Hanna Instruments Nitrite Low HI 707 User manual

Nitrite Low Range Preliminary examination:
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please notify your Dealer.
Each HI 707 meter is supplied complete with:
• Two Sample Cuvettes and Caps
• Six packets of Nitrite LR Reagent
• 1 x 1.5V AAA Battery
• Instruction Manual
HI 707
Functional description:
Errors and warnings:
Light High: There is too much light to perform a measurement.
Please check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement. Please
check the preparation of the zero cuvette.
Inverted Cuvettes: The sample and the zero cuvette are inverted.
Under Range: A blinking “0” indicates that the sample absorbs less
light than the zero reference. Check the procedure and make sure you
use the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Battery Low: The battery must be replaced soon.
Dead Battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the
meter.
1. Dust cover.
2.Cuvette with cap.
3.Cuvette holder.
4.Liquid Crystal Display.
5.Button.
For more details about spare parts and accessories
see “Accessories”.
www.hannainst.com
Dear Customer,
Thank you for choosing a Hanna Instruments Product.
Please read this instruction manual carefully before using the instrument. If you need
additional technical information, do not hesitate to e-mail us at [email protected] or
view our worldwide contact list at www.hannainst.com.
Technical specifications:
Range 0 to 600 ppb
Resolution 1 ppb
Accuracy ±20 ppb ±5% of reading @ 25 °C / 77 °F
Light Source Light Emitting Diode @ 470 nm
Light Detector Silicon Photocell
Method Adaptation of the
EPA Diazotization method 354.1
. The
reaction between nitrite and the reagent causes a pink tint
in the sample.
Environment 0 to 50 °C (32 to 122 °F);
max 95% RH non-condensing
Battery Type 1 x 1.5V AAA
Auto-Shut off After 10 minutes of non-use
Dimensions 81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5”)
Weight 64 g (2.25 oz.)
[email protected]
www.ponpe.com
•Turn the meter on by pressing the button. All
segments will be displayed. When the display
shows “Add”, “C.1” with “Press” blinking,
the meter is ready.
•Fill the cuvette with 10 mL of unreacted
sample and replace the cap. Place the cuvette
into the meter and close the meter’s cap.
•Press the button. When the display shows
“Add”, “C.2” with “Press” blinking the meter
is zeroed.
•Remove the cuvette from the meter and
unscrew the cap. Add the content of one packet
of HI 707-0 reagent. Replace the cap and
shake gently for 15 seconds. Place the cuvette
back into the meter.
•Press and hold the button until the timer is
displayed on the LCD (the display will show
the countdown prior to the measurement) or,
alternatively, wait for 15 minutes and press
the button.
Note: Between 15 and 10 minutes the timer
will be displayed in minutes and under
10 minutes timer is expressed in minutes
and seconds.
•The instrument displays concentration in ppb
of NO2¯-N.
•To convert the displayed concentration to the
nitrite ion concentration (NO2¯), multiply the
reading by 3.29. To convert the NO2¯-N
concentration to sodium nitrite (NaNO2)
multiply the reading by 4.93. The meter
automatically turns off after 10 minutes.
10 mL
•It is important that the sample does not contain any debris.
•Whenever the cuvette is placed into the measurement cell, it must be dry
outside, and completely free of fingerprints, oil and dirt. Wipe it thoroughly with
HI 731318 or a lint-free cloth prior to insertion.
•Shaking the cuvette can generate bubbles, causing higher readings. To obtain
accurate measurements, remove bubbles by swirling or by gently tapping the
cuvette.
•Do not let the reacted sample stand for too long after reagent is added, as
accuracy will be affected.
•After the reading it is important to immediately discard the sample, otherwise
the glass might become permanently stained.
REAGENT SETS
HI 707-25 Reagent for 25 Nitrite Low Range tests
OTHER ACCESSORIES
HI 707-11 Nitrite LR Certified Standard Kit
HI 731318 Cloth for wiping cuvettes (4 pcs.)
HI 731321 Glass cuvettes (4 pcs.)
HI 731225 Cuvette black cap for checker HC (4 pcs.)
HI 731353 Cuvette seal cap for checker HC (4 pcs.)
HI 740028 1.5V AAA batteries (4 pcs.)
HI 93703-50 Cuvette cleaning solution (230 mL)
Recommendations for Users
Before using this product, make sure that it is entirely suitable for your specific application and for the
environment in which it is used.
Operation of this instrument may cause unacceptable interferences to other electronic equipment,
thus requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damage or burns, do not put the instrument in microwave oven. For the safety of you and the
instrument do not use or store the instrument in hazardous environments.
For additional information, contact your
dealer or the nearest
Hanna Customer Service Center.
To find a Hanna Office in your area,
visit our web site:
www.hannainst.com
Hanna Instruments reserves the right to modify the design, construction, or
appearance of its products without advance notice.
Battery management
To save the battery, the instrument shuts down after 10 minutes of non-use.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the
instrument will display “bAd” then ”bAt” for 1 second and then turns off.
To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
•Turn the instrument off by holding the button until the meter shuts off.
•Turn the instrument upside down and remove the battery cover with a screwdriver.
•Remove the battery from its location and replace it with a new one, inserting the
negative end first.
•Insert the battery cover and replace the screw with a screwdriver.
IST707 03/13
Measurement procedure: Accessories:Tips for an accurate measurement
[email protected]
www.ponpe.com

