  9. Hanna Instruments HI 93748 User manual

Hanna Instruments HI 93748 User manual

Instruction Manual
HI 93748
Manganese Low Range
ISM
http://www.hannainst.com
This Instrument is in
Compliance with the CE Directives
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
WARRANTYWARRANTY
WARRANTYWARRANTY
WARRANTY
HI 93748 is warranted for two years against defects in workman-
ship and materials when used for its intended purpose and maintained
according to instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of pur-
chase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service de-
partment and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete protec-
tion.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instru-
ments Inc., Woonsocket, Rhode Island, 02895 , USA.
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at [email protected].
This instrument is in compliance with directives
EN 50081-1 and EN 50082-1.
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each Ion Specific Meter is supplied complete with
• 9V Battery
• Two Sample Cuvets and Caps
• One Transport Cap
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
The HI 93748 meter measures the manganese content in
water and wastewater in the 0 to 300 µg/L (ppb) range.
The meter uses an exclusive positive-locking system to ensure
that the cuvet is in the same place every time it is placed
into the measurement cell.
The reagents are in liquid and powder form and are
supplied in bottles and in packets. The amount of reagent
is precisely dosed to ensure maximum repeatability.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Range 0 to 300 µg/L
Resolution 1 µg/L
Accuracy ±2 µg/L ±3% of reading
Typical EMC ±1 µg/L
Deviation
Light Source Light Emitting Diode @ 555 nm
Method Adaptation of the PAN Method. The reac-
tion between manganese and the reagents
causes an orange tint in the sample
Light Detector Silicon Photocell
Environment 0 to 50C (32 to 122F);
max 95% RH non-condensing
Battery Type/Life 1 x 9 volt/40 hours
Auto-Shut off After 10' of non-use
Dimensions 180 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight 290 g (10 oz.).
REQUIRED REAGENTS
Code Description Quantity
HI 93748A-0 Ascorbic acid 2 packets
HI 93748B-0 Alkaline-cyanide sol. 0.40 mL
HI 93748C-0 0.1% PAN indicator 2 mL
HI 93703-51 Dispersing Agent 4-6 drops (only
when necessary, see note)
REAGENT SETS
HI 93748-01 Reagents for manganese LR (100 pcs of
HI93748A-0, 20 mL of HI93748B-0, 100 mL
of HI93748C-0, 20 mL of HI93703-51)
HI 93748-03 Reagents for manganese LR (300 pcs of
HI93748A-0, 60 mL of HI93748B-0, 300 mL
of HI93748C-0, 60 mL of HI93703-51).
CUVET HOLDER
CUVET
ALIGNMENT INDICATOR
ON/OFF
ZERO
READ TIMED
READ DIRECT
ins
e
rum ns
ON
OFF
ZERO
READ
DIRECT
READ
TIMED
HI 93748
MANGANESE
LOW RANGE
Range: 0 to 300 ppb
This indicates that the meter is in a
ready state and zeroing can be per-
formed.
Sampling in Progress. This prompt ap-
pears each time the meter is performing
a measurement.
This indicates that the meter is in a
zeroed state and measurement can be
performed.
A zero reading was not taken. Insert a
sample before adding reagent and press
ZERO.
Under range. A blinking "0.00" indicates
that the sample absorbs less light than
the zero reference. Check the procedure
and make sure you use the same cuvet
for reference (zero) and measurement.
Over range. A flashing value higher than
the maximum concentration readable (see
specifications) indicates that the sample
absorbs too much light, meaning that
the concentration is too high. Dilute the
sample.
Light over range. The cuvet is not in-
serted correctly and an excess ambient
light is reaching the detector. If the cover
is properly installed, then contact your
dealer or the nearest Hanna Customer
Service Center.
Light under range. The zero sample is
too dark for proper zeroing. If this is not
the case, contact your dealer or the
nearest Hanna Customer Service Center.
The "V" indicates that the battery volt-
age is getting low and the battery needs
to be replaced.
This indicates that the battery is dead
and must be replaced.
Note: once this indication is displayed,
the meter will lockup. Change the bat-
tery to restart.
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDE
V
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
MEASUREMENT PROCEDURE
• Turn the meter on by pressing ON/OFF.
Note: for best results perform your tests between 20-24C.
• When the LCD displays "- - -", it is
ready.
• Fill one cuvet up to 1.5 cm (¾")
below the rim with 10 mL of deion-
ized water.
• Fill a second cuvet up to 1.5 cm (¾")
