Hawk DR060380B-2 User manual

1
VIDEO-CYSTOSCOPE
Model: DR060380B-2
USER MANUAL
MANUAL P/N:IA-MN-213-3(Version 3.1)
Hangzhou Hawk Optical Electronic
Instruments Co., Ltd.

2
Introduction
Thank you for your purchasing the HAWK’s equipment for endoscopy. This device
has been scientifically designed and manufactured to meet the international standards
required in the application related to endoscopic diagnosis and surgery. In order to use the
device easily, please read the user manual carefully before using this device. Operate the
device according to the specific procedures.
The engineer at Hawk is always prepared to assist in answering any of your questions
or in meeting your specific requirements. We also welcome any comments you may have
concerning about our equipment. Your comments and questions will receive immediate
Please see Customer Service for details or visit our Homepage
http://www.hawkendoscopes.com to learn about more information.
Manufacturer/Service: Hangzhou Hawk Optical Electronic Instruments Co.,Ltd.
Address:No.9 Xinda Road,Suoqian, Xiaoshan,Hangzhou,Zhejiang 311201P.R.China
TEL:+86-571-22811666 FAX:+86-571-22811228
European Union representative:Shanghai International Holding Gorp. Gbmh(Europe)
Add: Eiffestrasse 80,20537 Hamburg, Germany
Tel: +49-40-2513175 Fax:+49-40-255726

3
Conformance Standards
1.The following classifications are in accordance with the IEC/EN 60601-1:6.8.1:
1) According to 93/42/EEC Medical Device Directive, this is Class IIa medical Device.
2) According to IEC/EN 60601-1.
• Equipment is Class I, BF Part.
3) According to IEC 60529:2001-02.
The protection level of monitor is IPX0.
The protection level of endoscope is IPX7.
2. Certification
The general electrical medical system is CE, ISO9002, ISO9001 certified.
3. Original Documentation
The original document is written in English.
Note DISASSEMBLY AND REPAIR BY THE CUSTOMER ITSELF IS NOT
PERMITTED, IF THIS LEAD TO SOME DAMAGE, THE CUSTOMER
SHOULD TAKE RESPONSIBILITY OF IT.

4
Contents
1.Safety…………………………………………………………………………………………5
1.1 Introduction…………………………………………………………………………………5
1.2 About Safety…………………………………………………………………………………5
1.3 Clinial Operation……………………………………………………………………………8
1.4 Pre-use Inspection…………………………………………………………………………8
1.5 Using with other accessories ……………………………………………………8
1.6 Maintenance…………………………………………………………………………………8
1.7 Cleaning and disinfection ………………………………………………………………8
1.8 Compatibility…………………………………………………………………………………8
1.9 Complete set…………………………………………………………………………8
1.10 Label symbole description………………………………………………………………9
1.11 EMC (Electromagnetic compatibility) ………………………………………………10
1.12 WEEE………………………………………………………………………………………13
2 Products description……………………………………………………………………13
2.1 Products list……………………...…………………………………………………………13
2.2 Intended use ………………………………………………………………………13
2.3 Scope of use………………………………………………………………………………13
2.4 Feature ……………………………………………………………………………………14
2.5 Electrical connections……………………………………………………………………14
3 Instruction of product……………………………………………………………………14
3.1 3.5” LCD……………………………………………………………………………………14
3.2 8” LCD ……………………………………………………………………………………..15
3.3 Videoscope………………………………………………………………………………….16
3.4 Instruction………..……………………………………………………………………16
3.5 DVR……………..…………………………………………………………………….17
4 Operation ………..……………………………………………………………………20
4.1 Preparation before operation………..………………………………………………20
4.2 Checking before use……….………..………………………………………………22
4.3 Operation method…………………………………………………………………………23
5 Reprcessing…………………….…………….………………………………………………25
5.1 Reprcessing…………………….…………………………………………………25
5.2 Checking…………………….………………………………………………………………25
5.3 Malfunction and Solution ………………………………………………………………26
6 Appendix……………………………………………………………………………………27
6.1 Technical Specifications………………………………………………………………27

