HP 7683 User manual

Installation Guide
HP 7683 Automatic
Liquid Sampler

Little Falls Site
Hewlett-Packard Company
2850 Centerville Road
Wilmington, DE 19808-1610
Hewlett-Packard Company
1997
All Rights Reserved.
Reproduction, adaptation, or
translation without permission
is prohibited, except as allowed
under the copyright laws.
HP Part No. G2612-90100
First edition, November 1997
Printed in USA
Warranty
The information contained in
this document is subject to
change without notice.
Hewlett-Packard makes no
warranty of any kind with
regard to this material,
including, but not limited to,
the implied warranties of
merchantability and fitness
for a particular purpose.
Hewlett-Packard shall not be
liable for errors contained
herein or for incidental or
consequential damages in
connection with the furnishing,
performance, or use of this
material.
Safety Information
The HP 7683 Automatic Liquid
Sampler meets the following
IEC (International
Electrotechnical Commission)
classifications: Safety Class 1,
Transient Overvoltage
Category II, and Pollution
Degree 2. This unit has been
designed and tested in
accordance with recognized
safety standards and designed
for use indoors. Whenever the
safety protection of the HP 7683
Automatic Liquid Sampler has
been compromised, disconnect
the unit from all power sources
and secure the unit against
unintended operation.
The recyclable carbon
monoflouride lithium battery is
BR-2/3 A 1,200 mAh. Fuses F001
and F002 are 3 A, 250 Vac, IEC
127 Type T. Fuses F201 and
F202 are 10 A, 250 Vac, IEC 127
Type T. Fuse F101 is a 0.5 A,
250 Vac.
Warnings in this manual or on
the instrument must be
observed during all phases of
operation, service, and repair of
this instrument. Failure to
comply with these precautions
violates safety standards of
design and the intended use of
the instrument. Hewlett-
Packard assumes no liability for
the customer’s failure to
comply with these
requirements.
Refer servicing to qualified
service personnel. Substituting
parts or performing any
unauthorized modification to
the instrument may result in a
safety hazard. Disconnect the
AC power cord before removing
covers. The customer should
not attempt to replace the
battery or fuses in this
instrument.
Safety Symbols
This manual contains safety
information that should be
followed by the user to ensure
safe operation.
WA RN IN G
A warning calls attention to a
condition or possible situation
that could cause injury to the
user.
CAUTION
A caution calls attention to a
condition or possible situation
that could damage or destroy
the product or the user's work.
Important User
Information for In Vitro
Diagnostic Applications
This is a multipurpose product
that may be used for qualitative
or quantitative analyses in many
applications. If used in
conjunction with proven
procedures (methodology) by a
qualified operator, one of these
applications may be In Vitro
Diagnostic Procedures.
Generalized instrument
performance characteristics
and instructions are included in
this manual. Specific In Vitro
Diagnostic procedures and
methodology remain the choice
and the responsibility of the
user, and are not included.
Sound Emission
Certification for Federal
Republic of Germany
If Test and Measurement
Equipment is operated with
unscreened cables and/or used
for measurements in open set-
ups, users have to assure that
under these operating
conditions the Radio
Interference Limits are still met
at the border of their premises.
The following information is
provided to comply with the
requirements of the German
Sound Emission Directive dated
January 18, 1991.
Sound pressure Lp
< 57 dB(A)
during normal operation
at the operator position
according to ISO 7779 (Type
Test).
When operating the
HP Automatic Liquid Sampler,
the sound pressure ≈68 dB(A)
during short burst injection
pulses.
Schallemission
Werden Meß-und Testgeräte mit
ungeschirmten Kabeln und/oder
in offenen Meßaufbauten
verwendet, so ist vom Betreiber
sicherzustellen, daß die Funk-
Eströbedingungen unter
Betriebsbedingungen an seiner
Grundstücksgrenze
eingehalten werden.
Diese Information steht im
Zusammenhang mit den
Anforderungen der
Maschinenlärminformationsv
erordnung vom 18 Januar 1991.
Schalldruckpegel LP < 57 dB(A)
Am Arbeitsplatz
Normaler Betrieb
Nach DIN 45635 T. 19
(Typprüfung)
Bei Betrieb des
HP Automatischer
Slüssigkeitsprobengeber treten
beim Oeffnen des Ventils
impulsfoermig Schalldrucke Lp
bis ca. 68 dB(A) auf.

