hivox PureCare PL-982 User manual

982
TENS EMS MASSAGE
Advanced Pro.
Digital Electric Stimulator
User’s Manual
2
2
SUPER CHANNELS
Handheld, Advanced
Please read this User’s Manual thoroughly
before using your blood glucose meter.
All rights reserved 2009. 450-2301A MADE IN TAIWAN
HIVOX BIOTEK INC. 8F,No.98,Shingde Rd.,
Sanchong City, Taipei County, Taiwan
MEC Electronics B.V.
De Run 4428, 5503 LR
Veldhoven, Netherlands

1
Dear Customer,
Thank you very much for your support on HIVOX’s products. HIVOX products stand for high
quality, reliable performance, and most of all, Innovations for Better Living.
Please read this instruction manual carefully prior to use and keep it for later reference.
CONTENT
1. Introduction
2. Important Information
3. Unit description
4. Start-up
5. Program overview
6. Applications
7. Cleaning and storage
8. Disposal
9. Problems and solutions
10. Technical Specifications
PL982-user180x128.indd 1PL982-user180x128.indd 1 2010/2/9 下午 02:12:072010/2/9 下午 02:12:07

Introduction
Chapter 1
2 3
EMS, Electrical muscle stimulation is a
widespread and generally recognized method
and has been used for years in sports and
rehabilitation medicine. In the sports and
fitness field, one of the uses of EMS is as a
supplement to conventional muscle training,
in order to increase the efficiency of muscle
groups and adapt physical proportions to the
desired aesthetic results. EMS application
goes in two directions. On the one hand, a
targeted strengthening of musculature can
be produced (activating application) and on
the other hand a relaxing, resting effect can
also be achieved (relaxing application).
Active application includes the
following:
• Muscle training to improve endurance
performance and/or
• Muscle training to support the
strengthening of certain muscles or muscle
groups in order to achieve desired changes
to body proportions.
Relaxing application includes the
following:
• Muscle relaxation in order to loosen up
muscular tension Massage
• Improving muscular fatigue symptoms
• Accelerating muscle regeneration after
high muscular performance (e.g. after a
marathon).
Massage, through integrated massage
technology, Digital TENS/EMS also offers the
possibility of reducing muscular tension and
combating fatigue symptoms using a program
based on real massage in terms of sensation
and effect. Using the positioning suggestions
and program charts in this manual, you can
determine the device setting fast and easily
for the respective application (depending on
the region of the body affected) in order to
ensure the desired effect.
Through the two separately adjustable
channels, the Digital TENS/EMS has the
advantage of adapting the intensity of the
pulses independently of each other to two
parts of the body to be treated, for example
in order to cover both sides of the body or
ensure even stimulation of larger areas of
tissue. The individual intensity setting of
each channel also allows you to treat two
different parts of the body at the same time,
whereby it is possible to save time compared
with individual sequential treatment.
TENS, transcutaneous electrical nerve
stimulation, means electrical stimulation
of nerves through the skin. TENS is
recognized as a clinically proven, effective,
non-medication method of treating pain from
certain causes. It is free from side-effects
when used properly, and can also thereby be
used as a simple means of self-treatment.
The pain-alleviating or suppressing effect is
also achieved by preventing pain from being
passed on into nerve fibers (above all through
high-frequency pulses) and increasing the
secretion of the body’s own endorphins
which reduce awareness of pain through
their effect on the central nervous system.
The method is scientifically underpinned and
medically approved.
Any symptoms indicating TENS application
must be clarified with the doctor in charge
of your treatment. The latter will also give
you information on the respective benefits of
TENS self-treatment.
TENS is clinically tested and approved
with the following applications:
• Back pain, particularly also lumbar and
cervical spine problems
• Joint pain (e.g. knee joint, hip joint,
shoulder)
• Neuralgia
• Headaches
• Women's period pains
• Pain after injuries to musculoskeletal
system
• Pain with circulatory problems
• Chronic pain through various causes
With HIVOX’s PureCare PL-982, you can find this product provides almost
everything you might need for daily usage.
There are 3 different modes for you to find the best program which fits your
needs as following.
PL982-user180x128.indd 2-3PL982-user180x128.indd 2-3 2010/2/9 下午 02:12:082010/2/9 下午 02:12:08

