Instruction Manual
HI 93729
Fluoride ISM
http://www.hannainst.com ThisInstrumentisin
Compliancewith the CE Directives
HannaInstrumentsreservestherighttomodifythedesign,construction
andappearanceofitsproductswithoutadvancenotice.
WARRANTYWARRANTY
WARRANTYWARRANTY
WARRANTY
HI93729iswarrantedfor two years against defects in workman-
shipand materialswhenusedforits intendedpurposeandmaintained
accordingtoinstructions.
Thiswarranty islimitedto repairorreplacement freeofcharge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenanceare not covered.
Ifserviceis required,contactthedealerfromwhom youpurchasedthe
instrument.Ifunderwarranty, report the model number, dateofpur-
chase,serialnumberandthe nature of the failure. Iftherepairisnot
coveredby the warranty,youwillbe notified ofthechargesincurred. If
theinstrumentistobereturnedtoHannaInstruments,firstobtaina
ReturnedGoodsAuthorizationNumber from the Customer Service de-
partmentand thensenditwith shipmentcostsprepaid.Whenshipping
anyinstrument, makesureitisproperly packagedforcompleteprotec-
tion.
Tovalidate yourwarranty,filloutandreturntheenclosedwarrantycard
within14 daysfromthedateofpurchase.
Allrightsare reserved. Reproduction inwholeor in part isprohibited
without the written consent of the copyright owner, Hanna Instru-
mentsInc.,Woonsocket, Rhode Island, 02895,USA.
DearCustomer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at [email protected].
This instrument is in compliance with directives
EN 50081-1 and EN 50082-1.
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each Ion Specific Meter is supplied complete with
• 9V Battery
• Two Sample Cuvets and Caps
• One Transport Cap
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
The HI 93729 meter measures the fluoride (F–) content in
water, wastewater and seawater in the 0.00 to 2.00 mg/
L (ppm) range.
Themeter usesan exclusivepositive-locking systemto ensure
that the cuvet is in the same place every time it is placed
into the measurement cell.
The reagents are in liquid form and are supplied in bottles.
The amount of reagent is precisely dosed to ensure maxi-
mum repeatability.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Range 0.00 to 2.00 mg/L
Resolution 0.01 mg/L
Accuracy ±5% of reading
TypicalEMC ±0.01 mg/L
Deviation
Light Source Light Emitting Diode @ 585 nm
Method Adaptation of the SPADNS method. The
reaction between fluoride and the liquid
reagent causes a red tint in the sample
Light Detector SiliconPhotocell
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type/Life1 x 9 volt/40 hours
Auto-Shut off After 10' of non-use
Dimensions 180 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight 290 g (10 oz.).
REQUIRED REAGENTS
Code Description Quantity
HI 93729-0 SPADNS Reagent 4 mL
REAGENT SETS
HI93729-01 Reagents for 100 tests
HI93729-03 Reagents for 300 tests
This indicates that the meter is in a
ready state and zeroing can be per-
formed.
Sampling in Progress. This prompt ap-
pears each time the meter is performing
a measurement.
This indicates that the meter is in a
zeroed state and measurement can be
performed.
A zero reading was not taken. Insert a
sample before adding reagent and press
ZERO.
Under range. A blinking "0.00" indicates
that the sample absorbs less light than
the zero reference. Check the procedure
and make sure you use the same cuvet
for reference (zero) and measurement.
Over range. A flashing value higher than
themaximumconcentrationreadable(see
specifications) indicates that the sample
absorbs too much light, meaning that
the concentration is too high. Dilute the
sample.
Light over range. The cuvet is not in-
serted correctly and an excess ambient
light is reaching the detector. If the cover
is properly installed, then contact your
dealer or the nearest Hanna Customer
Service Center.
Light under range. The zero sample is
too dark for proper zeroing. If this is not
the case, contact your dealer or the
nearest Hanna Customer Service Center.
The "V" indicates that the battery volt-
age is getting low and the battery needs
to be replaced.
This indicates that the battery is dead
and must be replaced.
Note: once this indication is displayed,
the meter will lockup. Change the bat-
tery to restart.
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDE
V
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
