11
Table of contents
1. Technical Specifications . . . . . . . . . . . . . . . . . . . 3
1.1. Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2. General information . . . . . . . . . . . . . . . . . . . . . . 5
2.1. Safety information . . . . . . . . . . . . . . . . . . . . . . 5
2.1.1. Use of hazard information. . . . . . . . 5
2.2. General product information. . . . . . . . . . . . . 6
2.3. Items supplied. . . . . . . . . . . . . . . . . . . . . . . . . . 6
3. Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.1. si628 C (wall mount). . . . . . . . . . . . . . . . . 7
3.1.1. Assembly. . . . . . . . . . . . . . . . . . . . . . . . 7
3.1.2. Connections . . . . . . . . . . . . . . . . . . . . . 8
3.2. si628 C (panel mount) . . . . . . . . . . . . . . . 10
3.2.1. Assembly. . . . . . . . . . . . . . . . . . . . . . . . 10
3.2.2. Connections . . . . . . . . . . . . . . . . . . . . . 11
3.3. Cell cables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.3.1. Cells with MP-5 connector . . . . . . . 12
3.3.2. Cells with connector S8 . . . . . . . . . 13
3.4. Protective wiring of switching contacts . . . . 13
4. Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.1. Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.2. Setting up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.3. Quick guide . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.4. Programming . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.5. Calibration.. . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.5.1. Recognized standards . . . . . . . . . . . . 18
4.6. Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.7. Access codes . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.1. Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.2. Fuse replacement. . . . . . . . . . . . . . . . . . . . . . . 21
6. Error messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
7. Replacement parts . . . . . . . . . . . . . . . . . . . . . . . . 23
8. Warranty and liability . . . . . . . . . . . . . . . . . . . . 24
Contact information
ENGLISH
si628 C_HACH LANGE_Ed. 0510.indd ANG:1si628 C_HACH LANGE_Ed. 0510.indd ANG:1 20/5/10 11:08:4020/5/10 11:08:40

si628 C_HACH LANGE_Ed. 0510.indd ANG:2si628 C_HACH LANGE_Ed. 0510.indd ANG:2 20/5/10 11:08:4020/5/10 11:08:40

3
1. Technical specifications
Measuring ranges
Conductivity * 0.01 µS/cm to 199.9 mS/cm
* with C=0.1 cm -1 cell
°C –20.0 a 150.0
Measuring error
Conductivity ≤1% (±1 digit)
Temperature ≤0.5°C (±1 digit)
Reproducibilidad
Conductivity ±0.1% (±1 digit)
Temperature ±0.1°C (±1 digit)
Automatic temperature compensation
With temperature probe Pt 1000 or by keypad
Temperature coefficient, TC
Fixed at 2 % / °C
Reference temperature, RT
Fixed at 25 °C
Accepted cells
2 pole-cell. Constant between 0.05 and 50 cm-1
Recognised standards
147 µS/cm, 1413 µS/cm and 12.88 mS/cm
(25°C)
Calibration possibilities
With 1 or 2 standards
Access code: 100
Accepted calibration, 2 standards
Diference <30 % between the obtained cell
constants
Control mode
ON / OFF.
A relay programmable as maximum or minimum
Activation relays lag 2 seconds
Alarm relay programmable between 1 and 99 ss,
min or hours
Language
English, German, Spanish, French, Italian, Portuguese
Keypad
LCD, alphanumeric, backlit,
2 lines of 16 characters
Inputs and outputs
2 pole cell, terminal strip
Pt 1000 ATC, terminal strip
Analogic output, galvanically isolated:
• 4 to 20 mA for measurement (R max.= 500 Ω)
• 21 mA or the last mA value measured for the instru-
ment in hold
• 22 mA for the alarm
Relés
1 limit relay, with no potential. NO contact.
1 alarm relay, with no potential. NC Contact.
Maximum load:
en CA < 24 V / < 3 A / < 72 VA
Maximum cable length
≤25 m
Power supply
230 or 24 VAC ±10 %, 45-65 Hz,
(115 VAC under demand)
Consumption 4 VA
Protection: Class II
Overvoltage category: II
EC Directives
Electrical security according to 2006/95/EC
Electromagnetic compatibility according to
2004/108/EC
Environmental conditions
Working temperature: 0 to 50 °C (32 to 122 °F)
Storage temperature: –20 to 65 °C (–4 to 149 °F)
Relative humidity < 80 % (non condensing)
Max. height 2.000 m. at 230 V; 3.000 at 24 V
Enclosure
si628 C (wall mount): Thermoplastic material,
protection IP55. Flame rating HB
si628 C (panel mount): Noryl,
protection IP54. Flame rating FV-1
Weight
si628 C (wall mount): 700 g
si628 C (panel mount): 600 g
Dimensions
si628 C (wall mount):
105 x 170 x 100 mm (4 x 6.7 x 3.9 in.)
si628 C (panel mount):
96 x 96 x 100 mm (3.8 x 3.8 x 3.9 in.)
Warranty
Instruments: 2 years
Specifications are subject to change without notice.
ENGLISH
si628 C_HACH LANGE_Ed. 0510.indd ANG:3si628 C_HACH LANGE_Ed. 0510.indd ANG:3 20/5/10 11:08:4020/5/10 11:08:40

