Hach BODTrak II User manual

DOC022.53.80363
BODTrak II™
11/2020, Edition 3
User Manual


Table of Contents
Section 1 Specifications.......................................................................................................................................................... 3
Section 2 General information............................................................................................................................................. 3
2.1 Safety information....................................................................................................................................................................... 3
2.1.1 Use of hazard information................................................................................................................................................. 4
2.1.2 Precautionary labels......................................................................................................................................................... 4
2.1.3 Certification....................................................................................................................................................................... 4
2.2 Product overview........................................................................................................................................................................ 5
2.3 Product components................................................................................................................................................................... 5
Section 3 Installation................................................................................................................................................................. 5
3.1 External connections................................................................................................................................................................... 5
3.2 Connect the RS232 interface...................................................................................................................................................... 6
3.3 Bottle connections....................................................................................................................................................................... 6
Section 4 User interface.......................................................................................................................................................... 6
4.1 Channel selection keys............................................................................................................................................................... 7
4.2 Arrow keys.................................................................................................................................................................................. 7
4.3 ON key........................................................................................................................................................................................ 7
4.4 OFF key...................................................................................................................................................................................... 7
Section 5 Startup......................................................................................................................................................................... 7
5.1 Turn the instrument on................................................................................................................................................................ 7
5.2 Set the clock............................................................................................................................................................................... 7
Section 6 Standard operation.............................................................................................................................................. 8
6.1 Download test results.................................................................................................................................................................. 8
6.2 Import data.................................................................................................................................................................................. 8
6.3 Data format................................................................................................................................................................................. 8
6.4 Print test results.......................................................................................................................................................................... 9
Section 7 BODTrak II™ procedures.................................................................................................................................. 9
7.1 Procedure notes.......................................................................................................................................................................... 9
7.2 Simplified procedure................................................................................................................................................................. 10
7.3 Hach Standard Method procedure............................................................................................................................................ 11
1

7.4 Completion steps for all procedures......................................................................................................................................... 12
Section 8 Determination of results................................................................................................................................. 14
8.1 Simplified sample results.......................................................................................................................................................... 14
8.2 Hach Standard Method results................................................................................................................................................. 14
Section 9 Typical curves....................................................................................................................................................... 15
Section 10 Special considerations................................................................................................................................. 15
10.1 Sample dilution....................................................................................................................................................................... 15
10.2 Sample seeding...................................................................................................................................................................... 16
10.3 Sample temperature............................................................................................................................................................... 16
10.4 Toxic materials........................................................................................................................................................................ 16
10.5 Chlorine................................................................................................................................................................................... 16
10.6 pH effect.................................................................................................................................................................................. 16
10.7 Supersaturation....................................................................................................................................................................... 16
Section 11 Maintenance.........................................................................................................................................................17
11.1 Cleaning.................................................................................................................................................................................. 17
11.2 Storage................................................................................................................................................................................... 17
Section 12 Troubleshooting................................................................................................................................................ 17
12.1 Incorrect BOD curves.............................................................................................................................................................. 17
12.2 High oxygen demand.............................................................................................................................................................. 18
12.3 Nitrification.............................................................................................................................................................................. 18
12.4 Excessive time lag.................................................................................................................................................................. 19
12.5 Sample temperature............................................................................................................................................................... 19
12.6 Bottle leak............................................................................................................................................................................... 19
Section 13 Replacement parts and accessories.................................................................................................... 19
13.1 Replacement parts.................................................................................................................................................................. 19
13.2 Required reagents.................................................................................................................................................................. 19
13.3 Optional reagents.................................................................................................................................................................... 19
13.4 Accessories............................................................................................................................................................................. 20
Table of Contents
2

