Instruction Manual
CB-200-CA
Circulation Pro
13-0095 CB200-CA_IM_E_vE.qxd:Layout 1 7/2/13 2:05 PM Page 1

2
E
QUICK START GUIDE
PLEASE NOTE – THIS DEVICE DOES NOT VIBRATE – IT USES ELECTRICAL IMPULSES, NOT VIBRATION!
For detailed operation of your Circulation Pro please refer to the comprehensive instructions within
this manual.
For a full explanation of setting the intensity, refer to page 13
Remove your Circulation Pro from the packaging. Take out the
Remote Control and remove the screw from the back door using the
screwdriver provided. Then insert 2 AAA batteries into the
compartment as per indication. Then screw the battery door closed.
Please refer to page 15 for a step-by-step guide on changing the
battery on the remote control.
Connect the DC adapter to a suitable power outlet and plug the
small DC socket into the device.
Turn on the power. The central display will light up orange and
go off.
Remove your footwear and socks or stockings. PLACE YOUR BARE
FEET ONTO THE FOOT PADS. YOUR RIGHT FOOT ON THE RIGHT FOOT
PAD AND YOUR LEFT FOOT ON THE LEFT FOOT PAD. BOTH FEET
NEED TO BE ON THE DEVICE FOR IT TO WORK.
Sit in a comfortable chair. Place your bare feet on the left and right
foot plates. Increase the intensity levels for the foot by pressing the
“SOLE+” or press“SOLE –“ to decrease the intensity. The intensity
level ranges from 0-99, slowly increase the level until you begin to
feel the micro-current stimulation.
1
2
3
4
5
4
13-0095 CB200-CA_IM_E_vE.qxd:Layout 1 7/2/13 2:05 PM Page 2

3
E
IMPORTANT SAFETY INFORMATION
1) Please read these instructions thoroughly before use.
2) Please check that you have all of the component parts as detailed in this user manual.
3) Take all parts out of the plastic bags and examine them to familiarize yourself with the components.
Notes on safety
Danger
This unit must not be used in combination with the following medical devices:
(1) Internally transplanted electronic medical devices, e.g. pacemakers
(2) Electronic life support equipment, such as respirators
(3) Electronic medical devices attached to the body, such as electrocardiographs
Using this unit with other electronic medical devices may cause erroneous operation of those devices.
13-0095 CB200-CA_IM_E_vE.qxd:Layout 1 7/2/13 2:05 PM Page 3

4
E
Warning
Persons with the following conditions must consult a doctor before using this unit:
1) acute disease
2) malignant tumor
3) infectious disease
4) pregnancy
5) cardiac dysfunction
6) high fever
7) abnormal blood pressure
8) skin sensory disorders or skin problems
9) receiving medical treatment, especially those feeling discomfort. May cause an accident or ill health.
Do not use this unit near the heart, above the neck, on the head, around the mouth or on diseased skin.
May cause an accident or ill health.
- Application of electrodes between the neck and diaphragm (chest area) may increase the risk of cardiac fibrillation.
Do not use this unit simultaneously with other therapeutic device or in combination with
ointments including spray type ointments.
May cause discomfort or ill health.
- Operation in close proximity (e.g. 1 m) to a shortwave or microwave therapy EQUIPMENT may produce
instability in the STIMULATOR output.
Do not use this unit for purposes other than treatment indicated in this manual.
May lead to accident, problems, or failure of the unit.
Do not insert the electrode cord plug into any place other than the electrode cord jack of the main unit.
May cause an electric shock or accident.
Do not disassemble or remodel this unit.
No user serviceable parts.
Caution
If the unit is not functioning properly or you feel discomfort, immediately stop using the unit.
If you feel any problems with your body or skin, consult a doctor and follow his/her instructions.
If you want to move the Electrode Pad to another region or your body during treatment, be sure to turn off the power first.
If not, you may receive a strong electrical shock.
Do not try to attach the Pads to any other person during the treatment.
You may receive strong electrical shock.
Do not start treatment while wearing an electronic device.
The settings and timings of the device may be affected.
Do not use this unit on infants or people not capable of expressing their intentions.
May cause an accident or ill health.
Do not use this unit in places with high humidity such as bathrooms or while taking a bath or shower.
You will receive a strong electrical shock.
Do not use this unit while sleeping.
The main unit may develop trouble, or the pad may move to an unexpected region and cause ill health.
Do not use this unit while driving.
Receiving sudden strong stimulation may lead to traffic accident.
Do not leave the Electrode Pad attached to the skin after treatment.
Prolonged attachment may cause skin irritation or infection.
Be careful not to allow any metal object, such as a belt buckle or necklace to come into contact with the Electrode Pad dur-
ing treatment.
You may receive a strong electrical shock.
Do not use cellular phones or other electronic devices near this unit.
Put the Long Life pads only on skin or on the Long Life pads holder to avoid damage of the adhesive
surfaces of the pads.
13-0095 CB200-CA_IM_E_vE.qxd:Layout 1 7/2/13 2:05 PM Page 4

