Hanna Instruments HI 93718 User manual

Instruction Manual
HI 93718
Iodine ISM
http://www.hannainst.com ThisInstrumentisin
Compliancewith the CE Directives
HannaInstrumentsreservestherighttomodifythedesign,construction
andappearanceofitsproductswithoutadvancenotice.
WARRANTYWARRANTY
WARRANTYWARRANTY
WARRANTY
HI93718iswarrantedfor two years against defects in workman-
shipand materialswhenusedforits intendedpurposeandmaintained
accordingtoinstructions.
Thiswarranty islimitedto repairorreplacement freeofcharge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenanceare not covered.
Ifserviceis required,contactthedealerfromwhom youpurchasedthe
instrument.Ifunderwarranty, report the model number, dateofpur-
chase,serialnumberandthe nature of the failure. Iftherepairisnot
coveredby the warranty,youwillbe notified ofthechargesincurred. If
theinstrumentistobereturnedtoHannaInstruments,firstobtaina
ReturnedGoodsAuthorizationNumber from the Customer Service de-
partmentand thensenditwith shipmentcostsprepaid.Whenshipping
anyinstrument, makesureitisproperly packagedforcompleteprotec-
tion.
Tovalidate yourwarranty,filloutandreturntheenclosedwarrantycard
within14 daysfromthedateofpurchase.
Allrightsare reserved. Reproduction inwholeor in part isprohibited
without the written consent of the copyright owner, Hanna Instru-
mentsInc.,Woonsocket, Rhode Island, 02895,USA.
DearCustomer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at [email protected].
This instrument is in compliance with directives
EN 50081-1 and EN 50082-1.
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each Ion Specific Meter is supplied complete with
• 9V Battery
• Two Sample Cuvets and Caps
• One Transport Cap
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
The HI 93718 meter measures the iodine content in water
and wastewater in the 0.0 to 12.5 mg/L (ppm) range.
Themeter usesan exclusivepositive-locking systemto ensure
that the cuvet is in the same place every time it is placed
into the measurement cell.
The reagents are in powder form and are supplied in
packets. The amount of reagent is precisely dosed to ensure
maximum repeatability.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Range 0.0 to 12.5 mg/L
Resolution 0.1 mg/L
Accuracy ±0.1 mg/L ±5% of reading
TypicalEMC ±0.1 mg/L
Deviation
Light Source Light Emitting Diode @ 555 nm
Method Adaptation of the Standard Methods for
the Examination of Water and Wastewa-
ter,18th edition,DPDmethod. Thereaction
between iodine and the reagent causes a
pink tint in the sample.
Light Detector SiliconPhotocell
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type/Life 1 x 9 volt/40 hours
Auto-Shut off After 10' of non-use
Dimensions 180 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight 290 g (10 oz.).
REQUIRED REAGENTS
Code Description Quantity
HI 93718-0 DPD Reagent 1 packet
REAGENT SETS
HI93718-01 Reagents for 100 tests
HI93718-03 Reagents for 300 tests
ins
e
rum ns
ON
OFF ZERO
READ
TIMED
READ
DIRECT
HI 93718
IODINE
Range: 0.0..12.5 ppm
READ DIRECT
READ TIMED
ZERO
ON/OFF
CUVET
ALIGNMENT INDICATOR
CUVET HOLDER
This indicates that the meter is in a
ready state and zeroing can be per-
formed.
Sampling in Progress. This prompt ap-
pears each time the meter is performing
a measurement.
This indicates that the meter is in a
zeroed state and measurement can be
performed.
A zero reading was not taken. Insert a
sample before adding reagent and press
ZERO.
Under range. A blinking "0.00" indicates
that the sample absorbs less light than
the zero reference. Check the procedure
and make sure you use the same cuvet
for reference (zero) and measurement.
Over range. A flashing value higher than
themaximumconcentrationreadable(see
specifications) indicates that the sample
absorbs too much light, meaning that
the concentration is too high. Dilute the
sample.
Light over range. The cuvet is not in-
serted correctly and an excess ambient
light is reaching the detector. If the cover
is properly installed, then contact your
dealer or the nearest Hanna Customer
Service Center.
Light under range. The zero sample is
too dark for proper zeroing. If this is not
the case, contact your dealer or the
nearest Hanna Customer Service Center.
The "V" indicates that the battery volt-
age is getting low and the battery needs
to be replaced.
This indicates that the battery is dead
and must be replaced.
Note: once this indication is displayed,
