Healthtree Medical JKS50A User manual

Operator’s Manual
Fingertip Pulse Oximeter
General Description
The measurement of oxygen saturation of arterial blood (also known
as pulse oxygen saturation, usually shortened as SpO2) adopts the
principles of light spectra and volume tracing. The LED emits lights with
two specific wavelengths, which are selectively absorbed by oxygenated
hemoglobin and deoxyhemoglobin. The optical receptor measures the
changes in the light intensity after the light passes the capillary network
and estimates the ratio of oxygenated hemoglobin and the total
hemoglobin.
Precautions for use
◼Explosion hazard. Do not use the oximeter in the presence of
flammable anesthetics mixture with air, oxygen, or hydrogen.
◼When the oximeter is in use, there should not be any great power
appliances as high voltage cables, X-ray machine, ultrasound
equipment and electrizer in use nearby.
◼Keep the oximeter away from dust, vibration, corrosive substances,
explosive materials, high temperature and moisture.
◼This oximeter for sports or aviation use only.
◼The oximeter should be handled with care so as to avoid shocks and
falls.
◼When the oximeter is in use, it must be ensured the batteries have
sufficient capacity; otherwise there might be such phenomena as
starting-up abnormalities or inaccurate measurement data, etc.
◼Please do not use such pointed objects as pen point or nails for
pressing operation, otherwise it might cause permanent damage to the
surface of the keyboard.
◼To ensure accurate performance and prevent device failure, do not
expose the oximeter to extreme moisture, such as direct exposure to
rain. Such exposure may cause inaccurate performance or device
failure.
◼Do not conduct SpO2measurement on the finger smeared with nail
polish, otherwise this will lead to unreliable measurement results.
◼Please do not open the enclosure. The enclosure shall only be opened
by the authorized person.
◼In order to have more accurate measurements of SpO2and PR, the
oximeter should be used in quiet and comfortable environment.
◼Follow local ordinances and recycling instructions regarding disposal or
recycling of the device and device components, including batteries.
Intended Use
This product is intended for sports and aviation use only. It should not
be used to diagnose or treat any medical condition .
Operation Instructions
1. Install two AAA batteries into battery cassette before closing its
cover.
2. Nip the oximeter, then insert one of fingers
into the rubber hole of the oximeter before
releasing the oximeter, and your nail surface
must be upward.
3. Press the function button once on front panel.
4. Your finger and body do not tremble during measuring.
5. Read corresponding data in only one direction on the display
screen.
①Indication of screen brightness
②Indication of pulse intensity
③SpO2Plethysmogram
④Indication of battery capacity
⑤Pulse rate reading
⑥SpO2 reading
⑦PI reading
Battery Installations
1. Press the button down on the back panel of oximeter and push the
battery cover horizontally along the arrow shown as below:
2. Install the two AAA batteries into battery cabin
in correct polarities.
3. Close the battery cover.
Notes:
◼Pay attention to the correct polarities terminals of the battery,
which are marked with plus and minus signs. Do not install the
battery in the direction of the spring.
◼Please remove the battery if the oximeter will not be used for long
time.
Maintenance
1. Use a soft cloth dampened with either a commercial, nonabrasive
cleaner, or a solution of 70% alcohol in water, lightly wipe the surfaces
of the oximeter.
2. The most commonly used hospital cleaning agent and non-corrosive
detergents can be used for cleaning the oximeter, but please be careful
that many types of detergents must be diluted before use; Please use
them according to the directions of the manufacturers of the
detergents.
3. Avoid using alcohol-based, amido or acetone-based detergents
4. The casing of the oximeter should be kept from the contamination of
filth and dirt, and it can be wiped with non-velvet soft cloth. When
cleaning, do not spill the liquid onto the instrument. Ensure no liquid is
allowed to enter the inside of the oximeter.
5. It is forbidden to use such grinding materials as wire brush or metal
polishing agent, because these materials may cause damage to the
panels of the oximeter.
6. Please do not soak the oximeter in liquid.
7. Under normal circumstances, it is unnecessary for the oximeter to have
special maintenance, and cautions must be exercised on the following
points during the use of the oximeter:
◼Please use the oximeter in the environment according to the
requirements of the performance criteria.
◼Avoid contacts with organic solutions, dusts or corrosive gases.
Product Specifications
◆Measurement specifications
SpO2
Measuring Range
0~100%
Resolution
1%
Accuracy
At 70~100%, ±2%;
At 0~69%, unspecified
Data update period
<13 s
PR
Measuring Range
25~250 bpm
Resolution
1 bpm
Accuracy
±1% or ± 1 bpm, whichever is greater
Data update period
< 13 s
%100
lobindeoxyhemogbinoxyhemoglo
hemoglobinoxygenated
%SpO2
+
=