5
E
Important information regarding Electro Magnetic Compatibility
With the increased number of electronic devices such as PC’s and mobile (cellular) telephones, medical
devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic
interference may result in incorrect operation of the medical device and create a potentially unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to prevent unsafe
product situations, the IEC 60601-1-2 standard has been implemented. This standard defines the levels of
immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for
medical devices.
This medical device manufactured by Homedics conforms to this IEC 60601-1-2 standard for both immunity
and emissions. Nevertheless, special precautions need to be observed:
Do not use mobile (cellular) telephones and other devices, which generate strong electrical or
electromagnetic fields, near the medical device. This may result in incorrect operation of the unit and create a
potentially unsafe situation.
Recommendation is to keep a minimum distance of 7 m. Verify correct operation of the device in case the
distance is shorter.
CB-200-CA needs special precautions regarding EMC and needs to be installed and put into service according
to the EMC information provided in the ACCOMPANYING DOCUMENTS.
Portable and mobile RF communications equipment can affect CB-200-CA.
WARNING that the use of accessories, transducers and cables other than those supplied with the exception of
the transducers and cables sold by the manufacturer of the CB-200-CA as replacement parts for internal
components, may result in increased EMISSIONS or decreased IMMUNITY of CB-200-CA.
WARNING the CB-200-CA should not be used adjacent to or stacked with other equipment.
Equipment not suitable for use in the presence of a flammable anaesthetic mixture with air or with oxygen or
nitrous oxide.
13-0095 CB200-CA_IM_E_vE.qxd:Layout 1 7/2/13 2:05 PM Page 5

6
E
WHAT IS ELECTRONIC NERVE STIMULATION?
INTENDED USE: Medical Purpose
This Electronic Nerve Stimulator is intended to be used as a massager to relieve (muscle) pain, increase blood
circulation, relax stiff muscles, reduce swollen feet, ankles and fatigue. The massaging effect is achieved by
electronic stimulation of the nerves through electrode pads placed on the skin. Various massage regions and
treatment programs can be selected.
Suitable Users: Please read“Notes on safety” before using the unit. (This unit should not be used by people
prohibited from doing so in“Notes on safety”.).
Environment: This unit is intended for home use only.
Effectiveness: Massager: relieve of (muscle) pain, stiffness and fatigue.
Precautions for use: Please read “Notes on safety” before using the unit.
Electronic Nerve Stimulation is a non-invasive, safe nerve stimulation intended to reduce pain. The Circulation
Pro uses proven neuromuscular electrical stimulation therapy to send micro current pulses through the soles
of your feet. This type of electrical stimulation is clinically proven to be safe and effective and can be used in
the comfort of your own home. The Circulation Pro improves muscles function by stimulating nerves and
increasing the flow of blood helping to reduce PAIN, SWELLING, TIRED AND ACHING LEGS.
13-0095 CB200-CA_IM_E_vE.qxd:Layout 1 7/2/13 2:05 PM Page 6

7
E
MACHINE OVERVIEW AND PART NAMES
A. Main Unit
B. AC Power Adapter
C. Remote Control
D. Electrode Gel Pad x 4 pcs.
E. Wire cable connecting the
Electrode Gel pads and device
F. Plastic Protector for Gel Pads
1. LCD Display Screen
2. Control Panel
3. Output for the feet – Electrode Area for Left Foot
4. Output for the feet – Electrode Area for Right
Foot
5. Cable Connecting the Electrode Gel Pads and
Device.
6. Cable Connecting the Electrode Gel Pads and
Device.
7. Remote Control Receiver Sensor
8. Silver colour decoration plate
9. Adaptor Jack
Top View
Side View
13-0095 CB200-CA_IM_E_vE.qxd:Layout 1 7/2/13 2:05 PM Page 7