4
1. Technical Specifications
1.1. Dimensions
si628 C (wall mount)
si628 C (panel mount)
105 mm
4.1 in 45 mm
1.8 in 50 mm
2 in
170 mm
6.7 in
90 mm
3.5 in
96 mm
3.8 in
96 mm
3.8 in
90 mm
3.5 in
80 mm
3.1 in
7 mm
0.3 in
si628 C_HACH LANGE_Ed. 0510.indd ANG:4si628 C_HACH LANGE_Ed. 0510.indd ANG:4 20/5/10 11:08:4020/5/10 11:08:40

5
2. General information
As a result of constant improvements to our products sometimes differences may exist between this manual
and the instructions supplied with the instrument.
2.1. Safety information
Please, read carefully this information before installing and using the instrument !
Pay attention to all danger and caution statements.
ENGLISH
2.1.1. Use of hazard information
This symbol references the instruction manual for operation or safety information.
This symbol indicates that a risk of electrical shock or electrocution exists.
This symbol indicates the connection for Protective Earth.
This symbol indicates ground connection.
This symbol identifies a fuse.
Electrical equipment marked with this symbol may not be disposed of in European public disposal sys-
tems after 12 August of 2005. In conformity with European local and national regulations (EU Directive
2002/96/EC), European electrical equipment users must now return old or end-of life equipment to the
Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how
to return end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper
disposal.
DANGER
Indicates a potentially or imminently hazardous situation that, if not avoided, will result in death
or serious injury.
WARNING
Indicates a potentially or imminently hazardous situation that, if not avoided, may result in death
or serious injury.
CAUTION
Indicates a potentially hazardous situation that, if not avoided, may result in minor or moderate injury.
Important note
Indicates a situation that, if not avoided, could lead to damage to the instrument. Important infor-
mation that requires special emphasis.
Note
Information that supplements points in the main text.
Precautionary labels
Read carefully all labels and tags attached to the instrument.
si628 C_HACH LANGE_Ed. 0510.indd ANG:5si628 C_HACH LANGE_Ed. 0510.indd ANG:5 20/5/10 11:08:4020/5/10 11:08:40

6
2. General information
DANGER
si628 C (wall mount)
• Instrument (with strips for connections).
• User manual.
si628 C (panel mount)
• Instrument (with strips for connections).
• Clamps to fix to the panel.
• User manual.
2.3. Items supplied
• Remember that the voltage across accessible
parts of the open instrument, may be dangerous
to life. Do not open the instrument when it is
connected to the power supply, there are areas
whose voltage could cause death.
• Always disconnect power to the instrument
before it is opened.
• Repair or adjustment of an opened instrument
under voltage shall be carried out only by
a qualified technician who is aware of the
hazards involved.
• Never work in an environment subject to
explosion hazards. The housing of the
instrument is not gas tight.
• The Company will not be responsible for any
physical damage caused by non authorised
work or manipulation.
• It is the responsibility of whoever uses this
instrument to consult and establish appropriate
safety and health practices and determine the
applicability of regulatory limitations prior to use.
• Before connecting the instrument to a power
supply unit, make sure that the mains voltage
lies within the range:
230 or 24 V AC ±10%, 45-65 Hz (standard
version).
115 V AC ±10%, 45-65 Hz (under demand).
Important notes
• Read carefully the manual of the instrument.
• Avoid direct sunlight over the instrument.
• Exclude the following environmental influences:
- vibrations
- atmospheric humidity higher than 80 %
- corrosive gases present
- temperature < 0 °C, or > 50 °C (< 32 °F, or >
122 °F).
- powerful electric and magnetic fields.
• Only use original accessories and spare parts.
• Have the instrument serviced only by the
Company Service.
2.2. General product information
Proper use
The si628 C is used for conductivity and temperature measurement in food industry, waste water, in the
environment, etc.
si628 C_HACH LANGE_Ed. 0510.indd ANG:6si628 C_HACH LANGE_Ed. 0510.indd ANG:6 20/5/10 11:08:4120/5/10 11:08:41