Section 1 Specifications
Specifications are subject to change without notice.
Table 1 General Specifications
Specification Details
Operating
temperature
5 to 40 ºC (41 to 104 ºF)
Altitude limit 2000 m (6500 ft)
Pollution degree 2
Installation
category
II
Storage/operatin
g humidity
Maximum relative humidity is 80% for temperatures
up to 31 ºC (87.8 ºF), decreases linearly to 50%
relative humidity at 40 ºC (104 ºF)
Location Laboratory / Indoor
Protection class 2
Range Selectable, 0 to 35, 0 to 70, 0 to 350, 0 to 700 mg/L
Dimensions 28.9 x 26 x 9.8 cm (11.375 x 10.25 x 3.875 in.)
External power
supply
Input: 100 to 240 VAC, 50/60 Hz, 1.5 A; Output:
24 VDC, 2.7 A, UL CSA and TUV approved.
Capacity Six 492 mL bottles
Shipping weight 4 kg (8.8 lb)
Warranty 1 year (EU 2 years)
Table 2 Method performance specifications
Specification Details
Precision Parameters:
• Standard: 150 mg/L each of glucose and
glutamic acid
• Number of samples: 44
• Number of analysts: 1
• Number of BodTrak II instruments: 6
Results:
• Mean of 235 mg/L BOD
• Distribution: 11 mg/L or range of 224 to
246 mg/L BOD
• 95% confidence limit
Drift Less than 3 mg/L BOD in 5 days
Resolution 1 mg/L BOD
Section 2 General information
In no event will the manufacturer be liable for direct, indirect, special,
incidental or consequential damages resulting from any defect or
omission in this manual. The manufacturer reserves the right to make
changes in this manual and the products it describes at any time, without
notice or obligation. Revised editions are found on the manufacturer’s
website.
2.1 Safety information
The manufacturer is not responsible for any damages due to
misapplication or misuse of this product including, without limitation,
direct, incidental and consequential damages, and disclaims such
damages to the full extent permitted under applicable law. The user is
soley responsible to identify critical application risks and install
English 3

appropriate mechanisms to protect processes during a possible
equipment malfunction.
Please read this entire manual before unpacking, setting up or operating
this equipment. Pay attention to all danger and caution statements.
Failure to do so could result in serious injury to the operator or damage
to the equipment.
Make sure that the protection provided by this equipment is not impaired.
Do not use or install this equipment in any manner other than that
specified in this manual.
2.1.1 Use of hazard information
DANGER
Indicates a potentially or imminently hazardous situation which, if not
avoided, will result in death or serious injury.
W A R N I N G
Indicates a potentially or imminently hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in minor or
moderate injury.
NOTICE
Indicates a situation which, if not avoided, may cause damage to the
instrument. Information that requires special emphasis.
2.1.2 Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or
damage to the instrument could occur if not observed. A symbol on the
instrument is referenced in the manual with a precautionary statement.
This symbol, if noted on the instrument, references the
instruction manual for operation and/or safety information.
Electrical equipment marked with this symbol may not be
disposed of in European domestic or public disposal
systems. Return old or end-of-life equipment to the
manufacturer for disposal at no charge to the user.
2.1.3 Certification
CAUTION
This equipment is not intended for use in residential environments and
may not provide adequate protection to radio reception in such
environments.
Canadian Radio Interference-Causing Equipment Regulation,
ICES-003, Class A:
Supporting test records reside with the manufacturer.
This Class A digital apparatus meets all requirements of the Canadian
Interference-Causing Equipment Regulations.
Cet appareil numérique de classe A répond à toutes les exigences de la
réglementation canadienne sur les équipements provoquant des
interférences.
FCC Part 15, Class "A" Limits
Supporting test records reside with the manufacturer. The device
complies with Part 15 of the FCC Rules. Operation is subject to the
following conditions:
1. The equipment may not cause harmful interference.
2. The equipment must accept any interference received, including
interference that may cause undesired operation.
Changes or modifications to this equipment not expressly approved by
the party responsible for compliance could void the user's authority to
operate the equipment. This equipment has been tested and found to
comply with the limits for a Class A digital device, pursuant to Part 15 of
4 English

