Hanna Instruments HI 93705 User manual

Instruction Manual
HI 93705
Silica ISM
www.hannainst.com This Instrument is in
Compliance with the CE Directives
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
WARRANTY
HI 93705 is warranted for two years against defects in workman-
ship and materials when used for its intended purpose and maintained
according to instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of pur-
chase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service de-
partment and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete protec-
tion.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instru-
ments Inc., Woonsocket, Rhode Island, 02895 , USA.
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at [email protected].
This instrument is in compliance with directives
EN 50081-1 and EN 50082-1.
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each Ion Specific Meter is supplied complete with
• 9V Battery
• Two Sample Cuvets and Caps
• One Transport Cap
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
PRELIMINARY EXAMINATION
GENERAL DESCRIPTION
The HI 93705 meter measures the silica (SiO2) content in
water and wastewater in the 0.00 to 2.00 mg/L (ppm)
range.
The meter uses an exclusive positive-locking system to ensure
that the cuvet is in the same place every time it is placed
into the measurement cell.
The reagents are in liquid and powder form and are
supplied in bottles and in packets. The amount of reagent
is precisely dosed to ensure maximum repeatability.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
SPECIFICATIONS
SPECIFICATIONS
Range 0.00 to 2.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.03 mg/L ±3% of reading
Typical EMC ±0.01 mg/L
Deviation
Light Source Light Emitting Diode @ 890 nm
Method Adaptation of the ASTM D859 method of
the heteropoly blue method. The reaction
between silica and reagents causes a blue
tint in the sample
Light Detector Silicon Photocell
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type/Life 1 x 9 volt/40 hours
Auto-Shut off After 10' of non-use
Dimensions 180 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight 290 g (10 oz.).
REQUIRED REAGENTS
Code Description Quantity
HI 93705A-0 Molybdate 6 drops
HI 93705B-0 Citric acid 1 packet
HI 93705C-0 Amino acid 1 packet
REAGENT SETS
HI 93705-01 Reagents for 100 tests
HI 93705-03 Reagents for 300 tests
This indicates that the meter is in a
ready state and zeroing can be per-
formed.
Sampling in Progress. This prompt ap-
pears each time the meter is performing
a measurement.
This indicates that the meter is in a
zeroed state and measurement can be
performed.
A zero reading was not taken. Insert a
sample before adding reagent and press
ZERO.
Under range. A blinking "0.00" indicates
that the sample absorbs less light than
the zero reference. Check the procedure
and make sure you use the same cuvet
for reference (zero) and measurement.
Over range. A flashing value higher than
the maximum readable concentration (see
specifications) indicates that the sample
absorbs too much light, meaning that
the concentration is too high. Dilute the
sample.
Light over range. The cuvet is not in-
serted correctly and an excess ambient
light is reaching the detector. If the cover
is properly installed, then contact your
dealer or the nearest Hanna Customer
Service Center.
Light under range. The zero sample is
too dark for proper zeroing. If this is not
the case, contact your dealer or the
nearest Hanna Customer Service Center.
The "V" indicates that the battery volt-
age is getting low and the battery needs
to be replaced.
This indicates that the battery is dead
and must be replaced.
Note: once this indication is displayed,
the meter will lockup. Change the bat-
tery to restart.
DISPLAY CODE GUIDE
V
OPERATIONAL GUIDE
MEASUREMENT PROCEDURE
• Turn the meter on by pressing ON/OFF.
• When the LCD displays “- - -”, it is
ready.
• Fill the cuvet with 10 mL of