Chapter 2
4 5
WARNING!
In order to prevent any damage to
health, we would urgently advise
against using the Digital TENS/EMS in
the following cases:
• With any implanted electrical devices
(such as pacemakers)
• If there are any metal implants
• When using an insulin pump
• In case of high fever (e.g. > 39°)
• In case of known or acute cardiac
arrhythmia and other disorders in
• stimulus formation and conduction of the
heart
• If suffering from attacks
(e.g. epilepsy)
• If pregnant
reddening, skin rashes (e.g. allergies),
burns, bruises, swellings, open wounds
and wounds in the process of healing, on
operation scars in the process of healing)
• In environments with high humidity such
as in the bathroom or when having a bath
or hower
• Do not use after consuming alcohol
• When a high frequency surgical device is
connected at the same time
Before using the device, you should
consult with the doctor in charge of
your treatment in the event of the
following:
• Acute diseases, especially if hypertension
is suspected or actually exists, or
blood-clotting disorders, tendency to
thromboembolic diseases or in case of
malignant new masses
• All skin diseases
• Chronic pain disorders which have not
been clarified, regardless of the region of
the body
• Diabetes
• All sensitivity disorders with reduced
pain awareness (such as metabolic
disturbances)
Important Information
• In case of cancer
• After operations where increased muscle
contractions might impair the healing
process
• Never use near the heart. Stimulation
electrodes should never be placed
anywhere on the front of the thorax
(marked by ribs and breastbone), but
above all not on the two large pectoral
muscles. Here it can increase the risk of
ventricular fibrillation and lead to cardiac
arrest
• On the bony part of the cranium, near the
mouth, throat or larynx
• Near the throat / carotid artery
• Near the genitals
• On acutely or chronically diseased (injured
or inflamed) skin (e.g. in the event
of painful and painless inflammation,
Use of the device does not replace medical consultation and treatment.
In the event of any type of pain or illness, you must therefore always first
ask your doctor.
• Medical treatments carried out
simultaneously
• Complaints arising as a result of the
stimulation treatment
• Constant skin irritation as a result of
prolonged stimulation at the same
electrode point.
PL982-user180x128.indd 4-5PL982-user180x128.indd 4-5 2010/2/9 下午 02:12:092010/2/9 下午 02:12:09

6 7
IMPORTANT!
Use the Digital TENS/EMS solely:
• On humans
• For the purpose for which it was
developed and in the manner specified in
these instructions for use. Any improper
use can be hazardous.
• For external application.
• With the original accessory parts which
are supplied and can be re-ordered,
otherwise the warranty becomes null and
void.
PRECAUTIONARY MEASURES:
• Always remove the electrodes from the
skin with a moderate pull in order to avoid
injury in the event of highly sensitive skin.
• Keep the device away from heat sources
and never use it near (~1 m) short-wave
or micro-wave equipment (e.g. mobile
phones), as this may lead to unpleasant
current spikes.
• Do not expose the device to direct
sunlight or high temperatures.
• Protect the device from dust, dirt and
humidity. Never submerge the device in
water or other liquids.
• The device is suitable for private use.
• For reasons of hygiene, the electrodes
should only be used on one person.
• If the device is not functioning properly, or
if you start feeling unwell or there is any
pain, stop the application immediately.
• To remove or move the electrodes, first
switch off the device or the appropriate
channel in order to avoid unwanted
irritation.
• Never modify any electrodes (e.g.
by cutting). This leads to a higher
current density and can be dangerous
(max. recommended output value for
electrodes: 9 mA/cm2, an effective cur-
rent density over 2 mA/cm2 requires
greater attention).
• Do not use when asleep, when driving a
vehicle or at the same time as operating
equipment.
• Never use with any activities where an
unforeseen reaction (e.g. increased
muscle contraction despite low intensity)
may be dangerous.
• Make sure that no metallic objects such
as belt buckles or necklaces can come
in contact with the electrodes during
stimulation. If you wear any jewellery or
piercings (e.g. belly piercing) near the
application site, you must remove these
before using the device as burning may
otherwise occur at some points.
• Keep the device away from children to
prevent any hazards.
• Do not confuse the electrode cables
and contacts with your headphones or
other devices, and do not connect the
electrodes to other devices.
• Do not use this device at the same time
as other equipment which sends electrical
pulses to your body.
• Do not use it near any highly flammable
substances, gases or explosives.
• Never use accumulators, always use the
same types of battery.
• In the initial minutes, carry out the
application sitting or lying down in order
to avoid any un-necessary risk of injury in
the rare event of a vagal reaction (feeling
of weakness). If you start feeling weak,
switch the device off immediately and put
your legs up (approx. 5–10 minutes).
• It is not advisable to apply any enriching
creams or ointments to the skin
beforehand, as this greatly increases
electrode wear or else unpleasant current
spikes may occur here.
Damage:
• If the device is damaged, do not use it if
you are in any doubt but rather
• Contact your dealer or the customer
service address indicated.
• Check the device at regular intervals
for signs of wear or damage. If you find
any such signs or if the device was used
improperly, you must bring it to the
manufacturer or dealer before it is used
again.
• Switch off the device immediately if it is
defective or if is not working properly.
• Do not, under any circumstances, attempt
to open or repair the device yourself! Only
have repairs carried out by the customer
service department or an authorized
dealer. If these instructions are not
heeded, the warranty becomes null and
void.
PL982-user180x128.indd 6-7PL982-user180x128.indd 6-7 2010/2/9 下午 02:12:092010/2/9 下午 02:12:09

