RecommendationsforUsers
Before using this product, make sure that it is entirely suitable for the environment in
whichitis used. Operationof thisinstrumentinresidential areas could causeunaccept-
ableinterferencestoradioandTVequipment.
Theglassbulbattheendoftheelectrodeissensitivetoelectrostaticdischarges.Avoidtouching
thebulbatalltimes.Duringoperation,ESDwriststrapsshouldbeworntoavoidpossible
damagetotheelectrodebyelectrostaticdischarges.
Anyvariationintroducedbytheusertothesuppliedequipmentmaydegradetheinstrument’s
EMCperformance.
Toavoidelectricalshock,donotusethisinstrumentwhenvoltageatthemeasurementsurface
exceeds24VACor60VDC.Toavoiddamageorburns,donotperformanymeasurementin
microwaveovens.
ACCESSORIESACCESSORIES
HI1286 Double junction, plastic body pH electrode
with 2 m (6.6’) cable and BNC connector
HI70004P pH 4.01solution, 20 mL sachet(25 pcs)
HI70007P pH 7.01solution, 20 mL sachet(25 pcs)
HI70031P 1.41EC (mS)solution,20 mL sachet(25pcs)
HI70442P 1500 ppmsolution, 20 mLsachet (25 pcs)
HI7004L pH 4.01 solution, 460 mL bottle
HI7007L pH 7.01 solution, 460 mL bottle
HI7031L 1.41 EC (mS/cm)solution, 460mL bottle
HI70442L 1500 ppm solution, 460 mL bottle
HI7061L Electrode cleaning solution, 460 mL bottle
HI710005 12 VDC power adapter, US plug
HI710006 12 VDC power adapter, European plug
HI710012 12 VDC power adapter, Australian plug
HI710013 12 VDC power adapter, Southern Africa plug
HI710014 12 VDC power adapter, UK plug
ISCOMBOR2 10/99PRINTED IN PORTUGAL
OPERATIONAL GUIDEOPERATIONAL GUIDE
pH ELECTRODE CONNECTION
•Slide the protective sheath down. Connect the pH elec-
trode to the BNC connector and then slide the protective
sheath back up all the way up to the casing. For
maximum waterproof protection, make sure the connector
iscompletelycovered.
•Do not be alarmed if white crystals appear around the
electrode protective cap. This is normal with pH electrodes
and they dissolve when rinsed with water.
In order to protect the instrument against vapors and humid-
ity,the BNC connector is shielded behind awaterproof sheath.
TAKING MEASUREMENTS
•Turn the meter on by connecting the 12 VDC power
adapter to the meter and the mains.
•Remove the protective cap from the pH electrode.
•Immerse the tips (4cm/1½”) of pH electrode, EC/TDS
probe and ground probe
inside the sample. In or-
derto ensureaccuracy of
measurements, probes
shouldnottouch orstand
close to the sample
vessel’swalls orbottom.
•The2 LCD’s willshowthe
pH and EC/TDS value .
Any initial measuring
variation maybe due to
pH electrode condition-
ing and EC/TDS
temperature compensation. Allow the readingsto stabilize.
ADJUSTING THE SETPOINT
With HI 981404 and HI 981405 you can select your own
setpoint and be alerted with the pH or EC/TDS LED alarm
when an abnormal situation arises.
•
Unscrew and remove the rear panel and gasket seal to access
the MEASURE/SET switch and move it to the left (SET).
•With a small screwdriver adjust the setpoint trimmer to
display the desired value on the LCD. You can choose to
adjust one or both setpoints at once inside the following
ranges:
pH:from3.0 to7.0
EC:from1.00to 3.50 mS/cm
TDS:from500to1600 ppm
pHLED blinkswhen measurement exceedsthesetpoint.
•Move the switch to MEASURE mode and secure back the
rear panel properly replacing the gasket.
pH ELECTRODE MAINTENANCE
•Whennot in use,rinse the electrodewith water tominimize
contamination and store it with a few drops of storage
(HI 70300) or pH 7 (HI 7007) solution in the protective
cap. Always replace the protective cap when not in use.
DO NOT USE DISTILLED OR DEIONIZED WATER FOR STOR-
AGE PURPOSES.
•If the electrode has been left dry, soak the tip in a storage
(HI 70300) or pH 7 (HI 7007) solution overnight to
reactivate it.
•To minimize clogging and provide longer life for the pH
electrode, it is recommended to clean it at least once a
month. Immerse thetip of theelectrode in HI7061forhalf
an hour and then rinse it with tap water.
EC/TDS PROBE MAINTENANCE
Thefollowing cleaningprocedure is recommendedat leastonce
a month:
•Immerse the tip of the electrode in HI 7061 Cleaning
Solution for half an hour.
•If a more thorough cleaning is required, brush the metal
pins with very fine sandpaper.
•After cleaning, rinse the probe with tap water.
The instrument must be recalibrated whenever:
a) After cleaning or replacing the probe.
b) Where high accuracy is required.
c) At least once a month
pH CALIBRATION
Pour small quantities of pH 7.0 (HI 7007) and pH 4.0
(HI 7004) solution into two clean beakers.
CALIBRATIONCALIBRATION
Foraccurate calibrationuse twobeakers for each buffer solution,
the first one for rinsing the tip of the electrode and the second
one for calibration.
•Turn the meter on and make sure that the MEASURE/SET
switch is on the MEASURE mode.
•Remove the protective cap from the electrode, rinse and
immerseit in a pH 7.0 buffer. Stir gently and thenwait for
the reading to stabilize.
Note: the electrode should be submerged approximately 4
cm (1½”) in the solution.
•Adjust the right hand trimmer (pH 7pH 7) with the calibra-
tion screwdriver until the LCD shows pH 7.0.
•Rinse and immerse the pH electrode in pH 4.0 buffer and
stir gently.
•Wait a couple of minutes and then adjust the left hand
trimmer(pH 4pH 4)until theLCD showsthe valueof thesecond
buffer.
The pH calibration is now complete.
EC/TDSCALIBRATION
•Turnthe meter on and make sure thatthe MEASURE/SET
switch is on the MEASURE mode.
•Pour a small quantity of calibration solution* in a beaker.
If possible, use plastic beakers to minimize any EMC
interference.
*1.41 mS/cm for HI981405;1500ppm for HI 981404.
•Immerse the EC/TDS probe in the solution, making sure
that the metal pins are completely submerged.
Note: in order not to affect the accuracy of measurements, it
is important that probe body does not touch nor stand
close to the side walls of the beaker.
•Waitfor a couple of minutes for thermal equilibrium tobe
reached and the temperature to be compensated.
•Tap the probe gently on the bottom, then shake it while
rotating to make sure no air bubbles have remained
trapped.
•Adjustthe calibration trimmer with a screwdriver untilthe display
shows1.41mS/cm(HI981405)or1500ppm(HI981404).
•Calibrationis now complete.
HI981404 HI981405