Hanna HI 98129 User manual

www.hannainst.com
Hanna Instruments reserves the right to modify the design, con-
struction and appearance of its products without advance notice.
WARRANTYWARRANTY
WARRANTYWARRANTY
WARRANTY
HI 98129 and HI 98130 are warranted for one
year against defects in workmanship and materials
when used for their intended purpose and maintained
according to instructions. The electrode is war-
ranted for a period of six months. This warranty is
limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack
of prescribed maintenance are not covered.
If service is required, contact the dealer from whom
you purchased the instrument. If under warranty,
report the model number, date of purchase, serial
number and the nature of the failure. If the repair is
not covered by the warranty, you will be notified of
the charges incurred. If the instrument is to be re-
turned to Hanna Instruments, first obtain a Returned
Goods Authorization Number from the Customer
Service department and then send it with shipment
costs prepaid. When shipping any instrument, make
sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the
enclosed warranty card within 14 days from the date
of purchase.
All rights are reserved. Reproduction in whole or in
part is prohibited without the written consent of the
copyright owner, Hanna Instruments Inc., Woonsocket,
Rhode Island, 02895 , USA.
Instruction Manual
HI 98129 - HI 98130
Waterproof pH, EC/TDS &
temperature meters
Dear Customer,
Thank you for choosing a Hanna product. This manual
will provide you with the necessary information for a
correct operation. Please read it carefully before
using the meter. If you need additional technical
information, do not hesitate to e-mail us at
[email protected].
These instruments are in compliance with the
directives.
Remove the instrument from the packing material
and examine it carefully. If any damage has occurred
during shipment, immediately notify your Dealer or
the nearest Hanna Customer Service Center.
The meters are supplied with:
• 4 x 1.5V batteries
• HI 73127 pH electrode
• HI 73128 Tool to remove the pH electrode
Note: Conserve all packing material until the instru-
ment has been observed to function correctly.
Any defective item must be returned in its
original packing.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
HI 98129 and HI 98130 are waterproof pH, EC/TDS
and temperature meters. The housing has been com-
pletely sealed against humidity and designed to float.
All pH and EC/TDS readings are automatically tem-
perature compensated (ATC), and temperature values
can be displayed in °C or °F units.
For EC/TDS readings, the EC/TDS conversion factor
(CONV) is selectable by the user, as well as the
temperature compensation coefficient β(BETA).
The meters can be calibrated at one or two points for
pH (with auto-buffer recognition and against five
memorized buffer values), and at one point for EC/
TDS.
Measurements are highly accurate with a unique
stability indicator right on the LCD.
A low battery symbol warns the user when the batter-
ies are to be replaced. In addition the Battery Error
Prevention System (BEPS) avoids erroneous reading
caused by low voltage level by turning the meter off.
The HI 73127 pH electrode, supplied with the meters,
is interchangeable and can be easily replaced by the
user.
The stainless steel encapsulated temperature sensor
facilitates faster and more accurate temperature mea-
surement and compensation.
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
1. Battery compartment
2. Liquid Crystal Display (LCD)
3. Stability indicator
4. Low battery indicator
5. ON/OFF/MODE button
6. HI 73127 pH electrode
7. Temperature sensor
8. EC/TDS probe
9. SET/HOLD button
10. Secondary LCD
11. Primary LCD
This Instrument is in
Compliance with CE
Directives
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
RangeTemperature: 0.0 to 60.0°C or 32.0 to 140.0°F
HI 98129 pH: 0.00 to 14.00
