HI758
Marine Calcium
INSTRUCTION MANUAL
Reagent Sets
HI758-26 Reagents for 25 Marine Calcium tests
Other Accessories
HI758-11 Marine Calcium Certified Standard Kit
HI731315 Glass Cuvettes and Caps for Checker®HC Colorimeters (2 pcs.)
HI731318 Cloth for wiping cuvettes (4 pcs.)
HI731339P 0.1 mL minipipette
HI731349P Tip for 0.1 mL minipipette (10 pcs.)
HI740028P 1.5V AAA batteries (12 pcs.)
HI70436M Deionized water (230 mL)
HI93703-50 Cuvette cleaning solution (230 mL)
Accessories
IST758 05/18
HI758 is warranted for a period of one year after date of purchase against defects in
workmanship and materials when used for their intended purpose and maintained according
to instructions. This warranty is limited to repair or replacement free of charge. Damage due
to accidents, misuse, tampering or lack of prescribed maintenance is not covered. If service
is required, contact your local Hanna Instruments Office. If under warranty, report the model
number, date of purchase, serial number and the nature of the problem. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If the instrument is to
be returned to Hanna Instruments, first obtain a Return Goods Authorization number from the
Technical Service department and then send it with shipping costs prepaid. When shipping any
instrument, make sure it is properly packed for complete protection.
Before using Hanna Instruments products, make sure that they are entirely suitable for
your specific application and for the environment in which they are used. Operation of
these instruments may cause unacceptable interferences to other electronic equipment,
thus requiring the operator to take all necessary steps to correct such interferences. Any
variation introduced by the user to the supplied equipment may degrade the instrument’s
EMC performance. To avoid damages or burns, do not put the instrument in microwave
oven. For yours and the instrument safety do not use or store the instrument in hazardous
environments.
Recommendations for Users
Warranty
To conserve battery, the instrument shuts down after
10 minutes of non-use and 2 minutes after reading. One fresh battery
lasts for a minimum of 5000 measurements. When the battery is
dead the instrument will display “bAd” then “bAt” for 1 second
and then turns off. To replace the battery, turn the instrument off
by holding the button until the meter shuts off. Turn the instrument
upside down and remove the battery cover with a screwdriver.
Remove the battery from its location and replace it with a new one,
inserting the negative end first. Replace the battery cover and screw.
It is important that the sample does not contain any debris.
Whenever the cuvette is placed into the measurement cell, it must be
dry outside, and completely free of fingerprints, oil and dirt. Wipe it
thoroughly with HI731318 or a lint-free cloth prior to insertion.
Shaking the cuvette can generate bubbles, causing higher readings.
To obtain accurate measurements, remove bubbles by swirling or by
gently tapping the cuvette. Do not let the reacted sample stand for
too long after reagent is added, as accuracy will be affected. After the
reading, it is important to immediately discard the sample, otherwise
the glass might become permanently stained.
• Attach the pipette tip.
• Press the button down to the first stop.
• Immerse the pipette tip in the liquid approximately 2-3 mm.
• Slowly let the button move back to the original position, wait 2 seconds.
• Remove the pipette tip from the liquid.
• To dispense the liquid, place the pipette tip on the inside wall of the container.
• Slowly press the button down to the first stop.
• Wait until all of the liquid has been dispensed.
• Press the button down to the second stop, this will allow any remaining liquid to be dis-
pensed. First stop Second stopOriginal position
Minipipette Instructions:
Tips for an Accurate Measurement
Battery Management
10
mL
Functional Description
Errors and Warnings
Cover
Cuvette with cap
Cuvette holder
LCD Liquid Crystal Display
Operation Button
Under Range: A blinking “200” indicates that the reading is under
range.
Over Range: A blinking value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Inverted Cuvette: The sample and the zero cuvette are inverted.
Low Battery: The battery is low and must be replaced soon.
Dead Battery: The battery is dead and must be replaced. Once this
indication is displayed, normal operation of the instrument will be
interrupted. Change the battery and restart the meter.
Light High: There is too much light to perform a measurement. Please
check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement.
Please check the preparation of the zero cuvette.
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please contact your local Hanna Instruments Office.
Each HI758 meter is supplied complete with:
• Sample Cuvettes and Caps (2 pcs.)
• Reagents for 25 tests
• 1 mL syringe with tip
• Minipipette with tips
• Plastic pipette
• 1.5V AAA battery
• Instruction manual and Quick Reference Guide
Preliminary Examination
Range 200 to 600 ppm
Resolution 1 ppm
Accuracy ±6 ppm ±5% of reading @25 °C/77 °F
Light Source Light Emitting Diode @610 nm
Light Detector Silicon Photocell
Method Adaptation of the zincon method.
Environment 0 to 50 °C (32 to 122 °F); max 95% RH non-condesing
Battery Type 1.5V AAA (1 pc.)
Auto-Shut off After 10 minutes of non-use
Dimensions 86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)
Weight 52 g (1.84 oz.)
Specifications
Thank You
Thank you for choosing a Hanna Instruments product. Please read this instruction
manual carefully before using the instrument.
For more information about Hanna Instruments and our products, visit
www.hannainst.com.
For technical support, contact your local Hanna Instruments Office or e-mail us at
[email protected].
Find your local Hanna Instruments Office on www.hannainst.com.
Measurement Procedure
Hanna Instruments reserves the right to modify the design, construction or appearance of
its products without advance notice.
• Turn the meter on by pressing the button.
All segments will be displayed. When the
display shows “Add”, “C.1” with “Press”
blinking, the meter is ready.
• Add 1 mL of HI7581 Reagent A to the
cuvette using the 1 mL syringe. Use
the plastic pipette to fill the cuvette
to the 10 mL mark with deionized
water and replace the cap. Invert
3-5 times to mix.
• Place the cuvette into the meter and close
the meter’s cap. Press the button. When the
display shows “Add”, “C2” with “Press”
blinking the meter is zeroed.
• Remove the cuvette from the meter and
unscrew the cap. Use the minipipette to
add 0.1 mL of sample to the cuvette.
• Add 1 packet of HI7582 Reagent B.
Replace the cap and shake vigorously for
15 seconds or until the powder is completely
dissolved. Allow air bubbles to dissipate for
15 seconds before taking a reading.
• Place the cuvette back into the meter and
close the meter’s cap. Press the button.
The meter displays concentration in ppm
of calcium. The meter automatically turns
off after 2 minutes.