the FCC rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is operated
in a commercial environment. This equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in
accordance with the instruction manual, may cause harmful interference
to radio communications. Operation of this equipment in a residential
area is likely to cause harmful interference, in which case the user will be
required to correct the interference at their expense. The following
techniques can be used to reduce interference problems:
1. Disconnect the equipment from its power source to verify that it is or
is not the source of the interference.
2. If the equipment is connected to the same outlet as the device
experiencing interference, connect the equipment to a different
outlet.
3. Move the equipment away from the device receiving the interference.
4. Reposition the receiving antenna for the device receiving the
interference.
5. Try combinations of the above.
2.2 Product overview
Respirometric Biological Oxygen Demand (BOD) is a test that measures
the quantity of oxygen consumed by bacteria that oxidize organic matter
in a water sample. The test is used to measure waste loadings at
wastewater treatment plants and to examine the efficiency of wastewater
treatment.
The instrument is sealed to prevent external atmospheric pressure
changes in the test bottle. The pressure in the sample bottles is
monitored. Bacteria in the sample use oxygen when they consume
organic matter. This oxygen consumption causes the pressure in the
bottle head space to drop. The pressure drop correlates directly to BOD.
During a test period, stir bars mix the sample and cause oxygen to move
from the air in the bottle to the sample. This helps simulate natural
conditions.
Carbon dioxide is a result of the oxidation process and can interfere with
a measurement. The instrument continuously removes carbon dioxide
from the system so that the monitored pressure difference stays
proportional to the quantity of oxygen used. Pressure changes in the
closed system are shown graphically in milligrams per liter (mg/L) on a
liquid crystal display. The instrument gives 360 uniform data points over
the selected time period.
The instrument adjusts for any negative errors produced when heat is
applied to a sample. The instrument does not start the test until the
temperature gets to equilibrium.
2.3 Product components
Make sure that all components have been received. If any of these items
are missing or damaged, contact the manufacturer or a sales
representative immediately.
• BODTrak™ II instrument
• A UL/CSA approved 115 VAC power cord with a NEMA 5-15P style
plug
• A 230 VAC harmonized power cord with a continental European plug
• Power supply, auto-switching between 115 V and 230 V
• Seal cups (6x)
• BODTrak II amber sample bottles (6x)
• BODTrak II magnetic stir bars (6x)
• Spatula scoop
• Nutrient buffer solution pillows (1 pkg)
• Potassium hydroxide pellets (1 container)
Section 3 Installation
3.1 External connections
Figure 1 shows the locations of the power switch and external
connections.
English 5

Figure 1 External connections
1 Power switch 3 Serial I/O port
2 DC power connector
3.2 Connect the RS232 interface
All RS232 connections are made through the serial I/O port. Connect the
9-pin D connector of a computer interface cable to the serial I/O port on
the instrument (Figure 1 on page 6). Connect the other end of the cable
to the computer serial I/O port (COM 1 or COM 2).
The instrument is equipped as Data Communication Equipment (DCE).
The instrument operates at 9600 baud with 8 data bits, no parity and one
stop bit. The computer or printer will not receive complete transmissions
if the device cannot continuously receive at 9600 baud.
Note: The use of the specified cable or an equivalent shielded cable is necessary
to meet radio frequency emissions requirements.
3.3 Bottle connections
Each bottle position/channel has the applicable tube numbered with a
plastic sleeve. The bottle positions are numbered 1 through 6 with
number 1 in the back left corner of the chassis. Use the channel
selection keys as a guide to the bottle positions Figure 2 on page 6.
Section 4 User interface
The instrument display and the keypad are shown in Figure 2.
Figure 2 Display and keypad
1 Display 4 Arrow keys
2 Channel selection keys 5 Power indicator
3 ON and OFF keys1
1The ON and OFF keys are used to start and stop a test. They do not power the instrument on and off.
6 English

4.1 Channel selection keys
Push the related channel selection key to show data for one of the six
bottles. The channel selection keys are also used in the instrument setup
menu to select a parameter to be edited. Refer to Figure 2 on page 6
and Table 3.
Table 3 Channel key setup parameters
Channel Parameter
1 Year (0–99)
2 Month (1–12)
3 Day (1–31)
4 Hour (0–24)
5 Minute (0–59)
6 Test length (5, 7 or 10 days)
4.2 Arrow keys
The display shows a graph of BOD values on the vertical axis and time
in days on the horizontal axis. Push the LEFT and RIGHT arrows to
move the cursor along the BOD curve to show the approximate
coordinates (time, BOD) of the selected data point.
The time interval and BOD value of the data point are shown in the lower
right of the display. The cursor is automatically placed at the most
recently collected data point in a channel display.
Push and hold the LEFT and RIGHTarrows at the same time to go into
the instrument setup menu. The arrow keys are also used to change the
time, date, test length and range.
4.3 ON key
To go to the range selection menu, push ON from a channel display
screen. Then push and hold ON to start the test for the selected channel.
4.4 OFF key
When a test is in DELAY or RUN modes, push and hold OFF to
manually end the test. The instrument will show END. The OFF key is
also used to exit the instrument setup menu or range selection menu.
The changes made before the menu is exited are saved.
Section 5 Startup
5.1 Turn the instrument on
Note: The ON and OFF keys are used to start and stop a test. They do not power
the instrument on and off.
1. Connect the power adaptor to the DC power connector (Figure 1
on page 6).
2. Toggle the power switch to set the instrument to on and off. (Figure 1
on page 6).
5.2 Set the clock
All the channels must show END or CLEAR before the clock can be set.
1. Push and hold the two arrow keys at the same time until the
instrument setup menu is shown.
2. Push the applicable channel key to select the clock parameter to be
adjusted.
3. Use the arrow keys to edit the selected parameter. Adjust each
parameter in the same manner.
4. When all the time adjustments are complete, push OFF to save and
go back to the data display screen.
English 7

