Helmer PC900i User manual

360096-1/P
Platelet Incubator Service Manual
i.Series®and Horizon Series™
Countertop
i.Series
PC100i, PC900i, PC1200i
PC100h, PC900h, PC1200h
Floor
i.Series
PC2200i, PC3200i, PC4200i
PC2200h, PC3200h, PC4200h

Document Updates
The document is furnished for information use only, is subject to change without notice and should not be construed as a commitment
by Helmer Scientic. Helmer Scientic assumes no responsibility or liability for any errors or inaccuracies that may appear in the
informational content contained in this material. For the purpose of clarity, Helmer Scientic considers only the most recent revision of
this document to be valid.
Notices and Disclaimers
Condential / Proprietary Notices
Use of any portion(s) of this document to copy, translate, disassemble or decompile, or create or attempt to create by reverse
engineering or otherwise the information from Helmer Scientic products is expressly prohibited.
Copyright and Trademark
Helmer®, i.Series®, i.Center®, Horizon Series™, AgiTrak™, and Rel.i™ are registered trademarks or trademarks of Helmer, Inc. in
the United States of America. Copyright © 2017 Helmer, Inc. All other trademarks and registered trademarks are the property of their
respective owners.
Helmer, Inc., doing business as (DBA) Helmer Scientic and Helmer.
Disclaimer
This manual is intended as a guide to provide the operator with necessary instructions on the proper use and maintenance of certain
Helmer Scientic products.
Any failure to follow the instructions as described could result in impaired product function, injury to the operator or others, or void
applicable product warranties. Helmer Scientic accepts no responsibility for liability resulting from improper use or maintenance of its
products.
The screenshots and component images appearing in this guide are provided for illustrative purposes only, and may vary slightly from
the actual software screens and/or product components.
Helmer Scientic
14400 Bergen Boulevard
Noblesville, IN 46060 USA
www.helmerinc.com
ISO 13485:2003 CERTIFIED Part No. 360096-1/ Rev P
Document History
Revision Date CO Supersession Revision Description
I 28 JUN 2013 8414 Supersedes A, B, C,
D, E, F, G, H Revised layout for ease of navigation and locating information.
J 23 DEC 2013 8965 J supersedes I Added 100 V information.
Added caution note for remote alarm interface.
Revised section I for consistency with existing manuals.
K 31 JAN 2014 9113 K supersedes J Changed voltage for remote alarm contacts.
L22 APR 2014* 9386 L supersedes K Revised voltage specication for remote alarm contacts.
M 26 MAY 2014* 9497 M supersedes L Added 0086 to CE certication.
Revised caption under photograph (chapter 16.1).
Revised power failure alarm test (chapter 19.2).
N 21 SEP 2015* 9754 N supersedes M
Changed part number for terminal board (chapters 12 and 21)
Revised schematic to correct labeling of J1 and J3 on the display board (page
78).
Changed test procedure for No Battery alarm test (chapter 8.8.7).
O 12 JUL 2016 12016 O supersedes N Updated Power Failure test for Horizon units
P 6 DEC 2017 13171 P supersedes O Reformatted content. Updated Emergo address. Updated condensing unit and
refrigerant charge information for PC2200, PC3200 and PC4200, 115 V and 230 V
models.
* Date submitted for Change Order review. Actual release date may vary.

Helmer Scientic Platelet Incubator Service Manual
360096-1/P ii
Contents
1 About this Manual................................................................................................ 5
1.1 Safety Precautions and Symbols ...............................................................................5
1.2 Product Labels .............................................................................................6
1.3 Control System .............................................................................................7
1.4 Temperature Probes ......................................................................................... 7
1.5 Chart Recorder .............................................................................................8
i.Series®Models ..............................................................9
2 InstallationandConguration ...................................................................................... 9
2.1 Install Batteries for Backup Power ..............................................................................9
2.2 Platelet Agitator and Platelet Incubator Compatibility ................................................................ 9
2.3 Install a Platelet Agitator in the Platelet Incubator.................................................................. 10
2.4 External Monitoring Devices ..................................................................................10
2.5 Congure Drawers (PC4200i) ..................................................................................11
2.6 Drawer Labels..............................................................................................11
3 TemperatureMonitorSettings ..................................................................................... 12
3.1 Home Screen .............................................................................................13
3.2 Main Screen ..............................................................................................14
3.3 Change Conguration Password ..............................................................................15
3.4 Calibrate Temperature Probes ................................................................................ 15
3.5 Factory Default Settings .....................................................................................16
3.6 Alarm Reference...........................................................................................16
3.7 Alarm Settings............................................................................................. 17
3.8 Alarm Tests ...............................................................................................20
3.9 AgiTrak™ System ..........................................................................................24
3.10 Additional System Settings ...................................................................................25
3.11 Event Log ................................................................................................29
3.12 System Firmware Upgrade ...................................................................................30
3.13 Reset the i.Center Monitoring System ..........................................................................30
3.14 View Manufacturer and Product Information...................................................................... 30
4 Temperature Controller Setpoints.................................................................................. 31
4.1 Level 1 Parameters and Values ............................................................................... 32
4.2 Level 2 Parameters and Values ............................................................................... 33
4.3 Level 3 Parameters and Values ............................................................................... 33
4.4 Level 4 Parameters and Values ............................................................................... 34
4.5 Platelet Incubator Setpoint ...................................................................................34
4.6 Calibrate the Temperature Controller Display ..................................................................... 35
5 Maintenance.................................................................................................... 36
5.1 Recharge Refrigerant .......................................................................................36
5.2 i.Center Batteries ..........................................................................................37
5.3 Motion Alarm Battery (PC4200i) ...............................................................................37
5.4 Clean the Platelet Incubator ..................................................................................38
5.5 Lubricate the Drawer Slides (PC4200i).......................................................................... 38
5.6 Lubricate and Service Internal Parts (PC4200i) ................................................................... 39