5
1.Safety
1.1 Introduction
Safety is everyone’s responsibility. The safe use of this equipment is largely depend upon the
installers, users, operators, and maintainers. It is imperative that personnel study and familiar with this
entire manual before attempting to install, use, clean, service or adjust this equipment and affiliated
accessories. It is paramount that the instructions contained in this manual are fully understood and
followed to enhance safety to the patient and the user/operator. It is for this reason that the following
safety notices have been places appropriately within the text of this manual to highlight safety related
information or information requiring special emphasis. All users, operators, and maintainers must be
familiar with and pay more attention to all Warnings and Notes incorporated herein(as Table1).
Table1:Warning and attention label meaning
Signal Meaning
Warning
“Warning “indicates in case of neglect of attention the presence
of a hazard that could result in severe personal injury, death or
substantial property damage.
Note
“Note” indicates informs the user in case of neglect of attention
to be improper installation, operation or maintenance, such
behavior can lead to device damage.
1.2 About Safety
Please study this manual carefully before using this equipment. You must following the proper
steps in training yourself for use of this equipment. Otherwise, there may be detrimental
consequence to patients or the operator. Furthermore, you should carefully study the operating
manuals of any other equipment and accessories that will be used in combination with this equipment.
Safety issue is discussed in comprehensive details in Chapter 1 SAFETY.
Any concern that may cause bodily harm in connection with the observation, diagnosis using the
endoscope will be set out in the “ Warning” symbol. Please study the content of each warning
carefully and follow the instruction set out in each warning. Furthermore, any concern that may
damage the endoscope will be set in the “ Note” symbol.
Warning
●Incorrect use or operation of endoscope is dangerous and may cause harm to the
patient, the operator or anybody in the vicinity.
●Please study this manual carefully before using this endoscope.
1.2.1 Safety Precaution
This equipment has been developed and tested according to safety standards as well as national
and international standards. This guarantees a very high degree of safety for this device. The legislator
expects us to inform the user expressively about the safety aspects in dealing with the device. The
correct operating this equipment is imperative for its safety. Therefore, please read carefully all
instructions before switching on this device. For more detailed information, please contact our
Customer Service Department or one of our authorized representatives.

6
Warning
●The operators who will do the endoscopic surgery with medical pressurizer must be
trained and qualified. Forbid unqualified personnel to use this equipment.
●This device must be installed, used, operated and maintained by or under supervision
the person who is properly trained and qualified.
●It should has reliable grounding with the power supply socket to control unit. The
power grid connected to the operating room must be grounded reliably, or doctors and
patients would result in injury.
Note
●Keep the User’s Manual in a place easily accessible at all times for persons operating
and maintaining the equipment.
●The manufacturer is only responsible of the safety, reliability and performance of the
device:
1) Electrical installation should be done in accordance with the provisions described in
this user manual.
2) This device should be installed, used, operated and maintained in accordance with
this user manual.
●Please do not use this device if it is suspected to have problems. Please contact
HAWK designated service providers and check the device before use. If any deficiency
is suspected or confirmed in the device, do not use the device until it has been checked
by the authorized service.
●Do not disassemble other parts by unauthorized person except the parts permitted by
this operation manual. The maintenance of this device should be done in accordance
with this user manual.Any additional maintenance and modification should be done only
by HAWK service technicians or authorized personnel.
●Do not modify this device without any authorization. The manufacturer is not liable for
damage caused by unauthorized tampering with the device(s). Such tampering will make
customer forfeit any rights to claim under warranty.
●This equipment may only be used together with accessories supplied by HAWK. If the
customer makes use of other brand accessories, use them only if their safe usability
under technical safety aspects has been proved and confirmed by HAWK or the
manufacturer of the accessory.
●Before insert plug into socket,it should check the plug is match the socket, do not force
to insert the plug into the socket. Do not pull the cable when turn off the power supply,
it should hold the plug and pull it out .
●Do not force cable connections. If a cable does not connect easily, be sure that the
connector (plug) is appropriate for the receptacle(socket). If you cause any damage to a
cable connector(s) or receptacle(s), let the damage(s) be repaired by an authorized
service technician.
●Don’t plug the AC power cord into the outlet before the connection between devices of
the equipment is completed which can generate the defect.
●Before every operation, visually check the equipment for exterior mechanical
damaged(s) and for proper function.