Section 6 Standard operation
6.1 Download test results
To transfer test results to a PC2:
1. Select
PROGRAMS>ACCESSORIES>COMMUNICATIONS>HYPERTERM
INAL.
2. In the Connection Description window, type in a name for the
connection and select an icon to represent the connection. Click OK.
3. In the Connect To window, use the drop-down menu to select the
COM port connected to the instrument. Click OK.
4. Configure the COM port properties: BPS = 9600, Data Bits = 8,
Parity = None, Stop Bits = 1, Flow Control = None.
5. Click OK. The connector indicator shows.
6. Select TRANSFER>CAPTURE TEXT.
7. In the Capture Text window, click START.
8. Power the instrument on. Push the key for the channel that has data
to be downloaded.
9. Type GA in the HyperTerminal window and push ENTER. The
transfer is complete when the screen stops adding new data.
10. Select TRANSFER>CAPTURE TEXT>STOP.
11. Select CALL>DISCONNECT. The disconnected indicator shows.
12. To end the HyperTerminal session, select FILE>EXIT.
6.2 Import data
To import the data from the captured text file:
1. Open a new or existing spreadsheet. Select DATA>IMPORT
EXTERNAL DATA>IMPORT DATA.
2. Select the text file captured in HyperTerminal. Click IMPORT.
3. In the Text Import Wizard, select Delimited as the file type, the start
row in the spreadsheet and Windows (ANSI) as the file origin. Click
NEXT.
4. Click the check boxes for Space delimiter and Treat consecutive
delimiters as one. Click NEXT.
5. Select General as the column data format, then click FINISH.
6. In the Import Data window, select Existing worksheet. Select the
starting cell, then click OK. The data will appear in the spreadsheet.
The spreadsheet data cannot be edited or formatted in
HyperTerminal or with the BODTrak II.
6.3 Data format
When a result array is downloaded to HyperTerminal, all of the data from
a test is sent without pause. The data flow cannot be stopped or paused.
The example shows the channel number, start date, start time and the
format of the downloaded data. BOD values in mg/L follow. Only the first
data points of a maximum of 360 equal distance points are shown. Each
line ends with a carriage return and a line feed. The end of the data
stream is shown by a message such as "Test Run to Completion" and a
dollar symbol ($).
If small negative BOD values are seen at the start of a test, refer to
Troubleshooting on page 17.
Example of the data format
BOD Log for Ch 1
Status: END
Full Scale: 700 mg/L
Tst length: 7 days
Start Date: 3/3/08
Time: 13:04
Days, Reading (mg/L)
0.00, 0
2Installed on Windows XP or older. On newer Windows versions the installation of a terminal software is necessary.
8 English

0.05, 10
0.11, 12
0.16, 12
0.22, 14
0.27, 14
0.33, 12
0.38, 8
0.44, 10
0.50, 12
0.55, 12
0.61, 14
-
-
-
Test Run to Completion
$
6.4 Print test results
The BODTrak II is applicable to use with the Citizen PD-24 printer.
1. Connect the printer cable to the serial I/O port on the instrument. Use
the gender adapter supplied with the printer to make the connection.
Make sure that the printer settings are correct (Connect the RS232
interface on page 6).
2. Power on the instrument.
3. Push and hold the applicable channel number for approximately
5 seconds at any time during a test.
The test results move from the instrument to the printer. The
instrument sends a copy of the graphical display and a truncated
data stream (127 data points).
Section 7 BODTrak II™ procedures
CAUTION
Chemical exposure hazard. Obey laboratory safety
procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled.
Refer to the current safety data sheets (MSDS/SDS) for
safety protocols.
CAUTION
Chemical exposure hazard. Dispose of chemicals and
wastes in accordance with local, regional and national
regulations.
7.1 Procedure notes
There are two procedure variations. Select the procedure that meets the
application requirements.
• The Simplified procedure is recommended when sample seeding,
extra nutrients or buffers are not necessary. It is also recommended
when accuracy requirements are not stringent.
• The Hach Standard Method procedure is recommended when
samples are seeded or extra nutrients or reagents are added. Use this
procedure for the Standard Methods for the Examination of Water and
Wastewater, 21st Edition, Method 5219 D Respirometric Method.
After any of the procedure variations, do the completion steps. Refer to
Completion steps for all procedures on page 12. It is possible to use a
combination of these procedures with one instrument but in different
bottles. Only one test length can be chosen.
Before the test:
English 9