Helmer Scientic Platelet Incubator Service Manual
360096-1/P iii
6 Troubleshooting ................................................................................................ 40
6.1 General Operation Problems .................................................................................40
6.2 Chamber Temperature Problems .............................................................................. 42
6.3 Alarm Activation Problems ...................................................................................42
6.4 Testing Problems...........................................................................................44
6.5 Condensation Problems .....................................................................................45
7 i.Series Parts ................................................................................................... 46
8 Schematics .................................................................................................... 53
8.1 PC100i, PC900i, PC1200i (100/115 V) .......................................................................... 53
8.2 PC100i, PC900i, PC1200i (230 V) .............................................................................54
8.3 PC2200i (115 V) ........................................................................................... 55
8.4 PC2200i (230 V) ...........................................................................................56
8.5 PC3200i (115 V) ........................................................................................... 57
8.6 PC3200i (230 V) ...........................................................................................58
8.7 PC4200i (115 V) ........................................................................................... 59
8.8 PC4200i (230 V) ...........................................................................................60
9 i.Center Screen Reference ........................................................................................ 61
Horizon Series™ Models ......................................................63
10 InstallationandConguration ..................................................................................... 63
10.1 Install Battery for Backup Power...............................................................................63
10.2 Platelet Agitator and Platelet Incubator Compatibility ............................................................... 63
10.3 Install a Platelet Agitator in the Platelet Incubator.................................................................. 64
10.4 External Monitoring Devices ..................................................................................64
10.5 Conguring Drawers (PC4200h) ...............................................................................65
10.6 Drawer Labels.............................................................................................66
11 AlarmSettings .................................................................................................. 67
11.1 Temperature and Power Failure Alarm Volume.................................................................... 67
11.2 Motion Alarm (PC4200h)..................................................................................... 68
11.3 Alarm Tests ............................................................................................... 69
12 Temperature Controller Setpoints .................................................................................. 72
12.1 Level 1 Parameters and Values ...............................................................................73
12.2 Level 2 Parameters and Values ...............................................................................74
12.3 Level 3 Parameters and Values ...............................................................................74
12.4 Level 4 Parameters and Values ...............................................................................75
12.5 Change the Platelet Incubator Setpoint ......................................................................... 75
12.6 Calibrate the Temperature Controller Display ..................................................................... 75
13 Maintenance.................................................................................................... 77
13.1 Recharge Refrigerant ....................................................................................... 77
13.2 Power Failure Alarm Battery ..................................................................................78
13.3 Motion Alarm Battery (PC4200h) .............................................................................. 78
13.4 Clean the Platelet Incubator .................................................................................. 79
13.5 Lubricate the Drawer Slides (PC4200h) ......................................................................... 79
13.6 Lubricate and Service Internal Parts (PC4200h)................................................................... 80

Helmer Scientic Platelet Incubator Service Manual
360096-1/P iv
14 Troubleshooting ................................................................................................ 81
14.1 General Operation Problems .................................................................................81
14.2 Chamber Temperature Problems ..............................................................................83
14.3 Alarm Activation Problems ................................................................................... 83
14.4 Condensation Problems ..................................................................................... 85
15 Horizon Parts................................................................................................... 86
16 Schematics .................................................................................................... 93
16.1 PC100h, PC900h, PC1200h (100/115 V) ........................................................................ 93
16.2 PC100h, PC900h, PC1200h (230 V) ........................................................................... 94
16.3 PC2200h (115 V)...........................................................................................95
16.4 PC2200h (230 V) ..........................................................................................96
16.5 PC3200h (115 V)...........................................................................................97
16.6 PC3200h (230 V) ..........................................................................................98
16.7 PC4200h (115 V)...........................................................................................99
16.8 PC4200h (230 V) .........................................................................................100
Appendix A: Compliance............................................................................................ 101
Appendix B: Warranty .............................................................................................. 102

