PARTS OF THE TONOMETER
CLEANING THE PROBE BASE
Theprobebasecanbereusedaercarefulcleaning.Makesure
thattheprobebaseiscompletelydrybeforeusingit.
1. Unscrewtheprobebasecollarandcarefullyremovetheprobe
basefromthetonometer.
2.Inverttheprobebaseoveracontainerlledwith70-100%
alcoholsoluon,allowingtheprobebasetodropintothe
containerandletsoakfor5-30minutes.
3.Drytheprobebasebyblowingsomecompressedairintoit.
4.Carefullyreinsertthecompletelydryprobebase,screwthe
collarbackintothetonometer.
Pleaserefertothemanualformoredetailedinstruconsof
changingandcleaningtheprobebase.
CHANGE/CLEAN THE PROBE BASE REGULARLY TO AVOID
UNNECESSARY SERVICE.