Hanna Foodcare HI99192 User manual

HI99192
pH / Temperature
Meter for Drinking Water
Analysis
INSTRUCTION MANUAL

Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before
using this meter.
This manual will provide you with the necessary
information for correct use of this meter, as well as a
precise idea of its versatility.
If you need additional technical information, do not
our worldwide contact list at www.hannainst.com.
Dear
Customer,
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments
Inc., Woonsocket, Rhode Island, 02895, USA.

3
TABLE OF CONTENTS
Table of Contents......................................................
Preliminary Examination ...........................................
General Description and Intended Use........................
Main Features..........................................................
Specifications...........................................................
Display Description...................................................
Operational Guide ....................................................
Meter Setup.............................................................
Battery Replacement.................................................
Accessories...............................................................
Electrode Maintenance..............................................
Certification..............................................................
Recommendations for Users ......................................
Warranty .................................................................
4
6
5
7
8
19
17
8
14
16
18
15
3
19

4
PRELIMINARY EXAMINATION
Remove the meter from the packing material and examine
it carefully to make sure that no damage has occurred
during shipment. If noticeable damage is evident, contact
your local Hanna Instruments Office.
Each meter is supplied complete with:
• FC2153 pH electrode with internal temperature sensor,
with DIN connector and 1 m (3.3’) cable
• pH 4.01 & 7.01 Buffer sachet
• HI700601 general electrode cleaning solution
(2 sachets)
• 100 mL beaker (1 pcs.)
• Alkaline batteries: 1.5V AAA (3 pcs.)
• Rugged carrying case
• Calibration certificate of meter
• Calibration certificate of probe
• Instruction manual
Note: Save all packing material until you are sure that
the meter functions correctly. All defective items must
be returned in the original packing together with the
supplied accessories.

5
GENERAL DESCRIPTION and INTENDED USE
HI99192 is a portable, lightweight pH meter with two
button operation that is simple to use. Features include a
waterproof and compact casing, large dual‑line display,
low battery detection, automatic pH calibration at one or
two points and selectable temperature unit (°C or °F). The
HI99192 meter comes with a FC2153 pH which is, de‑
signed specifically for measuring the pH of potable waters.
The pair are ideal for on‑site spot checks of drinking water.
The pH of potable water is fundamental to ensure safe water
quality. If the pH is too low, drinking water will be corrosive
to the distribution system and water pipes in homes. If it is
too high, it can reduce the effectiveness of disinfectants. The
pH of water also influences aesthetic or cosmetic properties
including taste, odor, and, clarity. Most public water opera‑
tions maintain pH between 6.5 and 8.5.
Many factors affect pH measurements. An accurate tem‑
perature measurement and temperature compensated pH,
along with at least two calibrated pH standards will improve
the quality of the pH measurement. Often pH electrodes
respond sluggishly in potable water samples due to the na‑
ture of the water, the sample temperature or the design of
the electrode.
The HI99192 together with the FC2153 pH electrode solves
all the problems found with standard pH systems. This spe‑
cialized electrode offers numerous features that improve pH
testing in drinking water. The spherical pH bulb features a
low resistance pH glass that responds quickly to the sample
(even at cold temperatures). It also has a refillable single
junction Ag/AgCl reference that is used with a KCl electrolyte
and has three ceramic junctions to ensure continuity and
provide quick and reproducible measurements (even in low
ionic strength waters).
FC2153 pH electrode has a built‑in temperature sensor for
temperature compensated pH and temperature readings
and contains an integral preamplifier to provide stable
measurements in electrically noisy environments.
.

6
Main features:
• Simultaneous pH and temperature measurements on a
large dual‑line LCD display
• Automatic pH calibration at one or two points within
two memorized buffer sets (standard or NIST)
• Selectable temperature unit (°C or °F)
• Electrode condition indicator
• mV of pH measurement for electrode check
• FC2153 dedicated pH with integral temperature sensor
• Probe quick connect system
• Battery life indication and low battery detection
• Keystroke confirmation tone
• Auto‑off function
• Waterproof casing IP67
MAIN FEATURES

