Hanna Instruments HI 93750 User manual

Instruction Manual
HI 93750
Potassium ISM
www.hannainst.com ThisInstrumentisin
Compliancewith the CE Directives
HannaInstrumentsreservestherighttomodifythedesign,construction
andappearanceofitsproductswithoutadvancenotice.
WARRANTYWARRANTY
WARRANTYWARRANTY
WARRANTY
HI93750iswarrantedfor two years against defects in workman-
shipand materialswhenusedforits intendedpurposeandmaintained
accordingtoinstructions.
Thiswarranty islimitedto repairorreplacement freeofcharge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenanceare not covered.
Ifserviceis required,contactthedealerfromwhom youpurchasedthe
instrument.Ifunderwarranty, report the model number, dateofpur-
chase,serialnumberandthe nature of the failure. Iftherepairisnot
coveredby the warranty,youwillbe notified ofthechargesincurred. If
theinstrumentistobereturnedtoHannaInstruments,firstobtaina
ReturnedGoodsAuthorizationNumber from the Customer Service de-
partmentand thensenditwith shipmentcostsprepaid.Whenshipping
anyinstrument, makesureitisproperly packagedforcompleteprotec-
tion.
Tovalidate yourwarranty,filloutandreturntheenclosedwarrantycard
within14 daysfromthedateofpurchase.
Allrightsare reserved. Reproduction inwholeor in part isprohibited
without the written consent of the copyright owner, Hanna Instru-
mentsInc.,Woonsocket, Rhode Island, 02895,USA.
DearCustomer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at [email protected].
This instrument is in compliance with directives
EN 50081-1 and EN 50082-1.
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each Ion Specific Meter is supplied complete with
• 9V Battery
• Two Sample Cuvets and Caps
• One Transport Cap
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
The HI 93750 meter measures the potassium (K) content in
water, wastewater and seawater in the 0.0 to 50.0 mg/L
(ppm) range.
Themeter usesan exclusivepositive-locking systemto ensure
that the cuvet is in the same place every time it is placed
into the measurement cell.
The reagents in liquid form are supplied in bottles and the
reagents in powder are supplied in packets. The amount of
reagent is precisely dosed to ensure maximum repeatability.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Range 0 to 50 mg/L
Resolution 0.05 mg/L from 0 to 9.95 mg/L
0.1 mg/L from 10.0 to 50.0 mg/L
Accuracy ±0.5 mg/L ±5% of reading from 0 to
10 mg/L
±1 mg/L ±5% of reading from 10 to
50 mg/L
TypicalEMC ±0.01 mg/L
Deviation
Light Source Light Emitting Diode @ 470 nm
Method Tetraphenylborate Method
Light Detector SiliconPhotocell
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type/Life1 x 9 volt/40 hours
Auto-Shut off After 10' of non-use
Dimensions 180 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight 290 g (10 oz.).
REQUIRED REAGENTS
Code Description Quantity
HI93750A-0 Potassium Reagent 6 drops
HI93750B-0 Powder Reagent 1 packet
REQUIRED REAGENTS
HI93750-01 Reagents for 100 tests
This indicates that the meter is in a
ready state and zeroing can be per-
formed.
Sampling in Progress. This prompt ap-
pears each time the meter is performing
a measurement.
This indicates that the meter is in a
zeroed state and measurement can be
performed.
A zero reading was not taken. Follow the
instructions for zeroing the meter.
Under range. A blinking "0.00" indicates
that the sample absorbs less light than
the zero reference. Check the procedure
and make sure you use the same cuvet
for reference (zero) and measurement.
Over range. A flashing value higher than
themaximum concentrationreadable (see
specifications) indicates that the sample
absorbs too much light, meaning that
the concentration is too high. Dilute the
sample.
Light over range. The cuvet is not in-
serted correctly and an excess ambient
light is reaching the detector. If the cover
is properly installed, then contact your
dealer or the nearest Hanna Customer
Service Center.
Light under range. The zero sample is
too dark for proper zeroing. If this is not
the case, contact your dealer or the
nearest Hanna Customer Service Center.
The "V" indicates that the battery volt-
age is getting low and the battery needs
to be replaced.
This indicates that the battery is dead
and must be replaced.
Note: once this indication is displayed,
