  9. Hanna Instruments HI 93726 User manual

Hanna Instruments HI 93726 User manual

Instruction Manual
HI 93726
Nickel ISM
http://www.hannainst.com ThisInstrumentisin
Compliancewith the CE Directives
HannaInstrumentsreservestherighttomodifythedesign,construction
andappearanceofitsproductswithoutadvancenotice.
WARRANTYWARRANTY
WARRANTYWARRANTY
WARRANTY
HI93726iswarrantedfor two years against defects in workman-
shipand materialswhenusedforits intendedpurposeandmaintained
accordingtoinstructions.
Thiswarranty islimitedto repairorreplacement freeofcharge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenanceare not covered.
Ifserviceis required,contactthedealerfromwhom youpurchasedthe
instrument.Ifunderwarranty, report the model number, dateofpur-
chase,serialnumberandthe nature of the failure. Iftherepairisnot
coveredby the warranty,youwillbe notified ofthechargesincurred. If
theinstrumentistobereturnedtoHannaInstruments,firstobtaina
ReturnedGoodsAuthorizationNumber from the Customer Service de-
partmentand thensenditwith shipmentcostsprepaid.Whenshipping
anyinstrument, makesureitisproperly packagedforcompleteprotec-
tion.
Tovalidate yourwarranty,filloutandreturntheenclosedwarrantycard
within14 daysfromthedateofpurchase.
Allrightsare reserved. Reproduction inwholeor in part isprohibited
without the written consent of the copyright owner, Hanna Instru-
mentsInc.,Woonsocket, Rhode Island, 02895,USA.
DearCustomer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at [email protected].
This instrument is in compliance with directives
EN 50081-1 and EN 50082-1.
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each Ion Specific Meter is supplied complete with
• 9V Battery
• Two Sample Cuvets and Caps
• One Transport Cap
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
The HI 93726 meter measures the nickel content in water
and wastewater in the 0.00 to 7.00 g/L (ppt) range.
Themeter usesan exclusivepositive-locking systemto ensure
that the cuvet is in the same place every time it is placed
into the measurement cell.
The reagents are in powder form and are supplied in
packets. The amount of reagent is precisely dosed to ensure
maximum repeatability.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Range 0.00 to 7.00 g/L
Resolution 0.01 g/L
Accuracy ±4% of reading
TypicalEMC ±0.02 g/L
Deviation
Light Source Light Emitting Diode @ 555 nm
Method Adaptation of the photometric method.
The reaction between nickel and the re-
agent causes a blue tint in the sample
Light Detector SiliconPhotocell
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type/Life1 x 9 volt/40 hours
Auto-Shut off After 10' of non-use
Dimensions 180 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight 290 g (10 oz.).
REQUIRED REAGENTS
Code Description Quantity
HI 93726-0 Powder reagent 1 packet
REAGENT SETS
HI93726-01 Reagents for 100 tests
HI93726-03 Reagents for 300 tests
This indicates that the meter is in a
ready state and zeroing can be per-
formed.
Sampling in Progress. This prompt ap-
pears each time the meter is performing
a measurement.
This indicates that the meter is in a
zeroed state and measurement can be
performed.
A zero reading was not taken. Insert a
sample before adding reagent and press
ZERO.
Under range. A blinking "0.00" indicates
that the sample absorbs less light than
the zero reference. Check the procedure
and make sure you use the same cuvet
for reference (zero) and measurement.
Over range. A flashing value higher than
themaximumconcentrationreadable(see
specifications) indicates that the sample
absorbs too much light, meaning that
the concentration is too high. Dilute the
sample.
Light over range. The cuvet is not in-
serted correctly and an excess ambient
light is reaching the detector. If the cover
is properly installed, then contact your
dealer or the nearest Hanna Customer
Service Center.
Light under range. The zero sample is
too dark for proper zeroing. If this is not
the case, contact your dealer or the
nearest Hanna Customer Service Center.
The "V" indicates that the battery volt-
age is getting low and the battery needs
