATURE SENSITIVE OSCILLATOR
Figure 5-2: TEMPERATURE SENSITIVE OSCILLATOR
Temperature changes are sensed by the thermistor RT1 connected with variable R2 across the 1:2
ratio autotransformer T101. As the temperature decreases, the thermistor resistance increases and
the voltage divider output to the base of Q101. As soon as this voltage exceeds half the
transformer output, the circuit gain becomes greater than one and oscillation occurs.
5.3. AMPLIFIER AND DETECTOR
Figure 5-3: AMPLIFIER AND DETECTOR
The output of the oscillator is amplified by Q103 and rectified by a voltage doubler, consisting of
C105, CR103, CR104 and C107. The resistor R115 gives current limiting protection for the next
stage.