Contents
Installation checklist ................................................................................. 1
Installing the HPG2614A Tray ................................................................ 2
Installing the HPG2613A Injector ........................................................... 4
Installing the HPG1926A Bar Code Reader ............................................ 6
Connecting the cabling ............................................................................. 8
Adapting for cool on-column injection ..................................................... 9
Configuring your GC .............................................................................. 11
Making a trial run ................................................................................... 12
Maintenance ............................................................................................ 14
Removing the turret ................................................................................ 15
Turret alignment ..................................................................................... 16
Replacing the needle support assembly .................................................. 18
Replacing the needle guide in the needle support foot ........................... 21
Correcting syringe problems ................................................................... 23
Correcting sample vial delivery problems .............................................. 24
Faults ....................................................................................................... 25
No light is on ................................................................................... 26
The Fault light is on ......................................................................... 26
The Fault and the Run lights are on ................................................. 26
Align Mode light is on ..................................................................... 27
All lights are on ............................................................................... 27
Error messages ........................................................................................ 28
Bottle in gripper ............................................................................... 28
Front (or back) door open or injector not mounted ......................... 28
Front (or back) injector com error ................................................... 29
Front (or back) injector incomplete injection .................................. 29
Front (or back) injector reset ........................................................... 29
Front (or back) plunger error ........................................................... 30
Front (or back) syringe error ............................................................ 30
Front (or back) turret error ............................................................... 31
Injector not present .......................................................................... 31
Injector offline ................................................................................. 32
No bar code reader ........................................................................... 32
No bottle in gripper .......................................................................... 33
Tray not present ............................................................................... 34
Tray offline ...................................................................................... 34

Installation Guide
Installation checklist
1
Installation checklist
The HP 7683 Automatic Liquid Sampler consists of the HP G2612A ALS Interface
Board, the HP G2613A Injector Module, the optional HP G2614A Tray, and an
optional HP G1926A Bar Code Reader for the tray. The tray is installed on the
HP 6890 Plus Gas Chromatograph (GC).
Use this checklist when installing your HP 7683 Automatic Liquid Sampler and
accessories.
❐Unpack the sampler components.
❒Install the ALS Interface Board, if purchased.
❒Install the tray, if purchased. See page 2.
❒Install the injector mounting post. See page 4.
❒Install the injector. See page 4.
❒Install the (8 sample position) turret, if desired. See page 16.
❒Install the bar code reader, if purchased. See page 6.
❒Connect the cables. See page 8.
❒When using the injector for cool on-column injection:
a. Check that the correct needle support assembly is installed. See
page 10.
b. Install the proper syringe for your column. See your Operation Guide.
❒Configure your GC. See page 11.
❒Test the installation by performing a trial run. See page 12.

Installation Guide
Installing the HP G2614A Tray
2
Installing the HP G2614A Tray
If you purchased HP accessory G2612A separately from your HP 6890 Plus GC,
install it now. Refer to the documentation provided with the accessory for
instructions. After it is installed, continue with the steps below to install the
optional HP G2614A Tray.
1. Unpack the injector module, tray, and bar code reader and place them on
the bench top.
2. Remove the 3 injection port cover Torx screws shown.
Figure 1 Installing the tray
Injection port cover
screws
Tray
Captive screw (install first)
Tray mounting
Remove screws from
cover