7
3. Installation
ENGLISH
3.1. si628 C (wall mount)
DANGER
Only qualified personnel should conduct the tasks described in this section of the manual.
3.1.1. Assembly
Wall mounting
Open the case by unscrewing the 4 Ascrews.
For wall mounting of the instruments, use the Bdrill holes indicated on the scheme.
Pipe or column mounting
Use the pipe adaptor, PN LZU9120.99.
Install the pipe adaptor on the rear part of the instrument using both C drill
holes as indicated on the figure. Fit the unit with two clamps to the column.
C
C
Cable
tie
Rear view Pipe adaptor Instrument mounted
B
50 mm
2 in 50 mm
2 in
25 mm
1 in
25 mm
1 in
115 mm
4.5 in
A
Front view Side view Rear view
si628 C_HACH LANGE_Ed. 0510.indd ANG:7si628 C_HACH LANGE_Ed. 0510.indd ANG:7 20/5/10 11:08:4120/5/10 11:08:41

8
3. Installation
System functions
1, 2 Voltage 230 or 24 VAC ± 10%.
(115 VAC under demand).
3 Ground connections, see Important
notes (page 9).
4, 5 Current output 4-20 mA (EC).
6, 7 Input for temperature probe.
8 Cell cable shielding.
9 to 12 Input for 2-pole cell.
13, 14 Limit relay contacts.
15, 16 Alarm relay contacts.
DANGER
Always disconnect power to the instrument when making any electrical connections.
3.1.2. Connections
When making any wiring connections to the instrument, any warnings and notes found throughout the
manual must be adhered to. For more safety information see “Safety information” on page 5.
Details
To open the transparent front door, press with two
fingers the part marked with ( ) in the scheme.
Cable glands.
On the lower part of the instrument there are 5
cable glands, 2 PG 7 and 3 PG 9.
PG 9 (3 units)
PG 7 (2 units)
DANGER
High-voltage wiring for the instrument is protected by a protective casing. The protective casing
must remain in place unless a qualified installation technician is installing wiring for power.
si628 C_HACH LANGE_Ed. 0510.indd ANG:8si628 C_HACH LANGE_Ed. 0510.indd ANG:8 20/5/10 11:08:4220/5/10 11:08:42

9
3. Installation
ENGLISH
Terminal assignment (interior panel)
Important notes:
• The si628 C does not require ground con-
nection (terminals 3).
• Connect the ground only when there could
be interference problems.
DANGER
Relays:
- Limit the current up to 3 A by using one fuse.
- Do not connect cables with voltage higher than 24 V.
4-20 mA
CELL
VCA
L
A
Pt 1000
WHEN IT IS ON, IT INDICATES A FUSED FUSE.
si628 C_HACH LANGE_Ed. 0510.indd ANG:9si628 C_HACH LANGE_Ed. 0510.indd ANG:9 20/5/10 11:08:4220/5/10 11:08:42

10
3. Installation
3.2. si628 C (panel mount)
DANGER
Only qualified personnel should conduct the tasks described in this section of the manual.
3.2.1. Assembly
For the si628 C mounting a drill hole is required according to the diagram.
After putting the instrument in the panel, it must be fitted using two clamps as indicated in the figure.
92 mm
3.7 in
92 mm
3.7 in
Clamp to fix to the panel
Panel
Drill hole
Side view
80 mm
3.1 in
21 mm
0.8 in 7 mm
0.3 in
Plug-in screw connector
si628 C_HACH LANGE_Ed. 0510.indd ANG:10si628 C_HACH LANGE_Ed. 0510.indd ANG:10 20/5/10 11:08:4320/5/10 11:08:43