8 9
3.1 Description of components
7
1
2
3
4
85 6bc d
e
f
a
Display (main menu):
1. Mode Select
2. Program Number
3. Treatment Timer
4. Intensity Indicators
5. Low Battery Indicator
6. Frequency Display
7. Pulse Width Display
8. Body Map Indicator
Keys:
a. UP and DOWN selection keys
b. ENTER key
c. Power key
d. MENU key
e. Intensity adjust dial (CH1/CH2)
f. Pause/Stop key
Accessories:
• 2x connection cable (with 2 independently
adjustable channels)
• 4x adhesive electrodes(45 x 45 mm)
• 1x belt clip
Chapter 3 Unit description
• The manufacturer assumes no liability for
damage caused through improper or wrong
use.
Information on ESD (Electrostatic
Discharge)
Please note that jacks marked with the
ESD warning label must not be touched.
ESD protective measures:
• Do not touch plugs/jacks marked with the
ESD warning label with fingers!
• Do not touch plugs/jacks marked with the
ESD warning label with handheld tools!
Further explanations on the ESD warming
label, as well as possible training courses
and their contents, are available on request
from the customer service department.
PL982-user180x128.indd 8-9PL982-user180x128.indd 8-9 2010/2/9 下午 02:12:102010/2/9 下午 02:12:10

10 11
1. Remove the belt clip from the device, if attached.
2. Press on the battery cover at the back of the device and push it to the side.
3. Insert 3x AAA 1.5 V alkaline batteries. Ensure that you insert the batteries with the correct
polarity as marked.
4. Replace the battery cover carefully.
5. Attach the connecting cable to the electrodes.
Note: The electrodes have a clip fastener to ensure easy connection.
6. Insert the connecting cable connectors into the socket at the bottom of the device
7. Do not pull, twist or make any sharp angles in the cables.
Chapter 4 Start-up
3.2 Key functions
When a key is pressed, this is acknowledged
by a signal tone in order to ensure that any
unintentional pressing of a key is detected.
This signal tone cannot be switched off.
Power key
1. Hit briefly to switch on the device. If the
key is held down for 10 seconds when
switching on, the unit automatically
switches off again.
2.Hold ON/OFF key down (approx. 3
seconds) to switch off the device.
UP and DOWN
1. Select between (A) treatment program, (B)
frequency and (C) treatment time.
2.The DOWN key is used to reduce pulse
intensity for both channels during
stimulation.
MENU
1. Navigation between the submenus TENS,
EMS and MASSAGE.
2. Return to (A) program selection window or
(B) main menu.
3. Press and hole to unlock when keys are
locked during stimulation.
ENTER
1. Confirm selection and move to next step.
Intensity Adjust Dial
1. When stimulating, rotate dials clockwise
to increase intensity and vise versa
2. In manual setting mode (at P0), CH1 dial
is used to adjust pulse width (μs) and CH2
dial is used to set frequency (Hz).
Pause/Stop
1. Press to interrupt stimulation
PL982-user180x128.indd 10-11PL982-user180x128.indd 10-11 2010/2/9 下午 02:12:102010/2/9 下午 02:12:10