Use the applicable sample volume tables for each procedure.
If power is interrupted when the instrument is in DELAY status, the
test will stop and the status will change to CLEAR when power
returns. Start the test again. If power is interrupted when the
instrument is in RUN status, the test will resume when power returns.
Keep deionized water overnight in an incubator at 20 ºC. Shake the
deionized water to saturate with air.
Let the seed become stable overnight in the BOD incubator at a
temperature of 20 ºC. Be careful not to disturb the settled solution.
Pipet seed solution from the top.
Dilution is necessary if samples have BOD values more than
700 mg/L.
At elevations higher than 5000 feet above sea level, the 0 to 35 mg/L
BOD range is decreased to 0 to 25 mg/L BOD. Adjustment is not
necessary for other test ranges.
Refer to Special considerations on page 15 for special
considerations including sample seeding and pretreatment.
Use only BODTrak II stir bars and bottles. These items are made
specifically to be used with the BODTrak II.
7.2 Simplified procedure
Items to collect:
• BODTrak II bottle
• Thermometer
• Blender (optional)
• Graduated cylinder
• Nutrient buffer pillow (1x)
Use the correct sample volume for the BOD range as shown in Table 4.
Table 4 Simplified sample volumes
BOD range (mg/L) Sample volume (mL)
0 to 35 420
0 to 70 355
0 to 350 160
0 to 700 95
1. Heat or cool the
sample to 19 to
21 ºC.
2. Homogenize the
sample in a blender
if it contains large
settleable or
floatable solids.
3. Select the correct
sample volume for
the BOD range
(Table 4). Measure
the sample into a
graduated cylinder.
10 English

4. Add the contents
of 1 nutrient buffer
pillow to the
graduated cylinder.
5. Move the
contents of the
graduated cylinder to
a BODTrak II bottle.
Continue to
Completion steps for
all procedures
on page 12. Repeat
steps 1 to 5 for
additional samples.
7.3 Hach Standard Method procedure
Items to collect:
• Thermometer
• BODTrak II bottle
• Blender (optional)
• Graduated cylinder
• Tensette pipet and pipet tips
• Wash water bottle
• Ampule breaker (for GGA)
Required reagents:
• 1 nutrient buffer pillow
• Additional nutrient or buffer (optional)
• Deionized water
• Hach GGA Solution (optional)
Before starting the test:
Use the sample volume table to select the correct sample size.
If samples are seeded, prepare a seed blank before a sample is
prepared. Treat the seed blank the same as any other sample and
omit step 5.
Use Hach BOD Standard Solution Ampules for Manometric Method
(3000 mg/L Glucose, 3000 mg/L Glutamic Acid) for GGA.
Refer to Special considerations on page 15 for special
considerations.
Use the correct volumes and dilution factor for the BOD range as shown
in Table 5.
Table 5 Hach Standard Method sample volumes
BOD range
(mg/L)
Sample
volume
(mL)
Seed
volume (mL)
Final
volume
(mL)
Dilution
factor
0 to 35 370 10 to 35 420 1.14
0 to 70 305 10 to 35 355 1.16
0 to 350 110 10 to 35 160 1.45
0 to 700 45 10 to 35 95 2.11
GGA
(0 to 350)
8 10 to 35 160 —3
3The GGA result calculation does not include a dilution factor. Refer to Hach Standard Method results on page 14, step 5 to calculate the GGA
result.
English 11