7
●Immediately turn off and unplug any equipment that gives off smoke, sparks, strange
noises, or odors .
1.2.2 Environment requirements
1.2.2.1 Avoid the following environments for operation or storage:
Where the device is exposed to water vapor. Don’t operate a device with a wet hand.
Where the device is exposed to direct sunlight.
Where the temperature exchanges extremely. Normal operating temperature range is from
10℃to 40℃,Humidity is from30% to 75%.
Where it is near the heat equipment.
Where the device is subject to excessive shocks or vibrations.
Where device is exposed to chemical material or explosive gas
Be careful not to be inserted dust, especially metal.
Where the humidity is extremely high or there is a ventilation problem.
Be careful not to close the fan located on the lateral or be opposite of the equipment.
Warning
●This equipment must not be used in such areas:
1) An area that is in danger of explosions.
2) An area that in the presence of flammable, explosive, or volatile solvent such as
alcohol, benzene or similar chemicals.
Note
●Do not put or use this device in humid rooms. Humidity should be kept between 30%
and 75% for normal operation. Do not expose the device to water splashes, dripping
water, or sprayed water. Do not place containers containing fluids, liquids, or gases on

8
top of any electrical equipment or devices.
1.2.2.2 Avoid places where the ambient temperature falls below +10℃or above 40℃for normal
operation;
1.2.2.3 Avoid environments where the equipment is exposed to excessive shocks or vibrations;
1.2.2.4 Storage envirement:
Ambient temperature -40℃~+55℃
Humidity will not exceed 95%
Atmosphere pressure 500hpa~1060hpa
None corrosive air
Keep ventilation
Warning
●Common injuries that may happen during use of the endoscope are: irritation to
mucosal, bleeding, infection, or others. If you do not follow the operating procedure in
this manual, then injury or damage to the equipment may result.
1.3 Clinical operation
The device must be operated by trained and qualified person, and operator must have skilled
clinical technology. This manual does not explain or discuss clinical technique, the operator should
consult the professionals about clinical endoscopic operating skills.
1.4 Pre-use inspection
To ensure proper use, before you use the product, please read the 4.2 Pre-use Inspection of
user manual, and check each function.
1.5 Using with other accessories
To use other accessories together with this videoscope. When using a controller or a display, in
order to prevent electric shock, use a properly grounded electrical outlet to power or charge.
1.6 Maintenance
After long-term use, the endoscope may be damaged, please be sure to regularly check. Even a
slight abnormality was found, please be cautious, careful examination. Be sure not to disassemble
or modify this videoscope.
1.7 Cleaning and disinfection
In the operation, endoscope will touch upon the patient's mucous membranes, in order to minimize
the risk of virus infection between patients, before and after each using, the videoscope is required
to be thoroughly disinfected or sterilized. Before disinfection or sterilization, the videoscope should
be cleaned carefully, thoroughly cleaned and microbe and the organic matter, the existence of
these substances may limit the effectiveness of disinfection or sterilization. Please read these
instructions carefully and strictly follow the cleaning, disinfection and sterilization rules.
1.8 Compatibility
The safety using of videoscope not only depends on videoscope itself, but also depends on other
devices and accessories which use with this videoscope. To ensure compatibility with these
devices and accessories, we recommend that the company uses only HAWK videoscope and
accessories.
1.9 Complete Set
A complete endoscope system includes an videoscope, monitors and accessories
Customers who bought this product, please open the packaging carefully and follow the packing
list to check each of the sets. Please keep all packing materials in case there is any damage to the
claimant's use transportation. If found damaged or missing parts, contact with the company or
distributor.