12 13
Neck Pain
Trapezium
Lumbar Region
External Abdominal Obligue
Deltoid
Triceps
Large Dorsals
Plantar Flexors
Buttocks
Hamstrings
(Biceps Femoris)
Upper Back Pain
Sciatica
Sciatica
n
ts
Knee Flexors
Adductors
Qudriceps
Ankle
Pectoral Muscles
Deltoid
Large Dorsals
Biceps
Elbow
Wrist
Abdominal Muscl
e
Sore Shoulder
Migraine
PMS
Arthritis At Joints
Pain Types
Sore Muscles
5.1 Basic information
The Digital TENS/EMS unit has a total 30
programs:
• 10 TENS programs
• 10 EMS programs
• 10 MASSAGE programs
With all programs, you can set the length
of application and (for each of the two
channels) pulse intensity separately. Besides
this, with TENS and EMS programs P0, you
can also change the pulse frequency & pulse
width of the individual cycles in order to
adapt the stimulation effect to the structure
of the site of application.
Cycles are the different sequences making
up the programs. They automatically run one
after the other and increase the efficacy of
stimulation on different types of muscle fiber
and combat pre-mature muscle fatigue. You
will find the standard stimulation parameter
settings and information on placing the
electrodes in the following program charts
for TENS, EMS and MASSAGE.
Chapter 5 Program overview
PL982-user180x128.indd 12-13PL982-user180x128.indd 12-13 2010/2/9 下午 02:12:112010/2/9 下午 02:12:11

14 15
PL982-user180x128.indd 14-15PL982-user180x128.indd 14-15 2010/2/9 下午 02:12:132010/2/9 下午 02:12:13

16
SPORTS PROGRAMS
On-Time (sec.) = Cycle switch-on time in seconds (contraction) -
Off-Time (sec.) = Cycle switch-off time in seconds (relaxation)
Note: Electrode position is supposed to surround the treated area.
With painful muscle groups, the electrodes are grouped round the
muscle affected. In case of joint pain, the joint on the front/rear side
of the joint and, if electrode distances permit, on the right and left
sides of the joint, should be surrounded with electrodes. Minimal
electrode distance should not be less than 5 cm and not exceed 15
cm. Burst programs are suitable for all sites which should be
treated with an alternating signal pattern (to ensure minimum
habituation).
250 4 30 0 250 4 30 0 250 5 30 0 250 5 30 0
250 6 20 0 250 6 20 0 250 8 30 0 250 8 30 0
250 6 30 0 250 6 30 0 250 8 20 0 250 8 20 0
250 10 20 0 250 10 20 0 - - - - - - - -
250 2 10 0 250 4 8 0 250 6 6 0 - - - -
250 60 20 0 250 70 20 0 250 80 30 0 250 80 30 0
- - - - - - - - - - - - - - - -
250 80 20 0 250 80 20 0 250 75 4 0 250 10 20 0
250 70 4 0 250 65 4 0 - - - - - - - -
250 40 30 0 250 45 30 0 250 55 30 0 250 60 30 0
- - - - - - - - - - - - - - - -
250 4 30 0 250 4 20 0 250 6 30 0 250 6 20 0
250 8 30 0 250 8 20 0 250 10 30 0 250 10 20 0
250 40 5 0 250 6 10 0 250 50 5 0
- - - - - - - - - - - - - - -
250 75 0.25 0.25 250 2 0.5 0 - - - - - - - -
- - - - - - - - - - - - - - - -
250 100 0.25 0.25 - - - - - - - - - - - -
- - - - - - - - - - - - - - - -
Cycle 1 Cycle 2 Cycle 3 Cycle 4
Cycle 5 Cycle 6 Cycle 7 Cycle 8
Width Frequency On time Off Time
(μs) (Hz) Sec Sec
Width Frequency On time Off Time
(μs) (Hz) Sec Sec
Width Frequency On time Off Time
(μs) (Hz) Sec Sec
Width Frequency On time Off Time
(μs) (Hz) Sec Sec
1 Neck pain, tension headache 02,05
2 Back pain 10,11,12,13,14,19
3 Shoulder Pain 04,09,21
4 Pain as result of rheumatoid arthritisn See note
5 Lumber pain 05
6 menstrual pain 20
7 Pain program I See note
8 Knee pain ankle joint pain, capsular injuries 26,30
9 pain program II See note
(Burst)
10 Pain program III See note
(Burst)
Prog.
No Reasonable Areas
of Application
Indicator
Possible
Electrode
Position
TENS PROGRAMS
PL982-user180x128.indd 16-17PL982-user180x128.indd 16-17 2010/2/9 下午 02:12:142010/2/9 下午 02:12:14

