Instruction Manual
HI 93748
Manganese Low Range
ISM
http://www.hannainst.com
This Instrument is in
Compliance with the CE Directives
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
WARRANTYWARRANTY
WARRANTYWARRANTY
WARRANTY
HI 93748 is warranted for two years against defects in workman-
ship and materials when used for its intended purpose and maintained
according to instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of pur-
chase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service de-
partment and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete protec-
tion.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instru-
ments Inc., Woonsocket, Rhode Island, 02895 , USA.
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at [email protected].
This instrument is in compliance with directives
EN 50081-1 and EN 50082-1.
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each Ion Specific Meter is supplied complete with
• 9V Battery
• Two Sample Cuvets and Caps
• One Transport Cap
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
The HI 93748 meter measures the manganese content in
water and wastewater in the 0 to 300 µg/L (ppb) range.
The meter uses an exclusive positive-locking system to ensure
that the cuvet is in the same place every time it is placed
into the measurement cell.
The reagents are in liquid and powder form and are
supplied in bottles and in packets. The amount of reagent
is precisely dosed to ensure maximum repeatability.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Range 0 to 300 µg/L
Resolution 1 µg/L
Accuracy ±2 µg/L ±3% of reading
Typical EMC ±1 µg/L
Deviation
Light Source Light Emitting Diode @ 555 nm
Method Adaptation of the PAN Method. The reac-
tion between manganese and the reagents
causes an orange tint in the sample
Light Detector Silicon Photocell
Environment 0 to 50C (32 to 122F);
max 95% RH non-condensing
Battery Type/Life 1 x 9 volt/40 hours
Auto-Shut off After 10' of non-use
Dimensions 180 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight 290 g (10 oz.).
REQUIRED REAGENTS
Code Description Quantity
HI 93748A-0 Ascorbic acid 2 packets
HI 93748B-0 Alkaline-cyanide sol. 0.40 mL
HI 93748C-0 0.1% PAN indicator 2 mL
HI 93703-51 Dispersing Agent 4-6 drops (only
when necessary, see note)
REAGENT SETS
HI 93748-01 Reagents for manganese LR (100 pcs of
HI93748A-0, 20 mL of HI93748B-0, 100 mL
of HI93748C-0, 20 mL of HI93703-51)
HI 93748-03 Reagents for manganese LR (300 pcs of
HI93748A-0, 60 mL of HI93748B-0, 300 mL
of HI93748C-0, 60 mL of HI93703-51).
CUVET HOLDER
CUVET
ALIGNMENT INDICATOR
ON/OFF
ZERO
READ TIMED
READ DIRECT
ins
e
rum ns
ON
OFF
ZERO
READ
DIRECT
READ
TIMED
HI 93748
MANGANESE
LOW RANGE
Range: 0 to 300 ppb
This indicates that the meter is in a
ready state and zeroing can be per-
formed.
Sampling in Progress. This prompt ap-
pears each time the meter is performing
a measurement.
This indicates that the meter is in a
zeroed state and measurement can be
performed.
A zero reading was not taken. Insert a
sample before adding reagent and press
ZERO.
Under range. A blinking "0.00" indicates
that the sample absorbs less light than
the zero reference. Check the procedure
and make sure you use the same cuvet
for reference (zero) and measurement.
Over range. A flashing value higher than
the maximum concentration readable (see
specifications) indicates that the sample
absorbs too much light, meaning that
the concentration is too high. Dilute the
sample.
Light over range. The cuvet is not in-
serted correctly and an excess ambient
light is reaching the detector. If the cover
is properly installed, then contact your
dealer or the nearest Hanna Customer
Service Center.
Light under range. The zero sample is
too dark for proper zeroing. If this is not
the case, contact your dealer or the
nearest Hanna Customer Service Center.
The "V" indicates that the battery volt-
age is getting low and the battery needs
to be replaced.
This indicates that the battery is dead
and must be replaced.
Note: once this indication is displayed,
the meter will lockup. Change the bat-
tery to restart.
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDE
V
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
MEASUREMENT PROCEDURE
• Turn the meter on by pressing ON/OFF.
Note: for best results perform your tests between 20-24C.
• When the LCD displays "- - -", it is
ready.
• Fill one cuvet up to 1.5 cm (¾")
below the rim with 10 mL of deion-
ized water.
• Fill a second cuvet up to 1.5 cm (¾")
below the rim with 10 mL of sample.