9
1.10 Label and Symbol description
The Label Symbol Description is listed as below.
Table2:Label and Symbol
Label Icon Meaning
~AC Power
Power supply
Turn on (Main power)
Turn off(Main Power)
Type BF Applied Part
“Note”
“Warning”
Fuse
Manufacture Date
Manufacturer and Address
European Union representative and Address
Do not throw into wastebin
Non-Ionizing Radiation
Back / Convert
MENU MENU
Left
Right

10
Video
Picture
Follow instructions for use
1.11 EMC (Electromagnetic compatibility)
1.11.1 Product classification
The device is an engineering medical device, classify according to the standard EN55011
"industrial, scientific and medical (ISM) radio frequency electromagnetic equipment harassment the
limits and the method of measurement" standard classification, it is belonging to one group
Classification A engineering medical equipment. The meaning of this classification as follows:
a) One set of equipment is defined as all engineering medical equipment intend to produce
coupled RF energy due to play to their own function.
b) Classification A equipment is defined as the equipment is not used at home and not connected
low-voltage power net.
1.11.2 Declaration of conformity
The device is made by Hawk company meet the requirements of EN60601-1-2: 2007.
Table 3:Guidance and manufacturers declaration -- Electromagnetic Emissions
The device is intended for use in the eietromagnetic environment specified below. The customer or the
user of the Irrigation Aspiration Unit should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11 Group 1
The device uses RF energrgy only for its internal
runction.Therfore its RF enissions are very low
and are not likely to cause any interference in
nearby electronic equipment
RF emissions
CISPR 11 Class B
The equipment is suitable for use in all
facilities, including home and connected
directly to the household residential public
low-voltage power supply network
Harmonic emissions
IEC 61000-3-2 Class A
Voltage
fluctuations/flicker
emissions
IEC61000-3-3
Complies
Table 4:Guidance and manufacturers declaration: Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified below.The customer or the
user of the device should assure that it is used in such an environment
lMMUNITY test test level Compliance level Electromag etic environmemt -
guidance
Electrostatic
discharge(ESD)
IEC 61000-4-2
±6 kv contact
±8 kv air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile lf floors are covered with
synthetic material, the relative

11
humidity should be at least 30%
Electrical fast
transient/burst
lEC 61000-4-4
±2 kv for power
supply lines
±1 kv for
input/output lines
±2 kV for power
supply lines
±1 kV for
input/output lines
Mains power quality should be that of
a typical commerdial or hospital
environment
Surge
lEC 61000-4-5
±1 kV line (s)to
line(s)
±2 kV line(s) to
earth
±1 kV differential
mode
±2 kV common
mode
Mains power quality should be that of
a typical commercial of hospital
environment
Voltage dips,shot
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% UT
(>95% dip in
UT)
for o,5 cycle
40% UT
(60% dip in UT)
for 5 cycles
<70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in
UT)
for 5 s
<5% UT
(>95% dip in UT)
for o,5 cycle
40% UT
(60% dip in UT)
for 5 cycles
<70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 s
Mains piwer quality should be that of
a typical commercial of hospital
environment lf the user of the
Ele Pneumatic Lithotriptor image
intensifier requires
continued operation during power
mains interruptions it is
recommended
that the Ele Pneumatic Lithotriptor
lmage lntensifier be powered from an
uninterruptible power supply or a
battery
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
3A/m 0.3A/m
lf image distortion occurs,it may
be necessary to position the Ele
Pneumatic Lithotriptor image
intensifier further from sources of
power frequency magnetic fields or to
install magnetic shielding The power
frequency magnetic field should be
measured in the intended installation
location to assure that it is sufficiently
low
NOTE UTis the a,c,mains voltage prior to application of the test level.
Table 5:Guidance and manufacturers declaration -- Electromagnetic Interference Resistance
The Model is intended for use in the electromagnetic environment specifled below The customer should
assure that it is used in such an environment.
IMMUNITY
test TEST LEVEL Compliance
level Electromagnetic environment guidance