Note: If the seed strength is unknown, use 20 mL. Adjust the seed volume as
necessary to achieve optimum results.
1. Heat or cool the
sample to 19 to
21 ºC (66 to 70 ºF).
2. Homogenize the
sample in a blender
if it contains large
settleable or
floatable solids.
3. Select the correct
GGA or sample
volume for the BOD
range (Table 5).
Measure the sample
into a graduated
cylinder.
4. Add the contents
of 1 nutrient buffer
pillow to the
graduated cylinder.
5. If the sample is
seeded, use a
tensette pipet to add
the correct quantity
of seed to the
graduated cylinder
(Table 5).
6. If necessary, add
more nutrient or
buffer. Do not add
more than a total
volume of 50 mL
(seed, nutrient,
buffer).
7. Fill to the final
test range volume if
necessary with a
deionized wash
water bottle
(Table 5).
8. Move the
prepared sample
from the graduated
cylinder to a
BODTrak II bottle.
Continue to
Completion steps for
all procedures
on page 12.
Note: Repeat steps 1 to
8 for additional
samples.
7.4 Completion steps for all procedures
Items to collect:
• BODTrak II
• Spatula scoop
• BOD incubator
• Seal cup
• Stir bar
Required reagents:
• 2 potassium hydroxide pellets
12 English

1. Put a stir bar into
the bottle.
2. Put a seal cup
into the neck of the
bottle.
3. Use a spatula
scoop to add 2
potassium hydroxide
pellets to the seal
cup. Repeat steps 1
to 3 for each sample
bottle.
4. Put the bottles on
the BODTrak II
chassis. Connect the
applicable tube to
the sample bottle
and tighten the cap.
5. Put the
instrument in the
incubator. The
incubator
temperature must be
20 ± 1 ºC.
Note: Instrument
performance has not
been tested at other
temperatures.
6. Plug in and
power on the
instrument. Make
sure that all the stir
bars are rotating. If
not, lift the bottle up
then set it down
again.
7. Push and hold
the LEFT and
RIGHT arrows at the
same time to go to
the instrument setup
menu.
Note: Set the time and
date if necessary (Set
the clock on page 7).
8. Push the channel
6 key to go to the
test length
parameter. Use the
arrow keys to select
a 5, 7 or 10 day test.
Note: The selected test
length is for all six
channels.
9. Push OFF to
save selections and
exit the menu.
10. To start the test,
push the channel
number for the
bottle.
11. Push ON. The
range selection
menu is shown.
12. Use the arrows
to select the test
range.
Note: Use the left arrow
key for the 0 to 35 and
0 to 70 mg/L ranges.
Use the right arrow key
for the 0 to 350 and 0 to
700 mg/L ranges.
English 13

13. Push and hold
ON to start a test. A
graph will show.
Note: To cancel a test,
push and hold OFF.
There is a built-in 1
hour instrument/sample
equilibration period
before data collection.
The display shows
"DELAY" during this
period.
14. Do steps 10
through 13 again to
set the test range
and start each of the
six channels. It is not
necessary to operate
all six channels if
less than six
samples are
available.
Section 8 Determination of results
After the end of the selected test period (5, 7 or 10 days), the display
shows "END". The results are given based on the selected procedure:
Simplified or Hach Standard Method.
8.1 Simplified sample results
The simplified procedure results are shown on the BODTrak II display.
Push the applicable channel selection key to show the results.
Note: If the sample was pre-diluted, apply a dilution factor to the instrument
reading. Refer to Sample dilution on page 15.
8.2 Hach Standard Method results
1. Push the channel selection key for the Hach Standard Method
sample bottles, and the seed blank bottle and GGA bottle when
applicable. The results are shown.
Note: Treat the seed blank the same as all other samples. If the sample was
pre-diluted, apply a dilution factor to the instrument reading (Sample dilution
on page 15).
2. Find the dilution factor based on the selected range (Table 5
on page 11). For example, if the sample range selected is 0 to
350 mg/L BOD, the dilution factor is 1.45.
3. Calculate the corrected results:
BOD mg/L = BOD mg/L (instrument reading) x dilution factor
Example:
• Instrument Reading = 200 mg/L
• BOD dilution factor = 1.45
200 mg/L x 1.45 = 290 mg/L BOD (corrected result)
4. When the samples are seeded (not GGA), calculate the results with
the equation that follows and the corrected results:
BOD (mg/L) = A – [B x (SA÷SB)]
Where:
• A = corrected BOD of the seeded sample
• B = corrected BOD of the seed blank
• SA = volume of seed in seeded sample
• SB = volume of sample in seeded sample
Example:
• A = 290 mg/L BOD
• B = 120 mg/L BOD
• SA = 20 mL
• SB = 110 mL
BOD (mg/L) = 290 mg/L - [120 mg/L x (20 mL÷110 mL)]
BOD mg/L = 268 mg/L
5. For GGA, calculate the results of the GGA with the equation that
follows and the corrected BOD of the seed blank:
BOD (mg/L) = C – [B x (SA÷SC)]
Where:
• C = instrument reading on seeded GGA
14 English