◆Battery specifications
Type
Voltage
two AAA alkaline battery
1.5 Volts DC (per battery)
The oximeter uses two 1.5 V A AA type batteries and a set of new
batteries can be used for more than 18 hours, depending on
concrete battery types.
◆Environmental specifications
Operation
Temperature
+5℃~+40℃
Atmospheric Pressure
700hPa~1060hPa
Relative Humidity
≤85%
Transport and Storage
Temperature
-20℃~+55℃
Atmospheric Pressure
500hPa~1060hPa
Relative Humidity
≤93%
◆Physical specifications
Weight
about 21g (exclude battery)
about 54g (include battery)
Dimensions
57mm(length) × 33mm(width) ×30mm(height)
◆Sensors specifications
Wavelength
Pulse oximetry sensors contain LEDs that emit
red light at a wavelength of approximately 660
nm and infrared light at a wavelength of
approximately 905 nm.
The total optical output power of the sensor
LEDs is less than 15 mW.
This information may be useful to clinicians,
such as those performing photodynamic
therapy.
Note: Sensor LED light emissions fall within
Class 1 level, according to IEC 60825-1:2001.
No special safety precautions are required.
Possible Problems and resolutions
Problems
Possible causes
Solution
There is no
response to
the function
button.
The button can not be
pressed to its position
Ensure that the button is
fully depressed.
Battery capacities are
low
The batteries may be
missing, discharged, or
oriented incorrectly.
Replaced them with new
ones.
The Pulse
search time
is too long
Perfusion may be too
low
Change the measuring
site. Try another
oximeter.
person movement
Interference due to
person activity may be
preventing the oximeter
from tracking the pulse.
Keep the person still, if
possible.
Electromagnetic
interference may be
preventing the oximeter
from tracking the pulse.
Remove the source of
interference.
There may be
interference due to
ambient light, or the
oximeter may be on an
extremity with a blood
pressure cuff, arterial
catheter, or
intravascular line.
Reposition oximeter, as
necessary.
Display is
dark-or-bright
Battery capacities are
low.
Replace the batteries.
Symbols Definitions
Symbol
Definition
Type BF equipment (Refer to IEC 60601-1:1995)
Attention! Please refer to this manual.
%SpO2
Oxygen saturation of arterial blood
/Min
Pulse rate
Non-Alarm indication (The device does not have alarm
function)
IPX1
Enclosure degree of ingress protection.
SN
Serial number
Symbol for the marking of electrical and electronics devices
according to Directive 2002/96/EC.
The device, accessories and the packaging have to be
disposed of waste correctly at the end of the usage. Please
follow Local Ordinances or Regulations for disposal.
Note: The Oximeter is applied to this regulation.
Guidance and manufacture’s declaration–electromagnetic
ermissions-for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration – electromagnetic
emission
The Fingertip Pulse Oximeter is intended for use in the electromagnetic
environment specified below. The customer of the user of the Fingertip
Pulse Oximeter should assure that it is used in such and environment.
Emission
test
Compliance
Electromagnetic environment –
guidance
RF
emissions
CISPR 11
Group 1
The Fingertip Pulse Oximeter uses RF
energy only for its internal function.
Therefore, its RF emissions are very low
and are not likely to cause any
interference in nearby electronic
equipment.
RF
emission
CISPR 11
Class B
The Fingertip Pulse Oximeter is suitable
for use in all establishments, including
domestic establishments and those
directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes.
Applicable Models
JKS50B
Packing List
NO.
Item
Quantity
1
Oximeter
1
2
AAA battery
2
3
Cord
1
4
User’s manual
1
ALL RIGHTS RESERVED
Guangdong HealthTree Medical Technology Co., Ltd.
ADD: 710, Area B, Dazhou Science and Technology Park, Xiangzhou
District, Zhuhai City,Guangdong Province,China
Tel:+86-13631235424
E-mail:[email protected]om
Website: www.healthtreeltd.com

FCC Caution.
§ 15.19 Labelling requirements.
This device complies with part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) This device may not cause harmful interference, and (2) this device must accept
any interference received, including interference that may cause undesired operation.
§ 15.21 Information to user.
Any Changes or modifications not expressly approved by the party responsible for compliance
could void the user's authority to operate the equipment.
§ 15.105 Information to the user.
Note: This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation. This equipment generates
uses and can radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If this equipment does cause
harmful interference to radio or television reception, which can be determined by turning the
equipment off and on, the user is encouraged to try to correct the interference by one or more of
the following measures:
-Reorient or relocate the receiving antenna.
-Increase the separation between the equipment and receiver.
-Connect the equipment into an outlet on a circuit different from that to which the receiver is
connected.
-Consult the dealer or an experienced radio/TV technician for help.
.
Popular Medical Equipment manuals by other brands

Omron
Omron HJ-320-E quick start guide

SamanTree Medical
SamanTree Medical Histolog Scanner v2 Instructions for use

Arthrex
Arthrex Lift-Assist Beach Chair Instructions for use

Nidek Medical
Nidek Medical Nuvo8 user guide

BioSphere Medical
BioSphere Medical Sequitor manual

Metrohm
Metrohm 940 Professional IC Vario ONE/ChS/LPG manual