to be replaced.
This indicates that the battery is dead
and must be replaced.
Note: once this indication is displayed,
the meter will lockup. Change the bat-
tery to restart.
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDE
V
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
MEASUREMENT PROCEDURE
• Turn the meter on by pressing ON/OFF.
• When the LCD displays "- - -", it is
ready.
• Fill the cuvet up to 1.5 cm (¾")
belowtherim with10 mLof sample.
• Placethe cuvetinto theholder
and ensure that the notch on
the cap is positioned securely
into the groove.
• Press ZERO and "SIP" will appear on the display.
• Wait for a few seconds and the dis-
playwillshow "-0.0-".Nowthe meter
is zeroed and ready for measurement.
• Remove the cuvet and add
the content of one packet of
HI 93726 reagent. Replace
the cap and shake gently.
• Reinsert the cuvet into the in-
strument.
1.5 cm
ZERO
ON
OFF
• Press READ TIMED and the display will show the count-
down prior to the measurement, or, alternatively wait
for 1 minute and press READ DIRECT. In both cases
"SIP" will appear during measurement.
• The instrument directly displays concentration in g/L of
nickel on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by copper.
12
3
6
9
1’
READ
DIRECT
READ
TIMED
The instruction listed below should be carefully followed
during testing to ensure best accuracy.
• Do not touch the cuvet walls with hands.
• In order to maintain the same conditions during the
zeroing and the measuring phases, it is necessary to
close the cuvet to prevent any contamination.
• Do not let the test sample stand too long after reagent
is added or accuracy will be lost.
• Whenever the cuvet is placed into the measurement cell,
it must be completely free of fingerprints, oil or dirt.
Wipe it thoroughly with HI 731318 or a lint-free cloth
prior to insertion.
• It is important that the sample does not contain any
debris. This would corrupt the readings.
• It is possible to take multiple readings in a row, but it
is recommended that a zero reading be taken for each
sample and that the same cuvet is used for zeroing and
measurement.
• It is important to discard the sample immediately after
the reading is taken because the glass might become
permanently stained.
• Shaking the cuvet can generate bubbles in the sample,
causing higher readings. To obtain accurate measure-
ments, remove such bubbles by swirling or by gently
tapping the vial.
• All the reaction times reported in this manual are
referred to 20°C (68°F). As a general rule of thumb,
they should be doubled at 10°C (50°F) and halved at
30°C(86°F).
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATE
MEASUREMENTMEASUREMENT
MEASUREMENTMEASUREMENT
MEASUREMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
Battery replacement must only take place in a non-hazard-
ous area using a 9V alkaline battery.
Simply slide off the battery cover on the back of the meter.
Detach the battery from the terminals and attach a fresh
9V battery while paying attention to the correct polarity.
Replace the battery and the cover.
9V
SLID
E
OFF
RecommendationsforUsers
Before using these products, make sure that they are entirely suitable for the environ-
mentinwhichtheyareused.
Operationoftheseinstrumentsinresidentialareacouldcauseunacceptableinterferences
toradioandTV equipments, requiring the operatortotakeallnecessary steps to correct
interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments'EMCperformance.
Toavoiddamagesorburns,donotperformanymeasurementinmicrowaveovens.
ACCESSORIESACCESSORIES
ACCESSORIESACCESSORIES
ACCESSORIES
REAGENT SETS
HI93726-01 Reagents for 100 tests
HI93726-03 Reagents for 300 tests
OTHER ACCESSORIES
HI710009 Blue rubber boot
HI710010 Orange rubber boot
HI721310 9V battery (10 pcs)
HI731318 Tissue for wiping cuvets (4 pcs)
HI731321 Glass cuvets (4 pcs)
HI731325 Caps for cuvets (4 pcs)
HI93703-50 Cuvets cleaning solution (230 mL).
CE DECLARATIONCE DECLARATION
CE DECLARATIONCE DECLARATION
CE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
IST93726R1 06/98
PRINTED IN PORTUGAL

