If pH 3.00 buffer
solution is used
as the first point
the value of the
buffer is recognized
and displayed
with the blinking
stability indicator.
If pH 7.01 buffer solution is used as
the first point the buffer is recognized
with the blinking stability indicator.
When the reading is stable, the
stability indicator will disappear
and pH 7.01 will be calibrated.
If pH 7.01 is the only calibration point,
finish one-point procedure at right.
If using pH 3.00 as a second point,
continue two-point procedure at right.
For one-point calibration using pH 3.00 buffer go to procedure B
For one or two-point calibration using pH 7.01 buffer go to procedure A
A
B
When the
reading is stable,
the stability
indicator will
disappear.“Sto”
will be displayed
when the
calibration
is saved.
Meter will
exit to
measurement
mode.
Meter will exit
to measurement
mode.
Two-Point
Use pH 3.00 to
perform a two-
point calibration.
The value is
automatically
recognized and
displayed with the
blinking stability
indicator.
When the reading
is stable, the
stability indicator
will disappear.
“Sto” will be
displayed when
the calibration
is saved.
Meter will exit
to measurement
mode.
Thank You
Thank you for choosing a Hanna Instruments product. Please read this instruction
manual carefully before using this instrument.
For technical support, contact your local Hanna Instruments Office or email us
at tech@hannainst.com.
To find your local Hanna Instruments Office or for additional information on
Hanna Instruments products, visit www.hannainst.com Meter Calibration
From measurement
mode, press and
hold the ON/OFF
button until “CAL”
is displayed.
When “7.01” blinks
on the display, place
the tip of the probe
into a pH 7.01 or
3.00 buffer solution.
One-Point
One-Point Calibration with pH 3.00
B
AOne or Two-Point Calibration with pH 7.01
pH 3.00 will then
blink on the display.
Ignore it and press
ON/OFF button.
Operation
Press the ON/OFF button to turn the meter on. All tags
will be displayed.
The meter will go into measurement mode: current
reading is displayed.
Range 0.0 to 14.0 pH
Resolution 0.1 pH
Accuracy ±0.2 pH @25 °C/77 °F
Calibration Automatic, one or two-point
Electrode Built-in probe for specific application
Battery Type CR2032 Li-ion
Battery Life Approximately 1000 hours of continuous use
Auto-off 8 minutes, 60 minutes or can be disabled
Environment 0 to 50 °C (32 to 122 °F); RH 95% max
Dimensions 51 x 157 x 21 mm (2 x 6.2 x 0.9“)
Weight 46 g (1.6 oz.)
Preliminary Examination
Remove the meter from the packing material and examine it carefully to make sure that
no damage has occurred during shipment. If noticeable damage is evident, contact your
local Hanna Instruments Office.
Each meter is supplied with:
• pH 3.00 buffer solution liquid sachet (2 pcs.)
• pH 7.01 buffer solution liquid sachet (2 pcs.)
• Cleaning solution for wine deposits
• Cleaning solution for wine stains
• Electrode storage solution, 13 mL dropper
• Electrolyte fill solution
• Pipette
• Instruction manual
• Quality certificate
Note: Save all packing material until you are sure that the instrument functions
correctly. All defective items must be returned in the original packaging with the
supplied accessories.
Meter Overview
Preparation:
The pH electrode is shipped with a protective cap containing storage solution. Before
using the meter, remove the protective cap and condition the electrode by soaking the
tip (bottom 3 cm (1.2”)) in pH 7.01 buffer solution for several minutes. Then follow the
calibration procedure.
• Do not be alarmed if white crystals appear around the cap. This is normal with pH
electrodes and they dissolve when rinsed with water.
• Turn the meter on by pressing ON/OFF button.
• Remove the protective cap and immerse the tip of the electrode in the sample
to be tested.
DO NOT IMMERSE THE ELECTRODE OVER THE MAXIMUM IMMERSION LEVEL.
• Stir gently and wait for a stable reading.
• For best results, periodically calibrate at the temperature of the wine.
• After use, rinse the electrode with water and store it with a few drops of storage
solution in the protective cap.
• Reattach the protective cap after each use.
DO NOT USE DISTILLED OR DEIONIZED WATER FOR STORAGE PURPOSES.
Intended use
The pH of wine is important because it will affect the quality of the final product in
terms of taste, color, oxidation, chemical stability and other factors. Generally in
winemaking, the higher the pH reading, the lower the amount of acidity in the wine.
Three important factors in determining the pH of wine include the ratio of malic acid to
tartaric acid, the amount of potassium, and the total amount of acid present.
Most wines optimally have a pH between 2.9 and 4.0, with values differing based on
the type of wine. Values above pH 4.0 indicate that the wine may spoil quickly and be
chemically unstable. Lower pH values allow the wine to stay fresher for a longer period
and retain its original color and flavor. High pH wine is more likely to breed bacteria
and become unsuitable to drink. For finished white wines, the ideal pH is between pH
3.00 and pH 3.30, while the final pH for red wine is ideally between pH 3.40 and pH
3.50. The optimal pH before the fermentation process is between pH 2.9 and pH 4.0.
The pH of wine therefore not only affects the color of wine, but also the oxidation, yeast
fermentation, protein stability and bacteria growth and fermentation.
Probe Features
CPS™ Sleeve Junction
Clogging Prevention System (CPS™) technology improves pH measurements in samples
with a high solids content. Conventional pH electrodes use ceramic junctions that can clog
quickly when used in samples that have a high solids content such as wine must or juice.
When the junction is clogged, the electrode does not function. CPS technology utilizes the
porousnessof groundglass coupled with a PE sleeve to prevent clogging of the junction. The
ground glass allows proper flow of the liquid, while the PE sleeve repels solids. As a result,
pH electrodes with CPS technology take up to 20 times longer to be fouled as compared
to conventional electrodes.
Liquid Crystal
Display
ON/OFF
Button
Maximum
Immersion
Level
Minimum
Recommended
Immersion
pH Probe
Minimum level
of electrolyte
solution
Low Temperature Glass Formulation
The glass tip uses a special low temperature (LT)
glass formulation. This is beneficial since many
food products are at low temperatue.
Spherical Glass Tip
The domed tip design allows for a wide area
of contact with the measured sample. This permits
a faster electrode response with a higher degree of
stability.
Calibration
Battery
indicator
Unstable
tag
Measurement
unit
Digits
Specifications
"Sto" will be
displayed when
the calibration
is saved.