Installation Guide
Installing the HP G2614A Tray
3
Caution Do not move or manipulate the robotic arm or gripper. Moving the arm or gripper
back and forth can cause damage. If necessary, rotate the arm assembly a few
degrees until there is enough clearance.
3. Align the tray over the 3 holes in the injection port cover and locate in place.
Tighten the captive screw in the center of the tray mounting arm to hold the
tray in place. Install the 2 mounting screws into the tray and tighten until
the tray is secure.
4. Install the tray quadrants. Slide the leading edge of each quadrant under the
guides on the tray, then snap the front tab into the slot on the tray. See
Figure 2 below.
Figure 2 Install the tray quadrants
Guide tab
Slot
Tray quadrant
Front tab

Installation Guide
Installing the HP G2613A Injector
4
Installing the HP G2613A Injector
Note Do not mount the HP G2613A Injector on a mounting post for an HP 7673
Injector. Remove the old post and replace it with the new one.
1. Install the mounting post on the injection port cover in the front or rear
location, as desired.
2. Open the injector door. Slide the packing material and syringe carriage
down, then remove the packing.
3. Mount the injector. Line up the hole in the base of the injector that is nearest
the cable with the mounting post. Lower the injector about an inch (2.5 cm)
onto the post.
Figure 3 Mounting the injector
Alignment hole
Alignment pin
Mounting post
Cable
Hole in injector base
Inlet cover

Installation Guide
Installing the HP G2613A Injector
5
4. Front location:
Turn the injector so that the turret is facing the front of the GC. Lower the
injector until the alignment pin in the base enters the alignment hole in the
injection port cover.
Back location:
Turn the injector so the turret faces the left side of the GC. Lower the injector
until the alignment pin in the base enters the alignment hole in the injection
port cover.
5. Check your work:
• The injector must be vertical.
• The alignment pin must be properly seated in the alignment hole.
• The injector feet should touch the injection port cover.
If the injector will not sit upright on the GC, check that the plumbing and
cabling under the injection port cover are properly routed in their channels.
Choosing the turret type to use
Your HP G2613A Injector comes with 2 types of turrets, a standard turret for use
with or without a tray, and an alternate 8 sample position turret which is used
without a tray. If you are using the optional HP G2614A Tray, the injector comes
pre-configured with the proper turret type.
If you are not using a tray, you may use 1 to 3 sample vials in the standard turret.
If you wish to use up to 8 sample vials, you may remove the standard turret and
install the alternate 8 sample position turret. See Removing the turret for details.
Note After changing turrets, it is recommended that you perform the turret alignment
procedure described on page 16.

Installation Guide
Installing the HP G1926A Bar Code Reader
6
Installing the HP G1926A Bar Code Reader
1. Install the HP G1926A Bar Code Reader support bracket under the front of
the tray as shown below. Leave the screws slightly loose. You will tighten
them later.
Figure 4 The Bar Code Reader support bracket
2. Mount the Bar Code Reader onto the front of the tray base by inserting the
two mounting clips on the Bar Code Reader into the mounting slots on the
edge of the tray base. See Figure 5.
Figure 5 Mounting location for the Bar Code Reader
Adjust this space
Mounting screws (2)
Height adjustment screws
Tray
Support Bracket
GC
Mounting clips
Bar code reader

Installation Guide
Installing the HP G1926A Bar Code Reader
7
3. Connect the Bar Code Reader cable to the Bar Code Reader connection on
the tray.
Figure 6 Connecting the Bar Code Reader
4. Slide the support bracket towards or away from the tray until the top of the
bar code reader is parallel with the tray surface. Tighten the mounting
screws.
5. Adjust the Bar Code Reader’s height. Slightly loosen the height adjustment
screw on the side of the support bracket and raise or lower the bracket until
there is a very small gap—about the thickness of a piece of paper—between
the mounting clips on the bar code reader and the tray. See Figure 4. Tighten
the height adjustment screw.
Connect Bar Code Reader