the meter will lockup. Change the bat-
tery to restart.
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDE
V
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
MEASUREMENT PROCEDURE
• Turn the meter on by pressing ON/OFF.
• When the LCD displays “- - -”, it is
ready.
• Fill the cuvet with 10 mL of unre-
acted sample up to the mark.
• Add six drops of HI 93750A-0, replace the cap and
swirl the solution.
• Place the cuvet into the holder
and ensure that the notch on
the cap is positioned securely
into the groove.
• Press ZERO and “SIP” will appear on the display.
• Wait for a few seconds and the
display will show “-0.0-”. Now the
meter is zeroed and ready for measurement.
• Remove the cuvet.
ON
OFF
ZERO
• Add one packet of HI 93750B-0 reagent. Replace the
cap and mix very gently for one minute by turning the
cuvet upside down.
• Reinsert the cuvet into the instru-
ment.
• Press READ TIMED and the display will show the count-
down prior to the measurement or, alternatively, wait
for 2 minutes and press READ DIRECT. In both cases
“SIP” will appear during measurement.
• The instrument directly displays concentration in mg/L
(ppm) of potassium (K).
INTERFERENCES
The following ions do not interfere below the concentrations
reported:
Ammonium 10 ppm
Calcium 10000 ppm as CaCO3
Chloride 12000 ppm
Magnesium 8000 ppm as CaCO3
Sodium 8000 ppm
Color or suspended matter in large amounts will cause
interference. Therefore the turbidity should be removed by
prior filtration.
For best results samples should be tested at room tempera-
ture (20°C or 68°F). Lower temperatures will result in
higher readings.
The instruction listed below should be carefully followed
during testing to ensure best accuracy.
• Do not touch the cuvet walls with hands.
• In order to maintain the same conditions during the
zeroing and the measuring phases, it is necessary to
close the cuvet to prevent any contamination.
• Do not let the test sample stand too long after reagent
is added or accuracy will be lost.
• Whenever the cuvet is placed into the measurement cell,
it must be completely free of fingerprints, oil or dirt.
Wipe it thoroughly with HI 731318 or a lint-free cloth
prior to insertion.
• It is important that the sample does not contain any
debris. This would corrupt the readings.
• It is possible to take multiple readings in a row, but it
is recommended that a zero reading be taken for each
sample and that the same cuvet is used for zeroing and
measurement.
• It is important to discard the sample immediately after
the reading is taken because the glass might become
permanently stained.
• Shaking the cuvet can generate bubbles in the sample,
causing higher readings. To obtain accurate measure-
ments, remove such bubbles by swirling or by gently
tapping the vial.
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATE
MEASUREMENTMEASUREMENT
MEASUREMENTMEASUREMENT
MEASUREMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
Battery replacement must only take place in a non-hazard-
ous area using a 9V alkaline battery.
Simply slide off the battery cover on the back of the meter.
Detach the battery from the terminals and attach a fresh
9V battery while paying attention to the correct polarity.
Replace the battery and the cover.
9V
SLID
E
OFF
RecommendationsforUsers
Before using these products, make sure that they are entirely suitable for the environ-
mentinwhichtheyareused.
Operationoftheseinstrumentsinresidentialareacouldcauseunacceptableinterferences
toradioandTV equipments, requiring the operatortotakeallnecessary steps to correct
interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments'EMCperformance.
Toavoiddamagesorburns,donotperformanymeasurementinmicrowaveovens.
ACCESSORIESACCESSORIES
ACCESSORIESACCESSORIES
ACCESSORIES
REAGENT SETS
HI93750-01 Reagents for 100 tests
OTHER ACCESSORIES
HI710009 Blue rubber boot
HI710010 Orange rubber boot
HI721310 9V battery (10 pcs)
HI731318 Tissue for wiping cuvets (4 pcs)
HI731321 Glass cuvets (4 pcs)
HI731325 Caps for cuvets (4 pcs)
HI93703-50 Cuvets cleaning solution (230 mL).
CE DECLARATIONCE DECLARATION
CE DECLARATIONCE DECLARATION
CE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
IST93750 09/99
PRINTED IN PORTUGAL
READ
TIMED