Helmer Scientic Platelet Incubator Service Manual
360096-1/P 5
1 About this Manual
This manual covers all platelet incubators, which may be identied singly, by their size, or by their respective “Series” and
provides information on how to use the device. It is intended for use by end users of the platelet incubator and authorized service
technicians.
Generic references are used throughout this manual to group models that contain similar features. For example, “PC100 models”
refers to all models of that size (PC100i, PC100h).
1.1 Safety Precautions and Symbols
Symbols found in this document
The following symbols are used in this manual to emphasize certain details for the user:
Task Indicates procedures which need to be followed.
Note Provides useful information regarding a procedure or operating technique when using Helmer
Scientic products.
NOTICE Advises the user against initiating an action or creating a situation which could result in damage to
equipment; personal injury is unlikely.
CAUTION Advises the user against initiating an action or creating a situation which could result in damage to
equipment or impair the quality of the products or cause minor injury.
WARNING Advises the user against initiating an action or creating a situation which could result in damage to
equipment and serious personal injury to a patient or the user.
Authorized representative in the European Community
Symbols found on the units
The following symbols may be found on the refrigerator or refrigerator packaging:
CE Mark (European units only) Caution: Unlock all casters
Caution: Risk of damage to equipment or
danger to operator Earth / ground terminal
Caution: Hot surface Protective earth / ground terminal
Caution: Shock / electrical hazard Product falls under the scope of the WEEE
(Waste Electrical and Electronic Equipment)
directive.
EC REP

Helmer Scientic Platelet Incubator Service Manual
360096-1/P 6
Avoiding Injury
Review safety instructions before installing, using, or maintaining the equipment.
♦Before moving unit, remove contents from the drawers (if applicable).
♦Do not open multiple drawers at the same time (if applicable).
♦Before moving unit, ensure door(s) are closed and casters (if applicable) are unlocked and free of debris.
♦Before moving unit, disconnect the AC power cord and secure the cord.
♦When moving unit, use assistance from a second person.
♦Never physically restrict any moving component.
♦Avoid removing electrical service panels and access panels unless so instructed.
♦Keep hands away from pinch points when closing the door or when agitation motion is enabled (if applicable).
♦Avoid sharp edges when working inside the electrical compartment.
♦Ensure biological materials are stored at recommended temperatures determined by standards, literature, or good
laboratory practices.
♦Proceed with caution when adding and removing samples from the platelet incubator.
♦Use supplied power cord only.
♦ Using the equipment in a manner not specied by Helmer Scientic may impair the protection provided by the equipment.
♦Ensure biological materials are stored safely, in accordance with all applicable organizational, regulatory, and
legal requirements.
♦ The platelet incubator is not considered to be a storage cabinet for ammable or hazardous materials.
NOTICE
Decontaminate parts prior to sending for service or repair. Contact Helmer Scientic or your distributor for
decontamination instructions and a Return Authorization Number.
1.2 Product Labels
Note
Service information varies depending on the model and power requirements.
This information appears on the product specication label, on the back of the platelet incubator.
REF
SN
PC2200i
i.Series®
1000000
Voltage 115 V
HZ 50/60
Amps 14.5 A
Power 1.61kW
Fuse 10.0A
Platelet Storage System
Weight 363 lb / 165 kg
Helmer, Inc.
14400 Bergen Boulevard
Noblesville, IN USA
www.helmerinc.com
2017-12-05
Made in USA
0086
Ref Type Amount Design Pressures GWP CO2 Equ
R134A 8oz/.23kg 88 psig/186 psig 1430 324kg
C
A
B
Sample Product Specication label.
Label Description
A Model
B Serial number (SN)
C Power requirements

Helmer Scientic Platelet Incubator Service Manual
360096-1/P 7
1.3 Control System
Note
Service information varies depending on the control system.
i.Center Control System
i.Series platelet incubators are equipped with the i.Center monitoring system. A separate temperature controller is located behind
the control panel on the platelet incubator.
i.Center monitoring system. Temperature controller.
Horizon Series Control System
Horizon Series platelet incubators feature the combined temperature monitor and controller. The Horizon Series system controls
chamber temperature and monitors and displays operational information.
Horizon Series temperature controller and display.
1.4 Temperature Probes
Number and location of probes varies by model:
Model Location
PC100 Front, top right corner
PC900 Back, right bottom corner
PC1200 Back, right bottom corner
PC2200 Front, right, between upper shelf and top of cabinet
PC3200 Front, right, between upper shelf and top of cabinet
PC4200 Front, right, next to middle shelf

