HyquestSolutionsPtyLtd
CMC20Ainstructionmanual3Issue2:19/5/2016
©Copyright
1. SPECIFICATIONS
Housing: Weatherproof powder coated aluminium die cast case with nylon
webbing adjustable neck strap.
Display: 4 Digit liquid crystal display rated to 80oC.
Accuracy: + 0.1 seconds (Timed from first contact closure)
+ 1 pulse
Timer: Quartz crystal accuracy, better than 1 in 20,000.
Preset Facility: Time intervals of 10, 20, 30, 40, 50, 80, 100, 200 and infinity
seconds.
Infinity Mode: INT. This mode is used for taking discharge measurements by the
integration method, for use with ground-feeler weights or for manual
operation of counter.
Contact Debounce: The current meter pulse is debounced in software. The debounce
time is user selectable. (1, 2, 4, 6, 8, 10, 12, 14, 16, 18, 20, 25, 30,
35, 40, 45mS) This is useful for a meter with a mechanical “cat
whisker” contact wire.
Buzzer: Instantaneous beep for each current meter pulse except in the
integrate mode. In the integrate mode, the buzzer gives a continuous
beep while the current meter switch is closed.
Power Supply: 9V DC - 6 AA Alkaline batteries. The actual battery voltage is
displayed for 2 seconds each time the unit is switched on. Low
voltage is also indicated by all decimal points showing on the display
at 5.8 volts.