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI97701 User manual

Hanna Instruments

Hanna Instruments HI97701 User manual

Hanna Instruments HI 91410 User manual

Hanna Instruments

Hanna Instruments HI 91410 User manual

Hanna Instruments HI96786 User manual

Hanna Instruments

Hanna Instruments HI96786 User manual

Hanna Instruments HI 755 User manual

Hanna Instruments

Hanna Instruments HI 755 User manual

Hanna Instruments Color of Water HI 727 User manual

Hanna Instruments

Hanna Instruments Color of Water HI 727 User manual

Hanna Instruments HI 991002 User manual

Hanna Instruments

Hanna Instruments HI 991002 User manual

Hanna Instruments HI981400 User manual

Hanna Instruments

Hanna Instruments HI981400 User manual

Hanna Instruments HI96751C User manual

Hanna Instruments

Hanna Instruments HI96751C User manual

Hanna Instruments HI 707 User manual

Hanna Instruments

Hanna Instruments HI 707 User manual

Hanna Instruments HI 96706C User manual

Hanna Instruments

Hanna Instruments HI 96706C User manual

Hanna Instruments HI 96832 User manual

Hanna Instruments

Hanna Instruments HI 96832 User manual

Hanna Instruments HI97750 User manual

Hanna Instruments

Hanna Instruments HI97750 User manual

Hanna Instruments HI 758 User manual

Hanna Instruments

Hanna Instruments HI 758 User manual

Hanna Instruments HI 991401 User manual

Hanna Instruments

Hanna Instruments HI 991401 User manual

Hanna Instruments HI 121 User manual

Hanna Instruments

Hanna Instruments HI 121 User manual

Hanna Instruments HI 504920 User manual

Hanna Instruments

Hanna Instruments HI 504920 User manual

Hanna Instruments HI 2300 User manual

Hanna Instruments

Hanna Instruments HI 2300 User manual

Hanna Instruments HI 98509 User manual

Hanna Instruments

Hanna Instruments HI 98509 User manual

Hanna Instruments Silica ISM 93705 User manual

Hanna Instruments

Hanna Instruments Silica ISM 93705 User manual

Hanna Instruments HI 96104C User manual

Hanna Instruments

Hanna Instruments HI 96104C User manual

Hanna Instruments HI 747 User manual

Hanna Instruments

Hanna Instruments HI 747 User manual

Hanna Instruments HI 9126 User manual

Hanna Instruments

Hanna Instruments HI 9126 User manual

Hanna Instruments HI 93733 User manual

Hanna Instruments

Hanna Instruments HI 93733 User manual

Hanna Instruments HI 96715C User manual

Hanna Instruments

Hanna Instruments HI 96715C User manual

Popular Measuring Instrument manuals by other brands

NARDA nbm-520 operating manual

NARDA

NARDA nbm-520 operating manual

Kedida CT-6020A instruction manual

Kedida

Kedida CT-6020A instruction manual

Quantum 980 user guide

Quantum

Quantum 980 user guide

M-system 47DV manual

M-system

M-system 47DV manual

TechnipFMC Smith Meter Ultra C Series Replacement manual

TechnipFMC

TechnipFMC Smith Meter Ultra C Series Replacement manual

Draper 34280 Instructions for

Draper

Draper 34280 Instructions for

Keysight Technologies MY53280000 Installation notes

Keysight Technologies

Keysight Technologies MY53280000 Installation notes

AR DC7128A Operating and service manual

AR

AR DC7128A Operating and service manual

Thermo Scientific Orion Star A200 Series SOFTWARE UPDATE INSTRUCTIONS

Thermo Scientific

Thermo Scientific Orion Star A200 Series SOFTWARE UPDATE INSTRUCTIONS

AutoAqua TDS-100 manual

AutoAqua

AutoAqua TDS-100 manual

U-F-M KATflow 230 operating instructions

U-F-M

U-F-M KATflow 230 operating instructions

FLIR CM57-2 quick start

FLIR

FLIR CM57-2 quick start

socomec DIRIS A80 operating instructions

socomec

socomec DIRIS A80 operating instructions

Multi 200 instruction manual

Multi

Multi 200 instruction manual

Milltronics MultiRanger Plus instruction manual

Milltronics

Milltronics MultiRanger Plus instruction manual

Ex-baltic BA444DF-P PROFIBUS PA instructions

Ex-baltic

Ex-baltic BA444DF-P PROFIBUS PA instructions

Rhosonics 8500 manual

Rhosonics

Rhosonics 8500 manual

Storch ProDist compact plus Original instruction

Storch

Storch ProDist compact plus Original instruction

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.