18 19
BEAUTY PROGRAMS
250 30 5 1 250 10 15 1 250 50 5 1 - - - -
- - - - - - - - - - - - - - - -
250 4 30 1 250 4 20 1 250 5 30 1 250 5 20 1
250 6 30 1 250 6 20 1 - - - - - - - -
250 2 10 1 250 4 10 1 250 6 10 1 - - - -
- - - - - - - - - - - - - - - -
250 4 30 1 250 4 30 1 250 4 30 1 250 5 30 1
250 5 30 1 - - - - - - - - - - - -
250 6 15 1 250 8 15 1 250 10 15 1 - - - -
- - - - - - - - - - - - - - - -
250 2 20 1 250 2 20 1 250 1 30 1 250 1 30 1
- - - - - - - - - - - - - - - -
250 4 30 1 250 4 20 1 250 6 30 1 250 6 20 1
250 8 30 1 250 8 20 1 - - - - - - - -
250 20 5 1 250 6 5 1 250 30 5 1 - - - -
- - - - - - - - - - - - - - - -
250 20 5 1 250 6 8 1 250 25 5 1 - - - -
- - - - - - - - - - - - - - - -
250 25 5 1 250 6 8 1 250 35 5 1 - - - -
- - - - - - - - - - - - - - - -
Cycle 1 Cycle 2 Cycle 3 Cycle 4
Cycle 5 Cycle 6 Cycle 7 Cycle 8
Width Frequency On time Off Time
(μs) (Hz) Sec Sec
Width Frequency On time Off Time
(μs) (Hz) Sec Sec
Width Frequency On time Off Time
(μs) (Hz) Sec Sec
Width Frequency On time Off Time
(μs) (Hz) Sec Sec
1 Neck Pain, Tension, Headache 04,09,21
2 Middle and lower trapezius muscle, lattisimus
dorsi muscle, neck muscles 02,09,10,11,12,16,17
3 Back musclature near spinal column, prog. 1 14,15,19
4 Front and back upper arm musculature (including
biceps), front and back lower arm musculature 05,06,18,22
5 Straight and transverse abdominal musculature 07
6 Back musculature near spinal column, Prog. II 14,15,19
7 Back musculature near spinal column, Prog. III 14,15,19
8 Posterior musculature 28
9Front and back upper thigh musculature 29
(Burst)
10 Front and back lower thigh musculature 25,27
(Burst)
Prog.
No Reasonable Areas
of Application
Indicator
Possible
Electrode
Position
EMS Programs
PL982-user180x128.indd 18-19PL982-user180x128.indd 18-19 2010/2/9 下午 02:12:152010/2/9 下午 02:12:15

20 21
MASSAGE PROGRAMS
Electrodes must be placed so that they surround the muscle
segments in question. For optimal effect, electrode distance must
not be greater than approx. 15 cm. The electrodes should not be
applied to the anterior wall of the thorax, i.e. never massage the left
or right large pectoral muscle.
250 8 60 - 250 60 250ms 250ms 250 15 60 - 250 60 250ms 250ms
- - - - - - - - - - - - - - - -
250 110 60 - 250 70 250ms 250ms 250 80 250ms 250ms 250 90 60 -
- - - - - - - - - - - - - - - -
250 5 15 0 250 10 15 0 250 15 15 0 250 20~80~20HZ
- - - - - - - - - - - - - - - -
250 20 5 0 250 30 5 0 250 40 5 0 250 50 5 0
250 60 5 0 250 75 5 0 250 90 5 0 250 110 5 0
250 20 60 0 250 30 60 0 250 40 60 0 250 50 60 0
250 60 60 0 250 75 60 0 250 90 60 0 250 110 60 0
20Hz 40μs~250μs~40μs200 8 30 0 - - - - - - - -
- - - - - - - - - - - - - - - -
250 6 - 0 - - - - - - - - - - - -
- - - - - - - - - - - - - - - -
250 25 - 0 - - - - - - - - - - - -
- - - - - - - - - - - - - - - -
250 60 - 0 - - - - - - - - - - - -
- - - - - - - - - - - - - - - -
250 100 - 0 - - - - - - - - - - - -
- - - - - - - - - - - - - - - -
Cycle 1 Cycle 2 Cycle 3 Cycle 4
Cycle 5 Cycle 6 Cycle 7 Cycle 8
Width Frequency On time Off Time
(μs) (Hz) Sec Sec
Width Frequency On time Off Time
(μs) (Hz) Sec Sec
Width Frequency On time Off Time
(μs) (Hz) Sec Sec
Width Frequency On time Off Time
(μs) (Hz) Sec Sec
1 Tap and grasp massage
2 Knead and grasp massage
3 Tap massage
4 Side of hand/pressing massage
5 Side of hand/pressing massage
6 Vibratory massage
7 Tap massage (switch between electrodes)
8 Massage jet (switch between electrodes)
9 Pressure massage jet (switch between electrodes)
10 Combination program (switch between electrodes)
Prog.
No Reasonable Areas
of Application
Indicator
Massage Programs
PL982-user180x128.indd 20-21PL982-user180x128.indd 20-21 2010/2/9 下午 02:12:152010/2/9 下午 02:12:15

