Controlling the FAM through the CS2000
HYDAC Filter Systems GmbH en Page 10 / 16
BeWa FAM Zusatz CS2000 3189702b en 2009-10-16.doc 2009-10-16
HYDAC Contamination Sensor offline!
Measuring cycle stopped
x - Restart Measurement
p - Enter Parameter Mode
Enter key (x, p) and Return!
8. Press the "p" key and then Return to switch to Parameter Mode. The following
display appears after Parameter mode starts:
Parameter Mode:
x - Go back
t - Measuring cycle time
m - Measuring mode
2 - Limits Mode M2
3 - Limits Mode M3
4 - Limits Mode M4
b - Set DIN-Bus Address and Name (for CS models with DIN-Bus module only)
p - Pump protection
o - PLC Output Format
r - RS232-Display Output Format
i - Current Output Format (for CS models with analog module only)
c - Check Current Output (for CS models with analog module only)
f - Flow check
s - Sensor check
d - Recover Factory Default Settings
Enter key (x, t, m, 2, 3, 4, b, p, o, r, i, c, f, s ,d) and Return!
The gray-highlighted items are relevant for controlling the FAM. They are
described below.
9. t = Measuring cycle time is used to set the measurement interval (20 to 120
seconds). The current cleanliness class is shown in the display after each
measurement interval. A default value of 60 seconds should be set.
Now press "t" and then Return. The following then appears, for example:
actual measuring cycle time (s) set to: 20
New value for measuring cycle time (s) (enter 'Return' to keep setting): 60
actual measuring cycle time (s) set to: : 60
The measurement interval has been increased in the example from 20s to 60s.
10. Only Measuring Mode M2 (measurement and switching) is used for controlling
the FAM.
Now press "m" and then Return. The following then appears, for example:
actual Mode: 3
1=M1, 2=M2, 3=M3, 4=M4
Enter key (1, 2, 3, 4) and Return (enter 'Return' to keep setting) 2