EC: 0 to 3999 µS/cm
TDS: 0 to 2000 ppm
HI 98130 pH: 0.00 to 14.00
EC: 0.00 to 20.00 mS/cm
TDS: 0.00 to 10.00 ppt
Resolution 0.1°C or 0.1°F
HI 98129 0.01 pH; 1 µS/cm ; 1 ppm
HI 98130 0.01 pH, 0.01 mS/cm ; 0.01 ppt
Accuracy (@20°C/68°F) Temperature ±0.5°C or ±1°F
EC/TDS ±2% f.s.
pH ±0.01
TypicalEMC Deviation Temperature ±0.5°C or ±1°F
pH ±0.02 pH
EC/TDS ±2% f.s.
Temp.Compensation pH: Automatic
EC/TDS: with β=0.0 to 2.4%/ºC
Environment 0 to 50°C (32 to 122°F); RH 100%
EC/TDS Conversion Factor 0.45 to 1.00 (CONV)
Calibration pH: at 1 or 2 points with 2 sets of memo
rized buffers (pH 4.01/7.01/10.01 or
pH 4.01/6.86/9.18)
EC/TDS: automatic, at 1 point
EC/TDS Calibration Solutions
HI 98129 HI7031 (1413 µS/cm)
HI7032 (1382 ppm; CONV=0.5)
HI70442 (1500 ppm; CONV=0.7)
HI 98130 HI7030 (12.88 mS/cm)
HI70038 (6.44 ppt; CONV=0.5
or 9.02 ppt; CONV=0.7)
Electrode HI 73127
pH electrode (includedincluded
includedincluded
included)
BatteryType/Life 4 x 1.5V with BEPS / typical 100 hours
Auto-off After 8 min.
Dimensions 163 x 40 x 26 mm (6.4 x 1.6 x 1.0”)
Weight 85 g (3.0 oz)
Recommendations for Users
Before using this product, make sure that it is entirely suit-
able for the environment in which it is used. Operation of
this instrument in residential areas could cause unaccept-
able interferences to radio and TV equipment.
The glass bulb at the end of the electrode is sensitive to elec-
trostaticdischarges. Avoid touching thisglassbulb at all times.
During operation, ESD wrist straps should be worn to avoid
possible damage to the electrode by electrostatic discharges.
Any variation introduced by the user to the supplied equip-
ment may degrade the instrument’s EMC performance.
To avoid electrical shock, do not use this instrument when
voltages at the measurement surface exceed 24 VAC or 60
VDC. To avoid damages or burns, do not perform any mea-
surement in microwave ovens.
ISTR98129R2 01/02
www.tools.in.th
www.neonics.co.th
Taking measurementsTaking measurements
Taking measurementsTaking measurements
Taking measurements
Select either EC or TDS mode with the SET/HOLD
button.
Submerge the probe in the solution to be tested. Use
plastic beakers to minimize any electromagnetic inter-
ferences.
The measurements should be taken when the stability
symbol on the top left of the LCD disappears.
The EC (or TDS) value automati-
cally compensated for temperature
is shown on the primary LCD
while the secondary LCD shows
the temperature of the sample.
pH ELECTRODEpH ELECTRODE
pH ELECTRODEpH ELECTRODE
pH ELECTRODE MAINTENANCEMAINTENANCE
MAINTENANCEMAINTENANCE
MAINTENANCE
• When not in use, rinse the electrode with water to
minimize contamination and store it with a few drops
of storage (HI 70300) or pH 7 (HI 7007) solution in
the protective cap after use. DO NOT USE DISTILLED
OR DEIONIZED WATER FOR STORAGE PURPOSES.
• If the electrode has been left dry, soak in a storage
or pH 7 solution for at least one hour to reactivate it.
• To prolong the life of the pH electrode, it is
recommended to clean it monthly by immersing it in
the HI 7061 cleaning solution for half an hour.
Afterwards, rinse it thoroughly with tap water and
recalibrate the meter.
• The pH electrode can be easily replaced by using
the supplied tool (HI 73128). Insert the tool into the
probe cavity as shown below.
ToTo
ToTo
To
turnturn
turnturn
turn
thethe
thethe
the
metermeter
metermeter
meter
onon
onon
on
andand
andand
and
toto
toto
to
checkcheck
checkcheck
check
batterybattery
batterybattery
battery
statusstatus
statusstatus
status
Press and hold the
MODE button for 2-3 seconds. All
the used segments on the LCD will be visible for a few
seconds, followed by a percent indication of the
remaining battery life. Eg. % 100 BATT.
ToTo
ToTo
To
change the temperature unitchange the temperature unit
change the temperature unitchange the temperature unit
change the temperature unit
To change the temperature unit (from °C to °F), from
measurement mode, press and hold the MODE but-
