Part Number - 350100004 Revision – A00-000
Place the chair / bed mat into its required position (refer to the placement
instructions for the appropriate chair / bed mat). Place the Fall Detection Unit on
the chair or at the bedside.
Activating / Deactivating the Fall Detection Unit
Insert the batteries in order to activate the Fall Detection Unit. Loosen the screw
located on the back lower panel of the Fall Detection Unit and ently remove the
cover. Insert the required batteries into the battery holder. The LED indicator
illuminates and a sin le beepin sound is heard. Gently replace the cover and
ti hten the screw holdin the back lower panel.
Once in active mode, the Fall Alert Unit automatically sends a transmission (every
20 seconds) indicatin the current battery status and current location information.
This status messa e is sent as lon as the Fall Detection Unit is activated.
To deactivate the Fall Detection Unit, simply remove the batteries from the unit.
Setting the Fall Alarm Volume
The level of the audible fall alarm depends on the level defined on the Alarm
Volume panel. Holdin the Fall Alert Unit ti htly, press the Alarm button until the
desired level is selected. The alarm level options are; silent, low or loud. The
illuminated LED indicates the alarm level selected.
Setting the Fall Alarm Delay
The delay time of the audible fall alarm depends on the level defined on the Alarm
Delay panel. Holdin the Fall Alert Unit ti htly, press the Delay button until the
desired level is selected. The delay level options are; 2-second delay (non-
illuminated LED and sin le beepin will sound) or 5-second delay (illuminated LED
and sin le beepin will sound). The LED status indicates the delay level selected.
Initiating a Fall Alarm
The Fall Detection Unit be ins to monitor for falls after the resident sits or lies on
the chair / bed mat for more than 15 consecutive-seconds. At the end o the 15
seconds the LED indicator illuminates and a single beeping sound is heard.
If the resident ets up from the chair or bed once the unit has one into monitorin
mode, an audible fall alarm is activated.
Once in alarm mode, the Fall Detection Unit sends an alarm transmission (every 20
seconds) to the monitorin station. This messa e is sent until the fall alarm is reset.
To reset the fall alarm, hold the Fall Detection Unit ti htly and press the Reset
button for approximately 5 seconds. The audible alarm is deactivated. The Fall