12
Conducted RF
IEC
61000-4-6
Radiated RF
IEC
61000-4-3
3 Vrms
150kHz to 80 MHz
3v/m
80 MHz to 2.5 GHz
3Vrms
3V/m
Portable and mobile RF communications equipment
should be used no closer to any part of Model
006,including cables,than recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d=1.2 P
d=1.2 P80MHz~800MHz
d=2.3 P800Hz~2.5GHz
where p is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufactuer and ais the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,as
deter-mined by an electromagnetic site survey,ashould
be less than the compliance level in each frequency
range. b
lnterference may occur in the vicinity of equipment
marked witn the following symbol:
NOTE 1 At 80 MHz and 800 MHz,the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a) Field strengths from fixed transmitters,such as base stations for radio (cellular/cordless) telephones and
landmobile radios, amateur radio, AM and FM radio bradcast and TV broadcast cannot be predicted
theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. lf the measuraed field strength in the location in which
the Model 006 is used exceeds the applicable RF compliance level above, the Model 006 shold be
observed to verify normal operation. lf abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the Model 006.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
Note
●It should need special preventive measures to this device,install and service
according to explanation of “EMC”.
●Portable and movable RF communication equipment can influent medical electronic
equipment.
●In addition to the transducer and cables as spare parts of the internal components
sold by manufacturers, it will probably increase eradiation or decrease the interference
if using other unspecified accessories, transducer and cable.
●This equipment or system can not be closed or piled up other equipment. If it must

13
close or piled up to other equipment, it should observe and test the system can be run
normally.
●Although the device has been designed to interfere with the protection of such
electromagnetic radiation. When it connects with RF surgical set, you should restart the
device as soon as electromagnetic interference occurs. Furthermore, If it is still
unresolved, please increase the distance with the RF.
1.12 WEEE
The following description conforms to the stipulation of 2002/96/CE which stipulate the measures
to deal with such electronic products . Please comply with the WEEE standards when deal with such
equipment including it’s parts and material.
.
This label is printed on products, package or instructor manual. This label means this
products enters into the market after August 13th,2005. You should deal with each
component separately according to local stipulation of waste disposal.
Mishandling or throwing away this equipment will be harmful to environment or human
healthy because this equipment contains special material.
According to WEEE standard, it must not deal with this equipment like domestic
waste, it should be divided and disposaled.
You can contact local waste disposal collecting department to get detailed information.
Illegal disposal of waste will be subject to legal sanctions.
2 Products description
2.1 Products list
1)Flexible Video- cystoscope
2)3.5” monitor or 8” monitor
3)Leakage tester, Cover for gas sterilization, SD card, Cable, AC connector.
※Our company can configure the instruments according to customers’ requirements.
※The equipment which produced by our company only can use with our original parts and
accessories.
2.2 Intended use
An endoscope with a flexible inserted portion intended for the visual examination and treatment of
the bladder and the urinary tract. It is inserted into the urethra (the tube that carries urine from the
bladder to the outside of the body) through the natural urinary orifice during cystoscopy. Anatomical
images are transmitted to the user by a video system with a charge-coupled device (CCD) chip at the
distal end of the endoscope and the images showing on a monitor. It is used to examine/diagnose
incontinence, the inability to pass urine (retention), recurrent urinary tract infections, and to remove
tissue samples/stones/small tumours from the bladder. This is a reusable device.
2.3 Scope of use