• B = corrected BOD of the seed blank
• SA = volume of seed in seeded GGA
• SC = final volume of seeded GGA (160 mL)
Example:
• C = 250 mg/L BOD
• B = 120 mg/L BOD
• SA = 20 mL
• SC = 160 mL
BOD (mg/L) = 250 mg/L - [120 mg/L x (20 mL÷160 mL)]
BOD mg/L = 235 mg/L
The GGA result is 235 ± 11 mg/L BOD.
Section 9 Typical curves
Refer to the expanded version of the manual for information about
specific procedures.
Figure 3 shows typical curves through a 10 day test period. For incorrect
curves, refer to Figure 4 on page 18.
Figure 3 Typical curves
1 Typical with substrate variation 3 Typical with time lag
2 Typical
Section 10 Special considerations
10.1 Sample dilution
Unknown sample BOD effluent is typically in the 0 to 70 mg/L range.
Unknown sample BOD influent is typically in the 0 to 700 mg/L range.
English 15

When the oxygen requirement of a sample is more than 700 mg/L, dilute
the sample with high-quality distilled or deionized water.
Calculate the results to include the additional dilution factor. Example: if
the BOD of the sample is 1000 mg/L, dilute the sample 1:1 with distilled
or deionized water. The estimated BOD is now 500 mg/L. Use the
sample volume specified in the table for the 0 to 700 mg/L range of the
selected procedure. Multiply the instrument reading result by 2. If the
Hach Standard Method procedure is used, continue with the rest of the
calculations.
10.2 Sample seeding
Some types of BOD samples do not contain sufficient bacteria to oxidize
the organic matter in the sample. Many industrial wastes are of this type.
Some sewage treatment plant effluents are chlorinated and essentially
sterile. A BOD test cannot be done in the absence of viable bacteria. To
test such samples, seed each bottle from a source known to contain a
viable bacterial population.
Settled domestic wastewater plant influent or primary clarifier effluent are
the preferred sources of seed for most samples. Mixed liquor or
undisinfected effluent can be used for seed, but it is recommended to
include a nitrification inhibitor. Commercial seed sources are sometimes
suitable. To prepare, see the instructions from the manufacturer.
10.3 Sample temperature
Standard Methods for the Examination of Water and Wastewater, 21st
Ed., 2005 5210 D recommends an incubation temperature of 20 ±1 ºC
(68 ºF) for the BOD test. Put the instrument in an incubator that is
adjusted to 20 ±1 ºC. Warm or cool samples to 20 ±1 ºC.
Instrument performance has not been validated at temperatures other
than 20 ºC.
10.4 Toxic materials
Industrial and chlorinated samples often contain toxic substances and
require special considerations when running BOD tests. Toxic materials
in the sample will cause decreased BOD values. Dilute the sample to
minimize the toxic materials or their effects. Refer to Standard Methods
for the Examination of Water and Wastewater, 21st edition, 5210 D.
10.5 Chlorine
Chlorine in the sample must be removed prior to testing. Keep the
sample at room temperature for 1 to 2 hours before a test to dissipate
low chlorine concentrations. If any chlorine remains after sitting for
2 hours, or if the chlorine concentration is high, add sodium thiosulfate to
remove the chlorine.
1. In a 250-mL Erlenmeyer flask, add 100 mL of sample.
2. Add 10 mL of 100 g/L potassium iodide solution and 10 mL of 0.02 N
sulfuric acid standard solution to the Erlenmeyer flask.
3. Add 3 droppers of starch indicator solution and swirl to mix.
4. Titrate from dark blue to colorless with 0.025 N Sodium Thiosulfate
standard solution.
5. Calculate the quantity of sodium thiosulfate standard solution
necessary to dechlorinate the remaining sample:
mL of sodium thiosulfate = (mL used)×(mL sample to be
dechlorinated)÷100
6. Add the necessary quantity of 0.025 N sodium thiosulfate standard
solution to the sample and mix fully. After 10 to 20 minutes, do the
BOD test.
10.6 pH effect
Low BOD test results occur when sample pH is outside the range of 6 to
8. Keep this pH to simulate source sample conditions or adjust the pH to
neutrality (buffered at pH 7). Use 1.0 N (or weaker) sulfuric acid to
neutralize caustic samples. Use 1.0 N (or weaker) sodium hydroxide to
neutralize acidic samples. When samples are pH adjusted, they should
also be seeded.
10.7 Supersaturation
Equilibrate supersaturated cold samples (samples that contain more
than 9 mg/L of dissolved oxygen at 20 ºC) to saturation.
16 English