Helmer Scientic Platelet Incubator Service Manual
360096-1/P 8
1.5 Chart Recorder
If installed, refer to the Temperature Chart Recorder Operation and Service Manual.
The chart recorder has a battery system, enabling a period of continuous operation if power is lost. Battery life varies by
manufacturer as well as voltage level remaining. Providing full power is available, backup power for the temperature chart recorder
is available for up to 14 hours.
Prior to use:
♦Install battery.
♦Add paper.
♦Calibrate chart recorder to match upper chamber temperature.
Open door by pressing and releasing.
Install Chart Paper
Note
For accurate temperature reading, ensure that current time is aligned with time line groove when chart knob is tightened.
1. Press and hold C button. When stylus begins to move left, release button. The LED ashes to indicate current
temperature range.
2. When stylus stops moving, remove chart knob then move knob up and away.
3. Place chart paper on chart recorder.
4. Gently lift stylus and rotate paper so current time line corresponds to time line groove.
5. Hold chart paper and reinstall chart knob.
6. Conrm temperature range is set to the correct value.
7. Press and hold C button. When stylus begins to move right, release button.
8. Conrm stylus is marking temperature correctly.

Helmer Scientic Platelet Incubator Service Manual
360096-1/P 9
i.Series®Models
2 InstallationandConguration
2.1 Install Batteries for Backup Power
The monitoring system and chart recorder each have a battery system, enabling a period of continuous operation if power is lost.
Battery life varies by manufacturer as well as voltage level remaining. Providing full power is available and no battery-related
alarms are active, backup power for the monitoring system is available for up to two hours.
The batteries are located on the top of the platelet incubator, under a removable access panel.
Batteries are disconnected for shipping. Reconnect batteries to provide monitoring system with backup power in the event of AC
power failure.
Notes
• The monitoring system will start on battery power alone. If the platelet incubator was previously not connected toAC
power and the batteries are connected, the monitoring system will begin running on battery power.
• IfAC power is lost, the monitoring system will continue to collect data until battery power is depleted.
CAUTION
Before installing or replacing batteries, switch the power OFF and disable backup power to the i.Center monitoring system
by switching the ON/OFF key switch OFF (located behind the control door). Disconnect the platelet incubator from AC
power.
NOTICE
Use only manufacturer supplied battery when installing a replacement battery.
2.2 PlateletAgitatorandPlateletIncubatorCompatibility
i.Series platelet incubators may be tted with i.Series platelet agitators. When installed in an i.Series platelet incubator, the i.Series
platelet agitator can be monitored by the platelet incubator’sAgiTrak™ system.
The AgiTrak system receives alarm information through a direct connection from the platelet agitator’s alarm system to the platelet
incubator’s alarm system.
Note
Only i.Series platelet incubators can monitor i.Series platelet agitators.
Platelet Incubator CompatibleHelmerPlateletAgitator
PC100i PF15i
PC900i PF48i
PC1200i, PC2200i, PC3200i PF96i
i.Series Information

Helmer Scientic Platelet Incubator Service Manual
360096-1/P 10
2.3 InstallaPlateletAgitatorinthePlateletIncubator
Notes
• This section is not applicable to the PC4200i platelet incubator, which has a built-in platelet agitator.
• For additional instruction on conguring a platelet agitator in a platelet incubator, refer to the platelet agitator service manual.
PC900i platelet incubator with a PF48i platelet agitator installed.
Helmer platelet agitators can be installed inside Helmer platelet incubators. i.Series platelet agitators can be connected to i.Series
platelet incubators via a data cable. The connection allows the platelet incubator to monitor the platelet agitator motion, and to
sound an alarm when agitation motion stops.
PC2200i and PC3200i platelet incubators hold 1 agitator per shelf. Each shelf features a corresponding power connector. In
PC2200i and PC3200i platelet incubators, each shelf includes a data port. Through the data port and data cable, agitation motion
is communicated from i.Series platelet agitators to the i.Series platelet incubator.
When installing a platelet agitator on a shelf, use the power connector for that shelf to prevent the power cord from obstructing the
agitation motion. When installing an i.Series platelet agitator on a shelf in an i.Series platelet incubator, use the data port for that
shelf so agitation motion is communicated appropriately.
2.4 ExternalMonitoringDevices
The remote alarm interface is a relay switch with three terminals:
♦Common (COM)
♦Normally Open (NO)
♦Normally Closed (NC)
Terminals are dry contacts and do not supply voltage. Interface circuit is either normally open or normally closed, depending on
terminals used.
Requirements for your alarm system determine which alarm wires must connect to terminals.
CAUTION
• The interface on the remote alarm monitoring system is intended for connection to the end user’s central alarm
system(s) that uses normally-open or normally-closed dry contacts.
• If an external power supply exceeding 30 V (RMS) or 60 V (DC) is connected to the remote alarm monitoring system’s
circuit, the remote alarm will not function properly; may be damaged; or may result in injury to the user.
NOTICE
• For any platelet incubator (except PC4200i) without a platelet agitator installed, a bypass plug is needed for a normally-
closed interface circuit.
• The bypass plug is needed if multiple platelet agitators are installed, and not all are Helmer i.Series platelet agitators.
For each position in which a non-Helmer i.Series platelet agitator is installed, the bypass plug must be installed in the
data port.
• To obtain a bypass plug, contact Helmer Technical Service.
The terminals on the remote alarm interface have the following maximum load capacity:
♦0.5 A at 30 V (RMS)
♦1.0 A at 24 V (DC)
i.Series Information