8
E
FUNCTION OF CONTROL PANEL
LCD showing the
intensity level for Sole
– maximum 99 levels
LCD showing the
program in AUTO or
in MODE.
LCD showing the timer
LCD showing the
intensity level for body
– maximum 99 levels
On the device the black colour area on the device which
is the electrode area for the sole. (see. the fig. A).
On the gel pad, the black colour
area on the sticky part is the
electrode area for the body, size is
5 cm x 9 cm. (see. the fig. B)
ON/OFF switch button
BODY – Decrease the output intensity of body (Available from 1 - 99 levels)
BODY + Increase the output intensity of body (Available from 1 - 99 levels)
SOLE - Decrease the output intensity of Sole (Available from 1 - 99 levels)
SOLE + Increase the output intensity of sole (Available from 1 - 99 levels)
TIME - Decrease the operation time (available from 1-60 minutes)
TIME + Increase the operation time (available from 1-60 minutes)
AUTO/MODE
Auto - is the preset program with 14 pattern cycles running for Foot and 10 pattern cycles running
for body
Mode - user can fix the program to the exiting massage pattern on the rest of the time
Electrode area of the unit and the gel pad
Fig. A
Fig. B
13-0095 CB200-CA_IM_E_vE.qxd:Layout 1 7/2/13 2:05 PM Page 8

9
E
CIRCULATION PRO OUTPUT WAVE FORMS
FOOT ELECTRO THERAPY MASSAGE: We will examine the operation in more detail later in your instructions
but the principle is relatively easy to understand. Place your feet on the electrode areas, turn on Circulation
Pro with the central on/off switch, then increase the intensity for the foot. There are 99 different levels. When
you start to feel the mild electro-therapy will depend on your own nerve sensitivity. Certain individuals will
feel nothing until the intensity is up at a high level, others will feel the stimulation at relatively low levels. This
is completely normal.
BODY TONING: if you choose to tone muscle groups or target pain in other areas of your body the Circulation
Pro comes with four gel pads. These can be used to tone arms, hips, thighs, abs or buttocks or target neck
muscle or back pain.
OUTPUT WAVEFORM
***THE OUTPUTS HAVE NO DC COMPONENT
Program Output
1Pulse rate 12.2Hz with 8.5 seconds and off time in
900mS, A cycle repeating for 1 minute
2Pulse rate 16.13Hz with2.8 seconds and off time in
900mS, A cycle repeating for 1 minute
3Pulse rate 20.0Hz with 8.4 seconds and off time in
900mS, A cycle repeating for 1 minute
4Pulse rate 16.13Hz with 5.8 seconds and off time in
900mS, A cycle repeating for 1 minute
5Pulse rate 16.16Hz with 7.0 seconds and off time in
900mS, A cycle repeating for 1 minute
6Pulse rate 33.33Hz with 2.3 seconds and off time in
900mS, A cycle repeating for 1 minute
7Pulse rate 12.50Hz with 4.6 seconds and off time in
900mS, A cycle repeating for 1 minute
8Pulse rate 55.56Hz with 11.5 seconds and off time in
900mS, A cycle repeating for 1 minute
9Pulse rate 23.32Hz with 5.6 seconds and off time in
900mS, A cycle repeating for 1 minute
10 Pulse rate 20.0Hz with 4.5 seconds and off time in
900mS, A cycle repeating for 1 minute
11 Pulse rate 10Hz with 5.3 seconds and off time in 900mS,
A cycle repeating for 1 minute
12 Pulse rate 16.13Hz with 5.60 seconds and off time in
900mS, A cycle repeating for 1 minute
13 Pulse rate 26.32Hz with 3.5 seconds and off time in
900mS, A cycle repeating for 1 minute
14 Pulse rate 25Hz with 7.0 seconds and off time in 900mS,
A cycle repeating for 1 minute
Program Output
1Pulse rate 25.00Hz with 5.8 seconds and off time in
900mS, A cycle repeating for 1 minute
2Pulse rate 16.67Hz with 11.6 seconds and off time in
900mS, A cycle repeating for 1 minute
3Pulse rate 12.5hz 9.7 seconds and off time in 900mS, A
cycle repeating for 1 minute
4Pulse rate 12.50Hz with 4.4 seconds and off time in
900mS, A cycle repeating for 1 minute
5Pulse rate 25.00Hz with 13 seconds and off time in
900mS, A cycle repeating for 1 minute
6Pulse rate 16.67Hz with 10.2 seconds and off time in
900mS, A cycle repeating for 1 minute
7Pulse rate 12.5Hz with 5.6 seconds and off time in
900mS, A cycle repeating for 1 minute
8Pulse rate 12.5Hz with 18.2 seconds and off time in
900mS, A cycle repeating for 1 minute
9Pulse rate 16.67Hz with 5.1 seconds and off time in
900mS, A cycle repeating for 1 minute
10 Pulse rate 10Hz with 21.8 seconds and off time in
900mS, A cycle repeating for 1 minute
SOLE MASSAGER (during 1 kΩload)
The auto mode will cycle through the 14 programs
during the units operation, repeating automatically.
BODY MASSAGER (during 1 kΩload)
During the operation, the unit will cycle through the
10 programs, repeating automatically.
13-0095 CB200-CA_IM_E_vE.qxd:Layout 1 7/2/13 2:05 PM Page 9