ton until TEMP and the current temperature unit are
displayed on the lower LCD. Eg. TEMP °C.
Use the SET/HOLD button to change the temperature
unit, and then press MODE button twice to return to
normal measurement mode.
ToTo
ToTo
To
freezefreeze
freezefreeze
freeze
thethe
thethe
the
displaydisplay
displaydisplay
display
Press the SET/HOLD button for 2-3 seconds until
HOLD appears on the secondary display.
Press either button to return to normal mode.
ToTo
ToTo
To
turnturn
turnturn
turn
thethe
thethe
the
metermeter
metermeter
meter
offoff
offoff
off
Press the
MODE button while in normal measurement
mode. OFF will appear on the lower part of the
display. Release the button.
NotesNotes
NotesNotes
Notes:
• Before taking any measurement make sure the
meter has been calibrated.
• To clear a previous calibration, press the MODE
button after entering the calibration mode. The lower
LCD will display ESC for 1 second and the meter will
return to normal measurement mode. The CAL symbol
on the LCD will disappear. The meter will be reset to
the default calibration.
• If measurements are taken in different samples
successively, rinse the probe thoroughly to eliminate
cross-contamination; and after cleaning, rinse the
probe with some of the sample to be measured.
Taking measurementsTaking measurements
Taking measurementsTaking measurements
Taking measurements
Select the pH mode with the SET/HOLD button.
Submerge the electrode in the solution to be tested.
The measurements should be taken when the stability
symbol on the top left of the LCD disappears.
The pH value automatically com-
pensated for temperature is
shown on the primary LCD
while the secondary LCD shows
the temperature of the sample.
Calibration buffer setCalibration buffer set
Calibration buffer setCalibration buffer set
Calibration buffer set
• From measurement mode, press and hold the
MODE button until TEMP and the current tempera-
ture unit are displayed on the lower LCD. Eg. TEMP
°C.
• Press the MODE button again to show the current
buffer set: pH 7.01 BUFF (for 4.01/7.01/10.01 cali-
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
ACCESSORIESACCESSORIES
ACCESSORIESACCESSORIES
ACCESSORIES
pp
pp
pH MEASUREMENTS & CALIBRATIONH MEASUREMENTS & CALIBRATION
H MEASUREMENTS & CALIBRATIONH MEASUREMENTS & CALIBRATION
H MEASUREMENTS & CALIBRATION
bration) or pH 6.86 BUFF (for NIST 4.01/6.86/9.18
calibration).
• Press the SET/HOLD button to change the buffer
value.
• Press the MODE button to return to normal mea-
suring mode.
Calibration procedureCalibration procedure
Calibration procedureCalibration procedure
Calibration procedure
From measurement mode, press and hold the MODE
button until CAL is displayed on the lower LCD.
Release the button. The LCD will display pH7.01 USE
or pH 6.86 USE (if you have selected the NIST buffer
set). The CAL tag blinks on the LCD.
• For a single-point pH calibration, place the electrode
in any buffer from the selected buffer set (eg. pH 7.01
or pH 4.01 or pH 10.01). The meter will recognize
the buffer value automatically.
If using pH 4.01 or pH 10.01, the meter will display
OK for 1 second and then return to measurement
mode.
If using pH 7.01, after recognition of the buffer the
meter will ask for pH 4.01 as second calibration
point. Press the MODE button to return to measure-
ment mode or, if desired, proceed with the 2-point
calibration as explained below.
NoteNote
NoteNote
Note: It is always recommended to carry out a two-
point calibration for better accuracy.
• For a two-point pH calibration, place the electrode
in pH 7.01 (or 6.86 if you have selected the NIST
buffer set). The meter will recognize the buffer value
and then display pH 4.01 USE.
Rinse the electrode thoroughly to eliminate cross-
contamination.
Place the electrode in the second buffer value (pH
4.01 or 10.01, or, if using NIST, pH 4.01 or 9.18).
When the second buffer is recognized, the LCD will
display OK for 1 second and the meter will return to
