Hanna Instruments HI709 User manual

HI709
Manganese High Range
INSTRUCTION MANUAL
Before using Hanna Instruments products, make sure that they are entirely suitable for
your specific application and for the environment in which they are used. Operation of
these instruments may cause unacceptable interferences to other electronic equipment, thus
requiring the operator to take all necessary steps to correct such interferences. Any variation
introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance. To avoid damages or burns, do not put the instrument in microwave oven. For
yours and the instrument safety do not use or store the instrument in hazardous environments.
Recommendations for Users
Accessories
Reagent Sets
HI709-25 Reagents for 25 Manganese High Range tests
Other Accessories
HI709-11 Manganese HR Certified Standard Kit
HI731225 Cuvette Black Cap for Checker®HC Colorimeters (4 pcs.)
HI731318 Cloth for wiping cuvettes (4 pcs.)
HI731321 Glass cuvettes and Seal Cap for Checker®HC Colorimeters (4 pcs.)
HI731353 Cuvette Seal Cap for Checker®HC Colorimeters (4 pcs.)
HI740028P 1.5V AAA batteries (12 pcs.)
HI93703-50 Cuvette cleaning solution (230 mL)
IST709 05/18
HI709 is warranted for a period of one year after date of purchase against defects in
workmanship and materials when used for their intended purpose and maintained according
to instructions. This warranty is limited to repair or replacement free of charge. Damage due
to accidents, misuse, tampering or lack of prescribed maintenance is not covered. If service
is required, contact your local Hanna Instruments Office. If under warranty, report the model
number, date of purchase, serial number and the nature of the problem. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If the instrument is to
be returned to Hanna Instruments, first obtain a Return Goods Authorization number from the
Technical Service department and then send it with shipping costs prepaid. When shipping any
instrument, make sure it is properly packed for complete protection.
Warranty
• It is important that the sample does not contain any debris.
• Whenever the cuvette is placed into the measurement cell, it
must be dry outside, and completely free of fingerprints, oil and
dirt. Wipe it thoroughly with HI731318 or a lint-free cloth prior
to insertion.
• Shaking the cuvette can generate bubbles, causing higher
readings. To obtain accurate measurements, remove such bubbles
by swirling or by gently tapping the cuvette.
• Do not let the reacted sample stand for too long after reagent
is added, as accuracy will be affected.
• After the reading it is important to immediately discard the
sample, otherwise the glass might become permanently stained.
Tips for an Accurate Measurement
To save the battery, the instrument shuts down after 10 minutes of non-use.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the
instrument will display “bAd” then ”bAt” for 1 second and then turns off.
To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
• Turn the instrument off by holding the button until the meter shuts off.
• Turn the instrument upside down and remove the battery cover with a screwdriver.
• Remove the battery from its location and replace it with a new one, inserting the negative
end first.
• Insert the battery cover and replace the screw with a screwdriver.
Battery Management
• Turn the meter on by pressing the button. All
segments will be displayed. When the display
shows “Add”, “C.1” with “Press” blinking,
the meter is ready.
• Fill the cuvette with 10 mL of unreacted
sample and replace the cap. Place the cuvette
into the meter and close the meter’s cap.
• Press the button. When the display shows
“Add”, “C.2” with “Press” blinking the
meter is zeroed.
• Remove the cuvette from the meter and
unscrew the cap. Add the content of one
packet of HI709A-0 reagent. Replace the
cap and shake gently for 2 minutes.
• Unscrew the cap and add the content of one
packet of HI709B-0 reagent. Replace the
cap and invert gently for 2 minutes. Place
the cuvette back into the meter.
• Press and hold the button until the timer is
displayed on the LCD (the display will show
the countdown prior to the measurement)
or, alternatively, wait for 1 minute and
30 seconds and press the button.
• The instrument displays the concentration in
ppm of manganese. The meter automatically
turns off after 10 minutes.
Light High: There is too much light to perform a measurement. Please
check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement. Please
check the preparation of the zero cuvette.
Inverted Cuvettes: The sample and the zero cuvette are inverted.
Under Range: A blinking “0.0” indicates that the sample absorbs less
light than the zero reference. Check the procedure and make sure you use
the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
an over range condition. The concentration of the sample is beyond the
programmed range: dilute the sample and re-run the test.
Battery Low: The battery must be replaced soon.
Dead Battery: This indicates that the battery is dead and must
be replaced. Once this indication is displayed, normal operation of
the instrument will be interrupted. Change the battery and restart
the meter.
Thank You
Thank you for choosing a Hanna Instruments product. Please read this instruction
manual carefully before using the instrument.
For more information about Hanna Instruments and our products, visit
www.hannainst.com.
For technical support, contact your local Hanna Instruments Office or e-mail us at
[email protected].
Find your local Hanna Instruments Office on www.hannainst.com.
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please contact your local Hanna Instruments Office.
Each HI709 meter is supplied complete with:
• Sample Cuvettes and Caps (2 pcs.)
• Reagents for 6 tests
• 1.5V AAA Battery (1 pc.)
• Instruction Manual and Quick Reference Guide
Preliminary Examination
Range 0.0 to 20.0 ppm
Resolution 0.1 ppm
Accuracy ±0.2 ppm ±5% of reading @25 °C/77 °F
Light Source Light Emitting Diode @525 nm
Light Detector Silicon Photocell
Method
Adaptation of the Standard Methods for the Examination of Water
and Wastewater, 18th edition, Periodate method. The reaction
between manganese and the reagent causes a pink tint in the
sample.
Environment 0 to 50 °C (32 to 122 °F); max 95% RH non-condesing
Battery Type 1.5V AAA (1 pc.)
Auto-Shut off After 10 minutes of non-use
Dimensions 86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)
Weight 52 g (1.84 oz.)
Specifications
Cover
Cuvette with cap
Cuvette holder
LCD Liquid Crystal Display
Operation Button
Functional Description
Errors and Warnings
Measurement Procedure
Hanna Instruments reserves the right to modify the design, construction or appearance of
its products without advance notice.

