manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Hach
  6. •
  7. Measuring Instrument
  8. •
  9. Hach Sension4 Reference guide

Hach Sension4 Reference guide

51775-23
Laboratory pH/ISE Meter Quick Reference Card
Changing Meter Setup Options
Setup options 1–6 and 10 have default settings (underlined in the table). To change the default
setting, use the table below. Settings for options 7–9 must be entered from the setup menu.
Setup How To Get There Description of Option
1 Turn the meter on and press
SETUP. Press ENTER.Choose between using the BNC or 5-pin connector. ENTER
toggles between the options. Press EXIT when selection is
complete.
2 Turn the meter on and press
SETUP. Press the up arrow once.
Press ENTER.
Turn display lock On or Off. ENTER toggles between the
options. Press EXIT when selection is complete.
3 Turn the meter on and press
SETUP. Press the up arrow twice.
Press ENTER.
Choose between °C or °F for displayed temperatures.
ENTER toggles between the options. Press EXIT when
selection is complete.
4 Turn the meter on and press
SETUP. Press the up arrow three
times. Press ENTER.
Select measurement resolution from tenths, hundredths, or
thousandths. ENTER toggles between the options. Press
EXIT when selection is complete.
5 Turn the meter on and press
SETUP. Press the up arrow four
times. Press ENTER.
Select pH 6.86 or 7.00 as the pH buffer the meter will
automatically recognize during calibration. ENTER toggles
between the options. Press EXIT when selection
is complete.
6 Turn the meter on and press
SETUP. Press the up arrow until
the setup number is 6.
Press ENTER.
Enter the stability criteria (stability drift rate) using the
number keys (the default value is 0.500 mV). Press ENTER
to accept the value. Press EXIT when selection is complete.
7 Turn the meter on and press
SETUP. Press the up arrow until
the setup number is 7.
Press ENTER.
Select the time of day by pressing the desired number key.
The meter uses a 24-hour clock. All four places must have a
value. Use a 0 if the left-most digit does not have a value
(i.e., 03:08). Press ENTER to accept the value. Press EXIT
when selection is complete.
8 Turn the meter on and press
SETUP. Press the up arrow until
the setup number is 8. Press
ENTER.
Set the month and day by pressing the desired number
keys. The month uses the two places left of the slash; the
day of the month occupies the two places right of the slash.
Enter a 0 if the left-most digit does not have a value
(i.e., 02/06). Press ENTER to accept the value. Press EXIT
when selection is complete.
9 Turn the meter on and press
SETUP. Press the up arrow until
the setup number is 9.
Press ENTER.
Set the year by using the number keys. Every place must
have a value. Press ENTERto accept the value. Press EXIT
when selection is complete.
10 Turn the meter on and press
SETUP. Press the up arrow until
the setup number is 10.
Press ENTER.
Select the print interval by scrolling through the options with
the arrow keys (default is Off). When the desired option is
displayed, press ENTER to accept the value. Press EXIT
when selection is complete.
pH Calibration
1. Prepare up to five pH buffers (choose from pH 1.69, 4.01, 6.86, 7.0, 10.01 and 12.45 buffers).
The meter can use other buffers for calibration, if desired.
2. Turn the instrument on. Select the desired channel. Select the 5-pin connector (in Setup 1). If
necessary, press pH/mV to put the channel in pH mode.
3. Press CAL. The display will prompt for standard 1.
4. Place the pH electrode in one of the buffers. Go to step 6 if using auto-recognized buffers.
5. If not using auto-recognized buffers, press READ/ENTER. Use the number keys to enter the
pH value of the buffer. Press ENTER. It is not necessary that all four places have values
entered in them.
6. When the reading is stable, the meter will prompt for standard 2. Rinse the electrode with
deionized water.
7. Place the electrode in the second buffer. Press READ/ENTER.
8. Repeat steps 4–6for the third buffer or press EXIT.
9. When the reading is stable, the slope and Store ? will appear.
10. To save the calibration, press ENTER. To exit without saving the calibration, press EXIT.
ISE Calibration
1. Prepare two to seven standards. Pour 25 mL of each standard into a beaker.
2. Turn the instrument on. Select the channel. Select the BNC connector. If necessary, press
ISE/mV to put the channel in ISE mode.
3. Press CAL. Use the arrow keys to change the units of measurement, if desired.
4. If necessary, add ISA to the lowest standard. Place the electrode in the standard and stir.
5. Press ENTER. Enter or change the standard value using the number keys. Press ENTER to
accept the displayed value.
6. When the reading is stable, the meter will prompt for standard 2.
7. Rinse the electrode with deionized water. Repeat steps 4–6 for the remaining standards.
8. After the last standard is measured, press EXIT. Store ? will appear.
9. To save the calibration, press ENTER. To exit without saving the calibration, press EXIT.
© Hach Company, 1999. All rights reserved. Printed in the U.S.A hm/dk 5/99