7
SPECIFICATIONS
Range*
‑2.00 to 16.00 pH / ‑2.0 to 16.0 pH
±825 mV (pH‑mV)
‑5.0 to 105.0 °C/23.0 to 221.0 °F
Resolution
0.01 pH / 0.1 pH
1 mV
0.1 °C/0.1 °F
Accuracy
@ 20°C / 68°F
±0.02 pH / ±0.1 pH
±1 mV (pH‑mV)
±0.5 °C up to 60 °C; ±1.0 °C outside
±1.0 °F up to 140 °F; ±2.0 °F outside
Temperature
compensation
Automatic
‑5.0 to 105.0 °C/23.0 to 221.0 °F
pH Calibration
Automatic, 1 or 2 point selectable buffer
set Standard: 4.01;7.01;10.01 or
NIST: 4.01;6.86;9.18
Probe (included)
FC2153 pH electrode with internal
temperature sensor, with DIN connector
and 1 m (3.3’) cable
Battery type / life 1.5V AAA (3 pcs.)
approx. 1400 hours of continuous use
Auto‑Off user selectable: after 8 min, 60 min or
disabled
Environment 0 to 50 °C (32 to 122 °F)
RH max. 100%
Meter Dimensions 154 x 58 x 30 mm (6.0 x 2.3 x 1.2”)
Meter Mass
(with batteries) 205 g (7.2 oz.)
Case Ingress
Protection Rating IP67
* the FC2153 is limited to be used from 0 to 12 pH and
from 0 to 70 °C temperature (32 to 158 °F).

8
OPERATIONAL GUIDE
1. Automatic Temperature Compensation indicator
2. Stability indicator
3. Battery percentage
4. Electrode condition indicator
5. Low battery indicator
6. Measurement unit
7. Primary LCD
8. Temperature unit
9. Secondary LCD
10. Meter modes indicator
11. pH calibration buffer(s) used
Each meter is supplied with batteries. Before using the
meter for the first time, open the battery compartment
and insert batteries, observing the polarity (see “Battery
Replacement“).
TO CONNECT THE ELECTRODE
With the meter turned off, connect the FC2153 probe to the
DIN socket on the bottom of the meter by aligning the pins
and pushing in the plug firmly. Remove the protective cap
from the probe before taking any measurements.
DISPLAY DESCRIPTION

9
OPERATIONAL GUIDE
TO TURN THE METER ON
To turn the meter ON, press the button
on the front of the meter. If it does not turn
on, make sure that the batteries are properly installed
in their place. The meter is provided with an active
acoustic signal when a key is pressed. At start‑up the
meter displays all LCD segments for a few seconds,
followed by the percentage indication of the remaining
battery life, displaying “WAIT“ until electrode check is in
process then the meter enters the normal measurement mode.
Note: The meter detects the presence and the type of the
probe at its input.
- If the probe is not connected the message “NO”
”PROBE“ appears alternatively on the secondary LCD
with “---“ blinking on the first LCD line.
- If the probe is not compatible “WRONG” ”PROBE“
message appears alternatively on the secondary LCD
with “---“ blinking on the first LCD line.
- If the readings are out of range, the nearest range
limits are displayed blinking (E.g. -2.00 pH -5.0 °C).
TO SELECT THE MEASUREMENT RANGE
While in measurement mode, press the SET button to select
pH or pH‑mV measurement on the first LCD line.
TO FREEZE THE MEASUREMENT VALUES
While in measurement mode, press and hold the SET
button until “HOLD“ appears on the secondary LCD. The
“HOLD“ remains for 1 second and reading of pH, mV and
temperature will be frozen on the LCD with “H” blinking.
Battery life
Measurement
mode
LCD
check up
→ →
→
Electrode
check

10
OPERATIONAL GUIDE
→
Press any button to resume active measurements.
TO ENTER CALIBRATION MODE
Press and hold the button until ”POWER” and tag
is replaced by “STD“ and tag. Release the button.
TO ENTER SETUP MODE
Press and hold button until “STD“ and tag
is replaced by “SETUP“ and tag.
Release the button.
TO TURN THE METER OFF
While in measurement mode, press the button.
”POWER” and tag will appear.
Release the button.
pH MEASUREMENT & CALIBRATION
Make sure the meter has been
calibrated before use.
If the probe is dry, soak it in
HI70300 storage solution for
30 minutes to reactivate it. If
fouled, clean the electrode by soak‑
ing in cleaning solution for 20 minutes, then rinse the tip
and soak in storage solution at least 30 minutes before use.
Rinse the electrode off well and shake off excess
droplets. Recalibrate before using. Submerge the probe in
the sample to be tested while stirring it gently. Wait until
the tag on the LCD disappears.
The LCD displays the pH value (automatically compensated
for temperature) on the primary LCD, while the second‑
ary LCD displays the sample temperature. If measurements
are taken in different samples successively, rinse the probe
tip thoroughly†to eliminate cross‑contamination.