This manual suits for next models

1

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI96725C User manual

Hanna Instruments

Hanna Instruments HI96725C User manual

Hanna Instruments HI 717 User manual

Hanna Instruments

Hanna Instruments HI 717 User manual

Hanna Instruments HI 99181 User manual

Hanna Instruments

Hanna Instruments HI 99181 User manual

Hanna Instruments HI99162 User manual

Hanna Instruments

Hanna Instruments HI99162 User manual

Hanna Instruments HI749 User manual

Hanna Instruments

Hanna Instruments HI749 User manual

Hanna Instruments HI97738 User manual

Hanna Instruments

Hanna Instruments HI97738 User manual

Hanna Instruments HI 9813-5 User manual

Hanna Instruments

Hanna Instruments HI 9813-5 User manual

Hanna Instruments HI 255 User manual

Hanna Instruments

Hanna Instruments HI 255 User manual

Hanna Instruments HI 9143 User manual

Hanna Instruments

Hanna Instruments HI 9143 User manual

Hanna Instruments HI 96832 User manual

Hanna Instruments

Hanna Instruments HI 96832 User manual

Hanna Instruments HI 84532 User manual

Hanna Instruments

Hanna Instruments HI 84532 User manual

Hanna Instruments HI 761 User manual

Hanna Instruments

Hanna Instruments HI 761 User manual

Hanna Instruments HI 98186 User manual

Hanna Instruments

Hanna Instruments HI 98186 User manual

Hanna Instruments HI9064 User manual

Hanna Instruments

Hanna Instruments HI9064 User manual

Hanna Instruments HI706 User manual

Hanna Instruments

Hanna Instruments HI706 User manual

Hanna Instruments HI 96831 User manual

Hanna Instruments

Hanna Instruments HI 96831 User manual

Hanna Instruments HI 84100 User manual

Hanna Instruments

Hanna Instruments HI 84100 User manual

Hanna Instruments HI 8564 User manual

Hanna Instruments

Hanna Instruments HI 8564 User manual

Hanna Instruments HI 746 User manual

Hanna Instruments

Hanna Instruments HI 746 User manual

Hanna Instruments HI 93720 User manual

Hanna Instruments

Hanna Instruments HI 93720 User manual

Hanna Instruments HI 4107 User manual

Hanna Instruments

Hanna Instruments HI 4107 User manual

Hanna Instruments HI 93749 User manual

Hanna Instruments

Hanna Instruments HI 93749 User manual

Hanna Instruments HI98163 User manual

Hanna Instruments

Hanna Instruments HI98163 User manual

Hanna Instruments HI83325 User manual

Hanna Instruments

Hanna Instruments HI83325 User manual

Popular Measuring Instrument manuals by other brands

HT 64 user manual

HT

HT 64 user manual

Endress+Hauser FMU 40 Description of Functions

Endress+Hauser

Endress+Hauser FMU 40 Description of Functions

Flow-tronic Raven-Eye Installation & operation manual

Flow-tronic

Flow-tronic Raven-Eye Installation & operation manual

Geo FL 45 user manual

Geo

Geo FL 45 user manual

Given Imaging PillCam DR3 quick start guide

Given Imaging

Given Imaging PillCam DR3 quick start guide

uher 4000 REPORT-L Service

uher

uher 4000 REPORT-L Service

Winland Electronics WaterBug WB-200 Operating instruction guide

Winland Electronics

Winland Electronics WaterBug WB-200 Operating instruction guide

Fowler Euro-Caliper IV quick start guide

Fowler

Fowler Euro-Caliper IV quick start guide

ATAGO PAL-mOsm. BLT/i instruction manual

ATAGO

ATAGO PAL-mOsm. BLT/i instruction manual

Precia Molen i5 user manual

Precia Molen

Precia Molen i5 user manual

Dish TV A2 quick start guide

Dish TV

Dish TV A2 quick start guide

Leica Geosystems iCON gps 30 user manual

Leica Geosystems

Leica Geosystems iCON gps 30 user manual

Primes Cube L1 operating manual

Primes

Primes Cube L1 operating manual

Parkside PMSHM 2 A1 user manual

Parkside

Parkside PMSHM 2 A1 user manual

MOBILE-ALERTS MA10350 quick reference

MOBILE-ALERTS

MOBILE-ALERTS MA10350 quick reference

Keysight E4991B troubleshooting guide

Keysight

Keysight E4991B troubleshooting guide

Dimetix FLS-C Application note

Dimetix

Dimetix FLS-C Application note

Multimode Vigil II user manual

Multimode

Multimode Vigil II user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.