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI 83749 User manual

Hanna Instruments

Hanna Instruments HI 83749 User manual

Hanna Instruments HI 98140 User manual

Hanna Instruments

Hanna Instruments HI 98140 User manual

Hanna Instruments HI96729 User manual

Hanna Instruments

Hanna Instruments HI96729 User manual

Hanna Instruments HI 9033 User manual

Hanna Instruments

Hanna Instruments HI 9033 User manual

Hanna Instruments HI 709 User manual

Hanna Instruments

Hanna Instruments HI 709 User manual

Hanna Instruments HI 141 Series User manual

Hanna Instruments

Hanna Instruments HI 141 Series User manual

Hanna Instruments HI 93734 User manual

Hanna Instruments

Hanna Instruments HI 93734 User manual

Hanna Instruments HI 720 User manual

Hanna Instruments

Hanna Instruments HI 720 User manual

Hanna Instruments HALO2 HI9810442 User manual

Hanna Instruments

Hanna Instruments HALO2 HI9810442 User manual

Hanna Instruments HI 93702 User manual

Hanna Instruments

Hanna Instruments HI 93702 User manual

Hanna Instruments HI 221 User manual

Hanna Instruments

Hanna Instruments HI 221 User manual

Hanna Instruments HALO 2 HI9810412 User manual

Hanna Instruments

Hanna Instruments HALO 2 HI9810412 User manual

Hanna Instruments 96728 User manual

Hanna Instruments

Hanna Instruments 96728 User manual

Hanna Instruments HI 84102 User manual

Hanna Instruments

Hanna Instruments HI 84102 User manual

Hanna Instruments HI 931101 User manual

Hanna Instruments

Hanna Instruments HI 931101 User manual

Hanna Instruments Alkalinity HI775 User manual

Hanna Instruments

Hanna Instruments Alkalinity HI775 User manual

Hanna Instruments HI 96715C User manual

Hanna Instruments

Hanna Instruments HI 96715C User manual

Hanna Instruments HI 83746 User manual

Hanna Instruments

Hanna Instruments HI 83746 User manual

Hanna Instruments HI 84102 User manual

Hanna Instruments

Hanna Instruments HI 84102 User manual

Hanna Instruments HI981954 User manual

Hanna Instruments

Hanna Instruments HI981954 User manual

Hanna Instruments HI 96715C User manual

Hanna Instruments

Hanna Instruments HI 96715C User manual

Hanna Instruments Checker Marine HI784 User manual

Hanna Instruments

Hanna Instruments Checker Marine HI784 User manual

Hanna Instruments HI701 User manual

Hanna Instruments

Hanna Instruments HI701 User manual

Hanna Instruments HALO FC2142 User manual

Hanna Instruments

Hanna Instruments HALO FC2142 User manual

Popular Measuring Instrument manuals by other brands

NARDA nbm-520 operating manual

NARDA

NARDA nbm-520 operating manual

Kedida CT-6020A instruction manual

Kedida

Kedida CT-6020A instruction manual

Quantum 980 user guide

Quantum

Quantum 980 user guide

M-system 47DV manual

M-system

M-system 47DV manual

TechnipFMC Smith Meter Ultra C Series Replacement manual

TechnipFMC

TechnipFMC Smith Meter Ultra C Series Replacement manual

Draper 34280 Instructions for

Draper

Draper 34280 Instructions for

Keysight Technologies MY53280000 Installation notes

Keysight Technologies

Keysight Technologies MY53280000 Installation notes

AR DC7128A Operating and service manual

AR

AR DC7128A Operating and service manual

Thermo Scientific Orion Star A200 Series SOFTWARE UPDATE INSTRUCTIONS

Thermo Scientific

Thermo Scientific Orion Star A200 Series SOFTWARE UPDATE INSTRUCTIONS

AutoAqua TDS-100 manual

AutoAqua

AutoAqua TDS-100 manual

U-F-M KATflow 230 operating instructions

U-F-M

U-F-M KATflow 230 operating instructions

FLIR CM57-2 quick start

FLIR

FLIR CM57-2 quick start

socomec DIRIS A80 operating instructions

socomec

socomec DIRIS A80 operating instructions

Multi 200 instruction manual

Multi

Multi 200 instruction manual

Milltronics MultiRanger Plus instruction manual

Milltronics

Milltronics MultiRanger Plus instruction manual

Ex-baltic BA444DF-P PROFIBUS PA instructions

Ex-baltic

Ex-baltic BA444DF-P PROFIBUS PA instructions

Rhosonics 8500 manual

Rhosonics

Rhosonics 8500 manual

Storch ProDist compact plus Original instruction

Storch

Storch ProDist compact plus Original instruction

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.