Installation Guide
Connecting the cabling
8
Connecting the cabling
1. Connect the injector(s) to the GC. Use the tray cable, HP part number
G2614-60610, to connect the tray to the use GC. See Figure 7.
Figure 7 HP 7683 Automatic Liquid Sampler cabling
2. When the cables are connected, turn on the GC power. The tray will beep
twice. After the start-up process ends, the Ready light on the injector tower
should be on.
If the Align Mode light is on, see Turret alignment on page 16.
If either the Fault or Run light is on, see Faults on page 25.
HP 6890 Plus GC
G2613A
Injector
(Front)
G2613A
Injector
(Back)
Injector cables
Tray cable
G2614A
Tray

Installation Guide
Adapting for cool on-column injection
9
Adapting for cool on-column injection
The HP 7683 Automatic Liquid Sampler can be adapted to inject samples directly
onto 250-µm, 320-µm, and 530-µm columns.
If you wish to adapt the injector and GC for cool on-column use, follow these
steps:
1. Select the on-column syringe needed for the column size. See your
HP Chemical Analysis Columns and Supplies catalog and Volume 2 of your
GC Operating Manual for a list of parts.
2. Prepare the GC inlet. (See Volume 2 of your GC Operating Manual for
instructions).
• Check your needle to column size.
• Verify that the insert matches the needle size.
• If necessary, replace the septum.
3. Verify that the correct needle support assembly is installed in the injector.
Replace if necessary. See Replacing the needle support assembly on
page 18.
For a list of consumable needed to perform these injections, see Volume 2 of
your GC Operating Manual.

Installation Guide
Adapting for cool on-column injection
10
Figure 8 Needle support assemblies
4. If necessary, reinstall the injector onto the GC.
5. Install the syringe. See your Operation Guide.
6. Rotate the turret clockwise until it stops, then verify the installation by
manually sliding the syringe carriage down until the needle enters the inlet.
530-µm
(standard)
250-µm/320-µm
(HP accessory 18599T)

Installation Guide
Configuring your GC
11
Configuring your GC
When installation is complete, configure your HP 6890 GC for use with the
HP 7683 Automatic Liquid Sampler.
Refer to the section on injector configuration in your GC Operating Manual for
instructions to set the following:
• front and/or back injector location
• nanoliter enable
• tray enable
• bar code reader position (position 3, enable, and type)

Installation Guide
Making a trial run
12
Making a trial run
Once installation and configuration are complete, make a quick injection using
the sampler to verify that it works properly.
1. Install an empty syringe in the injector.
2. Place empty bottles in the Solvent A and Waste A turret positions. Place an
empty capped sample vial in the tray 1 position (or the turret sample 1
position, if not using the tray).
3. Set these following parameters on the GC:
4. Set the GC oven program to 25°C with a 0°C/min ramp, a hold time of
0.1 minutes, an equilibrium time of 0.3 minutes, and an initial time of
0.3 minutes.
Parameter Setting
Injection volume 1
# Sample pumps 1
Viscosity delay 0
# Sample washes 1
# Solve A washes 1
# Solv B washes 0
Slow plunger off
Pre-dwell time 0
Post-dwell time 0
Sampling offset off (on if cool on-column inlet used)
# Solv A pre-wash 1
# Solv B pre-wash 0
# Injections/vial 1
Samples 1–1 (101–101 for turret with HP ChemStation control)

Installation Guide
Making a trial run
13
5. Store, load, then run the sequence.
• If there are no faults, the injector will make 1 "injection" from the first
vial position.
• If any faults occur, see Faults on page 25, Error messages on page 28,
Correcting syringe problems on page 23, or Correcting sample vial
delivery problems on page 24.