Helmer Scientic Platelet Incubator Service Manual
360096-1/P 11
Connect to Remote Alarm Interface
1. Switch the Battery ON/OFF key switch OFF. Switch the AC ON/OFF switch OFF.
2. On the alarm panel, locate the remote alarm terminals.
3. Connect remote alarm wires to appropriate terminals, according to requirements for your
alarm system.
4. Use a cable tie to relieve strain on alarm wires (as necessary).
5. Switch the Battery ON/OFF switch ON. Switch the AC ON/OFF switch ON.
6. Touch MUTE to disable the high temperature alarm while platelet incubator reaches
operating temperature.
2.5 CongureDrawers(PC4200i)
CAUTION
• Before moving drawers, ensure they are completely empty for safe lifting.
• Open the drawer using the drawer handle; not the label holder.
Remove / Install Drawer
Drawer in the full open position. Drawer with the tab in front of the drawer stop.
1. Pull drawer out until it stops (when the tabs on the drawer contact the spring-loaded stops).
2. Depress the spring-loaded drawer stops.
3. Pull the drawer outward until the tabs on the drawer have passed the drawer stops.
4. Pull drawer free of the drawer slides.
5. Align drawer with the drawer slides.
6. Push drawer in until it stops (when the tabs on the drawer contact the spring-loaded stops).
7. Depress the spring-loaded drawer stops.
8. Push the drawer into the storage frame until it stops.
2.6 Drawer Labels
Drawer label holders may be installed on platelet agitator drawers.
Installed label holder.
Install Label Holders:
1. Insert the tabs on the label holder into the slots on the drawer.
2. Pivot the holder around the drawer handle. Align the hole on the label holder with the corresponding hole on the drawer.
3. Insert a screw through the hole in label holder and through the hole in the drawer.
4. Install the nut on the screw, on the back side of the label holder.
Remote alarm interface.
i.Series Information

Helmer Scientic Platelet Incubator Service Manual
360096-1/P 12
3 TemperatureMonitorSettings
Through the i.Center monitoring system, current settings may be viewed and changed. The settings that can be viewed or changed
are as follows:
Setting Description
Clock Mode Time format used on the HOME screen and the Event Log
Date Format Date format used on the HOME screen and on the Event Log
Door Ajar Timeout Number of minutes the door can be open before an alarm activates
Pwr Failure Timeout Number of minutes the platelet incubator can be without AC power before an alarm activates
High Alarm Setpoint Temperature at which the High Temperature Alarm activates
Low Alarm Setpoint Temperature at which the Low Temperature Alarm activates
Cond. Alarm Setpoint Temperature at which the Condenser Temperature Alarm activates
Alarm Volume Volume for audible alarms
Alarm Pulse Pattern for audible alarms
Chart Paper Timer; or Chart
Paper Days Left Setting for the chart paper timer (disabled); or the number of days remaining before chart paper
must be changed (if temperature chart recorder is installed)
Temperature Graph The setting for the temperature graph feature (Enabled or Disabled)
Agitator Alarms Settings for the motion alarm for each Helmer i.Series agitator installed in the platelet incubator, or
built-in agitator (PC4200i)
• All platelet incubators are programmed to monitor three positions, even though some models
accommodate fewer platelet agitators.
• Positions that are not in use are set to OFF.
• Positions that are in use are set to ON and display the number of minutes the agitator can be
stopped before an alarm activates.
Viewsettings
Note
The Conguration screen is password protected. The factory default password is “1234”.
1. On the Home screen, press the MAIN button.
2. Press the DOWN button to highlight View Conguration. Press the SELECT button.
3. Press the UP or DOWN button to highlight the desired option. Press the SELECT button.
4. Press the BACK button to return to the View Conguration screen, or press the HOME button to exit.
NOTICE
Changing temperature monitor settings affects operation of the platelet incubator. Do not change settings unless
instructed in product documentation or by Helmer Technical Service.
i.Series Information