normal measurement mode.
The CAL symbol on the LCD means that the meter is
calibrated.
EC/TDSEC/TDS
EC/TDSEC/TDS
EC/TDS MEASUREMENTS & CALIBRATIONMEASUREMENTS & CALIBRATION
MEASUREMENTS & CALIBRATIONMEASUREMENTS & CALIBRATION
MEASUREMENTS & CALIBRATION
To change the EC/TDS conversion factor (CONV)To change the EC/TDS conversion factor (CONV)
To change the EC/TDS conversion factor (CONV)To change the EC/TDS conversion factor (CONV)
To change the EC/TDS conversion factor (CONV)
and the temperature compensation coefficientand the temperature compensation coefficient
and the temperature compensation coefficientand the temperature compensation coefficient
and the temperature compensation coefficient ββ
ββ
β
(BETA)(BETA)
(BETA)(BETA)
(BETA)
• From measurement mode, press and hold the
MODE button until TEMP and the current tempera-
ture unit are displayed on the lower LCD. Eg. TEMP
°C.
• Press the MODE button again to show the current
conversion factor. Eg. 0.50 CONV.
• Press the SET/HOLD button to change the conver-
sion factor.
• Press the MODE button to show the current tem-
perature compensation coefficient β. Eg. 2.1 BETA.
• Press the SET/HOLD button to change the tempera-
ture compensation coefficient β.
• Press the MODE button to return to normal mea-
suring mode.
Calibration procedureCalibration procedure
Calibration procedureCalibration procedure
Calibration procedure
• From measurement mode, press and hold the
MODE button until CAL is displayed on the lower
LCD.
• Release the button and immerse the probe in the
proper calibration solution: HI7031 (1413 µS/cm) for
HI98129 and HI7030 (12.88 mS/cm) for HI98130.
• Once the calibration has been automatically per-
formed, the LCD will display OK for 1 second and the
meter will return to normal measurement mode.
• Since there is a known relathionship between EC
and TDS readings, it is not necessary to calibrate the
meter in TDS
The CAL symbol on the LCD means that the meter is
calibrated.
Remove the probe by rotating it counterclockwise and
then pulling it out.
Insert a new pH electrode following the above instruc-
tions in reverse order.
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
The meter displays the remaining battery percentage
every time it is switched on. When the battery level is
below5%, the
- +
symbol onthe bottomleft of theLCD
lightsupto indicate a lowbattery condition. The batter-
iesmust be immediatelyreplaced. If howeverthe battery
levelis so lowas tocause erroneous reading,the Battery
Error Prevention System (BEPS) will automatically turn
themeter off.
Tochange thebatteries,remove the4screws locatedon
the top of the meter.
Oncethe top hasbeen removed, carefullyreplace the 4
batterieslocated inthecompartment whilepaying atten-
tion to their polarity.
Replacethetop, making sure thatthe gasket is properly
seated in place, and tighten the screws.
HI 73127 Replaceable pH electrode
HI 73128 Tool to remove the electrode
HI 70004P pH 4.01 solution, 20 mL sachet (25 pcs)
HI 70006P pH 6.86 solution, 20 mL sachet (25 pcs)
HI 70007P pH 7.01 solution, 20 mL sachet (25 pcs)
HI 70009P pH 9.18 solution, 20 mL sachet (25 pcs)
HI 70010P pH 10.01 solution, 20 mL sachet (25 pcs)
HI 77400P pH 4 & 7 solutions, 20 mL sachet (5 each)
HI 7004M pH 4.01 solution, 230 mL bottle
HI 7006M pH 6.86 solution, 230 mL bottle
HI 7007M pH 7.01 solution, 230 mL bottle
HI 7009M pH 9.18 solution, 230 mL bottle
HI 7010M pH 10.01 solution, 230 mL bottle
HI 70030P 12.88 mS/cm @25ºC calibration solu-
tion, 20 mL sachet (25 pcs)
HI 70031P 1413 µS/cm @25ºC calibration solu-
tion, 20 mL sachet (25 pcs)
HI 70032P 1382 ppm @25ºC calibration solution,
20 mL sachet (25 pcs)
HI 70038P 6.44 ppt @25ºC calibration solution, 20
mL sachet (25 pcs)
HI 70442P 1500 ppm @25ºC calibration solution,
20 mL sachet (25 pcs)
HI 7061M Electrode cleaning solution, 230 mL bottle
HI 70300M Electrode storage solution, 230 mL bottle
www.tools.in.th
www.neonics.co.th