below the rim with 10 mL of sample.
• Add the content of one packet of
HI 93748A ascorbic acid to each
cuvet, replace the caps and shake
gently.
• Add 0.2 mL of the HI
93748B alkaline-cya-
nide reagent solution to
each cuvet, replace the
caps and shake gen-
tly.
• Add 1 mL of the HI
93748C 0.1% PAN in-
dicator solution to each
cuvet, replace the caps
and shake gently.
• Place the cuvet with the re-
acted deionized water (blank)
into the holder and ensure
that the notch on the cap is
positioned securely into the
groove.
ON
OFF
1.5 c
m
1.5 c
m
HI93748A
B
C
• Press READ TIMED and the display will show the count-
down prior to zeroing the blank. Alternatively wait for 2
minutes and press ZERO. In both cases "SIP" will
appear during zeroing.
• Wait for a few seconds and the dis-
play will show "-0.0-". Now the meter
is zeroed and ready for measure-
ment.
• Insert the second cuvet with
the reacted sample into the
instrument.
• Press READ DIRECT. "SIP" will appear during measure-
ment.
• The instrument directly displays concentration in µg/L of
manganese on the display.
Note: a temperature above 30C may cause turbidity. In
this case, before zeroing and taking readings, add
2-3 drops of Dispersing Agent (HI93703-51) to each
cuvet and swirl until complete dissolution of the
turbidity.
INTERFERENCES
Interference may be caused by:
Aluminum above 20 mg/L
Cadmium above 10 mg/L
Calcium above 200 mg/L as CaCO3
Cobalt above 20 mg/L
Copper above 50 mg/L
Iron above 10 mg/L
Lead above 0.5 mg/L
Magnesium above 100 mg/L as CaCO3
Nickel above 40 mg/L
Zinc above 15 mg/L.
ZERO
12
3
6
9
2’
READ
TIMED
READ
DIRECT
The instruction listed below should be carefully followed
during testing to ensure best accuracy.
• Do not touch the cuvet walls with hands.
• In order to maintain the same conditions during the
zeroing and the measuring phases, it is necessary to
close the cuvet to prevent any contamination.
• Do not let the test sample stand too long after reagent
is added or accuracy will be lost.
• Whenever the cuvet is placed into the measurement cell,
it must be completely free of fingerprints, oil or dirt.
Wipe it thoroughly with HI 731318 or a lint-free cloth
prior to insertion.
• It is important that the sample does not contain any
debris. This would corrupt the readings.
• It is possible to take multiple readings in a row, but it
is recommended that a zero reading be taken for each
sample and that the same cuvet is used for zeroing and
measurement.
• It is important to discard the sample immediately after
the reading is taken because the glass might become
permanently stained.
• Shaking the cuvet can generate bubbles in the sample,
causing higher readings. To obtain accurate measure-
ments, remove such bubbles by swirling or by gently
tapping the vial.
• All the reaction times reported in this manual are
referred to 20C (68F). As a general rule of thumb,
they should be doubled at 10C (50F) and halved at
30C(86F).
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATE
MEASUREMENTMEASUREMENT
MEASUREMENTMEASUREMENT
MEASUREMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
Battery replacement must only take place in a non-hazard-
ous area using a 9V alkaline battery.
Simply slide off the battery cover on the back of the meter.
Detach the battery from the terminals and attach a fresh
9V battery while paying attention to the correct polarity.
Replace the battery and the cover.
9V
SLIDE
OFF
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environ-
ment in which they are used.
Operation of these instruments in residential area could cause unacceptable interferences
to radio and TV equipments, requiring the operator to take all necessary steps to correct
interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments' EMC performance.
To avoid damages or burns, do not perform any measurement in microwave ovens.
ACCESSORIESACCESSORIES
ACCESSORIESACCESSORIES
ACCESSORIES
REAGENT SETS
HI 93748-01 Reagents for manganese LR (100 pcs of
HI93748A-0, 20 mL of HI93748B-0, 100 mL
of HI93748C-0, 20 mL of HI93703-51)
HI 93748-03 Reagents for manganese LR (300 pcs of
HI93748A-0, 60 mL of HI93748B-0, 300 mL
of HI93748C-0, 60 mL of HI93703-51)
OTHER ACCESSORIES
HI 710009 Blue rubber boot
HI 710010 Orange rubber boot
HI 721310 9V battery (10 pcs)
HI 731318 Tissue for wiping cuvets (4 pcs)
HI 731321 Glass cuvets (4 pcs)
HI 731325 Caps for cuvets (4 pcs)
HI 93703-50 Cuvets cleaning solution (230 mL).
CE DECLARATIONCE DECLARATION
CE DECLARATIONCE DECLARATION
CE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
IST93748R1 06/98
PRINTED IN PORTUGAL

