ELECTRODE MAINTENANCE
PREPARATION
• Remove the protective cap. DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. Rinse with water.
• Shake the electrode down as you would do with a clinical thermometer to eliminate any air bubbles
inside the glass bulb.
• Remove the fill hole cover to ensure the reference junctions are flowing. Set aside for storage.
• Top off the electrolyte filling solution using HI7071 solution.
• If the bulb and/or junction are dry, soak the electrode in HI70300 Storage Solution for at least
30 minutes.
• Rinse with water.
• Calibrate before using.
Loss of shielding due to low electrolyte level: Empty electrolyte with a syringe and refill with fresh HI7071.
STORAGE
• To ensure a quick response, the glass bulb and the junction should be kept moist and not allowed to dry.
• Replace protective cap with a few drops of HI70300 Storage Solution. Follow PREPARATION above
before taking measurements.
• Replace the fill hole cover.
Note: NEVER STORE THE ELECTRODE IN DISTILLED WATER.
PERIODIC MAINTENANCE
• INSPECT the electrode for any scratches or cracks. If any present, replace the electrode.
• Rinse off any salt deposits with water.
• Follow the STORAGE procedure above.
CLEANING PROCEDURE
A frequent cleaning of the pH electrode is strongly recommended to ensure correct calibration and reliable
readings.
Hanna Instruments has developed a complete series of cleaning solutions dedicated to specific applications
and kind of dirt that has to be removed from the electrode.
In soil measurements you can choose between two different solutions accordingly to the type of tested soil:
• HI700663 is indicated for inorganic soil deposits (as minerals, limestone, adsorbed clays)
• HI700664 is specific for organic soil deposits (humus)
If cleaning is performed frequently, soak the electrode in the specific solution for a few minutes.
If the electrode has not be cleaned for a while, for a complete removal of soil deposits, proceed as follows:
• wipe the electrode body (not bulb) with paper or soft tissue
• rub the reference with abrasive paper
• immerse into cleaning solution for at least 20 minutes
• rinse the sensor with water and place in storage solution for a minimum of 30 minutes recalibrate
the probe before using.
IMPORTANT: After performing any of the cleaning procedures rinse the electrode thoroughly with
distilled water. Soak the electrode in HI70300 storage solution for at least of 30 minutes. Rinse with
water. Calibrate before using.
TROUBLESHOOTING
• pH Meter: Follow the meter’s operating and calibration procedures.
• Electrode: Evaluate your electrode performance based on the SENSOR CHECK procedure on page 10.