Installation Guide
Maintenance
14
Maintenance
This section contains some suggestions for ensuring good performance of your
HP 7683. The maintenance interval varies with the use of the instrument.
Caution Do not use any lubricants on the HP 7683 Automatic Liquid Sampler. They may
effect the chemical performance of the GC and damage the instrument.
On an occasional basis:
❒Clean the surface of the tray arm, gripper, gripper jaws, and tray quadrants.
❒Clean the needle guide and nearby surfaces of the injector. Dust and dirt
accumulate in these areas and can be picked up by the syringe needle and
carried into the inlet.
❒Clean the surface of the injector. Vacuum off any dust that builds up around
the vents.
❒Make sure the injector mounting post is tight.
❒Make sure the tray mounting screws are tight.
❒Make sure the knurled nut on the top of the turret is tight.
❒Be sure all the cables are securely attached.

Installation Guide
Removing the turret
15
Removing the turret
If you need to remove or replace your turret, use the following instructions for
proper replacement.
1. Disconnect the injector cable from the GC.
2. Open the injector door. Unscrew and remove the knurled nut from the top
of the turret. See Figure 9.
3. Rotate the turret so that the open section faces the back of the tower and
the arrow on top of the turret points directly back (see Figure 9). Push the
stripper arm toward the back, then lift the turret up so that it clears the
center shaft.
4. To install the replacement turret, align the tab on the motor hub with the
slot in the underside of the turret.
5. Push the stripper arm to the back of the tower, insert the turret with the
open section facing the back of the tower and the arrow pointing back, and
install the turret with the tab seated in the slot. Seat the turret on the hub,
rotating it gently if necessary until it drops completely into position. The top
of the threaded shaft should be almost level with the top of the turret.
6. Replace the knurled nut and tighten it firmly (finger-tight).
7. Close the injector door.
8. Plug the injector cable into the GC and check for the Ready light.
Note For increased accuracy of the needle sampling depth, perform the turret
alignment procedure each time the turret is changed. See Turret alignment on
page 16 .

Installation Guide
Turret alignment
16
Figure 9 Turret removal
Turret alignment
If you changed turrets in the injector and desire increased needle sampling depth
accuracy, or if the Align Mode light is on, perform this alignment procedure.
When the Align Mode light is on, the injector will not operate until this procedure
has been performed.
1. When the injector is not operating, open the injector tower door.
2. Slide the syringe carriage up until it stops.
3. Remove the syringe. See your Operation Guide for details.
4. Carefully remove the needle support assembly from the injector. See
Replacing the needle support assembly on page 18 for details.
5. Use a pen to press the recessed alignment button above the indicator lights,
then close the door. See Figure 10.
Knurled nut
Turret
Alignment arrow
Stripper arm
Motor hub
tab

Installation Guide
Turret alignment
17
Figure 10 Aligning the turret
6. The injector goes through the following steps:
• The turret rotates to verify that the needle support assembly was
removed, then rotates to determine what type of turret is installed.
• The syringe carriage moves all the way down, then moves back up again
and clears the turret.
• The syringe carriage steps down until it touches the turret. This sets its
position relative to the turret.
• The plunger moves to calibrate stops.
7. When the alignment is complete, the Ready light turns on and the sampler
returns to the ready state. Install the needle support assembly and syringe.
If the injector fails alignment, see Faults on page 25.
Press the alignment button
Align Mode
Fault
Run
Ready
Table of contents
Other HP Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Sarstedt
Sarstedt Mini Centrifuge MC 6 Series operating instructions

Metrohm
Metrohm 810 manual

Agilent Technologies
Agilent Technologies 5890 manual

Agilent Technologies
Agilent Technologies QuickSwap G3185B Installation and setup

bioMerieux
bioMerieux VITEK 2 Service manual

Fisher Scientific
Fisher Scientific accumet BASIC AB 15 manual

Judo
Judo JUV 20 G Installation and operating instructions

lbx
lbx H03D user manual

Thermo Scientific
Thermo Scientific TSQ 8000 Preinstallation manual

Leica BIOSYSTEMS
Leica BIOSYSTEMS HistoCore BIOCUT Instructions for use

IKA
IKA MF 10 operating instructions

Grant-bio
Grant-bio Intelli-Stirrer MSH-300i operating instructions