MEASUREMENT PROCEDURE
• Turn the meter on by pressing ON/
OFF.
• When the LCD displays "- - -", it is
ready.
• Fill the cuvet up to 1.5 cm (¾")
below the rim with 10 mL of distilled
water.
• Fill the other cuvet up to 1.5 cm
(¾") below the rim with 10 mL of
sample.
• Add to each cuvet 2 mL of HI
93729 SPADNS reagent, re-
place the caps and invert
several times to mix.
• Wait for one minute.
• Place the cuvet with the re-
acted distilled water into the
holder and ensure that the
notchonthecap is positioned
securely into the groove.
• Press ZERO and "SIP" will appear on the display.
• Wait for a few seconds and the dis-
playwillshow "-0.0-".Nowthe meter
is zeroed and ready for measure-
ment.
ON
OFF
1.5 cm
1.5 cm
12
3
6
9
1’
ZERO
• Insert the other cuvet with the
reacted sample into the instru-
ment.
• Press READ and "SIP" will appear on the LCD during
measurement.
• Theinstrument directly displays concentration in mg/L of
fluoride on the Liquid Crystal Display.
Note: With wastewater and seawater samples, before per-
formingany measurement adistillation of the sample
is needed.
INTERFERENCES
Negative interferences may be caused by:
Alkalinity (as CaCO3) above 5000 mg/L
Aluminum above 0.1 mg/L
Iron, ferric above 10 mg/L
Positive interferences may be caused by:
Chloride above 700 mg/L
Phosphate, ortho above 16 mg/L
Sodium hexametaphosphate above 1.0 mg/L
Sulfate above 200 mg/L
Highly colored and turbid samples may require distillation
Highly alkaline samples can be neutralized with nitric acid.
The instruction listed below should be carefully followed
during testing to ensure best accuracy.
• Do not touch the cuvet walls with hands.
• In order to maintain the same conditions during the
zeroing and the measuring phases, it is necessary to
close the cuvet to prevent any contamination.
• Do not let the test sample stand too long after reagent
is added or accuracy will be lost.
• Whenever the cuvet is placed into the measurement cell,
it must be completely free of fingerprints, oil or dirt.
Wipe it thoroughly with HI 731318 or a lint-free cloth
prior to insertion.
• It is important that the sample does not contain any
debris. This would corrupt the readings.
• It is possible to take multiple readings in a row, but it
is recommended that a zero reading be taken for each
sample and that the same cuvet is used for zeroing and
measurement.
• It is important to discard the sample immediately after
the reading is taken because the glass might become
permanently stained.
• Shaking the cuvet can generate bubbles in the sample,
causing higher readings. To obtain accurate measure-
ments, remove such bubbles by swirling or by gently
tapping the vial.
• All the reaction times reported in this manual are
referred to 20°C (68°F). As a general rule of thumb,
they should be doubled at 10°C (50°F) and halved at
30°C(86°F).
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATE
MEASUREMENTMEASUREMENT
MEASUREMENTMEASUREMENT
MEASUREMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
Battery replacement must only take place in a non-hazard-
ous area using a 9V alkaline battery.
Simply slide off the battery cover on the back of the meter.
Detach the battery from the terminals and attach a fresh
9V battery while paying attention to the correct polarity.
Replace the battery and the cover.
9V
SLID
E
OFF
RecommendationsforUsers
Before using these products, make sure that they are entirely suitable for the environ-
mentinwhichtheyareused.
Operationoftheseinstrumentsinresidentialareacouldcauseunacceptableinterferences
toradioandTV equipments, requiring the operatortotakeallnecessary steps to correct
interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments'EMCperformance.
Toavoiddamagesorburns,donotperformanymeasurementinmicrowaveovens.
ACCESSORIESACCESSORIES
ACCESSORIESACCESSORIES
ACCESSORIES
REAGENT SETS
HI93729-01 Reagents for 100 tests
HI93729-03 Reagents for 300 tests
OTHER ACCESSORIES
HI710009 Blue rubber boot
HI710010 Orange rubber boot
HI721310 9V battery (10 pcs)
HI731318 Tissue for wiping cuvets (4 pcs)
HI731321 Glass cuvets (4 pcs)
HI731325 Caps for cuvets (4 pcs)
HI93703-50 Cuvets cleaning solution (230 mL).
CE DECLARATIONCE DECLARATION
CE DECLARATIONCE DECLARATION
CE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
IST93729R1 06/98
PRINTED IN PORTUGAL