unreacted sample up to the
mark.
• Add 6 drops of HI 93705A
molybdate reagent. Re-
place the cap and swirl
the solution.
• Wait for 4 minutes, add
the content of one packet
of HI 93705B citric acid
reagent and shake until it
is completely dissolved.
• Wait for 1 minute. This is the blank.
• Place the cuvet into the holder
and ensure that the notch on
the cap is positioned securely
into the groove.
• Press ZERO and “SIP” will appear on the display.
• Wait for a few seconds and the display
will show “-0.0-”. Now the meter is
zeroed and ready for measurement.
ON
OFF
12
3
6
9
1’
ZERO
READ
TIMED
12
3
6
9
1’
READ
DIRECT
• Remove the cuvet and add
the content of one packet of
HI 93705C amino acid re-
agent and shake until it has
dissolved.
• Reinsert the cuvet into the in-
strument.
• Press READ TIMED and the display will show the count-
down prior to the measurement. Alternatively, wait for
exactly 1 minute and press READ DIRECT. In both cases
“SIP” will appear during measurement.
• The instrument directly displays concentration in mg/L of
silica (SiO2) on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by:
Phosphate above 60 mg/L (causes a 2% reduction in
reading)
Phosphate above 75 mg/L (causes an 11% reduction in
reading)
Sulfide and high concentration of iron
Eliminate color and turbidity interferences by zeroing the
meter with the original water sample.
The instruction listed below should be carefully followed
during testing to ensure best accuracy.
• Do not touch the cuvet walls with hands.
• In order to maintain the same conditions during the
zeroing and the measuring phases, it is necessary to
close the cuvet to prevent any contamination.
• Do not let the test sample stand too long after reagent
is added or accuracy will be lost.
• Whenever the cuvet is placed into the measurement cell,
it must be completely free of fingerprints, oil or dirt.
Wipe it thoroughly with HI 731318 or a lint-free cloth
prior to insertion.
• It is important that the sample does not contain any
debris. This would corrupt the readings.
• It is possible to take multiple readings in a row, but it
is recommended that a zero reading be taken for each
sample and that the same cuvet is used for zeroing and
measurement.
• It is important to discard the sample immediately after
the reading is taken because the glass might become
permanently stained.
• Shaking the cuvet can generate bubbles in the sample,
causing higher readings. To obtain accurate measure-
ments, remove such bubbles by swirling or by gently
tapping the vial.
• All the reaction times reported in this manual are
referred to 20°C (68°F). As a general rule of thumb,
they should be doubled at 10°C (50°F) and halved at
30°C (86°F).
TIPS FOR AN ACCURATE
MEASUREMENT
BATTERY REPLACEMENT
Battery replacement must only take place in a non-hazard-
ous area using a 9V alkaline battery.
Simply slide off the battery cover on the back of the meter.
Detach the battery from the terminals and attach a fresh
9V battery while paying attention to the correct polarity.
Replace the battery and the cover.
9V
SLID
E
OFF
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environ-
ment in which they are used.
Operation of these instruments in residential area could cause unacceptable interferences
to radio and TV equipments, requiring the operator to take all necessary steps to correct
interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments' EMC performance.
To avoid damages or burns, do not perform any measurement in microwave ovens.
ACCESSORIES
REAGENT SETS
HI 93705-01 Reagents for 100 tests
HI 93705-03 Reagents for 300 tests
OTHER ACCESSORIES
HI 710009 Blue rubber boot
HI 710010 Orange rubber boot
HI 721310 9V battery (10 pcs)
HI 731318 Tissue for wiping cuvets (4 pcs)
HI 731321 Glass cuvets (4 pcs)
HI 731325 Caps for cuvets (4 pcs)
HI 93703-50 Cuvets cleaning solution (230 mL).
CE DECLARATION
OF CONFORMITY
IST93705R2 10/99
PRINTED IN PORTUGAL
10 mL
x 6