11
3. Installation
ENGLISH
When making any wiring connections to the instrument, any warnings and notes found throughout the
manual must be adhered to. For more safety information see “Safety information” on page 5.
3.2.2. Connections
System functions
DANGER
Always disconnect power to the instrument when making any electrical connections.
1 Ground connection, see Important
notes (page 12).
2, 3 Power supply 230 or 24 VAC ± 10%.
(115 VAC under demand).
4, 5 Alarm relay contacts.
6, 7 Limit relay contacts.
8, 9 Current output 4-20 mA (EC).
10 to 13 Input for 2-pole cell.
14 Cell cable shielding.
15, 16 Input for temperature probe.
DANGER
High-voltage wiring for the instrument is protected by a protective casing. The protective casing
must remain in place unless a qualified installation technician is installing wiring for power.
si628 C_HACH LANGE_Ed. 0510.indd ANG:11si628 C_HACH LANGE_Ed. 0510.indd ANG:11 20/5/10 11:08:4320/5/10 11:08:43

12
3. Installation
3.3. Cell cables
Compensation of the cable length
To minimize the effect of the cable resistance in a conductivity measurement, the cable length is compensa-
ted by the 4-wire system. With the Pt 1000 temperature sensor, the cable resistance has little importance.
However, in order to reduce the resistance of the conductor to a half, 2 cables have been connected to
the ends of the sensor.
3.3.1. Cells with MP-5 connector
Notas importantes:
• El si628 C does not require ground connec-
tion (terminal 1).
• Connect the ground only when there could
be interference problems.
DANGER
Relays:
- Limit the current up to 3 A by using
one fuse.
- Do not connect cables with voltage
higher than 24 V.
Terminal assignment (rear panel)
Transmitter si628 C
RED
VCA
CELL4-20 mA
L 1 L 2
S/N:
CRISON INSTRUMENTS, S.A.
E-08328 ALELLA - Barcelona
Made in Spain
IP 54
TERMINALS COLOUR
si628 C
(wall mount) si628 C
(panel mount)
616White
616Brown
715Pink
7 15 Yellow
814Shield
913Green
10 12 Blue
11 11 Gray
12 10 Red
MP-5 connector, 5 contacts
si628 C_HACH LANGE_Ed. 0510.indd ANG:12si628 C_HACH LANGE_Ed. 0510.indd ANG:12 20/5/10 11:08:4320/5/10 11:08:43

13
3. Installation
ENGLISH
3.4. Protective wiring of switching contacts
Relay contacts are subjected to electrical erosion.
Especially with inductive and capacitive loads, the service life of the contacts will be reduced.
For suppression of sparks and arcing, components such as RC combinations, nonlinear resistors, series
resistors and diodes should be used.
Important note:
Fit a jumper between terminals 9 and 10 (12 and 13 in panel mount instrument) and another between
the terminals 11 and 12 (10 and 11 in panel mount instrument).
3.3.2. Cells with connector S8
Typical protective wirings
AC applications with inductive load DC applications with
inductive load AC/DC with
capacitive load
Load
Contact
relay
Diode
1N4007
Typical RC at 230 VAC: Capacitor 0.1 µF/630 V
Resistor 100 Ω/1 W
RC
RC R
Load
R, for example:
5 Ω/ 1 W at
24 V / 0.4 A
Maximum relays load: AC < 24 V / < 3 A / < 72 VA.
TERMINALS COLOUR
si628 C
(wall mount) si628 C
(panel mount)
9 13 Middle
10 12 Middle
11 11 Shield
12 10 Shield
si628 C_HACH LANGE_Ed. 0510.indd ANG:13si628 C_HACH LANGE_Ed. 0510.indd ANG:13 20/5/10 11:08:4320/5/10 11:08:43

14
4. Operation
4.1. Description
Display
si628 C have an alphanumeric display that explains itself. One of the most significant screens is des-
cribed below as an example:
The instrument is already measuring conductivity,
with the following configuration:
• Temperature coefficient, CT, fixed at 2% / °C.
• Cell constant C=1 cm-1.
• Reference temperature, RT, fixed at 25°C.
• Limit relay is activated at 1000 µS/cm, and
desactivated at 800 µS/cm.
• Alarm relay is activated after 1minute with the
reading out of limit.
• Output current 4 mA = 200 µS/cm, 20 mA =
2000 µS/cm.
To modify the default configuration (except from
the temperature Coefficient and the reference tem-
perature) see page 16.
4.2. Setting up
Connect the si628 C to mains. At power up the first time only the following screens will appear:
Keys
ENTER.
• Horizontal movements to move forward in the “diagram”.
• Acceptance of numerical values.
ESCAPE.
• Horizontal movements to move back in the “diagram”.
• Vertical movements to move back in the “diagram”.
• To erase numerical values.
• Modification of numerical values.
• Cursor “>” movements on the screen.
si628_C
V1.0
Français
English
Deutsch
Português
LANGUAGE >Español
Italiano
MEASURE S/cm MEASURING 22.3°C
545 µS/cm @25°
Possibility to select other
language with keys.
MEASURING 23.2 °C
538 µS/cm @25 °C
Sample temperature
Resolution:
1 °C manual selection.
0.1 °C measuring with Pt 1000
Measured value.
Reference temperature
Measuring units.
si628 C_HACH LANGE_Ed. 0510.indd ANG:14si628 C_HACH LANGE_Ed. 0510.indd ANG:14 20/5/10 11:08:4420/5/10 11:08:44