11
OPERATIONAL GUIDE
For better accuracy, frequent calibration of the pH sensor
with the meter is recommended. In addition, the meter
must be recalibrated whenever:
a) The pH electrode is replaced.
b) After testing aggressive chemicals.
c) Where high accuracy is required.
d) At least once a month.
e) After cleaning the sensor.
†The probe tip should be rinsed with purified water (reverse osmosis,
distilled, or deionized) before and after placing in any solution (buffer,
storage, or sample).
pH calibration
Enter calibration mode while in
pH measurement mode. Place the
sensor into the first calibration
buffer. If performing a two‑point
calibration, use pH 7.01 (pH 6.86
for NIST) buffer first. The meter will
enter the calibration mode, displaying “pH 7.01 USE”
and tag blinking (or “pH 6.86 USE” for NIST).
Follow directions for single and two‑point calibration below:
Single-point calibration
1. Place the probe in any buffer from the selected buffer set.
The meter will automatically recognize the buffer value.
2. If the buffer is not recognized or the calibration offset is out
of the accepted range “---- WRONG“ is displayed.
3. If the buffer is recognized “REC“ is displayed then
“WAIT“ until the calibration is accepted.
If using pH 7.01 (or pH 6.86 for NIST), after acceptance
of the buffer press any key to exit. The ”SAVE” message is
displayed and the meter returns to pH measurement mode.
If using pH 4.01 or 10.01 (or pH 9.18 for NIST) buffer
the ”SAVE” message is displayed and meter returns to pH
measurement mode.

12
Two-point calibration
Proceed with steps 1 through 3 under single point
calibration using 7.01 (pH 6.86 for NIST) pH buffer
first. Then follow steps below:
The ”pH 4.01 USE” message is then displayed.
Place the probe in the second calibration buffer (pH 4.01
or 10.01, or, if using NIST, pH 4.01 or 9.18). When the
second buffer is accepted, the LCD will display “SAVE”
for 1 second and the meter will return to the normal
measurement mode.
If the buffer is not recognized or the slope is out of accepted
range ”--- WRONG” is displayed. Change the buffer, clean
the electrode or press any key to exit calibration.
It is always recommended to carry out a two‑point
calibration for better accuracy.
When the calibration procedure is completed, the tag
is turned on together with the calibrated points.
To exit calibration and reset default values
After entering the calibration mode and before the first point
is accepted, it is possible to quit the procedure and return
to the last calibration data by pressing the button. The
LCD displays “ESC“ for 1 second and the meter returns to
normal mode.
To reset the default values and clear a previous
calibration, press the SET button after entering the calibration
mode and before the first point is accepted.
The LCD displays “CLEAR“ for 1 second, the meter resets to
the default calibration and the tag with the calibrated
points on the LCD disappears.
OPERATIONAL GUIDE

13
ELECTRODE CONDITION
The display is provided with a probe icon (unless the feature
is disabled from setup) which indicates the electrode status
after calibration. The “condition” remains active for 12 hours
(unless the batteries are removed).
The electrode condition is evaluated only if the current
calibration has two points.
5 bars: excellent condition
4 bars: very good condition
3 bars: good condition
2 bars: fair condition
1 bar: poor condition
1 bar blinking: very poor condition
With 1 bar it is recommended to clean the electrode and
recalibrate. If there is still only 1 bar or 1 bar blinking
replace the probe.
Sensor Check
Setting the meter to pH‑mV range the user can check the
sensor status at any time. The offset value is the reading
in pH 7.01 buffer (@ 25 °C/77 °F). If this reading is
outside the range ±30 mV, the electrode is considered
“very poor”. The slope value of the sensor is the difference
between readings in pH 7.01 and in pH 4.01 buffers. When
the slope reaches the value of about 150 mV, the electrode
is considered “very poor“. When “poor“ or “very poor“, it
is recommended to replace it with a new one.
Note: To ensure reliable readings, the electrode must
be cleaned with cleaning solution and then hydrated in
storage solution for a minimum of 30 minutes before
calibrating the probe.
OPERATIONAL GUIDE