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI97751 User manual

Hanna Instruments

Hanna Instruments HI97751 User manual

Hanna Instruments HI 95729C User manual

Hanna Instruments

Hanna Instruments HI 95729C User manual

Hanna Instruments HI 216 User manual

Hanna Instruments

Hanna Instruments HI 216 User manual

Hanna Instruments HI 4110 User manual

Hanna Instruments

Hanna Instruments HI 4110 User manual

Hanna Instruments HI 84429 User manual

Hanna Instruments

Hanna Instruments HI 84429 User manual

Hanna Instruments HI97738 User manual

Hanna Instruments

Hanna Instruments HI97738 User manual

Hanna Instruments HI727 User manual

Hanna Instruments

Hanna Instruments HI727 User manual

Hanna Instruments HI 720 User manual

Hanna Instruments

Hanna Instruments HI 720 User manual

Hanna Instruments HI 9146 User manual

Hanna Instruments

Hanna Instruments HI 9146 User manual

Hanna Instruments HI 84102 User manual

Hanna Instruments

Hanna Instruments HI 84102 User manual

Hanna Instruments HI97750 User manual

Hanna Instruments

Hanna Instruments HI97750 User manual

Hanna Instruments HALO 2 HI9810402 User manual

Hanna Instruments

Hanna Instruments HALO 2 HI9810402 User manual

Hanna Instruments HI 38020 User manual

Hanna Instruments

Hanna Instruments HI 38020 User manual

Hanna Instruments HI 7635 User manual

Hanna Instruments

Hanna Instruments HI 7635 User manual

Hanna Instruments HI 762 User manual

Hanna Instruments

Hanna Instruments HI 762 User manual

Hanna Instruments HI 993310 User manual

Hanna Instruments

Hanna Instruments HI 993310 User manual

Hanna Instruments HI 8010 User manual

Hanna Instruments

Hanna Instruments HI 8010 User manual

Hanna Instruments HI 98713 User manual

Hanna Instruments

Hanna Instruments HI 98713 User manual

Hanna Instruments PRIMO 5 User manual

Hanna Instruments

Hanna Instruments PRIMO 5 User manual

Hanna Instruments HI97718 User manual

Hanna Instruments

Hanna Instruments HI97718 User manual

Hanna Instruments HI 98140 User manual

Hanna Instruments

Hanna Instruments HI 98140 User manual

Hanna Instruments HI 9142 User manual

Hanna Instruments

Hanna Instruments HI 9142 User manual

Hanna Instruments HI97737 User manual

Hanna Instruments

Hanna Instruments HI97737 User manual

Hanna Instruments HI 747 User manual

Hanna Instruments

Hanna Instruments HI 747 User manual

Popular Measuring Instrument manuals by other brands

Schweitzer Engineering Laboratories SEL-734 quick start guide

Schweitzer Engineering Laboratories

Schweitzer Engineering Laboratories SEL-734 quick start guide

Endress+Hauser Prothermo NMT81 operating instructions

Endress+Hauser

Endress+Hauser Prothermo NMT81 operating instructions

IMT Analytics FlowAnalyser Series user manual

IMT Analytics

IMT Analytics FlowAnalyser Series user manual

Larsen & Brusgaard ALTITRACK manual

Larsen & Brusgaard

Larsen & Brusgaard ALTITRACK manual

Novanta SYNRAD Pulstar p100 user manual

Novanta

Novanta SYNRAD Pulstar p100 user manual

LIMIT 1180-R operating manual

LIMIT

LIMIT 1180-R operating manual

Mahr MarCal 30 EWN operating instructions

Mahr

Mahr MarCal 30 EWN operating instructions

HOZO Design MEAZOR User instructions

HOZO Design

HOZO Design MEAZOR User instructions

Endress+Hauser StamoLys CA 70 operating instructions

Endress+Hauser

Endress+Hauser StamoLys CA 70 operating instructions

ITL MON-2697-COM Reference manual

ITL

ITL MON-2697-COM Reference manual

Ecolab EcoTrans II pH operating instructions

Ecolab

Ecolab EcoTrans II pH operating instructions

AEMC instruments MN185 user manual

AEMC instruments

AEMC instruments MN185 user manual

Amprobe VPC-10A user manual

Amprobe

Amprobe VPC-10A user manual

TFA 30.5033 instruction manual

TFA

TFA 30.5033 instruction manual

Testboy TV333 operating manual

Testboy

Testboy TV333 operating manual

ACL 395 Operation manual

ACL

ACL 395 Operation manual

AccuEnergy AcuRev 1300 user manual

AccuEnergy

AccuEnergy AcuRev 1300 user manual

Continental Refrigerator DTCO 1381 Process Guide

Continental Refrigerator

Continental Refrigerator DTCO 1381 Process Guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.