15
4. Operation
ENGLISH
4.3. Quick guide
This page shows the instrument’s operation by blocks. It is intended to provide a quick idea about all the
instrument’s capabilities and how to gain access to them
Security access
code.
It limits the access
to calibration
or programming
of the instrument
by non authorised
people.
When the “Code” screen has been shown for
more than 30 sec. the instrument will automatically
return to MEASURE.
MEASURING 22.3°C
545 µS/cm @25°
CÓDIGO 000
MEASURE S/cm
CALIBRATE
CELL
LIMIT
ALARM
HOLD
LANGUAGE
See “calibration... “
(page 17)
See “programming...”
(page 16)
Vary the underlined digit with
and accept with .
Authorised code = 100
Instrument in hold state.
When being in the grey zone, the
instrument is in hold state:
• Contact relays are opened.
• According to the configuration, the
current output is frozen at 21 mA or
at the mA corresponding to the last
measured value.
To quit the hold state it is necessary to
return to MEASURE. The current output
4 to 20 will activate in 10 sec.
OUTPUT 4-20 MA
si628 C_HACH LANGE_Ed. 0510.indd ANG:15si628 C_HACH LANGE_Ed. 0510.indd ANG:15 20/5/10 11:08:4520/5/10 11:08:45

16
4. Operation
4.4. Programming
The instrument is supplied with the default configuration specified in Setting up, page 14.
To modify the default configuration and adapt it to specific applications follow the diagram below.
Select the nominal value of cell
constant to be used. If the exact
value is known, it can be directly
entered, avoiding the calibration
procedure.
Selection of the functioning
values of the corresponding
relays.
Input “ON” (closed contact).
Output “OFF” (open contact).
Selectable values between
0.0 µS and 200 mS.
Select the time before the alarm
activation.
Selectable between 1 to 99 min
or hours. 0= deactivated.
Select the conductivity values
corresponding to the Current
Output (4 mA and 20 mA).
Selectable values between
0.0 µS and 200 mS.
Selection of the current output
when the instrument is in Hold
state: fixed at 21 mA or at the
mA corresponding to the last
measured value.
Selection of the working
language: English, German,
Spanish, French, Italian,
Portuguese.
(*) LIMIT and OUTPUT 4-20 mA:
Allows the introduction of five
digits, placing the decimal point
you want.
Note: When programming, the instrument is
in hold state:
• The contact relays are open.
• The current output is frozen at 21 mA.
To quit the hold state it is necessary to return
to MEASURE.
Vary the underlined digit with
and accept with
Vary the underlined digit with
and accept with
CELL
CALIBRATE
CODE 000
CONST:
1.000 cm –1
LIMIT (*) ON - - - - - - OFF
1000 µS 800 µS
LANGUAGE LANGUAGE >Español
Italiano
MEASURE S/cm MEASURING 22.3°C
545 µS/cm @25°
OUTPUT 4-20 mA 4 mA 20 mA
pH: 0.0 2.0 mS
MEASURING 22.3°C
545 µS/cm @25°
Enter code 100.
STORING DATA
HOLD
ALARM
>21 mA
last
ALARM:
01 min
Vary with
Vary with
si628 C_HACH LANGE_Ed. 0510.indd ANG:16si628 C_HACH LANGE_Ed. 0510.indd ANG:16 20/5/10 11:08:4520/5/10 11:08:45
Table of contents
Languages:
Other HACH LANGE Transmitter manuals
Popular Transmitter manuals by other brands

Moseley
Moseley DSP6000A user manual

Technalogix