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments BL 7917 User manual

Hanna Instruments

Hanna Instruments BL 7917 User manual

Hanna Instruments HI 93734 User manual

Hanna Instruments

Hanna Instruments HI 93734 User manual

Hanna Instruments HI 93700 User manual

Hanna Instruments

Hanna Instruments HI 93700 User manual

Hanna Instruments HI 718 User manual

Hanna Instruments

Hanna Instruments HI 718 User manual

Hanna Instruments HI759 User manual

Hanna Instruments

Hanna Instruments HI759 User manual

Hanna Instruments HI 93730 User manual

Hanna Instruments

Hanna Instruments HI 93730 User manual

Hanna Instruments HI97701 User manual

Hanna Instruments

Hanna Instruments HI97701 User manual

Hanna Instruments HI 9142 User manual

Hanna Instruments

Hanna Instruments HI 9142 User manual

Hanna Instruments HI 250 User manual

Hanna Instruments

Hanna Instruments HI 250 User manual

Hanna Instruments EC GroChek User manual

Hanna Instruments

Hanna Instruments EC GroChek User manual

Hanna Instruments HI 91410 User manual

Hanna Instruments

Hanna Instruments HI 91410 User manual

Hanna Instruments HI 99163 User manual

Hanna Instruments

Hanna Instruments HI 99163 User manual

Hanna Instruments HI 747 User manual

Hanna Instruments

Hanna Instruments HI 747 User manual

Hanna Instruments HI 84533 User manual

Hanna Instruments

Hanna Instruments HI 84533 User manual

Hanna Instruments HI 9811-5 User manual

Hanna Instruments

Hanna Instruments HI 9811-5 User manual

Hanna Instruments HI9812-51 User manual

Hanna Instruments

Hanna Instruments HI9812-51 User manual

Hanna Instruments HI 4521 User manual

Hanna Instruments

Hanna Instruments HI 4521 User manual

Hanna Instruments HI 88713 User manual

Hanna Instruments

Hanna Instruments HI 88713 User manual

Hanna Instruments 96711C User manual

Hanna Instruments

Hanna Instruments 96711C User manual

Hanna Instruments HI97739 User manual

Hanna Instruments

Hanna Instruments HI97739 User manual

Hanna Instruments HI97709 User manual

Hanna Instruments

Hanna Instruments HI97709 User manual

Hanna Instruments HI 96770 User manual

Hanna Instruments

Hanna Instruments HI 96770 User manual

Hanna Instruments HI96727 User manual

Hanna Instruments

Hanna Instruments HI96727 User manual

Hanna Instruments HI97721 User manual

Hanna Instruments

Hanna Instruments HI97721 User manual

Popular Measuring Instrument manuals by other brands

SellEton Scales OP-900 Series user manual

SellEton Scales

SellEton Scales OP-900 Series user manual

Siemens SENTRON 7KT PAC1600 Series Equipment manual

Siemens

Siemens SENTRON 7KT PAC1600 Series Equipment manual

Auper TM6400 Installation and programming guide

Auper

Auper TM6400 Installation and programming guide

Lovibond Scuba instruction manual

Lovibond

Lovibond Scuba instruction manual

Greenheck Sure-Aire 386719 Installation, operation and maintenance manual

Greenheck

Greenheck Sure-Aire 386719 Installation, operation and maintenance manual

Bender ISOMETER isoBAT425 manual

Bender

Bender ISOMETER isoBAT425 manual

Tektronix PA4000 Series Manual addendum

Tektronix

Tektronix PA4000 Series Manual addendum

vinmetrica SC-100A user manual

vinmetrica

vinmetrica SC-100A user manual

Veris Industries FSR Series installation guide

Veris Industries

Veris Industries FSR Series installation guide

Siemens SITRANS F Series quick start

Siemens

Siemens SITRANS F Series quick start

Eli Ezer ERK 7800 Operation manual

Eli Ezer

Eli Ezer ERK 7800 Operation manual

truflo TK Series manual

truflo

truflo TK Series manual

VICI CM-2016 Operation

VICI

VICI CM-2016 Operation

KMB SMY 133 operating manual

KMB

KMB SMY 133 operating manual

Rohde & Schwarz RT-ZP05S user manual

Rohde & Schwarz

Rohde & Schwarz RT-ZP05S user manual

UGO BASILE 47200 instruction manual

UGO BASILE

UGO BASILE 47200 instruction manual

Shinkawa VM-7 Series System manual

Shinkawa

Shinkawa VM-7 Series System manual

Agilent Technologies N4373A user guide

Agilent Technologies

Agilent Technologies N4373A user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.