the meter will lockup. Change the bat-
tery to restart.
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDE
V
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
MEASUREMENT PROCEDURE
• Turn the meter on by pressing ON/OFF.
• When the LCD displays "- - -", it is
ready.
• Fill the cuvet up to 1.5 cm (¾") below the
rim with 10 mL of unreacted sample and
replace the cap.
• Placethe cuvetinto theholder
and ensure that the notch on
the cap is positioned securely
into the groove.
• Press ZERO and "SIP" will appear on the display.
• Wait for a few seconds and the dis-
playwillshow "-0.0-".Nowthe meter
is zeroed and ready for measure-
ment.
• Remove the cap and add
the content of one packet of
HI 93718 reagent. Replace
the cap and shake gently.
• Reinsert the cuvet into the in-
strument.
1.5 cm
ON
OFF
ZERO
12
3
6
9
2’30"
READ
DIRECT
READ
TIMED
• Press READ TIMED and the display will show the count-
down prior to the measurement, or, alternatively wait
for 2 minutes and 30 seconds and press READ DIRECT.
In both cases "SIP" will appear during measurement.
• Theinstrument directly displays concentration in mg/L of
iodine on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by:
Bromine
Chlorine
Ozone
Oxidized forms of chromium and manganese
Alkalinity above 300 mg/L and acidity above 150 mg/L as
CaCO3require neutralization since the color could instantly
fade. To resolve this, neutralize the sample with diluted HCl
or NaOH.
In case of water with hardness greater than 500 mg/L
CaCO3, shake the sample for approximately 1 minute after
adding the reagent.
The instruction listed below should be carefully followed
during testing to ensure best accuracy.
• Do not touch the cuvet walls with hands.
• In order to maintain the same conditions during the
zeroing and the measuring phases, it is necessary to
close the cuvet to prevent any contamination.
• Do not let the test sample stand too long after reagent
is added or accuracy will be lost.
• Whenever the cuvet is placed into the measurement cell,
it must be completely free of fingerprints, oil or dirt.
Wipe it thoroughly with HI 731318 or a lint-free cloth
prior to insertion.
• It is important that the sample does not contain any
debris. This would corrupt the readings.
• It is possible to take multiple readings in a row, but it
is recommended that a zero reading be taken for each
sample and that the same cuvet is used for zeroing and
measurement.
• It is important to discard the sample immediately after
the reading is taken because the glass might become
permanently stained.
• Shaking the cuvet can generate bubbles in the sample,
causing higher readings. To obtain accurate measure-
ments, remove such bubbles by swirling or by gently
tapping the vial.
• All the reaction times reported in this manual are
referred to 20°C (68°F). As a general rule of thumb,
they should be doubled at 10°C (50°F) and halved at
30°C(86°F).
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATE
MEASUREMENTMEASUREMENT
MEASUREMENTMEASUREMENT
MEASUREMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
Battery replacement must only take place in a non-hazard-
ous area using a 9V alkaline battery.
Simply slide off the battery cover on the back of the meter.
Detach the battery from the terminals and attach a fresh
9V battery while paying attention to the correct polarity.
Replace the battery and the cover.
9V
SLID
E
OFF
RecommendationsforUsers
Before using these products, make sure that they are entirely suitable for the environ-
mentinwhichtheyareused.
Operationoftheseinstrumentsinresidentialareacouldcauseunacceptableinterferences
toradioandTV equipments, requiring the operatortotakeallnecessary steps to correct
interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments'EMCperformance.
Toavoiddamagesorburns,donotperformanymeasurementinmicrowaveovens.
ACCESSORIESACCESSORIES
ACCESSORIESACCESSORIES
ACCESSORIES
REAGENT SETS
HI93718-01 Reagents for 100 tests
HI93718-03 Reagents for 300 tests
OTHER ACCESSORIES
HI710009 Blue rubber boot
HI710010 Orange rubber boot
HI721310 9V battery (10 pcs)
HI731318 Tissue for wiping cuvets (4 pcs)
HI731321 Glass cuvets (4 pcs)
HI731325 Caps for cuvets (4 pcs)
HI93703-50 Cuvets cleaning solution (230 mL).
CE DECLARATIONCE DECLARATION
CE DECLARATIONCE DECLARATION
CE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
IST93718R1 06/98
PRINTED IN PORTUGAL