10
E
13-0095 CB200-CA_IM_E_vE.qxd:Layout 1 7/2/13 2:05 PM Page 10

11
E
13-0095 CB200-CA_IM_E_vE.qxd:Layout 1 7/2/13 2:05 PM Page 11

12
E
13-0095 CB200-CA_IM_E_vE.qxd:Layout 1 7/2/13 2:05 PM Page 12

13
E
HOW TO OPERATE
For Foot – SOLE
1. Place your bare feet onto Circulation Pro (do not wear socks).
2. Press the on/o button, the LCD screen will light up in Orange. The
program shows AUTO and both Bands show 00, which is in standby
mode (see Fig. 1).
3. Gently increase intensity setting by pushing the button “SOLE +”. Or
decrease intensity setting by pushing the button“SOLE -”. The intensity
level is adjustable between 0 and 99. The LCD display will show the
selected level (see Fig. 2).
4. You can adjust the auto o timer by pressing the“Time –“ or “Time +”.
Timer ranges from 1-60 minutes. The timer will begin to count down
from the time you selected (see Fig. 3). To terminate the massage period,
user can turn o the unit anytime by pressing the on/o Button once.
5. If you are satised with the current massage program, you can lock the
current massage program by pressing the Auto/Mode key. The rest of the
massage time will then only run the selected massage program
(see Fig. 4).
IMPORTANT INFORMATION:
a. The aim is not to get up to level ‘99’.
b Chose an itensity level that is comfortable for you! This level may very
from day to day.
c. Remember to drink plenty of uids – if you are dehyrdated, this will
reduce the eectiveness of the device
Fig. 1
Fig. 2
Fig. 3
Fig. 4
13-0095 CB200-CA_IM_E_vE.qxd:Layout 1 7/2/13 2:05 PM Page 13

14
E
For Body
USING THE GEL PADS
Wash and dry skin before use. Connect the output wire to the gel pads. Connect the other end of the output
wire to the output jack on Circulation Pro. Remove the protective film from the adhesive pads. Attach the gel
pads to the skin. Press the on/off button to turn on the unit and adjust the stimulating output intensity to the
desired level. (The display will show the mode and level that you selected and start to count down).
1. Plug in the 2 cables into the cable jack on the unit (see Fig. 5).
2. Connect pin of the cable to the gel pad properly (see Fig.6).
3. Remove the protective film on the gel pad, and attached the 4 gel pads to
the area of the body you wish to treat in accordance with warnings.
4. Repeat operation as in foot instructions, adjust the intensity using for body.
5. Gently increase intensity setting by pushing the button of “Body + ”. Or
decrease intensity setting by pushing the button of“Body - ”. The LCD will also show the level which you
have selected (see Fig. 7).
6. To terminate the massage period, user can turn off the unit anytime by pressing the on/off Button.
If you want to use with 2 gel pads only, then you must connect 1 gel pad to jack A and 1 gel pad to Jack B.
Care of your gel pads
Never stick two adhesive pads to each other. Keep the adhesive gel pads
clean, never expose them to high temperature or direct sunshine. If the
electrode gel pads are insufficiently adhesive or dirty, wipe with a wet cloth
or change for new ones, replacement parts will be available directly from
Homedics or your distributor.
Do not clean the electrode gel pads with any chemical.
ALWAYS try and protect the gel pads. Store the gel pads on the gel pad
protector when not in use, as in the illustration.
Fig. 5
Fig. 6
Fig. 7
13-0095 CB200-CA_IM_E_vE.qxd:Layout 1 7/2/13 2:05 PM Page 14