Helmer Scientic Platelet Incubator Service Manual
360096-1/P 13
3.1 Home Screen
The Home screen appears when the HOME button is touched from any other screen, or there is no interaction for two minutes on
any screen other than the password entry screen.
HOME screen on the monitoring system.
Home Screen Functions
Note
Refer to section 9 (i.Center Screen Reference) for a complete list of screens in the i.Center monitoring system.
♦View current temperature readings
♦View the current time and date
♦View detailed information about current or previous alarm events
♦View the remaining backup battery charge
♦View active alarms
♦Mute audible alarms
♦Adjust contrast
♦View 24-hour chamber temperature graph
♦Access Main screen to view and change settings
i.Series Information

Helmer Scientic Platelet Incubator Service Manual
360096-1/P 14
3.2 Main Screen
The Main screen displays functional options that allow access to all other screens in the system.
MAIN screen functional options.
Functions available from the Main screen:
Option Function
Event Log View historical information about alarms and operational events
System Alarm Test & Status Start or stop an automatic test for temperature alarms
View the number of days remaining before the paper for the temperature chart recorder
needs to be changed
View the current status of the door (OPEN or CLOSED)
View the current condenser temperature
Edit Conguration (password
required) Change the language used for text
Change date and time information
Change temperature units
Change the volume and pattern for audible alarms
Enable or disable the chart paper timer
Enable or disable the temperature graph display
Change alarm-related setpoints and timers
Calibrate the temperature probe reading
Change some settings to the factory default values
Change the password, preventing unauthorized changes
Congure the AgiTrak system to monitor agitation
View Conguration View the date and time formats
View alarm-related setpoints and timers
View the volume and pattern for audible alarms
View the setting for the chart paper timer
View the setting for the temperature graph display
View the settings for agitator alarms
Product/Company Information View the software versions for control and display components of the monitoring system
View information to contact Helmer
i.Help Access the on-board help system
View real-time agitation speed and cycle counts
View historical agitation event logs and details
Reset cycle counters for agitators
i.Series Information

Helmer Scientic Platelet Incubator Service Manual
360096-1/P 15
3.3 ChangeCongurationPassword
The default password is 1234. A new password must use four digits, ranging from 1 to 5.
Changethepassword:
1. On the Home screen, press the MAIN button.
2. Press the DOWN button to select Edit Conguration. Press the SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to select Change Password. Press the SELECT button.
5. Enter the new password, then re-enter the new password when prompted. If password entries match, the “update”
message is displayed. If password entries do not match, the “incorrect match” message is displayed.
6. Repeat the procedure to change the password.
3.4 Calibrate Temperature Probes
Verify the temperature probe(s) are reading chamber temperature correctly by comparing the chamber probe reading to
temperature read by an independent thermometer. If the chamber temperature probe(s) are not reading correctly, change the value
displayed on the temperature monitoring system.
Notes
• If the variance is within acceptable limits for your organization, changing probe settings is optional.
• Default setting for chamber temperature is 22.0 ºC. Value is factory-preset
• Value can be changed from -40.0 ºC to +40.0 ºC
CalibrateProbe(s)
Model Location Model Location
PC100i Front, top right corner PC3200i Upper reading: next to upper temperature probe
Lower reading: next to lower temperature probe
PC900i, PC1200i Back, right bottom corner PC4200i Upper or lower reading: in the middle shelf on the right side
PC2200i Next to upper temperature probe
1. Using the table above, place a calibrated independent reference thermometer traceable per national standards in the
specied location.
2. Close the door and allow the chamber temperature to stabilize for 30 minutes.
3. Observe and note the thermometer temperature.
4. On the Home screen, press the MAIN button.
5. Press the DOWN button to highlight Edit Conguration. Press the SELECT button.
6. Enter the password when prompted.
7. Press the DOWN button to highlight Temperature Calibration. Press the SELECT button.
8. Follow the system prompts for Model Type.
♦For models with two probes, press the UP or DOWN buttons to select the Upper Probe or Lower Probe options. Press
the SELECT button, then press the DOWN button to highlight the Temperature option.
♦For models with one probe, the Upper Probe option is displayed and the Temperature option is highlighted.
9. Press the INC or DEC buttons to change the displayed value to match the value measured by the
independent thermometer.
10.Press the DOWN button to highlight Store Calibration.
11. To save the new value, press the ENTER button. The “Calibration Memorized” message appears. New settings are saved.
12.To discard the new value, press the BACK button to return to the Edit Conguration screen, or press the HOME button to
exit. New settings are not saved.
13.Remove the thermometer from the chamber and close the door.
Notes
• The current temperature displayed by the monitoring system may change so that it no longer matches the new probe
calibration value. This is normal.
• If a new probe value is entered but not saved, the new value will appear when the calibration setting for the probe is
viewed. This is normal.
i.Series Information