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI 720 User manual

Hanna Instruments

Hanna Instruments HI 720 User manual

Hanna Instruments Checker HI 723 User manual

Hanna Instruments

Hanna Instruments Checker HI 723 User manual

Hanna Instruments edge pH User manual

Hanna Instruments

Hanna Instruments edge pH User manual

Hanna Instruments HI 93114 User manual

Hanna Instruments

Hanna Instruments HI 93114 User manual

Hanna Instruments HI 97500 User manual

Hanna Instruments

Hanna Instruments HI 97500 User manual

Hanna Instruments IC-HI98165 User manual

Hanna Instruments

Hanna Instruments IC-HI98165 User manual

Hanna Instruments HI 847492 User manual

Hanna Instruments

Hanna Instruments HI 847492 User manual

Hanna Instruments HI 96107 User manual

Hanna Instruments

Hanna Instruments HI 96107 User manual

Hanna Instruments HI97101 User manual

Hanna Instruments

Hanna Instruments HI97101 User manual

Hanna Instruments HI96738 User manual

Hanna Instruments

Hanna Instruments HI96738 User manual

Hanna Instruments HI 96831 User manual

Hanna Instruments

Hanna Instruments HI 96831 User manual

Hanna Instruments HI 991003 User manual

Hanna Instruments

Hanna Instruments HI 991003 User manual

Hanna Instruments HI 96108 User manual

Hanna Instruments

Hanna Instruments HI 96108 User manual

Hanna Instruments HI7630-28 User manual

Hanna Instruments

Hanna Instruments HI7630-28 User manual

Hanna Instruments HI96841 User manual

Hanna Instruments

Hanna Instruments HI96841 User manual

Hanna Instruments HI 9812-0 User manual

Hanna Instruments

Hanna Instruments HI 9812-0 User manual

Hanna Instruments HI727 User manual

Hanna Instruments

Hanna Instruments HI727 User manual

Hanna Instruments HI 9124 User manual

Hanna Instruments

Hanna Instruments HI 9124 User manual

Hanna Instruments HI96716C User manual

Hanna Instruments

Hanna Instruments HI96716C User manual

Hanna Instruments HI97752 User manual

Hanna Instruments

Hanna Instruments HI97752 User manual

Hanna Instruments HI775 User manual

Hanna Instruments

Hanna Instruments HI775 User manual

Hanna Instruments HALO 2 HI9810362 User manual

Hanna Instruments

Hanna Instruments HALO 2 HI9810362 User manual

Hanna Instruments HI 84429 User manual

Hanna Instruments

Hanna Instruments HI 84429 User manual

Hanna Instruments HI 93720 User manual

Hanna Instruments

Hanna Instruments HI 93720 User manual

Popular Measuring Instrument manuals by other brands

Waters ionKey/MS Guide

Waters

Waters ionKey/MS Guide

Powerex MH-C9000 instructions

Powerex

Powerex MH-C9000 instructions

Master Meter eLinx quick start guide

Master Meter

Master Meter eLinx quick start guide

R&S RT-ZC20 user manual

R&S

R&S RT-ZC20 user manual

BEKA A504G-SS-PM manual

BEKA

BEKA A504G-SS-PM manual

Emerson Daniel PT Series Installation and maintenance instructions

Emerson

Emerson Daniel PT Series Installation and maintenance instructions

Denver Instrument IR-200 Operation manual

Denver Instrument

Denver Instrument IR-200 Operation manual

Omega APH2 instruction sheet

Omega

Omega APH2 instruction sheet

Kaelus iVA0727A operating manual

Kaelus

Kaelus iVA0727A operating manual

ATAGO PAL-26S instruction manual

ATAGO

ATAGO PAL-26S instruction manual

ICP DAS USA iWSN-110X-RCT Series quick start

ICP DAS USA

ICP DAS USA iWSN-110X-RCT Series quick start

Rohde & Schwarz AMN6500 user manual

Rohde & Schwarz

Rohde & Schwarz AMN6500 user manual

Agilent Technologies E4980A Operation manual

Agilent Technologies

Agilent Technologies E4980A Operation manual

Dwyer Instruments FLMG Series instructions

Dwyer Instruments

Dwyer Instruments FLMG Series instructions

Spectris PARTICLE MEASURING SYSTEMS AirSentry II Operation manual

Spectris

Spectris PARTICLE MEASURING SYSTEMS AirSentry II Operation manual

National Instruments NI 5911 user manual

National Instruments

National Instruments NI 5911 user manual

Apator Powogaz JS operating instructions

Apator Powogaz

Apator Powogaz JS operating instructions

FLOWMAX 221 quick start guide

FLOWMAX

FLOWMAX 221 quick start guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.