15
E
CONNECT WITH THE SUPPLIED AC/DC POWER ADAPTER
Plug the DC plug of the power supply into the socket
on the side of Circulation Pro. (See Fig 8) Plug in the
power adapter to a suitable wall socket. (Make sure
that the input voltage of the wall socket is suitable for
the supplied adapter).
INSTALLING BATTERIES FOR MAIN UNIT
If you want to use the Circulation Pro with battery power instead of the supplied power adapter, the battery
compartment is located on the underside of the unit.
Remove the battery cover from the unit by removing the screw with a screwdriver. Insert 4 1.5V size AA
batteries with the + and -marks correctly aligned.
INSTALLING BATTERIES FOR REMOTE CONTROL
Remove the battery cover from the unit by removing the screw with the screwdriver provided. Insert 2 1.5V
size AAA batteries with the + and -marks correctly aligned.
The transmitter of the remote control is at the top in the colour black (Fig. 10).
When using the remote control, please remember to point the transmitter to the
receiver of the main unit, which is located between the gel pad jack of the main unit.
(Fig. 11).
Note on batteries:
Do not mix different types of batteries or an old battery with a new one. To prevent the risk of leakage or
explosions, never recharge the batteries, apply heat or take them apart.
When not using batteries, remove them to prevent battery drain. If liquid leaks from the batteries, throw them
away. Thoroughly clean the battery compartment with a dry cloth.
Fig. 8
Fig. 8 Fig. 9 Fig. 10
Fig. 11
13-0095 CB200-CA_IM_E_vE.qxd:Layout 1 7/2/13 2:05 PM Page 15

16
E
CLEANING
Electrode Gel pad
• When not in use, store the Electrode Gel pads on the plastic pad protectors provided at room temperature.
• Keep the Electrode Gel pads clean and dust free in a dry location, keep away from oily or sticky areas. The
life of the electrodes varies depending on skin conditions, storage, amount of use, type of stimulation, and
stimulation site. Usage may be extended by carefully cleaning the gel surface with water. Do not spill
liquid on the wire.
• Single patient use only.
• Do not apply to broken skin. Should a skin rash occur , discontinue use and contact your physician
• Do not use tissue, cloth etc. to wipe the electrode surface.
• Do not use finger nail, brush etc. to damage the electrode surface.
• Do not frequently clean the pads, and don’t use the detergent or hot water to clean the electrode gel pads
Main Device
• Switch off the power, remove the adapter and the electrode gel pad from the unit for storage in correct
way.
• Always keep clean the main device, by using a soft cloth to clean the surface of the unit.
• To clean the foot pedals, use a soft, damp, soapy cloth but make sure to squeeze the cloth dry and clean
the foot pedal area.
• If the device is very dirty, use a soft, damp soapy cloth but make sure to squeeze the cloth dry to clean the
unit.
• Do not spill liquid on the device
• Do not immerse the device in water
• Do not clean with chemicals
• Store in a dry, dust free location in a temperature between 10ºC to 40ºC and 30% to 90% relative humidity.
Safety Precautions
• Do not open the device or repair it yourself. This will invalidate your warranty and may cause serious harm.
• If the device malfunctions, disconnect it from the power source and contact your selling agent as soon as
possible.
• Use only the accessories supplied by the manufacturer.
• Use the device only for its intended purpose.
• Do not expose the device to extreme heat.
• Do not overload the electrical outlet.
• Do not stand on the machine. Use it when sitting down only.
• Do not spill liquid on the device or its accessories
The warranty is void if the product has been altered, misused or abused. HoMedics will not take any
responsibility.
13-0095 CB200-CA_IM_E_vE.qxd:Layout 1 7/2/13 2:05 PM Page 16