22 23
5.2 Information on placing
the electrodes
For the desired success of any stimulation
application, it is important that electrodes
are placed sensibly. We advise you to
coordinate with your doctor regarding the
best electrode positions for your intended
field of application. The electrode positions
suggested inside the cover (Figs. 1– 28)
serve as a guide. The following applies when
choosing electrode positions:
Electrode distance
The greater the electrode distance which is
selected, the greater will be the volume of
tissue that is stimulated. This applies to the
area and the depth of the tissue volume.
At the same time, however, the strength of
tissue stimulation decreases as the electrode
distance becomes greater, which means that,
if a greater electrode distance is chosen, a
greater volume is stimulated, but stimulation
is weaker. To increase stimulation, pulse
intensity must then be increased.
The following guideline holds when selecting
electrode distances:
• Most reasonable distance: approx. 5 –15 cm
• Below 5 cm, it is mainly surface structures
which are strongly stimulated
• Over 15 cm, largearea and deep structures
are stimulated very weakly
Electrode relationship to muscle fiber
orientation
Choice of current flow direction must be
adapted to the muscle fiber orientation
according to the desired muscle layer. If
surface muscles are to be reached, the
electrodes must be placed parallel to the
fiber orientation. If deep tissue layers are to
be reached, the electrodes must be placed
crosswise to the fiber orientation. The latter
arrangement can, for example, be achieved
via the cross electrode arrangement.
Note the color-code of the cables and the
channels. The white cable belongs to channel CH1
and the grey cable to channel CH2.
• In the treatment of pain (TENS) using the
Digital TENS/EMS device with its 2 separately
controllable channels and in each case
2 adhesive electrodes, it is advisable to
either position the electrodes of one chan-
nel so that the pain spot lies between the
electrodes or else you position one electrode
directly on the pain spot and the other at
least 2 – 3 cm away.
The electrodes of the second channel
can be used to treat other pain spots
simultaneously or else, however, together
with the electrodes of the first channel, to
surround the pain area (opposite). A cross
arrangement is again advisable here.
• Massage function tip: always use all 4
electrodes to ensure optimal treatment.
• In order to extend the durability of the
electrodes, use them on clean skin which is
as free as possible from hair and grease. If
necessary, clean the skin with water before
application and remove the hair.
• If an electrode becomes detached during
the application, the pulse intensity of both
channels goes to the lowest setting. Press the
ON/OFF key to enter pause mode, reposition
the electrode and continue application by
pressing the ON/OFF key again and setting
the required pulse intensity.
PL982-user180x128.indd 22-23PL982-user180x128.indd 22-23 2010/2/9 下午 02:12:162010/2/9 下午 02:12:16

24 25
6.1 Advice
* If the device is not used within 3 minutes, it switches off automatically (automatic switch-
off function). When it is switched on again, the LCD main menu screen appears, with the
last used submenu flashing.
* If an admissible key is pressed, there is one short beep. If an inadmissible key is pressed,
there are two short beeps.
6.2 Procedure for TENS, EMS and MASSAGE programs P1–
P9 (quick start)
1. From the program charts, select a program that is suitable for your purpose.
2.Place the electrodes in the area selected and connect these to the device. The
corresponding positioning suggestions may help you here.
3. Press the ON/OFF key to switch on the device.
4. Press MENU and navigate through the sub-menus (TENS/EMS/MASSAGE) and confirm your
selec-tion with ENTER (Fig. 1).
5. Using the UP/DOWN keys, select the program you want and confirm with ENTER (Fig. 2).
6. Using the UP/DOWN keys, select the total treatment time and confirm with ENTER. The
device is in stand by status (Fig. 3).
7. Press ON/OFF to start the stimulation treatment (Fig. 4). The operation status display starts
to change (Fig. 5).
8. Select the pulse intensity(Fig. 6) which is the most comfortable for you by pressing the key
CH1+/CH 2+. The pulse intensity display adapts accordingly.
9. All keys including intensity dials will be locked 20 seconds after last action done.
Chapter 6 Applications
Fig.1
Fig.2
Fig.3
Fig.4
Fig.5
Fig.6
PL982-user180x128.indd 24-25PL982-user180x128.indd 24-25 2010/2/9 下午 02:12:162010/2/9 下午 02:12:16