14
This equipment may be used for examination, diagnosis and operation. Do not use this equipment
for any other purposes or in any other settings.
2.4 Feature
2.4.1 The videoscope has a unique waterproof function, it can be cleaned thoroughly and disinfected.
The instrument channel can be cleaned in whole course, so that minimize the risk of infection among
patients with each other.
2.4.2 There are two directions, up and down direction are controlled by bending wheel.
2.4.3 Biopsy forceps, brush can be used.
2.4.4 It has fixed focus, large depth of field, wide angle, and get clean image within the range of 3 ~
50mm.
Note
●This equipment is according to the requirement of Application Part of Type BF,it
possess Type F isolation suspension application part, and B Type equipment have a
specific protective level for shock.
●And protective liquid level is IPX0.
●Clean, disinfect and sterilize this equipment in accordance with the manual.
●This equipment must not be used in setting where gaseous mixture of inflammable
anesthesia or nitrous oxide is in use.
2.5 Electrical connections
The input power supply is 100V-240V ~ 50Hz, only connect with a reliable grounding socket.
3. INSTRUCTION OF PRODUCT
3.1 3.5” LCD

15
1、Housing 2、Power Switch 3、Screen 4、Soft Protect Cover 5、AV output connector
6、Power input connector 7、USB connector 8、SD Slot 9、Decrease button
10、Increase button 11、Back / switch button 12、Scope Lock 13、MENU
14、Video button 15、Photo button 16、Power Indicator 17、Charging Indicator
3.2 8” LCD
18、Housing 19、AV output connector 20、USB connector 21、SD Slot 22、Screen
23、Standby button 24、Power lndicator 25、Decrease button 26、increase button
27、Back / switch button 28、Manu 29、Video button 30、photo button

16
34、Powerswitch 32、DC input connector 33、Charging Indicator
34、AV input connector 35、Support 36、Reset hole 37、Handle
3.3 Videoscope
(43) Leakage connector (44) Operating part (45) Insertion part (38) Bending part
(39) Distal part: (40) Connector (41) Bending wheel (42)Instruments connector
3.4 Instruction
1) Housing
2)Power switch: l=ON、O=OFF,Power lamp lights up when the power switch turned on.
3) Screen
4)Soft protection cover: Protection cover for dust.
5) AV output connector: connect AV output cable and output image to TV and monitor.
6) Power inpout connector: charge the monitor
7) USD connector: transfer image to computer
8) SD Slot: insert SD card
9) Decrease button: Decrease the parameter and up the menu, review the picture.
10) Increase button: Increase the parameter and down the menu, review the picture.
11) Back / switch button: switch contrast, brightness, saturation when in adjustment of parameter,
And back function in sub menu display
12) Scope Lock: connect and fix videosocope,
13) Menu: Display menu, out of menu, delete file
14) Video button: start and stop video
15) Photo/ confirm button: Take picture and confirm in menu
16) Power indicator: Indicator on means working, Indicator off means turn off the device
17) Charging indicator: Indicator on means charging, indicator off means full charged

17
18) Housing
19)AV output connector: connect AV output cable and output image to TV and monitor.
20) USD connector: transfer image to computer
21) SD Slot: insert SD card
22) Screen
23) Standby button: press this button for turn on the device
24) Power indicator: indicator on means switch on the device, indicator off means switch off the device
25) Decrease button: Decrease the parameter and up the menu, review the picture.
26) Increase button: Increase the parameter and down the menu, review the picture.
27) Back / switch button: switch contrast, brightness, saturation when in adjustment of parameter,
And back function in sub menu display
28) Menu: Display menu, out of menu, delete file
29) Video: start and stop video
30) Photo/ confirm button: Take picture and confirm in menu
31)Power switch: l=ON、O=OFF,Power lamp lights up when the power switch turned on.
32) Power input connector: charge the monitor
33) Charging indicator: Indicator on means charging, indicator off means full charged
34) AV output connector: connect AV output cable and output image to TV and monitor.
35) Support
36) Reset button: For reset
37) Holder: hold monitor
38) Bending part: Up and down direction
39) Distal part: imager senor and fiber equipped in the device, for provide image and light
40) Connector: connect with 3.5’and 8’ monitor, transfer power and signal
41) Bending wheel: operating bending part
42) Instruments connector: for insert instruments
43) Leakage connector: leakage test for videoscope when connect with leakage
44) Operating part: holder part, include bending wheel, light source and bending system
45) Insertion part: insert to human body
3.5 DVR
3.5.1 Image display: Turn on the power, the LCD will show our logo