14
METER SETUP
Setup mode allows the selection of the Temperature unit,
Auto‑off, Beep, the type of pH buffer set, the Resolution and
Information. To enter Setup mode press and hold button
until “STD“ and tag is replaced by “SETUP“ and
tag. Release the button.
• “TEMP“ is displayed on the secondary
LCD with the current temperature unit
(E.g. “TEMP °C“),for °C/°F selection,
use the SET button. After the temperature unit has been
selected, press to confirm and to enter the “A-OFF“
selection.
• Use the SET button, to cycle through
the auto‑off choices: 8 minutes (“8”,
default value), 60 minutes (“60”) or
disabled (“---”). Press to confirm
and to enter the “BEEP“ selection.
• To switch ON or OFF the beep
tone, press the SET button; press
to confirm and to enter
the calibration buffer selection
“pH 7.01 BUFF“.
• To change the type of calibration buffer set, the meter
will show the current buffer set:
“pH 7.01 BUFF” (for standard
buffer set: 4.01/7.01/ 10.01)
or “pH 6.86 BUFF” (for NIST
buffer set: 4.01/6.86/ 9.18).
Change the set with the SET button. Press
to confirm and to enter pH resolution selection
“RESOL“.
• To change the pH measurement
resolution between “0.1“ and “0.01“
use the SET button; then press
to confirm and to enter electrode
calibration information “INFO“
selection .

15
METER SETUP
• To switch ON or OFF the electrode
condition indicator on the LCD, press
the SET button; press to exit setup
options; Change the set with the SET
button, then press to confirm and
to return to normal mode.
BATTERY REPLACEMENT
When the remaining battery life is less
than 10% the battery tag blinks on the
display to warn the user.
Battery Error Prevention System (BEPS)
If the battery is too weak (“0%”) the display shows
“bAtt“, “DEAD“ for few seconds then the meter power off.
Immediately replace the batteries with new ones.
The batteries are accessed by opening the battery cover
on the back of the instrument. Remove protective boot
if present.
Replace the three 1.5V AAA alkaline batteries located in
the battery compartment, observing the indicated polarity.
Replace the battery cover
making sure that the gasket
is in place.
BATTERY REPLACEMENT

16
Code and serial
number on the
other side
Temperature
sensor
Quick connector with
Rubber protection, and
1 meter cable
FC2153
pH electrode with internal temperature
sensor, with DIN connector and 1 m
(3.3’) cable
HI7004L pH 4.01 buffer solution, 500 mL
HI7006L pH 6.86 buffer solution, 500 mL
HI7007L pH 7.01 buffer solution, 500 mL
HI7009LpH 9.18 buffer solution, 500 mL
HI7010L pH 10.01 buffer solution, 500 mL
HI70300L pH electrode storage solution, 500 mL
HI7061L pH electrode cleaning solution, 500 mL
HI7071 Electrolyte fill solution, 3.5M KCl +
AgCl (30 mL x 4 pcs.)
HI710029 Silicon rubber boot blue color
HI76405 Electrode holder
ACCESSORIES
3 reference
junctions
pH bulb