26 27
8. Using the UP/DOWN keys,select the desired treatment time and confirm with ENTER(Fig. 4).
9. The device is standing by for stimulation (Fig. 5).
10.Press Power to start the stimulation treatment. The operation status display starts to flash
(Fig. 6).
11.Adjust the pulse intensity which is the most comfortable for you by rotating intensity
adjust dials (CH 1/CH 2). The pulse intensity display adapts accordingly with number shown
(Fig. 7).
Note: Your customized program settings are stored and automatically retrieved when you select it
next.
6.3 Procedure for TENS/EMS program P0 (Only)
Program P0 is preset program which you can also customize. Here you can set pulse frequency
and pulse width.
1. From the program charts, select a program that is suitable for your purpose.
2.Place the electrodes in the area selected and connect these to the device. The
corresponding position suggestions may help you here.
3. Press the ON/OFF key in order to switch on the device.
4. Press MENU and navigate through the submenus (TENS/ EMS/MASSAGE) and confirm your
selection with ENTER (Fig. 1).
5. Using the UP/DOWN keys, select the program you want and confirm with ENTER (Fig. 2).
6. Using intensity dials to select the frequency and pulse width you want for Cycle 1 (Hz) and
confirm with ENTER (Fig. 3). Repeat this process for all subsequent cycles.
7. To terminate frequency setting, set the frequency (Hz) after the cycle last required “--0”
and press ENTER. All frequency settings of subsequent cycles are then automatically set at
“0” and thus not taken into account.
Tip: Hold down the ENTER key for 2 seconds in order to skip frequency settings and access ‘treat-
ment time setting’. In the event that you need all 8 cycles, you can also press ENTER for 2 seconds to
terminate selection of frequency settings.
Fig.1
Fig.2
Fig.3
Fig.4
Fig.5
Fig.6 Fig.7
PL982-user180x128.indd 26-27PL982-user180x128.indd 26-27 2010/2/9 下午 02:12:172010/2/9 下午 02:12:17

28 29
6.4 Changes to the settings
Automatic Key Lock Function
To avoid accidentally change of intensity,
PL-982 integrated auto key lock for safety
purpose. Before making any change during
stimulation, press and hold MENU for 3
seconds to unlock.
Changing intensity (during
stimulation)
• CH1 and CH2: Press corresponding channel
dial and rotate dial to increase/reduce
intensity. After intensity set, press the
dial again to prevent further change of
intensity or it will automatically lock after
20 seconds.
• DOWN key: the intensity of both channels
is reduced.
Interrupting stimulation
Press the pause key which locates between
CH1 and CH2 intensity adjust dial. To resume
stimulation, press Power key and adjust
intensity accordingly using dials.
Switching off a complete channel
Press and hold corresponding channel
intensity adjust dial (key unlocked). Graphic
of corresponding channel disappears. To
switch back on, unlock the keys and press
and hold the dial until the graphic appears.
Changing application (completely or
single parameters)
1.Interrupt stimulation by pressing Pause
key
2. Press MENU to return to previous menu
3.Set required parameters. Confirm
with ENTER. Press ON/OFF to continue
stimulation.
6.5 Doctor’s Function
Doctor’s Function is a specific setting to
allow you to call up your specific personal
program even much easier.
Your individual program setting is
immediately retrieved in stand by status
when you switch on and is activated by
simply pressing the ON/OFF key. Settings for
this individual program may, for example, be
on the advice of your doctor.
With the Doctor’s Function, only pulse
intensity can be changed during stimulation
treatment. All other parameters and
programs of the Digital TENS/EMS are
blocked in this case and cannot be modified
or called up.
Activating/De-activate the Doctor’s
Function:
1. Select your program and appropriate
settings from the list.
2. To activate or de-activate by pressing and
hold Power and MENU keys simultaneously
for 5 seconds. A long signal tone confirms
Doctor's Function activate status.
PL982-user180x128.indd 28-29PL982-user180x128.indd 28-29 2010/2/9 下午 02:12:182010/2/9 下午 02:12:18