17
E
TROUBLESHOOTING & MAINTENANCE
Hygiene
After using the product
Clean the device with a soft, damp cloth but make sure to squeeze the cloth dry, and to clean the foot pedal
area.
Store the Electrode Gel pads on the plastic pad protectors provided.
Storage
Keep the whole set clean and store it in a dust free and dry location under the following conditions
Storage temperature and humidity -10°C to 60°C, 10% to 95% RH
Operating temperature and humidity -5°C to 50°C, 30% to 90% RH
Problem Cause Rectification
Device will not turn on. - Batteries inserted in wrong direction.
- The adapter does not plug into device properly.
- Insert batteries in correct direction or check the
battery is in full power.
- check the connection of the adapter jack is well
connected; and also the DC adapter is well
connected to the main socket.
Power turns off too
soon
- Gel pads not attached correctly to the skin. - Attach Gel pads correctly to the skin
Power turns off while
using massager
- if you are using the batteries operation, then
the batteries weak/exhausted
- Fit 4 new identical 1.5V alkaline batteries type
AA
- Treatment period of 30 minutes is over and
power turns off automatically
- Restart treatment or turn off the massager.
- if you are using the body massage, the
electrode gel pad may be broken
-Replace electrode gel pad
It is difficult to attach
Gel pad to the skin
- Transparent film not peeled off
- Gel pad applied immediately after washing
- Adhesive surface of Gel pad damaged
- The gel pads get dirty and lost their
adhesive/stickiness
- Peel off film on the adhesive surface of Gel pad
- Sufficiently dry Gel pad
- Replace Gel pad
- Replace Gel pad or clean the gel pad with a
small drop of water onto the sticky side of the
electrode pad and rub into the surface
Adhesive surface of
Gel pad is not sticky
- Use of Gel pad while perspiring
- Gel pad washed too long and/or too frequently
- Gel pads stored under high temperature, high
humidity, direct sunshine
Leave Gel pad in freezer for overnight
It is difficult to feel
stimulation
- You sole is too dry, not enough moisture
- Your sole is not placed on the foot pedal
properly
- Gel pads not attached correctly to the skin
- Gel pads overlap each other
- Electrode cord not connected correctly
- Applied intensity too weak
- Put some water to your sole to moisture your
sole
- Ensure both of your soles are placed on each
pedal properly.
- Attach Gel pad firmly to the skin
- Reattach Long Life pads with no overlap
- Connect electrode cord correctly
- Increase the intensity by pressing the + button.
The skin turns red or
the skin feels irritated
- Adhesive surface of Gel pads dirty or dry
- Adhesive surface of Gel pads damaged
- Wash adhesive surface of Gel pads softly with
your fingertips for about 3 seconds under slow
running water
- Replace Gel pads
13-0095 CB200-CA_IM_E_vE.qxd:Layout 1 7/2/13 2:05 PM Page 17

18
E
TECHNICAL SPECIFICATIONS
Product Name Circulation Pro
Model CB-200-CA
Power supply 6V DC or 4x1.5V alkaline batteries type AA* for the main unit
2x1.5V alkaline batteries type AAA for the remote control
Supplier of Adapter
Model no. of Adapter GPE038-060050-1
Adapter Input AC 100-240V~50-60Hz 0.1A
Adapter Output DC 6V 500mA 3.0W
Battery life >350 minutes
Frequency generation Approx. 10 Hz to 55.56 Hz
Power consumption 1.05 W
Maximum Output Voltage U <54.8V (during 1 kΩ load)
Maximum Output Current I < 910 μA (during 1 kΩ load)
Operating temperature and humidity -5°C to 50°C, 30%to 90% RH
Storage temperature and humidity -10°C to 60°C, 10% to 95% RH
Main unit dimensions 338(L) x 324 (W) x 48(H) mm
Weight Approx. 1250 g
Package Contents Quantity Parts
1
1
1
2
4
2
1
2
Circulation Pro
AC/DC Adapter
Remote Control/Batteries
Cable Wire for Electrode Gel pads
Electrode Gel pads
Plastic Gel pads protector
Instruction manual
AAA batteries
Accessories :
• Only use original accessories.
Check that the contents of the delivery are complete.
* batteries not included.
13- 0095 CB200- CA_I M_E_vE. qxd: Layout 1 7/ 2/ 13 2: 05 PM Page 18

19
E
IMPORTANT INFORMATION
Electro Magnetic Compatibility (EMC)
1. CIRCULATION PRO needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided in the ACCOMPANYING DOCUMENTS;
2. Portable and mobile RF communications equipment can affect CIRCULATION PRO.
3. Warning that the use of accessories, transducers and cables other than those specified with the exception
of transducers and cables sold by the manufacturer of the CIRCULATION PRO as replacement parts for
internal components, may result in increased EMISSIONS or decreased IMMUNITY of the CIRCULATION PRO.
4. Warning that the CIRCULATION PRO should not be used adjacent to or stacked with other equipment.
Guidance and manufacturer’s declaration – electromagnetic emissions
The CIRCULATION PRO is intended for use in the electromagnetic environment specified below. The
customer or the user of the CIRCULATION PRO should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11 Group 1
The CIRCULATION PRO uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11 Class B
The CIRCULATION PRO is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2 Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
13-0095 CB200-CA_IM_E_vE.qxd:Layout 1 7/2/13 2:05 PM Page 19