17
ELECTRODE MAINTENANCE
PREPARATION
• Remove the protective cap. DO NOT BE ALARMED IF ANY
SALT DEPOSITS ARE PRESENT. Rinse with water.
• Shake the electrode down as you would do with a clini‑
cal thermometer to eliminate any air bubbles inside the
glass bulb.
• Remove the fill hole cover to ensure the reference junctions
are flowing. Set aside for storage.
• Top off the electrolyte filling solution using HI7071 solution.
• If the bulb and/or junction are dry, soak the electrode
in HI70300 Storage Solution for at least 30 minutes.
• Rinse with water.
• Calibrate before using.
Loss of shielding due to low electrolyte level:
Empty electrolyte with a syringe and refill with fresh HI7071.
STORAGE
• To ensure a quick response, the glass bulb and the junction
should be kept moist and not allowed to dry.
• Replace protective cap with a few drops of HI70300
Storage Solution. Follow PREPARATION above before taking
measurements.
• Replace the fill hole cover.
Note: NEVER STORE THE ELECTRODE IN DISTILLED WATER
PERIODIC MAINTENANCE
• INSPECT the electrode for any scratches or cracks. If any
present, replace the electrode.
• Rinse off any salt deposits with water.
• Follow the STORAGE procedure above.
CLEANING PROCEDURE
• Soak in Hanna HI7061 general cleaning solution for
20 minutes. Rinse with water.
IMPORTANT: After performing any of the cleaning proce‑
dures rinse the electrode thoroughly with distilled water. Soak
the electrode in HI70300 Storage Solution for a minimum
of 30 minutes. Rinse with water. Calibrate before using.
TROUBLESHOOTING
• pH Meter: Follow the meter’s operating and calibration
procedures.
• Electrode: Evaluate your electrode performance based on
the SENSOR CHECK procedure on page 13.

18
MAN99192 05/18
All Hanna Instruments conform to the CE European Directives.
Disposal of Electrical & Electronic Equipment. The
product should not be treated as household waste. Instead
hand it over to the appropriate collection point for the
recycling of electrical and electronic equipment which will
conserve natural resources.
Disposal of waste batteries. This product contains
batteries, do not dispose of them with other household
waste. Hand them over to the appropriate collection point
for recycling.
Ensuring proper product and battery disposal prevents
potential negative consequences for the environment and
human health. For more information, contact your city,
your local household waste disposal service, the place
of purchase or go to www.hannainst.com.
RoHS
compliant
CERTIFICATION

19
HI99192 is warranted for two years against
defects in workmanship and materials when
used for its intended purpose and maintained
according to instructions. This warranty is
limited to repair or replacement free of
charge. Damage due to accidents, misuse,
tampering or lack of prescribed maintenance
is not covered.
If service is required, contact your local Hanna
Instruments Office. If under warranty, report
the model number, date of purchase, serial
number (see engraved in the back of the
meter) and the nature of the problem. If the
repair is not covered by the warranty, you will
be notified of the charges incurred. If the meter
is to be returned to Hanna Instruments, first
obtain a Returned Goods Authorization number
from the Technical Service department and
then send it with shipping costs prepaid. When
shipping any meter, make sure it is properly
packed for complete protection.
Warranty
Hanna Instruments reserves the right to modify the design,
construction or appearance of its products without advance notice.
Before using this product, make
sure it is entirely suitable for
your specific application and
for the environment in which it
is used. Any variation introduced
by the user to the supplied
equipment may degrade the
meters’ performance. For yours
and the meter’s safety do not
use or store the meter in haz‑
ardous environments.
Recommendations
for Users

20
Printed in ROMANIA
World
Headquarters
Hanna Instruments Inc.
Highland Industrial Park
584 Park East Drive
Woonsocket, RI 02895 USA
www.hannainst.com
MAN99192
Local Office
Hanna Instruments Inc.
270 George Washington Highway
Smithfield, RI 02917
Phone: 800.426.6287
Fax: 401.765.7575
e‑mail: [email protected]
Table of contents
Other Hanna Measuring Instrument manuals

Hanna
Hanna HI 96785C User manual

Hanna
Hanna HI 9811 User manual

Hanna
Hanna HI 720 User manual

Hanna
Hanna HI759 User manual

Hanna
Hanna HI 96736 User manual

Hanna
Hanna SALINTEST HI98203 User manual

Hanna
Hanna HI 9142 User manual

Hanna
Hanna HI 96745 User manual

Hanna
Hanna HI 5221 User manual

Hanna
Hanna HI 84502 User manual

Hanna
Hanna HI 5522 User manual

Hanna
Hanna HI 991300 User manual

Hanna
Hanna HI 9835 User manual

Hanna
Hanna edge blu User manual

Hanna
Hanna HI981032 User manual

Hanna
Hanna HI981033 User manual

Hanna
Hanna HI 83203 User manual

Hanna
Hanna HI 991300 User manual

Hanna
Hanna HI 110 User manual

Hanna
Hanna HI 98129 User manual