This manual suits for next models

1

Other Hanna Measuring Instrument manuals

Hanna edge blu User manual

Hanna

Hanna edge blu User manual

Hanna HI 96736 User manual

Hanna

Hanna HI 96736 User manual

Hanna HI 720 User manual

Hanna

Hanna HI 720 User manual

Hanna HI 5221 User manual

Hanna

Hanna HI 5221 User manual

Hanna HI 96745 User manual

Hanna

Hanna HI 96745 User manual

Hanna HI 9811 User manual

Hanna

Hanna HI 9811 User manual

Hanna HI 110 User manual

Hanna

Hanna HI 110 User manual

Hanna SALINTEST HI98203 User manual

Hanna

Hanna SALINTEST HI98203 User manual

Hanna HI 96734 User manual

Hanna

Hanna HI 96734 User manual

Hanna HI 96724 User manual

Hanna

Hanna HI 96724 User manual

Hanna HALO2 User manual

Hanna

Hanna HALO2 User manual

Hanna HI759 User manual

Hanna

Hanna HI759 User manual

Hanna HI 991300 User manual

Hanna

Hanna HI 991300 User manual

Hanna HI 9142 User manual

Hanna

Hanna HI 9142 User manual

Hanna HI981033 User manual

Hanna

Hanna HI981033 User manual

Hanna HI 83203 User manual

Hanna

Hanna HI 83203 User manual

Hanna HI 5522 User manual

Hanna

Hanna HI 5522 User manual

Hanna HI981032 User manual

Hanna

Hanna HI981032 User manual

Hanna HI 9835 User manual

Hanna

Hanna HI 9835 User manual

Hanna Foodcare HI99192 User manual

Hanna

Hanna Foodcare HI99192 User manual

Hanna HI 84502 User manual

Hanna

Hanna HI 84502 User manual

Hanna HI 991300 User manual

Hanna

Hanna HI 991300 User manual

Hanna HI 96785C User manual

Hanna

Hanna HI 96785C User manual

Popular Measuring Instrument manuals by other brands

Harbor Freight Tools 92573 operating instructions

Harbor Freight Tools

Harbor Freight Tools 92573 operating instructions

Alpha tools DM 15 operating instructions

Alpha tools

Alpha tools DM 15 operating instructions

Tektronix Keithley 4200A-SCS user manual

Tektronix

Tektronix Keithley 4200A-SCS user manual

Tektronix PA4000 Series instructions

Tektronix

Tektronix PA4000 Series instructions

Vitalograph micro 6300 Instructions for use

Vitalograph

Vitalograph micro 6300 Instructions for use

Campbell CS205 instruction manual

Campbell

Campbell CS205 instruction manual

AEMC instruments Clamp-On 3705 user manual

AEMC instruments

AEMC instruments Clamp-On 3705 user manual

Tanaka AKV-202 Maintenance manual

Tanaka

Tanaka AKV-202 Maintenance manual

Hanna Instruments HI 4104 instruction manual

Hanna Instruments

Hanna Instruments HI 4104 instruction manual

GE Kelman DGA 900 Plus Installation & commissioning manual

GE

GE Kelman DGA 900 Plus Installation & commissioning manual

Eagle Eye BDS-Pro installation manual

Eagle Eye

Eagle Eye BDS-Pro installation manual

DayTronic 3770 instruction manual

DayTronic

DayTronic 3770 instruction manual

LI-COR LI-250 instruction manual

LI-COR

LI-COR LI-250 instruction manual

IPS M5 user guide

IPS

IPS M5 user guide

Cadex Electronic C7000-C Series user manual

Cadex Electronic

Cadex Electronic C7000-C Series user manual

Cirrus Research Dosebadge 5 user manual

Cirrus Research

Cirrus Research Dosebadge 5 user manual

Novus LogBox BLE instruction manual

Novus

Novus LogBox BLE instruction manual

Ronde & Schwarz R&S FPL1007-P6 user manual

Ronde & Schwarz

Ronde & Schwarz R&S FPL1007-P6 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.