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI 931101 User manual

Hanna Instruments

Hanna Instruments HI 931101 User manual

Hanna Instruments HI 83208 User manual

Hanna Instruments

Hanna Instruments HI 83208 User manual

Hanna Instruments DiST 2 User manual

Hanna Instruments

Hanna Instruments DiST 2 User manual

Hanna Instruments HI 96750 User manual

Hanna Instruments

Hanna Instruments HI 96750 User manual

Hanna Instruments HI 93731 User manual

Hanna Instruments

Hanna Instruments HI 93731 User manual

Hanna Instruments pHep HI 98107 User manual

Hanna Instruments

Hanna Instruments pHep HI 98107 User manual

Hanna Instruments HI 99131 User manual

Hanna Instruments

Hanna Instruments HI 99131 User manual

Hanna Instruments LP 2000 User manual

Hanna Instruments

Hanna Instruments LP 2000 User manual

Hanna Instruments HI 96731 User manual

Hanna Instruments

Hanna Instruments HI 96731 User manual

Hanna Instruments HI96719C User manual

Hanna Instruments

Hanna Instruments HI96719C User manual

Hanna Instruments HI9064 User manual

Hanna Instruments

Hanna Instruments HI9064 User manual

Hanna Instruments HI 9811-5 User manual

Hanna Instruments

Hanna Instruments HI 9811-5 User manual

Hanna Instruments Checker HI726 User manual

Hanna Instruments

Hanna Instruments Checker HI726 User manual

Hanna Instruments HI96722 User manual

Hanna Instruments

Hanna Instruments HI96722 User manual

Hanna Instruments HI 9126 User manual

Hanna Instruments

Hanna Instruments HI 9126 User manual

Hanna Instruments edge EC HI2003 User manual

Hanna Instruments

Hanna Instruments edge EC HI2003 User manual

Hanna Instruments HI 93750 User manual

Hanna Instruments

Hanna Instruments HI 93750 User manual

Hanna Instruments HI 98186 User manual

Hanna Instruments

Hanna Instruments HI 98186 User manual

Hanna Instruments Checker HI 721 User manual

Hanna Instruments

Hanna Instruments Checker HI 721 User manual

Hanna Instruments HI759 User manual

Hanna Instruments

Hanna Instruments HI759 User manual

Hanna Instruments HI97769 User manual

Hanna Instruments

Hanna Instruments HI97769 User manual

Hanna Instruments HI 98128 User manual

Hanna Instruments

Hanna Instruments HI 98128 User manual

Hanna Instruments HI6221 User manual

Hanna Instruments

Hanna Instruments HI6221 User manual

Hanna Instruments PICCOLO User manual

Hanna Instruments

Hanna Instruments PICCOLO User manual

Popular Measuring Instrument manuals by other brands

Vutlan HAT-100Q1 Quick installation manual

Vutlan

Vutlan HAT-100Q1 Quick installation manual

ITRON Intelis Setup guide

ITRON

ITRON Intelis Setup guide

BRUEL & KJAER 2203 Instructions and applications

BRUEL & KJAER

BRUEL & KJAER 2203 Instructions and applications

Sequoia LISST-VSF user manual

Sequoia

Sequoia LISST-VSF user manual

TTI 1604 Service manual

TTI

TTI 1604 Service manual

Goobay 67000 instruction manual

Goobay

Goobay 67000 instruction manual

Mitchell Instrument WENS 900C user manual

Mitchell Instrument

Mitchell Instrument WENS 900C user manual

Syco SWR quick start guide

Syco

Syco SWR quick start guide

Pacific Data Systems DPM-RT-2 operating manual

Pacific Data Systems

Pacific Data Systems DPM-RT-2 operating manual

YOKOGAWA Power Series Plus IM 2493WVP-1 user manual

YOKOGAWA

YOKOGAWA Power Series Plus IM 2493WVP-1 user manual

Lutron Electronics MHT-381SD Operation manual

Lutron Electronics

Lutron Electronics MHT-381SD Operation manual

Myzone Inbody 230 Setup instructions

Myzone

Myzone Inbody 230 Setup instructions

Endress+Hauser Prosonic T FMU 130 operating manual

Endress+Hauser

Endress+Hauser Prosonic T FMU 130 operating manual

XS Instruments Temp70 PT100 vio instruction manual

XS Instruments

XS Instruments Temp70 PT100 vio instruction manual

RadComm RC2PLUS user manual

RadComm

RadComm RC2PLUS user manual

ZBM MINI OTDR user manual

ZBM

ZBM MINI OTDR user manual

Dostmann Electronic P705 Operation manuals

Dostmann Electronic

Dostmann Electronic P705 Operation manuals

Simrad PI50 -  REV A installation manual

Simrad

Simrad PI50 - REV A installation manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.