1. Heat or cool the sample to approximately 20 ºC.
2. Half fill a sample bottle with sample.
3. Shake for 2 minutes or aerate with filtered compressed air for
2 hours.
Section 11 Maintenance
DANGER
Multiple hazards. Only qualified personnel must conduct the
tasks described in this section of the document.
CAUTION
Chemical exposure hazard. Obey laboratory safety
procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled.
Refer to the current safety data sheets (MSDS/SDS) for
safety protocols.
11.1 Cleaning
Clean the instrument
Clean spills on the instrument with a soft cloth that has been dampened
with deionized or distilled water.
Clean the sample bottles
Clean the sample bottles and caps with a brush, water and a mild
detergent. Flush the sample bottles with fresh water followed by a
distilled water rinse.
Clean the stir bars and seal cups
Clean the stir bars with hot water and soap. Use a brush to remove
deposits. Rinse with fresh water and then rinse with distilled water.
Carefully empty and rinse the seal cups with water. Invert to dry.
11.2 Storage
The bottle fences prevent tipping of the bottles and provide tubing
management in storage. For storage, put the tubing in the opening in the
bottle fence. Move the tubing counter-clockwise and secure the bottle
cap inside the fence.
Section 12 Troubleshooting
12.1 Incorrect BOD curves
Figure 4 shows incorrect BOD curves for a 10 day test period. For typical
curves, refer to Typical curves on page 15.
English 17

Figure 4 Incorrect curves
1 High oxygen demand 4 Initial sample temperature
below 20 ºC or supersaturated
with oxygen
2 Nitrification 5 Bottle leak
3 Excessive time lag
12.2 High oxygen demand
Refer to Figure 4 on page 18. Samples that are above range (for
example, a BOD over 350 mg/L when a 160-mL sample is taken) will
cause results as shown in Curve 1. Dilute the sample or use a higher
BOD range and a different sample volume. Refer to the Sample dilution,
Simplified procedure or the Hach Standard method procedure for more
information.
When the BOD range of a sample is unknown:
• Use the results from the Chemical Oxygen Demand (COD test).
Multiply the COD by 0.68 to get an estimated BOD value.
• Use the results from a series of BOD tests that use the same sample
but different volumes.
• Use dilution ratios to select an applicable BOD range.
Typically, effluent is in the 0–70 mg/L range while influent is in the
0-700 mg/L range. when the BOD of the sample is more than 700 mg/L,
prepare a sample dilution. Refer to the Sample dilution section in the
expanded version of this manual for more information.
12.3 Nitrification
Refer to Figure 4 on page 18. The condition shown by Curve 2 is an
example of nitrification. Deviation from the typical curve (shown as the
dashed line) is apparent by the concave increase near the end of the
test period.
Biological oxidation of organic nitrogen usually occurs after 5 days with
typical domestic waste. Nitrifying bacteria develop more slowly than
other types of bacteria.
Some samples contain a high concentration of nitrifying bacteria and
nitrification results can occur sooner. Control nitrification problems with
Hach Nitrification Inhibitor. Dispense the inhibitor powder into an empty
sample bottle and then add the sample. With the Hach Dispenser cap,
dispense 6 shots (approximately 0.48 grams) into the empty bottle. Refer
to Optional reagents on page 19.
18 English
Other manuals for BODTrak II
3
Table of contents
Other Hach Measuring Instrument manuals

Hach
Hach EZ Series Quick manual

Hach
Hach POLYMETRON 83 series User manual

Hach
Hach H128 User manual

Hach
Hach MTC101 User manual

Hach
Hach QbD1200+ User manual

Hach
Hach A1000 Quick manual

Hach
Hach APA 6000 User manual

Hach
Hach 51 Series User manual

Hach
Hach CLF10sc User manual

Hach
Hach Polymetron 9610sc SiO2 User manual

Hach
Hach DR300 User manual

Hach
Hach ORBISPHERE G1100 User manual

Hach
Hach FT660 sc Quick manual

Hach
Hach TL2300 Quick manual

Hach
Hach sension1 User manual

Hach
Hach sensION+ pH3 User manual

Hach
Hach DR 3900 Quick manual

Hach
Hach Surface Scatter 7 User manual

Hach
Hach 5500sc Instruction Manual

Hach
Hach POCKET COLORIMETER II User manual