Other manuals for HI 758

3

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI 749 User manual

Hanna Instruments

Hanna Instruments HI 749 User manual

Hanna Instruments HI 96108 User manual

Hanna Instruments

Hanna Instruments HI 96108 User manual

Hanna Instruments HI8424NEW User manual

Hanna Instruments

Hanna Instruments HI8424NEW User manual

Hanna Instruments HI 255 User manual

Hanna Instruments

Hanna Instruments HI 255 User manual

Hanna Instruments HI736 User manual

Hanna Instruments

Hanna Instruments HI736 User manual

Hanna Instruments HI 250 User manual

Hanna Instruments

Hanna Instruments HI 250 User manual

Hanna Instruments HI733 User manual

Hanna Instruments

Hanna Instruments HI733 User manual

Hanna Instruments HI727 User manual

Hanna Instruments

Hanna Instruments HI727 User manual

Hanna Instruments HI96733C User manual

Hanna Instruments

Hanna Instruments HI96733C User manual

Hanna Instruments pH 210 User manual

Hanna Instruments

Hanna Instruments pH 210 User manual

Hanna Instruments HI 93719 User manual

Hanna Instruments

Hanna Instruments HI 93719 User manual

Hanna Instruments Checker HI716 Instruction Manual

Hanna Instruments

Hanna Instruments Checker HI716 Instruction Manual

Hanna Instruments HI84531 User manual

Hanna Instruments

Hanna Instruments HI84531 User manual

Hanna Instruments HI 99141 User manual

Hanna Instruments

Hanna Instruments HI 99141 User manual

Hanna Instruments HI96747C User manual

Hanna Instruments

Hanna Instruments HI96747C User manual

Hanna Instruments HI 96745 User manual

Hanna Instruments

Hanna Instruments HI 96745 User manual

Hanna Instruments HI6421 User manual

Hanna Instruments

Hanna Instruments HI6421 User manual

Hanna Instruments HI753 User manual

Hanna Instruments

Hanna Instruments HI753 User manual

Hanna Instruments HI 83740 User manual

Hanna Instruments

Hanna Instruments HI 83740 User manual

Hanna Instruments HI 96770 User manual

Hanna Instruments

Hanna Instruments HI 96770 User manual

Hanna Instruments HI 96762C User manual

Hanna Instruments

Hanna Instruments HI 96762C User manual

Hanna Instruments HI 720 User manual

Hanna Instruments

Hanna Instruments HI 720 User manual

Hanna Instruments HI96713C User manual

Hanna Instruments

Hanna Instruments HI96713C User manual

Hanna Instruments HI 762 User manual

Hanna Instruments

Hanna Instruments HI 762 User manual

Popular Measuring Instrument manuals by other brands

ORTEC 439 operating manual

ORTEC

ORTEC 439 operating manual

Metrol TFM1186 user manual

Metrol

Metrol TFM1186 user manual

Extech Instruments EX845 user guide

Extech Instruments

Extech Instruments EX845 user guide

Keithley 660A instruction manual

Keithley

Keithley 660A instruction manual

Getinge SafeStep Operator's manual

Getinge

Getinge SafeStep Operator's manual

MEI KC-326B instructions

MEI

MEI KC-326B instructions

BC Biomedical SPO-2000 user manual

BC Biomedical

BC Biomedical SPO-2000 user manual

Shortridge Instruments ADM-860C operating instructions

Shortridge Instruments

Shortridge Instruments ADM-860C operating instructions

Datakom DKM-407 manual

Datakom

Datakom DKM-407 manual

Bushnell PRIME 1500 owner's guide

Bushnell

Bushnell PRIME 1500 owner's guide

Acon Mission user manual

Acon

Acon Mission user manual

Ecolab Lubo-DryExx Operating	 instruction

Ecolab

Ecolab Lubo-DryExx Operating instruction

Trotec BX09 operating manual

Trotec

Trotec BX09 operating manual

Bosch GCL 25 Professional Original instructions

Bosch

Bosch GCL 25 Professional Original instructions

Cellbox Ground 2.0 Technical manual

Cellbox

Cellbox Ground 2.0 Technical manual

Femtochrome FR-103MN instruction manual

Femtochrome

Femtochrome FR-103MN instruction manual

PCE Instruments PCE-PMI 2 manual

PCE Instruments

PCE Instruments PCE-PMI 2 manual

Dräger PointGard 2000 Series Instructions for use

Dräger

Dräger PointGard 2000 Series Instructions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.