• Add the content of one packet of
HI 93748A ascorbic acid to each
cuvet, replace the caps and shake
gently.
• Add 0.2 mL of the HI
93748B alkaline-cya-
nide reagent solution to
each cuvet, replace the
caps and shake gen-
tly.
• Add 1 mL of the HI
93748C 0.1% PAN in-
dicator solution to each
cuvet, replace the caps
and shake gently.
• Place the cuvet with the re-
acted deionized water (blank)
into the holder and ensure
that the notch on the cap is
positioned securely into the
groove.
ON
OFF
1.5 c
m
1.5 c
m
HI93748A
B
C
• Press READ TIMED and the display will show the count-
down prior to zeroing the blank. Alternatively wait for 2
minutes and press ZERO. In both cases "SIP" will
appear during zeroing.
• Wait for a few seconds and the dis-
play will show "-0.0-". Now the meter
is zeroed and ready for measure-
ment.
• Insert the second cuvet with
the reacted sample into the
instrument.
• Press READ DIRECT. "SIP" will appear during measure-
ment.
• The instrument directly displays concentration in µg/L of
manganese on the display.
Note: a temperature above 30C may cause turbidity. In
this case, before zeroing and taking readings, add
2-3 drops of Dispersing Agent (HI93703-51) to each
cuvet and swirl until complete dissolution of the
turbidity.
INTERFERENCES
Interference may be caused by:
Aluminum above 20 mg/L
Cadmium above 10 mg/L
Calcium above 200 mg/L as CaCO3
Cobalt above 20 mg/L
Copper above 50 mg/L
Iron above 10 mg/L
Lead above 0.5 mg/L
Magnesium above 100 mg/L as CaCO3
Nickel above 40 mg/L
Zinc above 15 mg/L.
ZERO
12
3
6
9
2’
READ
TIMED
READ
DIRECT
The instruction listed below should be carefully followed
during testing to ensure best accuracy.
• Do not touch the cuvet walls with hands.
• In order to maintain the same conditions during the
zeroing and the measuring phases, it is necessary to
close the cuvet to prevent any contamination.
• Do not let the test sample stand too long after reagent
is added or accuracy will be lost.
• Whenever the cuvet is placed into the measurement cell,
it must be completely free of fingerprints, oil or dirt.
Wipe it thoroughly with HI 731318 or a lint-free cloth
prior to insertion.
• It is important that the sample does not contain any
debris. This would corrupt the readings.
• It is possible to take multiple readings in a row, but it
is recommended that a zero reading be taken for each
sample and that the same cuvet is used for zeroing and
measurement.
• It is important to discard the sample immediately after
the reading is taken because the glass might become
permanently stained.
• Shaking the cuvet can generate bubbles in the sample,
causing higher readings. To obtain accurate measure-
ments, remove such bubbles by swirling or by gently
tapping the vial.
• All the reaction times reported in this manual are
referred to 20C (68F). As a general rule of thumb,
they should be doubled at 10C (50F) and halved at
30C(86F).
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATE
MEASUREMENTMEASUREMENT
MEASUREMENTMEASUREMENT
MEASUREMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
Battery replacement must only take place in a non-hazard-
ous area using a 9V alkaline battery.
Simply slide off the battery cover on the back of the meter.
Detach the battery from the terminals and attach a fresh
9V battery while paying attention to the correct polarity.
Replace the battery and the cover.
9V
SLIDE
OFF
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environ-
ment in which they are used.
Operation of these instruments in residential area could cause unacceptable interferences
to radio and TV equipments, requiring the operator to take all necessary steps to correct
interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments' EMC performance.
To avoid damages or burns, do not perform any measurement in microwave ovens.
ACCESSORIESACCESSORIES
ACCESSORIESACCESSORIES
ACCESSORIES
REAGENT SETS
HI 93748-01 Reagents for manganese LR (100 pcs of
HI93748A-0, 20 mL of HI93748B-0, 100 mL
of HI93748C-0, 20 mL of HI93703-51)
HI 93748-03 Reagents for manganese LR (300 pcs of
HI93748A-0, 60 mL of HI93748B-0, 300 mL
of HI93748C-0, 60 mL of HI93703-51)
OTHER ACCESSORIES
HI 710009 Blue rubber boot
HI 710010 Orange rubber boot
HI 721310 9V battery (10 pcs)
HI 731318 Tissue for wiping cuvets (4 pcs)
HI 731321 Glass cuvets (4 pcs)
HI 731325 Caps for cuvets (4 pcs)
HI 93703-50 Cuvets cleaning solution (230 mL).
CE DECLARATIONCE DECLARATION
CE DECLARATIONCE DECLARATION
CE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
IST93748R1 06/98
PRINTED IN PORTUGAL

