Quick Guide Component R12 / R6
Cat. No. : CP-R12 / CP-R6
Setting Temperature
Setting temperature
-10 to 40℃ (R12) / -20 to 40℃ (R6)
Setting temp. limit
Lid OpenStopStart
Max. RPM : 4,800 rpm
Max. RCF : 7,470 xg
Max. RPM : 5,000 rpm
Max. RCF : 7,261 xg
Component
R12
Component
R6
Serial No. Installation Date
Display Screen
Status of RPM /
RCF / Time /
Temperature /
Rotor ID / ACC /
DEC / Program
Setting Speed
Press once : Setting RPM
Press twice : Setting RCF
Setting Time
Up to < 100 hr
Continuous run (00:00)
View Settings
Verifying the setting
parameters during
centrifugation
Program Save / Call
Saving a program
(up to 100)
Calling up a program
Setting the Rotor ID manually
Enter the rotor ID. manually, when
rotor id is not identified automatically
Setting ACC / DEC Rate
ACC (0 ~ 9 steps, Max.9)
DEC (0 ~ 9 steps, Natural DEC.0)
Display & Control Panel
Saving / Calling Up
a Program
Setting Run Parameters
(RPM / RCF / Time /
Temperature / ACC / DEC)
Temperature range that
stops operation when the
temperature is lower or
higher than the set
temperature.
Setting Temp. Limit
Press a parameter
to be changed
Enter the desired values
on the keypad
Press the
Enter button
Press the
Enter button
Enter the desired
program number on the keypad
Press the
Save or call button
Press the
Temp. limit button
Press the
Enter button
Enter the desired temperature
range on the keypad
Connect the device to voltage
sources, which correspond to the
electrical requirements on the label
attached to the device.
Check the wall outlet is earth-
grounded, surge-protected and
can supply enough power.
Connect the power cable to the
distribution box, or install the plug
and socket that are packed with
the device.
Switch off the device after use.
Unplug the power plug before
cleaning or unused for a long
period of time.
Check the Line Voltage Grounded Outlet Power Connection Switch Off the Device
Quick Guide Component R12 / R6
Cat. No. : CP-R12 / CP-R6
Allow a safety clearance of
30 cm around the centrifuge
during operation.
Rotors
Spread the grease lightly
on the bucket where it
contacts the rotor and the
pivot.
Remove any dust regularly
from the ventilation slits on
the left side of the device.
There is a drain hose at the
bottom of the device.
The condensation water is
automatically drained through
the hose. Check and remove it
regularly.
Component R6Component R12
30 cm
30 cm
30 cm
Emergency Lid Open
If the centrifuge lid cannot be opened,
insert the provided T-wrench into the
emergency release hole and turn it
clockwise. (Before activating emergency
open, switch off the device and wait for
the rotor to stop.)
Wipe up the moisture in the chamber with a
dry cloth after use. Keep the lid open when
not in use.
Clean the chamber and rotor by wiping
with a cloth soaked in mild detergent and
dry completely.
Keep the Chamber Dry
Cleaning
Check the rotor, buckets and tubes are free of
scratchand crack before operation.
Check the Accessories for Abnormalities
Load tubes symmetrically into the rotor
with identical amount.
Load Tubes Symmetrically
Place the lid on the wind shield rotor before
centrifugation.
Check the Lid
After relocating the device or loading the rotor,
balance them by using a leveler. If necessary,
level the device by adjusting the ‘Leveling foot’
on the bottom of the device.
Balancing the Centrifuge and Rotor
3.0KVA (Component R12)
5.0KVA (Component R6)
220~240V 50/60Hz
Maintenance
Electrical Safety
Cautions for Use
Rev.1
Proudly made by Jiyoung Shim, Hyewon Jeong
Order & Inquiry l +82-2-3472-0727 l [email protected] l www.ihanil.com
Service l +82-2-3452-8966 l [email protected]
Before Operation After Operation
Maintenance of
Swing-Out Rotor
Remove Dust Empty the Condensation
Water
Allow a Safety Clearance
of 30 cm
WS1000-6WS2400-6 6 x 1,000 mL / Max. 5k 6 x 2,400 mL / Max. 4.8k

Other manuals for Component R12

1

This manual suits for next models

3

Other Hanil Control Panel manuals

Hanil Smart 15 Plus User manual

Hanil

Hanil Smart 15 Plus User manual

Hanil Supra R17 User manual

Hanil

Hanil Supra R17 User manual

Hanil Supra R22 User manual

Hanil

Hanil Supra R22 User manual

Hanil Combi 508 User manual

Hanil

Hanil Combi 508 User manual

Popular Control Panel manuals by other brands

SILENT KNIGHT 5700 Installation and operation manual

SILENT KNIGHT

SILENT KNIGHT 5700 Installation and operation manual

Biamp Apprimo Touch 8i Installation and operation guide

Biamp

Biamp Apprimo Touch 8i Installation and operation guide

LEGRAND bticino FL4653 manual

LEGRAND

LEGRAND bticino FL4653 manual

DSC Power Series PC1616 quick guide

DSC

DSC Power Series PC1616 quick guide

Honeywell NOTIFIER NCA-2 instruction manual

Honeywell

Honeywell NOTIFIER NCA-2 instruction manual

FAAC E145 quick guide

FAAC

FAAC E145 quick guide

EDM Solution 862 Operation manual

EDM

EDM Solution 862 Operation manual

Vaillant VWZ AI VWL X/2 A West installation instructions

Vaillant

Vaillant VWZ AI VWL X/2 A West installation instructions

Sensitron PL4+ Quick installation guide

Sensitron

Sensitron PL4+ Quick installation guide

Panasonic AWRP505 - MULTI HYBRID CTL PAN operating instructions

Panasonic

Panasonic AWRP505 - MULTI HYBRID CTL PAN operating instructions

Tyco AutoPulse Z-20 Operator's manual

Tyco

Tyco AutoPulse Z-20 Operator's manual

Chamberlain 880LMW manual

Chamberlain

Chamberlain 880LMW manual

Risco GT 601 Quick user guide

Risco

Risco GT 601 Quick user guide

Gemini Gemini GEM-P1664 datasheet

Gemini

Gemini Gemini GEM-P1664 datasheet

Whale i VAN quick guide

Whale

Whale i VAN quick guide

Velux KLR 200 quick guide

Velux

Velux KLR 200 quick guide

Simplex TrueAlarm 4100 Series manual

Simplex

Simplex TrueAlarm 4100 Series manual

CAME Z Series manual

CAME

CAME Z Series manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.