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI97707 User manual

Hanna Instruments

Hanna Instruments HI97707 User manual

Hanna Instruments HI 991001 User manual

Hanna Instruments

Hanna Instruments HI 991001 User manual

Hanna Instruments HI 758 User manual

Hanna Instruments

Hanna Instruments HI 758 User manual

Hanna Instruments HI 98171 User manual

Hanna Instruments

Hanna Instruments HI 98171 User manual

Hanna Instruments HI 141 Series User manual

Hanna Instruments

Hanna Instruments HI 141 Series User manual

Hanna Instruments HI749 User manual

Hanna Instruments

Hanna Instruments HI749 User manual

Hanna Instruments OHI 98140 User manual

Hanna Instruments

Hanna Instruments OHI 98140 User manual

Hanna Instruments HI705 User manual

Hanna Instruments

Hanna Instruments HI705 User manual

Hanna Instruments HI 991401 User manual

Hanna Instruments

Hanna Instruments HI 991401 User manual

Hanna Instruments HI98198 OPDO User manual

Hanna Instruments

Hanna Instruments HI98198 OPDO User manual

Hanna Instruments HI84531 User manual

Hanna Instruments

Hanna Instruments HI84531 User manual

Hanna Instruments HI 96801 User manual

Hanna Instruments

Hanna Instruments HI 96801 User manual

Hanna Instruments HI 96761C User manual

Hanna Instruments

Hanna Instruments HI 96761C User manual

Hanna Instruments HALO 2 HI9810352 User manual

Hanna Instruments

Hanna Instruments HALO 2 HI9810352 User manual

Hanna Instruments HI759 User manual

Hanna Instruments

Hanna Instruments HI759 User manual

Hanna Instruments PCA 310 User manual

Hanna Instruments

Hanna Instruments PCA 310 User manual

Hanna Instruments HI96000 Series User manual

Hanna Instruments

Hanna Instruments HI96000 Series User manual

Hanna Instruments HI 93738 User manual

Hanna Instruments

Hanna Instruments HI 93738 User manual

Hanna Instruments HI96785 User manual

Hanna Instruments

Hanna Instruments HI96785 User manual

Hanna Instruments HI97723 User manual

Hanna Instruments

Hanna Instruments HI97723 User manual

Hanna Instruments HI97701 User manual

Hanna Instruments

Hanna Instruments HI97701 User manual

Hanna Instruments HI96700C User manual

Hanna Instruments

Hanna Instruments HI96700C User manual

Hanna Instruments pHep+ HI 98108 User manual

Hanna Instruments

Hanna Instruments pHep+ HI 98108 User manual

Hanna Instruments HI97728 User manual

Hanna Instruments

Hanna Instruments HI97728 User manual

Popular Measuring Instrument manuals by other brands

wtw Cond 3110 quick start guide

wtw

wtw Cond 3110 quick start guide

Benning MM 5-2 operating manual

Benning

Benning MM 5-2 operating manual

BIRD 7027 Series Operation manual

BIRD

BIRD 7027 Series Operation manual

Colibri SMARTHUB user manual

Colibri

Colibri SMARTHUB user manual

Ametek WDG-V user manual

Ametek

Ametek WDG-V user manual

Citec LOOP420 instruction manual

Citec

Citec LOOP420 instruction manual

Mezon DSTM500 user manual

Mezon

Mezon DSTM500 user manual

Flexim FLUXUS F704**-NN Series user manual

Flexim

Flexim FLUXUS F704**-NN Series user manual

pH ELECTRODE PE-03 Operation manual

pH ELECTRODE

pH ELECTRODE PE-03 Operation manual

Honeywell DVC100 (M) MK2 installation guide

Honeywell

Honeywell DVC100 (M) MK2 installation guide

Milwaukee MW101 user manual

Milwaukee

Milwaukee MW101 user manual

Decagon Devices KD2 Pro Operator's manual

Decagon Devices

Decagon Devices KD2 Pro Operator's manual

Doepke DCTR B-X Installation and operating instructions

Doepke

Doepke DCTR B-X Installation and operating instructions

Smart Storm Universal Smart Meter instruction manual

Smart Storm

Smart Storm Universal Smart Meter instruction manual

M-system AS4M instruction manual

M-system

M-system AS4M instruction manual

Rixen SR-496 instruction manual

Rixen

Rixen SR-496 instruction manual

Vetek MO.DTR.526.R6 operating manual

Vetek

Vetek MO.DTR.526.R6 operating manual

ThermoWorks ThermaQ 2 operating instructions

ThermoWorks

ThermoWorks ThermaQ 2 operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.