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments DiST HI 98302 User manual

Hanna Instruments

Hanna Instruments DiST HI 98302 User manual

Hanna Instruments HI97751 User manual

Hanna Instruments

Hanna Instruments HI97751 User manual

Hanna Instruments HI97738 User manual

Hanna Instruments

Hanna Instruments HI97738 User manual

Hanna Instruments HI96752 User manual

Hanna Instruments

Hanna Instruments HI96752 User manual

Hanna Instruments HI 98186 User manual

Hanna Instruments

Hanna Instruments HI 98186 User manual

Hanna Instruments HI 2222 User manual

Hanna Instruments

Hanna Instruments HI 2222 User manual

Hanna Instruments Ammonia Medium Range User manual

Hanna Instruments

Hanna Instruments Ammonia Medium Range User manual

Hanna Instruments HI97771 User manual

Hanna Instruments

Hanna Instruments HI97771 User manual

Hanna Instruments HI 83414 User manual

Hanna Instruments

Hanna Instruments HI 83414 User manual

Hanna Instruments HI 4221 User manual

Hanna Instruments

Hanna Instruments HI 4221 User manual

Hanna Instruments HI 84431 User manual

Hanna Instruments

Hanna Instruments HI 84431 User manual

Hanna Instruments HI 99131 User manual

Hanna Instruments

Hanna Instruments HI 99131 User manual

Hanna Instruments pH 211R User manual

Hanna Instruments

Hanna Instruments pH 211R User manual

Hanna Instruments HI 83748 User manual

Hanna Instruments

Hanna Instruments HI 83748 User manual

Hanna Instruments HI96700C User manual

Hanna Instruments

Hanna Instruments HI96700C User manual

Hanna Instruments HI 83205 User manual

Hanna Instruments

Hanna Instruments HI 83205 User manual

Hanna Instruments HI 98703 User manual

Hanna Instruments

Hanna Instruments HI 98703 User manual

Hanna Instruments HI 702 User manual

Hanna Instruments

Hanna Instruments HI 702 User manual

Hanna Instruments HI 96813 User manual

Hanna Instruments

Hanna Instruments HI 96813 User manual

Hanna Instruments HI 9026 User manual

Hanna Instruments

Hanna Instruments HI 9026 User manual

Hanna Instruments C 215 User manual

Hanna Instruments

Hanna Instruments C 215 User manual

Hanna Instruments HI 83141V User manual

Hanna Instruments

Hanna Instruments HI 83141V User manual

Hanna Instruments HALO 2 HI9810422 User manual

Hanna Instruments

Hanna Instruments HALO 2 HI9810422 User manual

Hanna Instruments HI 142 Series User manual

Hanna Instruments

Hanna Instruments HI 142 Series User manual

Popular Measuring Instrument manuals by other brands

Tektronix 1502B Service manual

Tektronix

Tektronix 1502B Service manual

Endress+Hauser Proline Prosonic Flow 93 Function manual

Endress+Hauser

Endress+Hauser Proline Prosonic Flow 93 Function manual

Sealey TA203.V2 instructions

Sealey

Sealey TA203.V2 instructions

Keysight x series Installation notes

Keysight

Keysight x series Installation notes

KJ Instrument KJ6013 Operation manual

KJ Instrument

KJ Instrument KJ6013 Operation manual

KROHNE OPTIFLUX 2070 Handbook

KROHNE

KROHNE OPTIFLUX 2070 Handbook

Veris Industries E51C2A installation guide

Veris Industries

Veris Industries E51C2A installation guide

RTX 6001 user manual

RTX

RTX 6001 user manual

Siemens SENTRON PAC4200 manual

Siemens

Siemens SENTRON PAC4200 manual

MARTINDALE TEK100 instruction manual

MARTINDALE

MARTINDALE TEK100 instruction manual

MRC DI-900 Operation manual

MRC

MRC DI-900 Operation manual

Elcon SA10A user manual

Elcon

Elcon SA10A user manual

Schroeder Industries HY-TRAX Operating and maintenance instructions

Schroeder Industries

Schroeder Industries HY-TRAX Operating and maintenance instructions

MSI 2000 Programming manual

MSI

MSI 2000 Programming manual

WAGNER C575 instruction manual

WAGNER

WAGNER C575 instruction manual

Newport 1825-C Operator's manual

Newport

Newport 1825-C Operator's manual

jwLabs A3 Instruction manuals

jwLabs

jwLabs A3 Instruction manuals

Endress+Hauser Prosonic S FMU90 operating instructions

Endress+Hauser

Endress+Hauser Prosonic S FMU90 operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.