18
and green indicator in the low right is light. This process takes a
few seconds, please wait patiently. After power on, entering the
preview mode to show the current image. If shows on the
low left of screen, it can take photos and videos. Date and time
will show on the top right, video time and capacity show on the
low right, electricity capacity and show on the low left.
3.5.2 Take photos: Press can take picture and will
show on the top left of LCD screen. After about 2 seconds,
the icon disappears, the photo files will be saved automatically.
The photo resolution can be chosen in the menu as JPG
format.
3.5.3 Take videos: Press can take videos and will
show on the top left of LCD screen and flashing. Press the
button again to stop take video, the video files will be saved
automatically. The movie resolution can be chosen in the menu
as AVI format.
NOTE
10 minutes, 20 minutes, 30 minutes is supported when capturing video, when the
capacity of single file is full, recording will be stopped automatically and record next
video file until press video button to stop recording and enter into the preview mode.
3.5.4 Adjust parameters of screen: Press back / switch button , there are three optional items:
saturation, chromo, brightness, press and to decrease and increase value.
3.5.5 Main menu: Press the Menu button to enter the main menu in
preview mode. There are three sub-menus: Camera, My Files,
Setup, press and to move yellow strip and select, then
press Photo/confirm button to confirm and enter into sub-menus,
press menu button to back main menu and back to preview mode.
3.5.6 My Files: After enter into sub-menu of My Files, files name and
number of SD card will be showed, press and button to select,

19
then press photo/confirm button to enter into the file, press menu
button to back.
3.5.7 Setup: After enter into sub-menu of Setup,There are 6 sub
-menus: Video time, storage, Date time, Memory Info,
LCD Brightness, Language, press and button to select,
then press photo/confirm
button to confirm.
1)Video Time: 10Mins, 20Mins, 30Mins can be selected.
2)Storage: Flash, SD Card, we recommend you store
photos And videos in SD Card.
3) Date Time: Date Time stamp
and Date Time Setup can be selected.
Date time stamp can be selected.
Time can be set, use left and right button to set, and use photo/
confirm button to confirm.

20
4)Memory Info: To record inside memory and SD
Memory capacity.
5)Brightness LCD: Seven steps can be selected, use
and to decrease and increase the brightness.
6)Language: English and Chinese.
3.5.8 Review: In the preview mode, press menu and check photo files and Video file according to 5.i,
Press the Menu button to enter the main menu in preview mode, press Photo/confirm button to
confirm and play video, press Photo/confirm button or back/switch button to stop playing.
3.5.9 Naming Rule: Folder named according to the date, a file named according to the time.
3.5.10. Delete Files:Press Menu button to enter into delete menu when single file is displayed, 3
options can be selected: delete all of this, current and cancel, press left and right button to selected.
System will inform if file delete or not, select YES means delete all files or current files, if select No, it
will back to current file condition. Note: please delete all file before delete this folder. is back single file
state. Note: when deleted all files in the folder to delete the folder.
4. OPERATING INSTRUCTIONS
4.1. Preparation before Operation
Aim at red dot of LCD Screen and Video Endoscope, connect
them, then screw down lock nut and fix LCD Screen.
Table of contents
Popular Medical Equipment manuals by other brands

Covidien
Covidien Valleylab FX8 Series Quick reference guide

Tomey
Tomey TSL-4000Z Operator's manual

Inogen
Inogen One G2 user manual

ulrich medical
ulrich medical uCentrum Instructions for use

Diabetic Supply of Suncoast
Diabetic Supply of Suncoast Advocate HEATING PAD manual

Otto Bock
Otto Bock Michelangelo Hand Transcarpal 8E550 Instructions for use