Other Hach Measuring Instrument manuals

Hach LANGE DR 900 User manual

Hach

Hach LANGE DR 900 User manual

Hach ORBISPHERE K-M1100 Quick manual

Hach

Hach ORBISPHERE K-M1100 Quick manual

Hach DR/2010 User manual

Hach

Hach DR/2010 User manual

Hach sensION+ EC5 User manual

Hach

Hach sensION+ EC5 User manual

Hach H128 User manual

Hach

Hach H128 User manual

Hach POCKET COLORIMETER II User manual

Hach

Hach POCKET COLORIMETER II User manual

Hach TU5400 sc Quick manual

Hach

Hach TU5400 sc Quick manual

Hach TL2300 User manual

Hach

Hach TL2300 User manual

Hach HQ430d Quick manual

Hach

Hach HQ430d Quick manual

Hach Lange ORBISPHERE 3100 Quick manual

Hach

Hach Lange ORBISPHERE 3100 Quick manual

Hach FH950 Quick manual

Hach

Hach FH950 Quick manual

Hach ORBISPHERE 3650 Atex Quick manual

Hach

Hach ORBISPHERE 3650 Atex Quick manual

Hach DR/700 Reference guide

Hach

Hach DR/700 Reference guide

Hach sensION+ EC5 User manual

Hach

Hach sensION+ EC5 User manual

Hach pHD Sensor User manual

Hach

Hach pHD Sensor User manual

Hach ORBISPHERE 6110 User manual

Hach

Hach ORBISPHERE 6110 User manual

Hach IntelliCAL ISENH3181 Series User manual

Hach

Hach IntelliCAL ISENH3181 Series User manual

Hach 59530-16 User manual

Hach

Hach 59530-16 User manual

Hach REFO 3 User manual

Hach

Hach REFO 3 User manual

Hach sension 4 User manual

Hach

Hach sension 4 User manual

Hach 2200 PCX Quick manual

Hach

Hach 2200 PCX Quick manual

Hach DR/2010 User manual

Hach

Hach DR/2010 User manual

Hach sensION+ pH1 User manual

Hach

Hach sensION+ pH1 User manual

Hach DR 1900 Quick manual

Hach

Hach DR 1900 Quick manual

Popular Measuring Instrument manuals by other brands

Elster Instromet RVG instruction manual

Elster Instromet

Elster Instromet RVG instruction manual

Nidec-Shimpo FG-3000 Operation manual

Nidec-Shimpo

Nidec-Shimpo FG-3000 Operation manual

Hioki PW3365-20 quick start guide

Hioki

Hioki PW3365-20 quick start guide

BASETech Porty II operating instructions

BASETech

BASETech Porty II operating instructions

Keysight Technologies 16861A Service guide

Keysight Technologies

Keysight Technologies 16861A Service guide

Mobile Solutions PT-9A user manual

Mobile Solutions

Mobile Solutions PT-9A user manual

Storch CrossLine operating manual

Storch

Storch CrossLine operating manual

Wiltronics ME2100 user manual

Wiltronics

Wiltronics ME2100 user manual

Panasonic EYFR02 operating instructions

Panasonic

Panasonic EYFR02 operating instructions

Xylem FLYGT Start 151 Installation and operation manual

Xylem

Xylem FLYGT Start 151 Installation and operation manual

SMC Networks PFMV3 Series Operation manual

SMC Networks

SMC Networks PFMV3 Series Operation manual

Fuji Electric ZPB Insruction manual

Fuji Electric

Fuji Electric ZPB Insruction manual

Chauvin Arnoux MEMO MD32 quick start

Chauvin Arnoux

Chauvin Arnoux MEMO MD32 quick start

FlowLine EchoPod DL14 Series manual

FlowLine

FlowLine EchoPod DL14 Series manual

Schmidt ZT Series operating instructions

Schmidt

Schmidt ZT Series operating instructions

Hydronix Hydro-Probe installation guide

Hydronix

Hydronix Hydro-Probe installation guide

MICRO-EPSILON thermoMETER CS Assembly instructions

MICRO-EPSILON

MICRO-EPSILON thermoMETER CS Assembly instructions

MGP Instruments DRM-1 OPERATION & CALIBRATION MANUAL

MGP Instruments

MGP Instruments DRM-1 OPERATION & CALIBRATION MANUAL

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.