Helmer Scientic Platelet Incubator Service Manual
360096-1/P 16
3.5 FactoryDefaultSettings
Settings listed below may be simultaneously returned to factory default values.
Setting Restored Value
High Alarm Setpoint 24.0 °C
Low Alarm Setpoint 20.0 °C
Cond. Alarm Setpoint 50.0 °C
Door Ajar Timeout 3 minutes
Power Failure Timeout 3 minutes
Agitator 1 Setpoint ON, 3 minutes
Agitator 2 Setpoint ON, 3 minutes
Agitator 3 Setpoint ON, 3 minutes
RestoreFactoryDefaultSettings
1. Press the MAIN button.
2. Press the DOWN button to highlight Edit Conguration. Press the SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight Factory Default Settings. Press the SELECT button.
5. Press the ENTER button to restore the Factory default settings.
3.6 Alarm Reference
If an alarm condition is met, an alarm activates. Some alarms are visual only; others are visual and audible. Some alarms are sent
through the remote alarm interface.
The table indicates if an alarm is audible (A), visual (V), or sent through the remote alarm interface (R).
Alarm Alarm Type
High Temperature A, V, R
Low Temperature A, V, R
Condenser Temperature A, V, R (i.Series)
Door Open A, V, R (i.Series)
Low Battery V (i.Series)
No Battery A, V, R (i.Series)
AC Power Failure A, V, R
Change Chart Paper (1) V (i.Series)
Agitator 1 Motion Alarm (2) A, V, R (i.Series);A, R (Horizon Series)
Agitator 2 Motion Alarm (3, 5) A, V, R (i.Series); R (Horizon Series)
Agitator 3 Motion Alarm (4, 5) A, V, R (i.Series); R (Horizon Series)
(1) Applicable if a temperature chart recorder is installed. Additional alarms may be available on the stand-alone chart recorder. Refer to the
Temperature Chart Recorder operation and service manual.
(2) Applicable to PC4200 platelet incubators. For other models, this alarm applies if a Helmer i.Series platelet agitator is installed as Agitator 1.
(3) Applicable to PC2200 and PC3200 platelet incubators with a Helmer i.Series platelet agitator installed as Agitator 2.
(4) Applicable to PC3200 platelet incubators with a Helmer i.Series platelet agitator installed asAgitator 3.
(5) Not applicable to PC4200h platelet incubators.
i.Series Information

Helmer Scientic Platelet Incubator Service Manual
360096-1/P 17
3.7 AlarmSettings
The following alarm settings may be changed by the operator. The setpoint for temperature alarms may be changed (where
applicable), as well as the time delay between when the alarm condition commences and when the visual and audible alarms are
initiated.
Setting DefaultSetting Default Time Delay
Alarm Volume 10 n/a
Alarm Pulse Single n/a
High Alarm Setpoint 24.0 °C n/a
Low Alarm Setpoint 20.0 °C n/a
Condenser Alarm 50.0 °C n/a
Power Failure n/a 3 minutes
Door Ajar Timeout n/a 3 minutes
Motion On 3 minutes
Alarm Volume
The Alarm Volume controls volume for all audible alarms. The default setting is 10 and can be changed from 1 to 10 with 10 being
the maximum setting.
ChangetheAlarmVolume
1. Press the MAIN button.
2. Press the DOWN button to highlight Edit Conguration. Press the SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight System Options. Press the SELECT button.
5. Press the DOWN button to highlight Alarm Volume.
6. Press the INC or DEC buttons to change the setting.
7. Press the BACK button to return to Edit Conguration screen, or press the HOME button to exit. The new settings are saved.
Alarm Pulse
The alarm pattern can be changed to provide an easily distinguished sound. The default setting is Single, but may be changed to
Double, Triple or Constant.
ChangetheAlarmPulse
1. Press the MAIN button.
2. Press the DOWN button to highlight Edit Conguration. Press the SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight System Options. Press the SELECT button.
5. Press the DOWN button to highlight Alarm Pulse.
6. Press the INC or DEC buttons to change the setting.
7. Press the BACK button to return to the Edit Conguration screen, or press the HOME button to exit. The new settings
are saved.
HighChamberTemperatureAlarm
The High Alarm setpoint species the temperature at which the High Temperature Alarm activates. If the temperature detected
by the upper chamber probe is greater than or equal to this value, the alarm activates. The default setpoint is 24 °C and can be
changed from -40 °C to +40 °C.
ChangeHighTemperatureSetpoint:
1. Press the MAIN button.
2. Press the DOWN button to highlight Edit Conguration. Press the SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight Alarm Setpoints. Press the SELECT button.
5. Press the DOWN button to highlight High Alarm Setpoint.
6. Press the INC or DEC buttons to change the setting.
7. Press the BACK button to return to Edit Conguration screen, or press the HOME button to exit. The new settings are saved.
i.Series Information