30 31
Adhesive electrodes:
• In order to ensure that the electrodes
adhere for as long as possible, they should
be cleaned carefully with a damp, lint-
free cloth.
• After use, stick the electrodes onto the
backing film.
Cleaning the unit:
• Remove the batteries from the device
every time you clean.
• Clean the unit after use with a soft,
slightly moistened cloth. In case of more
extreme soiling you can also moisten the
cloth with mild soapy water.
• Ensure that no water penetrates into the
device. If this should ever happen, only
use the device again once it is completely
dry.
• Do not use any chemical cleaners or
abrasive agents for cleaning.
Chapter 7 Cleaning and storage
Storage:
• Remove the batteries from the unit if you
are not going to use it for a longer period.
Leaking batteries can damage the unit.
• Do not make any sharp kinks in the
connecting leads or electrodes.
• Detach the connecting cable from the
electrodes.
• After use, stick the electrodes onto the
backing film.
• Store the device in a cool, well-ventilated
place.
• Never place any heavy objects on the
device.
Used, fully discharged batteries must be disposed of in a specially labeled collection container,
at toxic waste collection points or through an electrical retailer. You are under legal obligation
to dispose of batteries correctly.
Chapter 8 Disposal
Note: You will find these markings on batteries containing harmful substances:
Pb=battery containing lead, Cd=battery containing cadmium, Hg=battery
containing mercury.
Please dispose of the device in accordance with the directive 2002/96/EC-WEEE
(Waste Electrical and Electronic Equipment). If you have any queries, please
refer to the local authorities responsible for waste disposal.
Pb Cd Hg
PL982-user180x128.indd 30-31PL982-user180x128.indd 30-31 2010/2/9 下午 02:12:182010/2/9 下午 02:12:18

32 33
The device does not switch on when
the ON/OFF button is pressed. What
should I do?
1. Make sure that the batteries have been
inserted correctly and have contact.
2. Change the batteries if necessary.
3. Contact customer service.
The electrodes are detaching
themselves from the body. What
should I do?
1.Clean the adhesive surface of the
electrodes with a damp, lint-free cloth.
Then let them dry in the air and attach
them again. Should the electrodes still not
adhere firmly, they must be replaced.
2. Before each application, clean the skin,
avoiding using any skin cream or oils
before treatment. Shaving can increase
adhesion of the electrodes.
Chapter 9 Problems and solutions
The device is giving out abnormal
signal tones during treatment. What
should I do?
1.Look at the display. Is any channel
flashing? ‡ Interrupt the program by
pressing the ON/OFF button. Make sure
that the connecting leads are properly
connected to the electrodes. Make sure
that there is a firm contact between the
electrodes and the treatment area.
2. Make sure that the plug of the connecting
lead is firmly connected to the device.
3.If the signal tones do not stop and the
channel is flashing, change the connecting
cable.
4. The display shows a flashing battery
signal. Renew the batteries.
There is no noticeable stimulation.
What should I do?
1. If a warning signal sounds, go through the
steps described above.
2. Press the ON/OFF key to start the program
again.
3. Check that the electrodes are positioned
properly and make sure that the adhesive
electrodes are not overlapping.
4. Increase pulse intensity in stages.
5. The batteries are almost empty. Replace
these.
You are aware that the electrodes
feel uncomfortable. What should I do?
1. The electrodes are badly positioned.
Check positioning and if necessary
reposition the electrodes.
2. The electrodes are worn. Due to the fact
that current distribution can no longer be
guaranteed evenly over the entire surface,
these may lead to skin irritation. You must
therefore replace these.
The skin in the treatment area
becomes red. What should I do?
Stop treatment immediately and wait
until the skin has returned to normal. If
any redness beneath the electrode soon
disappears, this is not dangerous and is
due to the increased circulation which has
been stimulated locally. If, however, the
skin irritation remains and there is possibly
itching or inflammation,
Please consult your doctor before further
use. It may possibly be caused by an allergy
to the adhesive surface.
The belt clip is not secure. What
should I do?
If the clip is fastened to the belt and only
then connected to the device, there is not
optimal hold. You should therefore take the
device in your hand. Position the belt clip on
the back of the device. You will notice the
correct position of the clip when you hear
it click into position. Now secure the device
with belt clip on the belt again.
PL982-user180x128.indd 32-33PL982-user180x128.indd 32-33 2010/2/9 下午 02:12:192010/2/9 下午 02:12:19

34
Name and model: PL-982
Initial curve shape: biphasic square-wave pulse
Pulse duration: 40– 300 μs
Pulse frequency: 1–120 Hz
Output voltage: max. 90 Vpp (at 500 Ohm)
Output current: max. 180 mApp (at 500 Ohm)
Power supply: 3x AAA batteries
Treatment time: adjustable from 5 to 90 minutes
Intensity: adjustable from 0 to 15
Operating conditions:10°C~40°C (50°F~104°F) with a relative humidity of 30~85%
Storage conditions-10°C~50°C (14°F~122°F) with a relative humidity of 10~95%
Explanation of symbols: Application part type BF
Caution! Read the instructions for Use.
Chapter 10 Technical Specifications
PL982-user180x128.indd 34PL982-user180x128.indd 34 2010/2/9 下午 02:12:192010/2/9 下午 02:12:19