Helmer Scientic Platelet Incubator Service Manual
360096-1/P 18
Low Chamber Temperature Alarm
The Low Alarm setpoint species the temperature at which the Low Temperature Alarm activates. If the temperature detected by
the upper chamber probe is less than or equal to this value, the alarm activates. The default setpoint is 20 °C and can be changed
from -40 °C to +40 °C.
ChangeLowTemperatureSetpoint:
1. Press the MAIN button.
2. Press the DOWN button to highlight Edit Conguration. Press the SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight Alarm Setpoints. Press the SELECT button.
5. Press the DOWN button to highlight Low Alarm Setpoint.
6. Press the INC or DEC buttons to change the setting.
7. Press the BACK button to return to the Edit Conguration screen, or press the HOME button to exit. The new settings
are saved.
Condenser Temperature Alarm
The Condenser Alarm setpoint species the temperature at which the Condenser Temperature Alarm activates. If the temperature
of the condenser discharge line is greater than or equal to this value, the alarm activates. The default setpoint is 50 °C and can be
changed from -40 °C to +80 °C.
NOTICE
Condenser TemperatureAlarm should not be changed unless directed by Helmer Technical Service.
ChangeCondenserTemperatureSetpoint:
1. Press the MAIN button.
2. Press the DOWN button to highlight Edit Conguration. Press the SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight Alarm Setpoints. Press the SELECT button.
5. Press the DOWN button to highlight Cond. Alarm Setpoint.
6. Press the INC or DEC buttons to change the setting.
7. Press the BACK button to return to the Edit Conguration screen, or press the HOME button to exit. The new settings
are saved.
Power Failure Alarm
The Power Failure Timeout species longest time the platelet incubator can be without AC power before the alarm activates. If the
time elapsed since the last power failure is greater than or equal to this value, the alarm activates. The default setting is three (3)
minutes and can be changed from 0 to 60 minutes.
ChangethePowerFailureAlarmDelay:
1. Press the MAIN button.
2. Press the DOWN button to highlight Edit Conguration. Press the SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight Alarm Setpoints. Press the SELECT button.
5. Press the DOWN button to highlight Power Failure Timeout.
6. Press the INC or DEC buttons to change the setting.
7. Press the BACK button to return to the Edit Conguration screen, or press the HOME button to exit. The new settings
are saved.
i.Series Information

Helmer Scientic Platelet Incubator Service Manual
360096-1/P 19
Door Ajar Timeout
The Door Ajar Timeout species longest time the platelet incubator door(s) can be open before the alarm activates. If the time
elapsed since the last door opening is greater than or equal to this value, the alarm activates. The default delay setting is three (3)
minutes and can be changed from 0 to 60 minutes.
ChangetheDoorAjarAlarmDelay:
1. Press the MAIN button.
2. Press the DOWN button to highlight Edit Conguration. Press the SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight Alarm Setpoints. Press the SELECT button.
5. Press the DOWN button to highlight Door Ajar Timeout.
6. Press the INC or DEC buttons to change the setting.
7. Press the BACK button to return to the Edit Conguration screen, or press the HOME button to exit. The new settings
are saved.
MotionAlarm(PC4200i)
Unlike other i.Series models, the PC4200i platelet incubator has a motion alarm delay control. When the platelet agitator motion
stops, this control delays when alarm information is sent to the remote alarm interface on the platelet incubator. The alarm delay
control does not delay motion information sent to the AgiTrak system. When theAgitator Alarm Setpoint is reached, the AgiTrak
system activates an alarm, which is also sent to the remote alarm interface on the platelet incubator.
Notes
• To prevent a motion alarm being communicated by an external monitoring device but not by the platelet incubator, set
the alarm delay control to a value greater than theAgitator 1 Setpoint.
• The alarm delay control for the motion alarm is located on the rear of the platelet incubator.
Top: Motion alarm volume control. Bottom: Alarm delay control for the motion alarm.
The default motion alarm delay is set at the halfway point (approximately four (4) minutes). To change the motion alarm delay, use
a small at-head screwdriver. The maximum alarm delay is eight (8) minutes.
The default alarm volume is set at the middle position (medium).
Note
Do not set the alarm delay to 0 minutes.
ChangeMotionAlarmDelay:
Using a small, at-head screwdriver, rotate the control to the left (counterclockwise) to shorten the motion alarm delay, or to
the right (clockwise) to extend the motion alarm delay.
ChangeMotionAlarmVolume:
1. On back of the platelet incubator, locate the alarm volume control.
2. Position the switch to correspond with the desired motion alarm volume.
i.Series Information
This manual suits for next models
11